Hach NA5600sc Online Sodium Analyser

1-channel, with Cation Kit & Autocalibration, panel mount

Product Overview

The Hach NA5600sc Online Sodium Analyser provides accurate, low-level sodium measurements and predictive diagnostics, ensuring uptime and optimized operation. Its automatic electrode reactivation uses non-hazardous chemicals, eliminating the need for manual reactivation or electrode etching.

Key Features

Specifications

Accuracy: 0.01 ppb - 40 ppb: ±2 ppb; 40 ppb - 200 ppm: ±5%

Acidity: < 250 ppm

Ambient temperature: 5 - 50 °C

Analogue outputs: 6 isolated, 0 - 20 mA or 4 - 20 mA; load impedance: 600 Ohm maximum

Connection: 0.644 - 1.29 mm² (24 - 16 AWG) wire; 0.644 - 0.812 mm² (24 - 20 AWG) recommended, twisted pair shielded wire

Calibration method: Automatic with known addition; Manual: 1 or 2 points

Detection limit: 0.01 ppb

Dimensions: 681 mm x 452 mm x 254 mm (H x W x D)

Display: Coloured 5.7" LCD

Electrode type: Sodium ISE (ion specific electrode) electrode and reference electrode with KCl electrolyte

Flow rate: 100 - 150 mL/min (6 - 9 L/h)

Fuse: Input power: T 1.6 A, 250 VAC; Relays: T 5.0 A, 250 V

Include Autocalibration?: Yes

Include Cation Kit?: Yes

Inlet: Sample line and sample bypass drain: 6 mm O.D. push-to-connect fitting for plastic tubing

Interference: Chemical and case drains: 7/16 inch I.D. slip-on fitting for soft plastic tubing

Load: 600 Ohm

Maintenance interval: Every 90 days: refill electrolyte, reactivation, conditioning, and calibration solution

Material: Polyol case, PC door, PC hinges and latches, 304/316 SST hardware

Max. concentration of suspended solids in sample: < 2 NTU, no oil, no grease (For boiler sample type install approx. 100 µm filter)

Measuring range: 0.01 ppb - 200 ppm

Mounting: Panel mount

Number of analog outputs: 6

Number of channels: 1

Number of relays: 6

Options: Analyser with Cation Kit and Autocalibration

Parameter: Sodium

pH Range: 2 - 10 pH

Pollution degree: 2

Power requirements (Hz): 50/60 Hz

Power requirements (Voltage): 100 - 240 V AC

Protection rating: IP65, PCBA housing

Relative humidity: 10 - 80%, non-condensing

Relay output: 6; type: not powered SPDT relays, each rated at 5 A resistive, 240 VAC maximum

Connection: 1.0 - 1.29 mm² (18 - 16 AWG) wire; 1.0 mm² (18 AWG) stranded recommended, 5 - 8 mm O.D. cable

Repeatability: < 0.02 ppb or 1.5% reading (whichever is greater) within ± 10 °C variation

Response time: From 0.1 ppb to 10 ppb: T90 ≤ 3 minutes, T95 ≤ 4 minutes; From < 1 ppb to 100 ppb: T90 < 2 minutes, T95 < 3 minutes (about 150 s)

Sample conditioner: DIPA (1 L/month) at 25 °C for a sample pH target of 10.5

Sample pressure: 0.2 - 6 bar

Sample temperature: 5 - 45 °C

Storage conditions: -20 - 60 °C

Weight: 14 kg with empty bottles

What's Included?

Hach NA5600sc Sodium Analyser, 1 channel, with 1 channel installation kit and user manuals, reference electrode sodium, sodium ion selective electrode, empty DIPA bottle, the cover for DIPA bottle and the tray for DIPA bottle.

Required Accessories

Models: NA5600sc, NA5600sc Online Sodium Analyser, NA5600sc, Online Sodium Analyser, Sodium Analyser, Analyser

PDF preview unavailable. Download the PDF instead.

LXV526.98.2311A Apache FOP Version 1.0 Apache FOP Version 1.0

Related Documents

Preview HACH BioTector B7000i Online TOC Analyzer User Manual | Installation, Operation & Maintenance Guide
Comprehensive user manual for the HACH BioTector B7000i Online Total Organic Carbon (TOC) Analyzer. Learn about safety, installation, operation, calibration, maintenance, and technical specifications for accurate water quality monitoring.
Preview Hach CL17sc Calibration Verification and Adjustment Recommendations
Guidance on verifying and adjusting the calibration of the Hach CL17sc analyzer, emphasizing factory calibration and recommended verification procedures.
Preview Hach Series 5000 to 5500sc Silica Analyzer Replacement Guide
This guide provides detailed instructions and specifications for replacing the Hach Series 5000 Silica Analyzer with the newer Hach 5500sc model, covering differences, removal, installation, and connection procedures.
Preview HACH SL1000 Portable Parallel Analyzer: Basic User Manual
Get started with the HACH SL1000 Portable Parallel Analyzer. This basic user manual covers essential information on specifications, installation, operation, maintenance, and troubleshooting for accurate water quality analysis.
Preview HACH CL17sc Analyzer User Manual - Chlorine Concentration Measurement
User manual for the HACH CL17sc analyzer, detailing specifications, installation, operation, maintenance, and troubleshooting for measuring total chlorine concentration in water. Connects to SC Controller.
Preview Hach A1000, A1000XP User Manual
Comprehensive user manual for Hach A1000 and A1000XP TOC analyzers, covering specifications, installation, operation, maintenance, and troubleshooting.
Preview Hach Replacement Oxygen Electrode (ppb) - Product 09182=A=1000
Information on the Hach Replacement Oxygen Electrode (ppb), without probe body. This electrode is a replacement sensor for the Polymetron 9582sc Dissolved Oxygen analyzer. Specifications include material, special notes, temperature sensor, and warranty.
Preview Hach HQ2200 Portable Multi-Meter: pH and Conductivity Analysis
Detailed overview of the Hach HQ2200 portable multi-meter, featuring rugged field pH and conductivity electrodes. Learn about its robust design, intuitive operation, Intellical sensor compatibility, and specifications for water quality professionals.