Hach Series 5000 to 5500sc Silica Analyzer Replacement Guide

Guide for switching from Hach Series 5000 to Hach 5500sc

1 Introduction

Twenty years after launching the Silica analyzer model 5000, Hach has introduced to the market a new analyzer with innovative solutions. The aim of this guide is to help you make the transition from the Series 5000 Analyzer to the new 5500sc Silica Analyzer.

2 Why switch from Hach 5000 to Hach 5500sc?

There are several reasons to switch from the Series 5000 to the 5500sc. Even if you are continuing to use the same measurement operations and principles, the new generation provides benefits that will prove advantageous.

Two analyzers are shown side-by-side: the Series 5000 Silica on the left and the Hach 5500sc on the right.

Series 5000 Silica Hach 5500sc
2.9 liters reagents every 30 days
1 bottle/month @ 8.8 min cycle time
Extended reagent replacement intervals and low reagent consumption
1 bottle (2-liter) every 90 days @ 15 min or 60 days @ 9.5 min
Fixed measurement cycle Configurable measurement cycle from continuous up to 240 mins (Interval Mode)
2.9 L PET bottles 2 L bottles with quick-connect, including pre-installed tubing and color coding - Reduces time maintaining the unit and reduced risk of contact with chemicals
No predictive diagnostic tool Predictive diagnostics (PROGNOSYS™) and built-in help function.
Reduces risk of unexpected failures and unit downtime
8-channel external sequencer Optional built-in sequencer, up to 6-channels
Lowers costs via sampling points and eliminates cost for an external sequencer (up to $ 6,000)
Grab Sample In Grab Sample In and Grab Sample Out
Enables quick QC validation in addition to normal process monitoring
Pressurized reagent delivery system NO PUMPS Same principle with improved design. Built-in air compressor requires no external air supply
Colorimetric analysis method at 810 nm. Same method
3 reagents and one standard Same number, but different reagent concentrations

3 Specification differences between Hach 5000 and Hach 5500sc

SPECIFICATION Hach 5000 Hach 5500sc
MEASUREMENT SPECIFICATIONS
Measurement range: 0.00 to 5000 µg/L as SiO2 0-5000 µg/L as SiO2
Accuracy (typical): 0.00–500 µg/L: ± 1.0 µg/L or ± 5% of reading, (the larger value); 500–5000 µg/L: ± 7% of reading 0–500 µg/L: ±1% or ±1 µg/L (the larger value), 500–5000 µg/L: ±5%
Minimum Detection Limit: 0.5 µg/L 0.5 µg/L
Precision/Repeatability: ± 0.5 µg/L or ± 1.0% (the larger value) ±0.5 ppb or ±1% (the larger value)
Step Response Time (sample temperature dependent): 8.8 minutes for 30 to 50 °C, 15 minutes for 5 to 40 °C, (field adjustable) Typical 9.5 minutes from 5 to 50 °C
Stabilization time After initial startup or annual maintenance: 5 meas. cycles
After standby: 1 meas. cycle
After calibration: 0 meas. cycles
After initial startup or annual maintenance: 5 meas. cycles
After standby: 1 meas. cycle
After calibration: 0 meas. cycles
Calibration time 15 min Tcal<30°C, 8 min Tcal>30°C Slope calibration: 10 minutes
Analyzer Sample Requirements Regulated to 5 ± 3 psig (34.5 ± 20.7 kPa). Flow rate: 100 to 300 mL/minute. Temperature: 5 and 50 °C. Accepted pressure: 2–87 psi (0.14 - 6 bar) in the provided preset regulators, pre-set to 2.0 psi.
Acceptable flow rate: 55–300 mL/min
Temperature: 5 to 50 °C (41 to 122 °F)
A sample pressure control kit is provided 1 1, 2, 4 or 6; programmable sequence
No. of sample streams yes yes
Grab sample in no yes
Grab sample out no yes
Calculations No Yes, between Si concentration of sample lines or external parameters
LIQUID CONNECTIONS
Sample line in ½ inch push-to-connect fitting, ½ inch only; rigid or semi rigid plastic tubing ½ inch or 6 mm OD push-to-connect fitting for plastic tubing
Sample bypass drain none ½ inch or 6 mm OD push-to-connect fitting for plastic tubing
Analyzed sample out drain fitting ½ inch hose barb fitting ½-inch NPT PVC female 11 mm (7/16 in.) ID slip-on fitting for soft plastic tubing
Case drains ½ inch hose fitting 11 mm (7/16 in.) ID slip-on fitting for soft plastic tubing
GASCONNECTIONS
Air Purge inlet: instrument- quality air 15-scfh (std cubic feet per hour) ½-inch OD stainless steel compression tubing fitting 15 scfh (0.425 m3/hour), 6 mm OD push-to-connect fitting for plastic tubing
Air Purge outlet none 11 mm (7/16 in.) ID slip-on fitting for soft plastic tubing, same tubing as analyzed sample out drain
Reagent Pressure Inlet Fitting and Source ½-inch OD stainless steel compression tubing fitting none (internal source)
ELECTRICALCONNECTIONS
Power Requirements: 115 or 230 VAC, 50/60 Hz, switch selectable; 52 VA, 32 W maximum 100–240 VAC, 50/60 Hz universal power supply
Analog/Recorder Outputs: One, selectable for 0–0.01/0.1/1 V, or 4–20 mA (500 Ω maximum). 4x; 4–20 mA outputs, Load impedance: 600 Ω max. Expandable to 8x or 12x with additional, 4-20ma SC1000 output module(s)
Serial I/O: RS232 and 20 mA current loop Optional modules for flow, ModBus,
Digital inputs none Profibus, *Hart, *Ethernet (*future, not yet availablet)
4-20 mA input none Six isolated inputs configurable as TTL or open-collector/relay
Alarms: 4x programmable relays, Functions: 2 x sample concentration alarms, analyzer system warning, analyzer system shutdown alarms Optional SC1000 4x 4-20 ma input module (2 modules Max)
4x programmable relays;
Functions: concentration alarm, scheduler, analyzer warning/error or event (measuring ch.x, zero cal, slope cal, shutdown, startup, grab sample or end of measuring cycle )
Dual monitor none Connection for a remote SC controller which becomes master
Smart probe none Connection for future external probe
Power out none 2x max current, 250 VAC, AC:5A, DC:1.6A, always on
REAGENTS
Reagents: 2.9 L Molybdate 3 (P/N 1995-03),
2.9 L Citric Acid/Surfactant(P/N 23470-03),
2.9 L Amino Acid F (P/N 23531-03),
2.9 L Silica Standard Solution, 500 µg/L (P/N. 21008-03) (250 mL required for standardization)
2 L Molybdate 3, 6774802
2 L Citric Acid/Surfactant, 6774902
2 L Amino Acid F, L 6775102
2 L Standard Silica, 500 µg/L , 67750022 (1 bottle) of standard for every 10 calibrations
Reagent consumption for a 15 minute cycle time 5.1 L (1.8 bottle) of each reagent every 90 days 2 L ( 1 bottle) of each reagent every 90 days (Measurement interval set to 15 min)
GENERAL
Enclosure: Rating: NEMA 4x/IP65
Molded ABS plastic cabinet with gasketed doors (for indoor use)
Rating: NEMA 4x/IP65
Material: PC/ABS case, PC door, PC hinges and latches, 316 SST hardware
Indoor use only.
Mounting Bench top or panel mounting only Wall, panel or table
Dimensions W x D x H
mm: 564 x 419 x 857
inches : 22 x 17 x 34
W x D x H
mm : 452 x 360 x 804
inches : 17.8 x 14.2 x 31.7
An adapter plate Model 6791000 is available to mount the Hach 5500sc on previous Hach 5000 panel location
Shipping Weight 36.7 kg (81 lb) without reagents and standards 20 kg (45 lb) without reagents and standards, 36.3 kg (80 lb) with reagents
Ambient Operating Conditions: 10 to 45 °C, 5 to 95% non-condensing humidity.
Suitable for general purpose, clean, indoor environments (not suitable for outdoor use).
5 to 45 °C (41 to 113 °F) 5 to 95% non-condensing humidity
Storage temperature:
−20 to 60 °C (−4 to 140 °F)
(Must not be in direct sunlight)
Screen Alphanumeric LCD display, membrane keyboard Display, LCD 5.7" color VGA (640x480), membrane keyboard

4 Reference documents

Below are shown the documents required to perform a quick, safe and correct replacement of analyzers. Each document can be downloaded directly from the web by clicking the description.

Code Description and link Status
Catalog 60000-18 Series 5000 Silica Analyzer, Instrument Manual old
DOC013.99.80240 5500sc SiO2 Analyzer Unpacking Guide Illustrated new
DOC273.99.80258 Illustrated user instructions to mount the analyzer on a wall, on panel or on table. new
DOC273.99.80259 Adapter Plate User Instructions for Polymetron Analyzer. new
DOC023.97.80240 Installation Manual 5500sc SiO2 new
DOC023.97.80267 Operations 5500sc SiO2 new

5 Step 1- Removing the Hach 5000 analyzer

5.1 Save the analyzer setup

Perform the following actions:

  • Perform an analyzer printout: (SETUP/PRINT SETUP/ENTER)
  • Back up your Hach 5000 analyzer setup, including:
    • High/Low Alarm values
    • Alarm relay identification
    • Recorder max concentration
    • 4-20mA output ranges
    • Digital com settings
  • Disconnect the analyzer from its power source.

A sample analyzer setup printout is displayed, showing alarm values, recorder settings, and date/time.

5.2 Remove bottles

  • Remove the 3 reagent bottles.
  • Follow regulations in place for the reagents' elimination/destruction.
  • Remove the silica standard solution.
  • You can keep the solution, but discard the bottle – the 5500 has different bottle types.

Two reagent bottles and a bottle of silica standard solution are shown, indicating they need to be removed.

5.3 Remove all electrical connections

  • 115/230 VAC power line
  • High voltage relays
  • Low voltage relays
  • Recorder output
  • RS 232 connection

A diagram illustrates the electrical connections of the Hach 5000 analyzer, showing power, relays, recorder output, and RS232 connections with corresponding wire colors.

5.4 Remove all hydraulic connections

  • Sample in/out
  • Cabinet/Sample Drain
  • Air purge
  • The sample heater, Model 48500, can be removed as long as the sample temperature is above 5°C (41° F).
  • The installed sample pressure conditioning kit (Model 45983-00) can be removed – the 5500 uses a different kit.

A diagram shows the hydraulic connections of the Hach 5000 analyzer, labeling sample inlet/outlet, drains, air purge, and indicating components like the sample heater and pressure conditioning kit.

5.5 Remove the Hach 5000 analyzer unit

  • Remove the six screws.
  • Remove the analyzer, being careful of its weight: 37kg (81 lb).

The Hach 5000 analyzer unit is shown, highlighting the need to remove six screws for removal.

6 Step 2 - Installing the Hach 5500sc

6.1 Wall mount installation

Follow instructions 1B to 4B on document: Illustrated user instructions to mount the analyzer on a wall, on panel or on table.

An illustration depicts the Hach 5500sc analyzer being mounted on a wall.

6.2 Panel mount installation

  • Install the optional adapter plate Model 6791000 with the six screws provided.
  • Follow instructions on document: Adapter Plate User Instructions for Polymetron Analyzer.

An illustration shows the optional adapter plate being installed for panel mounting the Hach 5500sc analyzer.

Follow instructions 1C to 4C on document: Illustrated user instructions to mount the analyzer on a wall, on panel or on table.

6.3 Panel mount

Follow instructions 1A on document: Illustrated user instructions to mount the analyzer on a wall, on panel or on table.

An illustration shows the Hach 5500sc analyzer mounted on a panel.

7 Connecting the Hach 5500sc hydraulics

7.1 Identify hydraulic connections

Identify the hydraulic inlet/outlet connections for one to six sample streams.

A diagram identifies the hydraulic inlet and outlet connections for up to six sample streams on the Hach 5500sc.

7.2 Connect samples and drain lines

  • Use the supplied tubing (6 mm), Y-strainer with filter and pressure regulator to plumb the drain and the sample to the analyzer.
  • The sample line tubing that goes into the plumbing access ports must be 6 mm.
  • Tubing of 1/4 in. may be used for the sample line up to the valve and y-strainer, but not into the analyzer's plumbing access ports.
  • See DOC023.97.80240 for more information.

A diagram illustrates the connection of sample and drain lines using supplied tubing, a Y-strainer with filter, and a pressure regulator.

7.3 Connect sample stream

Install each sample line into the center of a larger process pipe to minimize interference from air bubbles or bottom sediment. See figure on left.

An illustration shows how to connect the sample stream into a larger process pipe to minimize interference.

7.4 Set the bypass flow rate

  • The bypass flow can be adjusted when the analyzer is in shutdown mode. Adjust the flow rate of the sample bypass line with the flow valve as shown at left.
  • Use an external meter to measure the flowrate of the sample bypass line.
  • For faster response to changes in the process stream, increase the sample bypass line flow rate when the process stream is far from the analyzer.

A diagram shows how to adjust the bypass flow rate using a flow valve and suggests using an external meter.

8 Connecting the Hach 5500sc electrical connections

Follow procedure on pages 11-17 of the operation manual.

9 Installing analyzer bottles and stir bar on the Hach 5500sc

Follow procedure on pages 18-19 of the operation manual.

10 Training operators and supervisors on the new Hach 5500sc

The following table shows the topics to be presented to train the plant operators and supervisors.

Topics Operation Manual Page
Plant operators – Standard use of the analyzer
User interface and navigation 5
Grab sample measurement 9
Getting a grab sample from the analyzer 9
View data 14
View instrument information 16
Using SD card's 16
Instrumentation supervisor – startup and configuration
Reagent preparation 7
Analyzer setup 8, 10
Sequencer configuration 8
Reagents and standards setup 9
Calculation setup 11
4 – 20 mA setup 11
PID setup (optional) 12
Relay setup 12
Relay activation options 13
Updating the firmware 16
Analyzer calibration 17

PDF preview unavailable. Download the PDF instead.

2013-11 Replacement Guide 5000 Series to 5500sc 33-105 (Global) Microsoft Office Word 2007 Microsoft Office Word 2007

Related Documents

Preview Hach CL17sc Calibration Verification and Adjustment Recommendations
Guidance on verifying and adjusting the calibration of the Hach CL17sc analyzer, emphasizing factory calibration and recommended verification procedures.
Preview Hach NA5600sc Online Sodium Analyzer: Features, Specifications, and Included Components
Detailed information on the Hach NA5600sc Online Sodium Analyzer, a 1-channel instrument with cation kit and autocalibration. Covers features like automatic electrode reactivation, space-saving design, low maintenance, and predictive diagnostics, along with comprehensive specifications and included accessories.
Preview HACH BioTector B7000i Online TOC Analyzer User Manual | Installation, Operation & Maintenance Guide
Comprehensive user manual for the HACH BioTector B7000i Online Total Organic Carbon (TOC) Analyzer. Learn about safety, installation, operation, calibration, maintenance, and technical specifications for accurate water quality monitoring.
Preview Hach Replacement Oxygen Electrode (ppb) - Product 09182=A=1000
Information on the Hach Replacement Oxygen Electrode (ppb), without probe body. This electrode is a replacement sensor for the Polymetron 9582sc Dissolved Oxygen analyzer. Specifications include material, special notes, temperature sensor, and warranty.
Preview Hach A1000, A1000XP User Manual
Comprehensive user manual for Hach A1000 and A1000XP TOC analyzers, covering specifications, installation, operation, maintenance, and troubleshooting.
Preview HACH SL1000 Portable Parallel Analyzer: Basic User Manual
Get started with the HACH SL1000 Portable Parallel Analyzer. This basic user manual covers essential information on specifications, installation, operation, maintenance, and troubleshooting for accurate water quality analysis.
Preview HACH SC4500 User Manual: Installation, Operation, and Maintenance
Comprehensive user manual for the HACH SC4500 analytical controller. Covers installation, setup, operation, maintenance, and troubleshooting for optimal performance in process monitoring applications.
Preview HACH SC4500 Basic User Manual - Water Quality Controller Guide
User manual for the HACH SC4500 controller, detailing specifications, installation, operation, and maintenance for water quality monitoring in industrial and municipal applications.