Gilson PIPETMAN G Quick Guide

Product: PIPETMAN G Pipette

Language: English (EN)

Introduction

PIPETMAN® G is a fully adjustable air-displacement pipette designed for use with disposable tips. It builds upon the renowned accuracy, precision, and reliability of the PIPETMAN series, while significantly reducing pipetting and purge forces for enhanced user comfort during prolonged use.

The range includes eight single-channel models covering volumes from 0.2 μL to 10 mL, and eight multichannel models covering volumes from 1 μL to 300 μL.

? PIPETMAN G is identified by a visual blue ring.

Parts Checklist

Single Models

Multichannel Models

Good Laboratory Practice (GLP) Compliance

The serial number, engraved on the pipette body, provides unique identification and indicates the date of manufacture. For detailed information, consult the PIPETMAN G User's Guide LT801122.

Example Serial Number: AA10369

The barcode on the box and the Certificate of Conformity ensure the traceability of your pipette.

Figure 1: Serial Number Location

An illustration shows a pipette with a highlighted area indicating where the serial number is engraved on the body.

Description

Figure 2: PIPETMAN G multichannel and single models

Diagram illustrating the parts of the PIPETMAN G pipette. Key components include:

  • Color-coded push button
  • Tip ejector button
  • Thumbwheel
  • Blue ring
  • Body or handle
  • Clip tip ejector
  • Connecting nut
  • Ejector support
  • Ejector lock
  • Cover
  • Tip ejector
  • Tip holder
  • Ejector spacer
  • PIPETMAN® DIAMOND tip

Specifications

PIPETMAN G is a high-quality pipette offering excellent accuracy and precision. The performance data presented in the "Maximum Permissible Errors" table were obtained using genuine PIPETMAN® DIAMOND Tips, and these values are guaranteed only when such tips are used.

? With a precise pipetting technique, the P2G model can aspirate volumes as low as 0.1 μL, and the P10G model as low as 0.5 μL.

Maximum Permissible Errors

The following table details the maximum permissible errors, conforming to the ISO 8655-2 standard.

VOLUME (μL) GILSON ISO 8655
SYSTEMATIC ERROR (μL) RANDOM ERROR (μL) SYSTEMATIC ERROR (μL) RANDOM ERROR (μL)
PIPETMAN G SINGLE CHANNEL
P2G (P/N F144054M) WITH D10 OR DL10 TIPS
0.2± 0.024≤ 0.012± 0.08≤ 0.04
0.5± 0.025≤ 0.012± 0.08≤ 0.04
1± 0.025≤ 0.012± 0.08≤ 0.04
2± 0.030≤ 0.014± 0.08≤ 0.04
P10G (P/N F144055M) WITH D10 OR DL10 TIPS
1± 0.025≤ 0.012± 0.12≤ 0.08
5± 0.075≤ 0.030± 0.12≤ 0.08
10± 0.100≤ 0.040± 0.12≤ 0.08
P20G (P/N F144056M) WITH D200 TIPS
2± 0.10≤ 0.030± 0.2≤ 0.1
10± 0.10≤ 0.050± 0.2≤ 0.1
20± 0.20≤ 0.060± 0.2≤ 0.1
P100G (P/N F144057M) WITH D200 TIPS
10± 0.35≤ 0.10± 0.8≤ 0.3
50± 0.40≤ 0.12± 0.8≤ 0.3
100± 0.80≤ 0.15± 0.8≤ 0.3
P200G (P/N F144058M) WITH D200 TIPS
20± 0.50≤ 0.20± 1.6≤ 0.6
100± 0.80≤ 0.25± 1.6≤ 0.6
200± 1.60≤ 0.30± 1.6≤ 0.6
P1000G (P/N F144059M) WITH D1000 TIPS
100± 3.0≤ 0.6± 8.0≤ 3.0
500± 4.0≤ 1.0± 8.0≤ 3.0
1000± 8.0≤ 1.5± 8.0≤ 3.0
P5000G* (P/N F144066) WITH D5000 TIPS
500± 12≤ 3± 40≤ 15.0
2500± 15≤ 5± 40≤ 15.0
5000± 30≤ 8± 40≤ 15.0
P10mLG* (P/N F144067) WITH D10ML TIPS
1000± 30≤ 6± 60≤ 30.0
5000± 40≤ 10± 60≤ 30.0
10000± 60≤ 16± 60≤ 30.0
PIPETMAN G MULTICHANNEL
P8X10G (P/N F144068) P12X10G (P/N F144069) WITH D10 AND DL10 TIPS
1± 0.08≤ 0.05± 0.24≤ 0.16
5± 0.20≤ 0.10± 0.24≤ 0.16
10± 0.20≤ 0.10± 0.24≤ 0.16
P8X20G (P/N F144070) P12X20G (P/N F144071) WITH DL10 TIPS
2± 0.10≤ 0.08± 0.4≤ 0.20
10± 0.20≤ 0.10± 0.4≤ 0.20
20± 0.40≤ 0.20± 0.4≤ 0.20
P8X200G (P/N F144072) P12X200G (P/N F144073) WITH D200 TIPS
20± 0.50≤ 0.25± 3.2≤ 1.2
100± 1.00≤ 0.40± 3.2≤ 1.2
200± 2.00≤ 0.50± 3.2≤ 1.2
P8X300G (P/N F144074) P12X300G (P/N F144075) WITH D300 TIPS
30± 1.00≤ 0.35± 8.0≤ 3.0
150± 1.50≤ 0.60± 8.0≤ 3.0
300± 3.00≤ 1.00± 8.0≤ 3.0

* P5000G and P10mLG models do not have tip ejectors.

Setting the Volume

The volume of liquid to be aspirated is set using the volumeter. The dials are colored either black or red to indicate the position of the decimal point, depending on the model (refer to Figure 3).

The volume is set by turning the thumbwheel or the push button (refer to Figure 4). The push button offers a faster and easier way to set volumes, especially when wearing gloves. The thumbwheel allows for slow, precise adjustments using only one hand.

To achieve maximum accuracy when setting the volume:

Figure 3: Dial colors by model

Illustrations show the dial configurations for various PIPETMAN G models (P2G, P10G, P20G, P100G, P200G, P1000G, P5000G, P10mLG, Px10G, Px20G, Px200G, Px300G) indicating volume ranges and decimal point positions (e.g., 1.25 μL, 7.5 μL, 0.75 mL).

Figure 4: Location of push button and thumbwheel

A diagram shows the pipette body with the push button and thumbwheel clearly indicated.

MODEL COLOR OF VOLUMETER NUMBERS (BLACK) COLOR OF VOLUMETER NUMBERS (RED) INCREMENT
P2GμL0.01 μL0.002 μL
P10G - P20G - Px10G - Px20GμL0.1 μL0.02 μL
P100G - P200G - Px200G - Px300GμL0.2 μL0.2 μL
P1000G0.01 mLmL0.002 mL
P5000G0.01 mLmL0.002 mL
P10mLGmL0.1 mL0.02 mL

Fitting the Tips

To attach a new PIPETMAN DIAMOND Tip, push the tip holder into the tip with a slight twisting motion to ensure a firm, airtight seal.

The PIPETMAN G is equipped with a stainless steel tip ejector and can also be used with a plastic tip ejector, available as an accessory. Refer to the PIPETMAN G User's Guide LT801122 for accessories.

Fitting the Tips on P2G and P10G Models

ℹ️ For the P2G and P10G models, a dual-position adapter (plastic) is required to fit DL10 tips (long tips) or D10 tips (short tips). The metallic rod of the tip ejector is shaped to allow the adapter to be clipped in either position.

P2G and P10G models are supplied with the adapter pre-installed, ready for DL10 tips. If D10 tips are to be used, the adapter must be repositioned in the shorter slot following these steps:

  1. Pull the adapter down from the metallic rod.
  2. Turn the adapter 180°.
  3. Refit the adapter so the end of the metallic rod engages the shorter slot of the adapter.
Figure 5: Dual-position adapter for P2G and P10G with stainless steel tip ejector

An illustration shows the dual-position adapter. Text labels indicate "Short tips" and "Long tips" orientations for the adapter.

ℹ️ Plastic tip ejectors can be used and ordered separately (refer to PIPETMAN G User's Guide LT801122).

Ejector Spacer for PIPETMAN G 10 μL Multichannel Models

Depending on the tip used (D10 or DL10), you may need to exchange the ejector spacer. The broadly spaced one is for D10 tips, and the narrowly spaced one is for DL10 tips.

  1. Remove the tip ejector, keeping both ejector locks depressed. Pull the tip ejector down.
  2. Gently press the tabs from the ejector spacer and remove it from the tip ejector.
  3. Insert the alternative ejector spacer and click it to the tip ejector.
  4. To refit the tip ejector, gently re-insert it vertically into the rails of the ejector support.
Figure 6: Ejector spacer location

Diagram shows the tip ejector and two types of ejector spacers: one labeled "Ejector spacer D10" (broadly spaced) and another for "Long tips" (narrowly spaced).

PIPETMAN DIAMOND Tips are best fitted using the ROCKY RACK technique, invented by Gilson, available exclusively in their TIPACK and TOWERPACK products.

PIPETMAN DIAMOND Tip Compatibility for Multichannel Models
P8x10G, P12x10GD10*, DL10, DF10, DFL10
P8x20G, P12x20GDL10, DFL10, DF30, D200
P8x200G, P12x200GD200, D300, DF200, DF300
P8x300G, P12x300GD200, D300, DF200, DF300

* Using the broad ejector spacer D10, you can adapt a D10 tip.

Figure 7: ROCKY RACK technique

An illustration depicting the ROCKY RACK technique for fitting tips.

Pre-Wetting the Tips

Pre-wetting the tips before pipetting helps prepare them for optimal performance. Ideally, this process involves immersing the tip in the liquid and performing one pipetting step.

Pre-wetting ensures that the volumes you pipette will achieve the specified accuracy and precision.

Aspirating

  1. Press the push button to the first stop (this corresponds to the set volume of liquid).
  2. Hold the pipette vertically and immerse the tip in the liquid (refer to PIPETMAN G User's Guide LT801122).
  3. Release the push button slowly and smoothly (to the top position) to aspirate the set volume of liquid.
  4. Wait one second, and then withdraw the pipette tip from the liquid. You may wipe any droplets away from the outside of the tip using a medical wipe; however, if you do so, take care to avoid touching the tip's orifice.
Figure 1: Pipetting motion - aspirate and dispense

Diagram shows a multichannel pipette. Arrows indicate the motion for aspirating (push button to first stop, then release) and dispensing (push button to second stop). Labels indicate "Rest position", "First stop", and "Second stop".

Dispensing

  1. Place the end of the tip against the inside wall of the recipient vessel (at an angle of 10° to 40°).
  2. Press the push button slowly and smoothly to the first stop.
  3. Wait for at least one second, then press the push button to the second stop to expel any residual liquid from the tip. Keep the push button pressed fully down and (while removing the pipette) draw the tip along the inside surface of the vessel.
  4. Release the push button smoothly. Eject the tip by pressing firmly on the tip ejector button.
? It is recommended to regularly verify the volume displayed on the dial while pipetting.

Maintenance

Routine maintenance helps keep your pipette in good condition, ensuring a continued high level of performance. Maintenance is limited to cleaning or autoclaving the parts specified in the PIPETMAN G User's Guide LT801122, or to replacing the push button, connecting nut, tip ejector, tip holder, seal, and O-ring.

ℹ️ PIPETMAN P2G and P10G models should not be disassembled. You may only replace the push button, tip ejector, and its dual-position adapter. With these pipettes, if the tip holder is damaged, the piston may also be damaged. After replacing any parts, you should verify the performance of your pipette following the verification procedure available on the Gilson website (www.gilson.com). If the pipette needs to be readjusted, please contact your local authorized Gilson Service Center.
ℹ️ Decreased pipetting forces are due to the design of the piston assembly, which includes a very high-quality lubricant (part number 5440011070). The use of any other lubricant voids the warranty of this pipette.

Warranty

Gilson warrants this pipette against defects in material under normal use and service for a period of three years for both single channel and multichannel models from the date of purchase.

This warranty shall not apply to pipettes subjected to abnormal use and/or improper or inadequate maintenance (contrary to the recommendations in the user's guide), including, but not limited to, physical damage, improper handling, spillage, or exposure to corrosive environments. The warranty is void if pipettes are altered or modified by any party other than Gilson or its designates. Gilson's sole liability under this warranty is limited to, at Gilson's sole option, repair or replacement of defective components, or refund of the purchase price. Routine cleaning, control, and recalibration are not covered. The replacement of wearing parts such as seals, O-rings, broken piston assemblies, and broken tip holders are also not covered.

ℹ️ Yearly routine maintenance is highly recommended to keep your pipette in good condition, ensuring a continued high level of performance.

THE FOREGOING WARRANTY IS EXCLUSIVE AND GILSON HEREBY DISCLAIMS ALL OTHER WARRANTIES, WHETHER EXPRESS OR IMPLIED, INCLUDING ANY WARRANTIES OF MERCHANTABILITY AND ANY WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE. UNDER NO CIRCUMSTANCES SHALL GILSON BE LIABLE FOR ANY CONSEQUENTIAL, PUNITIVE, INDIRECT OR INCIDENTAL DAMAGES ARISING OUT OF ANY BREACH OF ANY EXPRESS OR IMPLIED WARRANTY.

Web Resources

Visit Gilson's website at www.gilson.com for:

Notes

This section is intentionally left blank for user notes.

Models: Pipetman G Pipette, Pipetman G, Pipette

File Info : application/pdf, 12 Pages, 869.67KB

PDF preview unavailable. Download the PDF instead.

QG PIPETMAN G-LT801624 D web

References

Adobe PDF Library 15.0 Adobe InDesign 16.1 (Windows)

Related Documents

Preview Gilson PIPETMAN Pipette Quick Guide
A concise guide to using Gilson PIPETMAN pipettes, covering specifications, operation, maintenance, and warranty information for single and multichannel models.
Preview Gilson PIPETMAN Classic Pipette Quick Guide: Features, Specifications, and Usage
A concise guide to the Gilson PIPETMAN Classic single-channel pipette, covering its features, specifications, maximum permissible errors, volume setting, tip fitting, pipetting techniques, maintenance, and warranty information.
Preview Gilson PIPETMAN Replacement Parts Guide
A comprehensive guide to replacement parts for Gilson PIPETMAN pipettes, including PIPETMAN Classic, Fixed, Neo, G, L, M, Microman, and Distriman models. Find part numbers and service kits for various pipette capacities.
Preview Gilson Product Guide: Liquid Handling, Purification, and Extraction Solutions
Explore Gilson's comprehensive range of laboratory instruments and solutions for liquid handling, sample purification, automation, and extraction, designed to enhance reproducibility and traceability in scientific discovery.
Preview Gilson 402 Dilutor Dispenser User's Guide
Comprehensive user guide for the Gilson 402 Dilutor Dispenser, a syringe-operated pump for precise liquid transfer in laboratory environments. Covers installation, operation, maintenance, and troubleshooting.
Preview Gilson EXTRACTMAN DNA Extraction System: High Yield, Rapid, Reproducible Application Note
This application note details the Gilson EXTRACTMAN system for rapid, reproducible DNA extraction with high yield. It highlights the Exclusion-based Sample Preparation (ESP™) technology, comparing its performance to traditional tube-based methods and demonstrating its versatility with various commercial kits. The document covers materials, methods, results, and sample integrity maintenance.