Lovibond Fluoride L / M170: SPADNS Method for Water Analysis

Fluoride L / M170

SPADNS Method for Fluoride Determination

Instrument Specific Information

The test can be performed on the following instruments. The required cuvette and the absorption range of the photometer are also indicated.

Instruments Cuvette Wavelength (λ) Measurement Range
MD 100, MD 600, MD 610, MD 640, MultiDirect, Spectro-Direct, XD 7000, XD 7500 Ø 24 mm 580 nm 0.05 - 2 mg/L F

Reagents

Required materials (partially optional):

Reagent Packaging Unit Order Number
SPADNS Reagent Solution 250 ml 250 mL 467481
SPADNS Reagent Solution 500 ml 500 mL 467482
Calibration Standard Fluoride 30 mL 205630

Application Areas

  • Drinking water treatment
  • Purification of polluted water

Preparation

  1. Before measurement, perform user calibration (refer to photometer instructions).
  2. Use the same SPADNS reagent solution batch for user calibration and sampling (refer to photometer instructions). The instrument must be calibrated for each new batch of SPADNS reagent solution (see Standard Methods 20th, 1991, APHA, AWWA, WEF 4500 F D., S. 4-82).
  3. For user calibration and measurement, the zero adjustment and the test must be performed using the same cuvette, as cuvettes have small tolerances relative to each other.
  4. The calibration solutions and water samples to be measured must have the same temperature (± 1 °C).
  5. The analysis result depends primarily on the exact sample and reagent volume. Use only a 10 ml or 2 ml graduated pipette (Class A) to dispense the volume of the sample and reagent.
  6. Seawater and wastewater samples must be distilled.
  7. It is recommended to use special cuvettes with a larger fill volume.

Execution of Fluoride Determination with Liquid Reagent

Select the method in the instrument. For the following instruments, a zero measurement is not required each time: XD 7000, XD 7500. Observe the notes!

Step 1: Sample Preparation

Fill a 24 mm cuvette with exactly 10 mL of sample.

Close the cuvettes. [close-icon]

Place the sample cuvette in the measuring chamber. Pay attention to positioning. [position-icon]

Step 2: Zero Adjustment

Press the ZERO button. [zero-button-icon]

Remove the cuvette from the measuring chamber.

For instruments that do not require a zero measurement, start here.

Step 3: Reagent Addition

Fill a 24 mm cuvette with exactly 2 mL of SPADNS Reagent Solution. Note: Fill the cuvette to the brim! [fill-icon]

Close the cuvettes. [close-icon]

Step 4: Mixing and Measurement

Mix the contents by swirling. [swirl-icon]

Place the sample cuvette in the measuring chamber. Pay attention to positioning. [position-icon]

Press the TEST (XD: START) button. [test-button-icon]

The display shows the result in mg/L Fluoride.

Chemical Method

SPADNS

Appendix

Calibration Function for Third-Party Photometers

Conc. = a + b•Abs + c•Abs² + d•Abs³ + e•Abs⁴ + f•Abs⁵

Coefficient Ø 24 mm 10 mm
a 8.44253 10+0 8.44253 10+0
b -1.41844 10+1 -3.04965 10+1
c 9.24803 10+0 4.2749 10+1
d -2.3046 10+0 -2.2904 10+1
e
f

Interferences

Permanent Interferences

1. Accuracy decreases above 1.2 mg/L fluoride. Although the results are sufficiently accurate for most applications, better accuracy can be achieved by diluting the sample 1:1 before use and multiplying the result by 2.

Interferences

Interference Interferes from
Cl₂ 5 mg/L

Literature References

Standard Methods 20th, 1992, APHA, AWWA, WEF 4500 F D, S. 4-82

Conformity

US EPA 13A
APHA Method 4500 F D

PDF preview unavailable. Download the PDF instead.

Gebruiksaanwijzing Enkele methode - M170 - Fluoride L - nl Apache FOP Version 2.5 Apache FOP Version 2.5

Related Documents

Preview Iron LR L (A) / M225: Method for Determining Iron Concentration
This document outlines the method for determining Iron LR L (A) using the Ferrozine / Thioglycolate chemical method with the M225 instrument. It includes instrument specific information, required materials, application lists, preparation steps, digestion procedures, and interference guidelines.
Preview Nitrite HR PP / M273 - Ferrous Sulfate Method
This document outlines the procedure for determining Nitrite HR PP using the Ferrous Sulfate Method with the M273 instrument. It includes instrument-specific information, reagents, application areas, step-by-step instructions, and evaluation data.
Preview Lovibond Scuba3s Smart Pooltester User Manual
Comprehensive user manual for the Lovibond Scuba3s Smart Pooltester, detailing its features, operation, maintenance, and technical specifications for pool water analysis.
Preview Lovibond Scuba3s Pooltester: Smart Pool Water Control
Discover the Lovibond Scuba3s, an advanced electronic pool tester for effortless pool water control. Features smart safety, wireless data transfer via NFC, and mobile analysis management with the PoolAssistant app.
Preview Lovibond MD200 Photometer: Bedienungsanleitung und technische Informationen
Umfassende Bedienungsanleitung für das Lovibond MD200 Photometer zur Wassertestung, einschließlich Funktionen, Methoden (wie COD Vario), Kalibrierung und technischer Daten in mehreren Sprachen.
Preview Lovibond Scuba III Smart Pooltester - User Manual and Technical Specifications
Comprehensive guide to the Lovibond Scuba III Smart Pooltester, covering its features, operation, maintenance, technical data, and accessories for pool water quality analysis.
Preview Lovibond PFXi-880/P Spectrophotometer with Heating - Technical Overview
Explore the Lovibond PFXi-880/P spectrophotometer, a precision instrument for color measurement. Features include RCMSi remote calibration, extended optical path, optional heating, and a 16-wavelength filter system. Ideal for industrial applications in petroleum, food, pharmaceuticals, and chemicals.