Gilson PIPETMAN G Quick Guide
Product: PIPETMAN G Pipette
Language: English (EN)
Introduction
PIPETMAN® G is a fully adjustable air-displacement pipette designed for use with disposable tips. It builds upon the renowned accuracy, precision, and reliability of the PIPETMAN series, while significantly reducing pipetting and purge forces for enhanced user comfort during prolonged use.
The range includes eight single-channel models covering volumes from 0.2 μL to 10 mL, and eight multichannel models covering volumes from 1 μL to 300 μL.
Parts Checklist
Single Models
- PIPETMAN G Pipette
- PIPETMAN G Quick Guide
- Safety bag
- Dual-position adapter (for P2G and P10G models only; refer to "Fitting the Tips" on page 6)
- Certificate of Conformity (including barcode sticker)
Multichannel Models
- PIPETMAN G Pipette
- PIPETMAN G Quick Guide
- Safety bag
- Ejector spacers for D10 tips (for Px10G models only; refer to "Ejector Spacers for PIPETMAN G 10 μL Multichannel Models" on page 7)
- Certificate of Conformity (including barcode sticker)
Good Laboratory Practice (GLP) Compliance
The serial number, engraved on the pipette body, provides unique identification and indicates the date of manufacture. For detailed information, consult the PIPETMAN G User's Guide LT801122.
Example Serial Number: AA10369
The barcode on the box and the Certificate of Conformity ensure the traceability of your pipette.
An illustration shows a pipette with a highlighted area indicating where the serial number is engraved on the body.
Description
Diagram illustrating the parts of the PIPETMAN G pipette. Key components include:
- Color-coded push button
- Tip ejector button
- Thumbwheel
- Blue ring
- Body or handle
- Clip tip ejector
- Connecting nut
- Ejector support
- Ejector lock
- Cover
- Tip ejector
- Tip holder
- Ejector spacer
- PIPETMAN® DIAMOND tip
Specifications
PIPETMAN G is a high-quality pipette offering excellent accuracy and precision. The performance data presented in the "Maximum Permissible Errors" table were obtained using genuine PIPETMAN® DIAMOND Tips, and these values are guaranteed only when such tips are used.
Maximum Permissible Errors
The following table details the maximum permissible errors, conforming to the ISO 8655-2 standard.
VOLUME (μL) | GILSON | ISO 8655 | ||
---|---|---|---|---|
SYSTEMATIC ERROR (μL) | RANDOM ERROR (μL) | SYSTEMATIC ERROR (μL) | RANDOM ERROR (μL) | |
PIPETMAN G SINGLE CHANNEL | ||||
P2G (P/N F144054M) WITH D10 OR DL10 TIPS | ||||
0.2 | ± 0.024 | ≤ 0.012 | ± 0.08 | ≤ 0.04 |
0.5 | ± 0.025 | ≤ 0.012 | ± 0.08 | ≤ 0.04 |
1 | ± 0.025 | ≤ 0.012 | ± 0.08 | ≤ 0.04 |
2 | ± 0.030 | ≤ 0.014 | ± 0.08 | ≤ 0.04 |
P10G (P/N F144055M) WITH D10 OR DL10 TIPS | ||||
1 | ± 0.025 | ≤ 0.012 | ± 0.12 | ≤ 0.08 |
5 | ± 0.075 | ≤ 0.030 | ± 0.12 | ≤ 0.08 |
10 | ± 0.100 | ≤ 0.040 | ± 0.12 | ≤ 0.08 |
P20G (P/N F144056M) WITH D200 TIPS | ||||
2 | ± 0.10 | ≤ 0.030 | ± 0.2 | ≤ 0.1 |
10 | ± 0.10 | ≤ 0.050 | ± 0.2 | ≤ 0.1 |
20 | ± 0.20 | ≤ 0.060 | ± 0.2 | ≤ 0.1 |
P100G (P/N F144057M) WITH D200 TIPS | ||||
10 | ± 0.35 | ≤ 0.10 | ± 0.8 | ≤ 0.3 |
50 | ± 0.40 | ≤ 0.12 | ± 0.8 | ≤ 0.3 |
100 | ± 0.80 | ≤ 0.15 | ± 0.8 | ≤ 0.3 |
P200G (P/N F144058M) WITH D200 TIPS | ||||
20 | ± 0.50 | ≤ 0.20 | ± 1.6 | ≤ 0.6 |
100 | ± 0.80 | ≤ 0.25 | ± 1.6 | ≤ 0.6 |
200 | ± 1.60 | ≤ 0.30 | ± 1.6 | ≤ 0.6 |
P1000G (P/N F144059M) WITH D1000 TIPS | ||||
100 | ± 3.0 | ≤ 0.6 | ± 8.0 | ≤ 3.0 |
500 | ± 4.0 | ≤ 1.0 | ± 8.0 | ≤ 3.0 |
1000 | ± 8.0 | ≤ 1.5 | ± 8.0 | ≤ 3.0 |
P5000G* (P/N F144066) WITH D5000 TIPS | ||||
500 | ± 12 | ≤ 3 | ± 40 | ≤ 15.0 |
2500 | ± 15 | ≤ 5 | ± 40 | ≤ 15.0 |
5000 | ± 30 | ≤ 8 | ± 40 | ≤ 15.0 |
P10mLG* (P/N F144067) WITH D10ML TIPS | ||||
1000 | ± 30 | ≤ 6 | ± 60 | ≤ 30.0 |
5000 | ± 40 | ≤ 10 | ± 60 | ≤ 30.0 |
10000 | ± 60 | ≤ 16 | ± 60 | ≤ 30.0 |
PIPETMAN G MULTICHANNEL | ||||
P8X10G (P/N F144068) P12X10G (P/N F144069) WITH D10 AND DL10 TIPS | ||||
1 | ± 0.08 | ≤ 0.05 | ± 0.24 | ≤ 0.16 |
5 | ± 0.20 | ≤ 0.10 | ± 0.24 | ≤ 0.16 |
10 | ± 0.20 | ≤ 0.10 | ± 0.24 | ≤ 0.16 |
P8X20G (P/N F144070) P12X20G (P/N F144071) WITH DL10 TIPS | ||||
2 | ± 0.10 | ≤ 0.08 | ± 0.4 | ≤ 0.20 |
10 | ± 0.20 | ≤ 0.10 | ± 0.4 | ≤ 0.20 |
20 | ± 0.40 | ≤ 0.20 | ± 0.4 | ≤ 0.20 |
P8X200G (P/N F144072) P12X200G (P/N F144073) WITH D200 TIPS | ||||
20 | ± 0.50 | ≤ 0.25 | ± 3.2 | ≤ 1.2 |
100 | ± 1.00 | ≤ 0.40 | ± 3.2 | ≤ 1.2 |
200 | ± 2.00 | ≤ 0.50 | ± 3.2 | ≤ 1.2 |
P8X300G (P/N F144074) P12X300G (P/N F144075) WITH D300 TIPS | ||||
30 | ± 1.00 | ≤ 0.35 | ± 8.0 | ≤ 3.0 |
150 | ± 1.50 | ≤ 0.60 | ± 8.0 | ≤ 3.0 |
300 | ± 3.00 | ≤ 1.00 | ± 8.0 | ≤ 3.0 |
* P5000G and P10mLG models do not have tip ejectors.
Setting the Volume
The volume of liquid to be aspirated is set using the volumeter. The dials are colored either black or red to indicate the position of the decimal point, depending on the model (refer to Figure 3).
The volume is set by turning the thumbwheel or the push button (refer to Figure 4). The push button offers a faster and easier way to set volumes, especially when wearing gloves. The thumbwheel allows for slow, precise adjustments using only one hand.
To achieve maximum accuracy when setting the volume:
- When decreasing the volume setting: Slowly approach the required setting, ensuring not to overshoot the mark.
- When increasing the volume setting: Pass the required value by approximately 1/3 of a turn, then slowly decrease to reach the target volume, ensuring not to overshoot the mark.
Illustrations show the dial configurations for various PIPETMAN G models (P2G, P10G, P20G, P100G, P200G, P1000G, P5000G, P10mLG, Px10G, Px20G, Px200G, Px300G) indicating volume ranges and decimal point positions (e.g., 1.25 μL, 7.5 μL, 0.75 mL).
A diagram shows the pipette body with the push button and thumbwheel clearly indicated.
MODEL | COLOR OF VOLUMETER NUMBERS (BLACK) | COLOR OF VOLUMETER NUMBERS (RED) | INCREMENT |
---|---|---|---|
P2G | μL | 0.01 μL | 0.002 μL |
P10G - P20G - Px10G - Px20G | μL | 0.1 μL | 0.02 μL |
P100G - P200G - Px200G - Px300G | μL | 0.2 μL | 0.2 μL |
P1000G | 0.01 mL | mL | 0.002 mL |
P5000G | 0.01 mL | mL | 0.002 mL |
P10mLG | mL | 0.1 mL | 0.02 mL |
Fitting the Tips
To attach a new PIPETMAN DIAMOND Tip, push the tip holder into the tip with a slight twisting motion to ensure a firm, airtight seal.
The PIPETMAN G is equipped with a stainless steel tip ejector and can also be used with a plastic tip ejector, available as an accessory. Refer to the PIPETMAN G User's Guide LT801122 for accessories.
Fitting the Tips on P2G and P10G Models
P2G and P10G models are supplied with the adapter pre-installed, ready for DL10 tips. If D10 tips are to be used, the adapter must be repositioned in the shorter slot following these steps:
- Pull the adapter down from the metallic rod.
- Turn the adapter 180°.
- Refit the adapter so the end of the metallic rod engages the shorter slot of the adapter.
An illustration shows the dual-position adapter. Text labels indicate "Short tips" and "Long tips" orientations for the adapter.
Ejector Spacer for PIPETMAN G 10 μL Multichannel Models
Depending on the tip used (D10 or DL10), you may need to exchange the ejector spacer. The broadly spaced one is for D10 tips, and the narrowly spaced one is for DL10 tips.
- Remove the tip ejector, keeping both ejector locks depressed. Pull the tip ejector down.
- Gently press the tabs from the ejector spacer and remove it from the tip ejector.
- Insert the alternative ejector spacer and click it to the tip ejector.
- To refit the tip ejector, gently re-insert it vertically into the rails of the ejector support.
Diagram shows the tip ejector and two types of ejector spacers: one labeled "Ejector spacer D10" (broadly spaced) and another for "Long tips" (narrowly spaced).
PIPETMAN DIAMOND Tips are best fitted using the ROCKY RACK technique, invented by Gilson, available exclusively in their TIPACK and TOWERPACK products.
PIPETMAN DIAMOND Tip Compatibility for Multichannel Models | |
---|---|
P8x10G, P12x10G | D10*, DL10, DF10, DFL10 |
P8x20G, P12x20G | DL10, DFL10, DF30, D200 |
P8x200G, P12x200G | D200, D300, DF200, DF300 |
P8x300G, P12x300G | D200, D300, DF200, DF300 |
* Using the broad ejector spacer D10, you can adapt a D10 tip.
An illustration depicting the ROCKY RACK technique for fitting tips.
Pre-Wetting the Tips
Pre-wetting the tips before pipetting helps prepare them for optimal performance. Ideally, this process involves immersing the tip in the liquid and performing one pipetting step.
Pre-wetting ensures that the volumes you pipette will achieve the specified accuracy and precision.
Aspirating
- Press the push button to the first stop (this corresponds to the set volume of liquid).
- Hold the pipette vertically and immerse the tip in the liquid (refer to PIPETMAN G User's Guide LT801122).
- Release the push button slowly and smoothly (to the top position) to aspirate the set volume of liquid.
- Wait one second, and then withdraw the pipette tip from the liquid. You may wipe any droplets away from the outside of the tip using a medical wipe; however, if you do so, take care to avoid touching the tip's orifice.
Diagram shows a multichannel pipette. Arrows indicate the motion for aspirating (push button to first stop, then release) and dispensing (push button to second stop). Labels indicate "Rest position", "First stop", and "Second stop".
Dispensing
- Place the end of the tip against the inside wall of the recipient vessel (at an angle of 10° to 40°).
- Press the push button slowly and smoothly to the first stop.
- Wait for at least one second, then press the push button to the second stop to expel any residual liquid from the tip. Keep the push button pressed fully down and (while removing the pipette) draw the tip along the inside surface of the vessel.
- Release the push button smoothly. Eject the tip by pressing firmly on the tip ejector button.
Maintenance
Routine maintenance helps keep your pipette in good condition, ensuring a continued high level of performance. Maintenance is limited to cleaning or autoclaving the parts specified in the PIPETMAN G User's Guide LT801122, or to replacing the push button, connecting nut, tip ejector, tip holder, seal, and O-ring.
Warranty
Gilson warrants this pipette against defects in material under normal use and service for a period of three years for both single channel and multichannel models from the date of purchase.
This warranty shall not apply to pipettes subjected to abnormal use and/or improper or inadequate maintenance (contrary to the recommendations in the user's guide), including, but not limited to, physical damage, improper handling, spillage, or exposure to corrosive environments. The warranty is void if pipettes are altered or modified by any party other than Gilson or its designates. Gilson's sole liability under this warranty is limited to, at Gilson's sole option, repair or replacement of defective components, or refund of the purchase price. Routine cleaning, control, and recalibration are not covered. The replacement of wearing parts such as seals, O-rings, broken piston assemblies, and broken tip holders are also not covered.
THE FOREGOING WARRANTY IS EXCLUSIVE AND GILSON HEREBY DISCLAIMS ALL OTHER WARRANTIES, WHETHER EXPRESS OR IMPLIED, INCLUDING ANY WARRANTIES OF MERCHANTABILITY AND ANY WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE. UNDER NO CIRCUMSTANCES SHALL GILSON BE LIABLE FOR ANY CONSEQUENTIAL, PUNITIVE, INDIRECT OR INCIDENTAL DAMAGES ARISING OUT OF ANY BREACH OF ANY EXPRESS OR IMPLIED WARRANTY.
Web Resources
Visit Gilson's website at www.gilson.com for:
- Service contact information
- PIPETMAN G User's Guide LT801122 pdf
- Information about accessories, spare parts, and related products
- Additional product information and special offers
Notes
This section is intentionally left blank for user notes.