Hach NA5600sc Online Sodium Analyser
1-channel, with Cation Kit & Autocalibration, panel mount
Product Overview
The Hach NA5600sc Online Sodium Analyser provides accurate, low-level sodium measurements and predictive diagnostics, ensuring uptime and optimized operation. Its automatic electrode reactivation uses non-hazardous chemicals, eliminating the need for manual reactivation or electrode etching.
Key Features
- Space-Saving Design: Features a smaller instrument footprint and streamlined layout for easy integration into existing or new sites.
- Low Maintenance: Requires reagent replenishment only every 90 days and annual replacement of reagent tubing and the sodium electrode. Clear step-by-step instructions simplify maintenance.
- Avoid Downtime: Predictive diagnostic tools, including Hach's proprietary Prognosys technology, warning LEDs, and high visibility notification screens help prevent unplanned downtime.
Specifications
Accuracy: 0.01 ppb - 40 ppb: ±2 ppb; 40 ppb - 200 ppm: ±5%
Acidity: < 250 ppm
Ambient temperature: 5 - 50 °C
Analogue outputs: 6 isolated, 0 - 20 mA or 4 - 20 mA; load impedance: 600 Ohm maximum
Connection: 0.644 - 1.29 mm² (24 - 16 AWG) wire; 0.644 - 0.812 mm² (24 - 20 AWG) recommended, twisted pair shielded wire
Calibration method: Automatic with known addition; Manual: 1 or 2 points
Detection limit: 0.01 ppb
Dimensions: 681 mm x 452 mm x 254 mm (H x W x D)
Display: Coloured 5.7" LCD
Electrode type: Sodium ISE (ion specific electrode) electrode and reference electrode with KCl electrolyte
Flow rate: 100 - 150 mL/min (6 - 9 L/h)
Fuse: Input power: T 1.6 A, 250 VAC; Relays: T 5.0 A, 250 V
Include Autocalibration?: Yes
Include Cation Kit?: Yes
Inlet: Sample line and sample bypass drain: 6 mm O.D. push-to-connect fitting for plastic tubing
Interference: Chemical and case drains: 7/16 inch I.D. slip-on fitting for soft plastic tubing
Load: 600 Ohm
Maintenance interval: Every 90 days: refill electrolyte, reactivation, conditioning, and calibration solution
Material: Polyol case, PC door, PC hinges and latches, 304/316 SST hardware
Max. concentration of suspended solids in sample: < 2 NTU, no oil, no grease (For boiler sample type install approx. 100 µm filter)
Measuring range: 0.01 ppb - 200 ppm
Mounting: Panel mount
Number of analog outputs: 6
Number of channels: 1
Number of relays: 6
Options: Analyser with Cation Kit and Autocalibration
Parameter: Sodium
pH Range: 2 - 10 pH
Pollution degree: 2
Power requirements (Hz): 50/60 Hz
Power requirements (Voltage): 100 - 240 V AC
Protection rating: IP65, PCBA housing
Relative humidity: 10 - 80%, non-condensing
Relay output: 6; type: not powered SPDT relays, each rated at 5 A resistive, 240 VAC maximum
Connection: 1.0 - 1.29 mm² (18 - 16 AWG) wire; 1.0 mm² (18 AWG) stranded recommended, 5 - 8 mm O.D. cable
Repeatability: < 0.02 ppb or 1.5% reading (whichever is greater) within ± 10 °C variation
Response time: From 0.1 ppb to 10 ppb: T90 ≤ 3 minutes, T95 ≤ 4 minutes; From < 1 ppb to 100 ppb: T90 < 2 minutes, T95 < 3 minutes (about 150 s)
Sample conditioner: DIPA (1 L/month) at 25 °C for a sample pH target of 10.5
Sample pressure: 0.2 - 6 bar
Sample temperature: 5 - 45 °C
Storage conditions: -20 - 60 °C
Weight: 14 kg with empty bottles
What's Included?
Hach NA5600sc Sodium Analyser, 1 channel, with 1 channel installation kit and user manuals, reference electrode sodium, sodium ion selective electrode, empty DIPA bottle, the cover for DIPA bottle and the tray for DIPA bottle.
Required Accessories
- Prognosys NA5600sc License Kit (Item 8428000)
Related Documents
![]() |
HACH BioTector B7000i Online TOC Analyzer User Manual | Installation, Operation & Maintenance Guide Comprehensive user manual for the HACH BioTector B7000i Online Total Organic Carbon (TOC) Analyzer. Learn about safety, installation, operation, calibration, maintenance, and technical specifications for accurate water quality monitoring. |
![]() |
Hach CL17sc Calibration Verification and Adjustment Recommendations Guidance on verifying and adjusting the calibration of the Hach CL17sc analyzer, emphasizing factory calibration and recommended verification procedures. |
![]() |
Hach Series 5000 to 5500sc Silica Analyzer Replacement Guide This guide provides detailed instructions and specifications for replacing the Hach Series 5000 Silica Analyzer with the newer Hach 5500sc model, covering differences, removal, installation, and connection procedures. |
![]() |
HACH SL1000 Portable Parallel Analyzer: Basic User Manual Get started with the HACH SL1000 Portable Parallel Analyzer. This basic user manual covers essential information on specifications, installation, operation, maintenance, and troubleshooting for accurate water quality analysis. |
![]() |
HACH CL17sc Analyzer User Manual - Chlorine Concentration Measurement User manual for the HACH CL17sc analyzer, detailing specifications, installation, operation, maintenance, and troubleshooting for measuring total chlorine concentration in water. Connects to SC Controller. |
![]() |
Hach A1000, A1000XP User Manual Comprehensive user manual for Hach A1000 and A1000XP TOC analyzers, covering specifications, installation, operation, maintenance, and troubleshooting. |
![]() |
Hach Replacement Oxygen Electrode (ppb) - Product 09182=A=1000 Information on the Hach Replacement Oxygen Electrode (ppb), without probe body. This electrode is a replacement sensor for the Polymetron 9582sc Dissolved Oxygen analyzer. Specifications include material, special notes, temperature sensor, and warranty. |
![]() |
Hach HQ2200 Portable Multi-Meter: pH and Conductivity Analysis Detailed overview of the Hach HQ2200 portable multi-meter, featuring rugged field pH and conductivity electrodes. Learn about its robust design, intuitive operation, Intellical sensor compatibility, and specifications for water quality professionals. |