Guide for switching from Hach Series 5000 to Hach 5500sc
1 Introduction
Twenty years after launching the Silica analyzer model 5000, Hach has introduced to the market a new analyzer with innovative solutions. The aim of this guide is to help you make the transition from the Series 5000 Analyzer to the new 5500sc Silica Analyzer.
2 Why switch from Hach 5000 to Hach 5500sc?
There are several reasons to switch from the Series 5000 to the 5500sc. Even if you are continuing to use the same measurement operations and principles, the new generation provides benefits that will prove advantageous.
Two analyzers are shown side-by-side: the Series 5000 Silica on the left and the Hach 5500sc on the right.
Series 5000 Silica | Hach 5500sc |
---|---|
2.9 liters reagents every 30 days 1 bottle/month @ 8.8 min cycle time |
Extended reagent replacement intervals and low reagent consumption 1 bottle (2-liter) every 90 days @ 15 min or 60 days @ 9.5 min |
Fixed measurement cycle | Configurable measurement cycle from continuous up to 240 mins (Interval Mode) |
2.9 L PET bottles | 2 L bottles with quick-connect, including pre-installed tubing and color coding - Reduces time maintaining the unit and reduced risk of contact with chemicals |
No predictive diagnostic tool | Predictive diagnostics (PROGNOSYS™) and built-in help function. Reduces risk of unexpected failures and unit downtime |
8-channel external sequencer | Optional built-in sequencer, up to 6-channels Lowers costs via sampling points and eliminates cost for an external sequencer (up to $ 6,000) |
Grab Sample In | Grab Sample In and Grab Sample Out Enables quick QC validation in addition to normal process monitoring |
Pressurized reagent delivery system NO PUMPS | Same principle with improved design. Built-in air compressor requires no external air supply |
Colorimetric analysis method at 810 nm. | Same method |
3 reagents and one standard | Same number, but different reagent concentrations |
3 Specification differences between Hach 5000 and Hach 5500sc
SPECIFICATION | Hach 5000 | Hach 5500sc |
---|---|---|
MEASUREMENT SPECIFICATIONS | ||
Measurement range: | 0.00 to 5000 µg/L as SiO2 | 0-5000 µg/L as SiO2 |
Accuracy (typical): | 0.00–500 µg/L: ± 1.0 µg/L or ± 5% of reading, (the larger value); 500–5000 µg/L: ± 7% of reading | 0–500 µg/L: ±1% or ±1 µg/L (the larger value), 500–5000 µg/L: ±5% |
Minimum Detection Limit: | 0.5 µg/L | 0.5 µg/L |
Precision/Repeatability: | ± 0.5 µg/L or ± 1.0% (the larger value) | ±0.5 ppb or ±1% (the larger value) |
Step Response Time (sample temperature dependent): | 8.8 minutes for 30 to 50 °C, 15 minutes for 5 to 40 °C, (field adjustable) | Typical 9.5 minutes from 5 to 50 °C |
Stabilization time | After initial startup or annual maintenance: 5 meas. cycles After standby: 1 meas. cycle After calibration: 0 meas. cycles |
After initial startup or annual maintenance: 5 meas. cycles After standby: 1 meas. cycle After calibration: 0 meas. cycles |
Calibration time | 15 min Tcal<30°C, 8 min Tcal>30°C | Slope calibration: 10 minutes |
Analyzer Sample Requirements | Regulated to 5 ± 3 psig (34.5 ± 20.7 kPa). Flow rate: 100 to 300 mL/minute. Temperature: 5 and 50 °C. | Accepted pressure: 2–87 psi (0.14 - 6 bar) in the provided preset regulators, pre-set to 2.0 psi. Acceptable flow rate: 55–300 mL/min Temperature: 5 to 50 °C (41 to 122 °F) |
A sample pressure control kit is provided | 1 | 1, 2, 4 or 6; programmable sequence |
No. of sample streams | yes | yes |
Grab sample in | no | yes |
Grab sample out | no | yes |
Calculations | No | Yes, between Si concentration of sample lines or external parameters |
LIQUID CONNECTIONS | ||
Sample line in | ½ inch push-to-connect fitting, ½ inch only; rigid or semi rigid plastic tubing | ½ inch or 6 mm OD push-to-connect fitting for plastic tubing |
Sample bypass drain | none | ½ inch or 6 mm OD push-to-connect fitting for plastic tubing |
Analyzed sample out drain fitting | ½ inch hose barb fitting ½-inch NPT PVC female | 11 mm (7/16 in.) ID slip-on fitting for soft plastic tubing |
Case drains | ½ inch hose fitting | 11 mm (7/16 in.) ID slip-on fitting for soft plastic tubing |
GASCONNECTIONS | ||
Air Purge inlet: instrument- quality air | 15-scfh (std cubic feet per hour) ½-inch OD stainless steel compression tubing fitting | 15 scfh (0.425 m3/hour), 6 mm OD push-to-connect fitting for plastic tubing |
Air Purge outlet | none | 11 mm (7/16 in.) ID slip-on fitting for soft plastic tubing, same tubing as analyzed sample out drain |
Reagent Pressure Inlet Fitting and Source | ½-inch OD stainless steel compression tubing fitting | none (internal source) |
ELECTRICALCONNECTIONS | ||
Power Requirements: | 115 or 230 VAC, 50/60 Hz, switch selectable; 52 VA, 32 W maximum | 100–240 VAC, 50/60 Hz universal power supply |
Analog/Recorder Outputs: | One, selectable for 0–0.01/0.1/1 V, or 4–20 mA (500 Ω maximum). | 4x; 4–20 mA outputs, Load impedance: 600 Ω max. Expandable to 8x or 12x with additional, 4-20ma SC1000 output module(s) |
Serial I/O: | RS232 and 20 mA current loop | Optional modules for flow, ModBus, |
Digital inputs | none | Profibus, *Hart, *Ethernet (*future, not yet availablet) |
4-20 mA input | none | Six isolated inputs configurable as TTL or open-collector/relay |
Alarms: | 4x programmable relays, Functions: 2 x sample concentration alarms, analyzer system warning, analyzer system shutdown alarms | Optional SC1000 4x 4-20 ma input module (2 modules Max) 4x programmable relays; Functions: concentration alarm, scheduler, analyzer warning/error or event (measuring ch.x, zero cal, slope cal, shutdown, startup, grab sample or end of measuring cycle ) |
Dual monitor | none | Connection for a remote SC controller which becomes master |
Smart probe | none | Connection for future external probe |
Power out | none | 2x max current, 250 VAC, AC:5A, DC:1.6A, always on |
REAGENTS | ||
Reagents: | 2.9 L Molybdate 3 (P/N 1995-03), 2.9 L Citric Acid/Surfactant(P/N 23470-03), 2.9 L Amino Acid F (P/N 23531-03), 2.9 L Silica Standard Solution, 500 µg/L (P/N. 21008-03) (250 mL required for standardization) |
2 L Molybdate 3, 6774802 2 L Citric Acid/Surfactant, 6774902 2 L Amino Acid F, L 6775102 2 L Standard Silica, 500 µg/L , 67750022 (1 bottle) of standard for every 10 calibrations |
Reagent consumption for a 15 minute cycle time | 5.1 L (1.8 bottle) of each reagent every 90 days | 2 L ( 1 bottle) of each reagent every 90 days (Measurement interval set to 15 min) |
GENERAL | ||
Enclosure: | Rating: NEMA 4x/IP65 Molded ABS plastic cabinet with gasketed doors (for indoor use) |
Rating: NEMA 4x/IP65 Material: PC/ABS case, PC door, PC hinges and latches, 316 SST hardware Indoor use only. |
Mounting | Bench top or panel mounting only | Wall, panel or table |
Dimensions | W x D x H mm: 564 x 419 x 857 inches : 22 x 17 x 34 |
W x D x H mm : 452 x 360 x 804 inches : 17.8 x 14.2 x 31.7 An adapter plate Model 6791000 is available to mount the Hach 5500sc on previous Hach 5000 panel location |
Shipping Weight | 36.7 kg (81 lb) without reagents and standards | 20 kg (45 lb) without reagents and standards, 36.3 kg (80 lb) with reagents |
Ambient Operating Conditions: | 10 to 45 °C, 5 to 95% non-condensing humidity. Suitable for general purpose, clean, indoor environments (not suitable for outdoor use). |
5 to 45 °C (41 to 113 °F) 5 to 95% non-condensing humidity Storage temperature: −20 to 60 °C (−4 to 140 °F) (Must not be in direct sunlight) |
Screen | Alphanumeric LCD display, membrane keyboard | Display, LCD 5.7" color VGA (640x480), membrane keyboard |
4 Reference documents
Below are shown the documents required to perform a quick, safe and correct replacement of analyzers. Each document can be downloaded directly from the web by clicking the description.
Code | Description and link | Status |
---|---|---|
Catalog 60000-18 | Series 5000 Silica Analyzer, Instrument Manual | old |
DOC013.99.80240 | 5500sc SiO2 Analyzer Unpacking Guide Illustrated | new |
DOC273.99.80258 | Illustrated user instructions to mount the analyzer on a wall, on panel or on table. | new |
DOC273.99.80259 | Adapter Plate User Instructions for Polymetron Analyzer. | new |
DOC023.97.80240 | Installation Manual 5500sc SiO2 | new |
DOC023.97.80267 | Operations 5500sc SiO2 | new |
5 Step 1- Removing the Hach 5000 analyzer
5.1 Save the analyzer setup
Perform the following actions:
- Perform an analyzer printout: (SETUP/PRINT SETUP/ENTER)
- Back up your Hach 5000 analyzer setup, including:
- High/Low Alarm values
- Alarm relay identification
- Recorder max concentration
- 4-20mA output ranges
- Digital com settings
- Disconnect the analyzer from its power source.
A sample analyzer setup printout is displayed, showing alarm values, recorder settings, and date/time.
5.2 Remove bottles
- Remove the 3 reagent bottles.
- Follow regulations in place for the reagents' elimination/destruction.
- Remove the silica standard solution.
- You can keep the solution, but discard the bottle – the 5500 has different bottle types.
Two reagent bottles and a bottle of silica standard solution are shown, indicating they need to be removed.
5.3 Remove all electrical connections
- 115/230 VAC power line
- High voltage relays
- Low voltage relays
- Recorder output
- RS 232 connection
A diagram illustrates the electrical connections of the Hach 5000 analyzer, showing power, relays, recorder output, and RS232 connections with corresponding wire colors.
5.4 Remove all hydraulic connections
- Sample in/out
- Cabinet/Sample Drain
- Air purge
- The sample heater, Model 48500, can be removed as long as the sample temperature is above 5°C (41° F).
- The installed sample pressure conditioning kit (Model 45983-00) can be removed – the 5500 uses a different kit.
A diagram shows the hydraulic connections of the Hach 5000 analyzer, labeling sample inlet/outlet, drains, air purge, and indicating components like the sample heater and pressure conditioning kit.
5.5 Remove the Hach 5000 analyzer unit
- Remove the six screws.
- Remove the analyzer, being careful of its weight: 37kg (81 lb).
The Hach 5000 analyzer unit is shown, highlighting the need to remove six screws for removal.
6 Step 2 - Installing the Hach 5500sc
6.1 Wall mount installation
Follow instructions 1B to 4B on document: Illustrated user instructions to mount the analyzer on a wall, on panel or on table.
An illustration depicts the Hach 5500sc analyzer being mounted on a wall.
6.2 Panel mount installation
- Install the optional adapter plate Model 6791000 with the six screws provided.
- Follow instructions on document: Adapter Plate User Instructions for Polymetron Analyzer.
An illustration shows the optional adapter plate being installed for panel mounting the Hach 5500sc analyzer.
Follow instructions 1C to 4C on document: Illustrated user instructions to mount the analyzer on a wall, on panel or on table.
6.3 Panel mount
Follow instructions 1A on document: Illustrated user instructions to mount the analyzer on a wall, on panel or on table.
An illustration shows the Hach 5500sc analyzer mounted on a panel.
7 Connecting the Hach 5500sc hydraulics
7.1 Identify hydraulic connections
Identify the hydraulic inlet/outlet connections for one to six sample streams.
A diagram identifies the hydraulic inlet and outlet connections for up to six sample streams on the Hach 5500sc.
7.2 Connect samples and drain lines
- Use the supplied tubing (6 mm), Y-strainer with filter and pressure regulator to plumb the drain and the sample to the analyzer.
- The sample line tubing that goes into the plumbing access ports must be 6 mm.
- Tubing of 1/4 in. may be used for the sample line up to the valve and y-strainer, but not into the analyzer's plumbing access ports.
- See DOC023.97.80240 for more information.
A diagram illustrates the connection of sample and drain lines using supplied tubing, a Y-strainer with filter, and a pressure regulator.
7.3 Connect sample stream
Install each sample line into the center of a larger process pipe to minimize interference from air bubbles or bottom sediment. See figure on left.
An illustration shows how to connect the sample stream into a larger process pipe to minimize interference.
7.4 Set the bypass flow rate
- The bypass flow can be adjusted when the analyzer is in shutdown mode. Adjust the flow rate of the sample bypass line with the flow valve as shown at left.
- Use an external meter to measure the flowrate of the sample bypass line.
- For faster response to changes in the process stream, increase the sample bypass line flow rate when the process stream is far from the analyzer.
A diagram shows how to adjust the bypass flow rate using a flow valve and suggests using an external meter.
8 Connecting the Hach 5500sc electrical connections
Follow procedure on pages 11-17 of the operation manual.
9 Installing analyzer bottles and stir bar on the Hach 5500sc
Follow procedure on pages 18-19 of the operation manual.
10 Training operators and supervisors on the new Hach 5500sc
The following table shows the topics to be presented to train the plant operators and supervisors.
Topics | Operation Manual Page |
---|---|
Plant operators – Standard use of the analyzer | |
User interface and navigation | 5 |
Grab sample measurement | 9 |
Getting a grab sample from the analyzer | 9 |
View data | 14 |
View instrument information | 16 |
Using SD card's | 16 |
Instrumentation supervisor – startup and configuration | |
Reagent preparation | 7 |
Analyzer setup | 8, 10 |
Sequencer configuration | 8 |
Reagents and standards setup | 9 |
Calculation setup | 11 |
4 – 20 mA setup | 11 |
PID setup (optional) | 12 |
Relay setup | 12 |
Relay activation options | 13 |
Updating the firmware | 16 |
Analyzer calibration | 17 |