Sealey 8 FUNCTION PROFESSIONAL AUTO-RANGING DIGITAL MULTIMETER

MODEL NO: TM102

Thank you for purchasing a Sealey product. Manufactured to a high standard, this product will, if used according to these instructions, and properly maintained, give you years of trouble free performance.

IMPORTANT: PLEASE READ THESE INSTRUCTIONS CAREFULLY. NOTE THE SAFE OPERATIONAL REQUIREMENTS, WARNINGS & CAUTIONS. USE THE PRODUCT CORRECTLY AND WITH CARE FOR THE PURPOSE FOR WHICH IT IS INTENDED. FAILURE TO DO SO MAY CAUSE DAMAGE AND/OR PERSONAL INJURY AND WILL INVALIDATE THE WARRANTY. KEEP THESE INSTRUCTIONS SAFE FOR FUTURE USE.

1. SAFETY

1.1. PERSONAL PRECAUTIONS

1.2. GENERAL SAFETY INSTRUCTIONS

1.3. SAFETY SYMBOLS

INPUT LIMITS

Function Maximum Input
VAC 600V AC
V DC 600V DC
MA DC/AC 200mA
A DC or A AC 10A DC/AC (30 seconds max every 15 minutes)
Resistance, Diode, Temp, Continuity 500V DC/AC

2. INTRODUCTION

High precision auto-ranging multimeter. Conforms to EN 61010-1 CATIII 600V safety requirements for electrical equipment for measurement, control and laboratory use. Features temperature measurement capability, max./data hold and auto-power-off function. Double moulded housing with soft grip case and large display for ease of use. Practical probe, lead storage and integral upright stand for use on the workbench.

MEASURES:

OVERVIEW

  1. LCD Display
  2. Function Switch
  3. 10A Jack
  4. COM Jack
  5. Positive Jack
  6. Mode Switch
  7. Range Switch
  8. Data Hold Switch
  9. MAX. Hold Switch

SYMBOLS AND ANNUNCIATORS

3. SPECIFICATION

NOTE: Accuracy specifications consist of two elements: 1.(% reading) - This is the accuracy of the measurement circuit. 2. (± digits) - This is the accuracy of the analog to digital converter.

NOTE: Accuracy is stated at 65°F to 83°F (18°C to 28°C) and less than 70% RH (Relative Humidity).

DC Voltage: (Auto-ranging)

Function Range Resolution ACCURACY
Input Impedance: 10MΩ 200.0mV 0.1mV ±0.5 Reading ±2 digits
Maximum Input: 600V dc or 600V ac rms (CATIII) 2.000V 1mV ±1.2 Reading ±2 digits
20.00V 10mV
200.0V 100mV ±1.5 Reading ±2 digits
600V 1V ±1.5 Reading ±2 digits

AC Voltage: (Auto-ranging except 200mV)

Function Range Resolution ACCURACY
Input Impedance: 10MΩ 200.0mV 0.1mV ±1.5 Reading ±30 digits
Frequency Range:50 to 60Hz 2.000V 1mV ±1.2 Reading ±3 digits
Maximum input: 600V dc or 600V ac rms (CATIII) 20.00V 10mV ±1.5 Reading ±3 digits
200.0V 100mV ±1.5 Reading ±3 digits
600V 1V ±2.0 Reading ±4 digits

DC Current (Auto-ranging for uA/mA)

Function Range Resolution ACCURACY
Overload Protection:0.2A/250V and 10A/250V Fuse 200.0μΑ 0.1μΑ ±1.0 Reading ±3 digits
Maximum Input:200mA dc or 200mA rms on uA/mA ranges, 10A dc or ac rms on 10 A range 2000μΑ 1μΑ ±1.5 Reading ±3 digits
20.00mA 10μΑ
200.0mA 100μΑ ±2.5 Reading ±5 digits
10A 10mA

AC Current (Auto-ranging for uA/mA)

Function Range Resolution ACCURACY
Overload Protection:0.2A/250V and 10A/250V Fuse 200.0μΑ 0.1μΑ ±1.5 Reading ±5 digits
Frequency Range:50 to 60Hz 2000μΑ 1μΑ ±1.8 Reading ±5 digits
Maximum Input:200mA dc or 200mA rms on uA/mA ranges, 10A dc or ac rms on 10 A range 20.00mA 10μΑ
200.0mA 100μΑ ±1.8 Reading ±5 digits
10A 10mA ±3.0 Reading ±7 digits

Resistance (Auto-ranging)

Function Range Resolution ACCURACY
Input Protection: 500V dc or 500V ac rms 200.0Ω 0.1Ω ±1.2 Reading ±4 digits
2.000ΚΩ ±1.0 Reading ±2 digits
20.00ΚΩ 10Ω ±1.2 Reading ±2 digits
200.0ΚΩ 100Ω
2000ΜΩ 1ΚΩ ±2.0 Reading ±3 digits
20.00ΜΩ 10ΚΩ

Audible Continuity Test

Function TEST CURRENT
Audible threshold: Less than 100Ω 0.3mA typical
Overload protection: 500V dc or ac rms

Diode Test

Function RANGE RESOLUTION ACCURACY
Open circuit voltage: 1.5V dc typical 1mV ±10 Reading ±5 digits
Overload protection: 500V dc or ac rms

Temperature

FUNCTION RANGE RESOLUTION ACCURACY
Sensor: Type K Thermocouple -50°C~+1000°C 1°C ±10 Reading ±5 digits
Overload protection:500V dc or ac rms -58°F~+1832°F 1°F ±10 Reading ±9 digits

4. OPERATION

WARNING! Risk of electrocution. High-voltage circuits, both AC and DC, are very dangerous and should be measured with great care.

ALWAYS turn the function switch to the OFF position when the meter is not in use. The meter has an AUTO POWER OFF function which will automatically switch the meter off, after 15 minutes of inactivity.

If "OL" appears in the display during a measurement, the value exceeds the range you have selected. Change to a higher range.

4.1. MODE SWITCH

Selects different Modes according to function selected.

4.2. RANGE SWITCH

4.3. DATA HOLD SWITCH

4.4. MAX HOLD SWITCH

This function allows the maximum reading taken to be held on the display - the maximum reading will continuously update upwards if readings change. Press the Max Hold switch to set the maximum reading. Press the Max Hold switch again and the display will return to normal. (Not applicable with Resistance Measurement, Diode Test, Continuity Test and Temperature Measurement).

NOTE: On some low AC and DC voltage ranges, with the test leads not connected to a device, the display may show a random reading. This is normal and is caused by the high-input sensitivity. The reading will stabilise and give a proper measurement when connected to a circuit.

4.5. DC VOLTAGE MEASUREMENT

CAUTION: DO NOT measure DC voltages if a motor on the circuit is being switched ON or OFF. Large voltage surges may occur that can damage the meter.

4.6. AC VOLTAGE MEASUREMENT

WARNING! Risk of Electrocution. The probe tips may not be long enough to contact the live parts inside some 240V outlets for appliances because the contacts are recessed deep in the outlets. As a result, the reading may show 0 volts when the outlet actually has voltage on it. Make sure the probe tips are touching the metal contacts inside the outlet before assuming that no voltage is present.

CAUTION: DO NOT measure AC voltages if a motor on the circuit is being switched ON or OFF. Large voltage surges may occur that can damage the meter. Set the function switch to the V AC position, AC and V will appear on the display.

4.7. DC CURRENT MEASUREMENT

CAUTION: DO NOT make current measurements on the 10A scale for longer than 30 seconds. Exceeding 30 seconds may cause damage to the meter and/or the test leads.

4.8. AC CURRENT MEASUREMENT

WARNING! To avoid electric shock, do not measure AC current on any circuit whose voltage exceeds 250V AC

CAUTION: DO NOT make current measurements on the 10A scale for longer than 30 seconds. Exceeding 30 seconds may cause damage to the meter and/or the test leads.

4.9. RESISTANCE MEASUREMENT

WARNING! To avoid electric shock, disconnect power to the unit under test and discharge all capacitors before taking any resistance measurements. Remove the batteries and unplug the line cords.

4.10. AUDIBLE CONTINUITY TEST

WARNING! To avoid electric shock, never measure continuity on circuits or wires that have voltage on them.

4.11. DIODE TEST

4.12. TEMPERATURE MEASUREMENT

WARNING! To avoid electric shock, disconnect any test probes from any source of voltage before making a temperature measurement.

5. MAINTENANCE

5.1. REPLACING THE BATTERY

WARNING! To avoid electric shock, disconnect the test leads from any source of voltage before removing the battery cover.

5.2. REPLACING THE FUSES

WARNING! To avoid electric shock, disconnect the test leads from any source of voltage before removing the fuse cover.

ENVIRONMENT PROTECTION

Recycle unwanted materials instead of disposing of them as waste. All tools, accessories and packaging should be sorted, taken to a recycling centre and disposed of in a manner which is compatible with the environment. When the product becomes completely unserviceable and requires disposal, drain any fluids (if applicable) into approved containers and dispose of the product and fluids according to local regulations.

WEEE REGULATIONS

Dispose of this product at the end of its working life in compliance with the EU Directive on Waste Electrical and Electronic Equipment (WEEE). When the product is no longer required, it must be disposed of in an environmentally protective way. Contact your local solid waste authority for recycling information.

BATTERY REMOVAL

Under the Waste Batteries and Accumulators Regulations 2009, Jack Sealey Ltd are required to inform potential purchasers of products containing batteries (as defined within these regulations), that they are registered with Valpak's registered compliance scheme. Jack Sealey Ltd Batteries Producer Registration Number (BPRN) is BPRN00705.

Note: It is our policy to continually improve products and as such we reserve the right to alter data, specifications and component parts without prior notice. Important: No Liability is accepted for incorrect use of this product. Warranty: Guarantee is 12 months from purchase date, proof of which is required for any claim.

Sealey Group, Kempson Way, Suffolk Business Park, Bury St Edmunds, Suffolk. IP32 7AR

01284 757500 sales@sealey.co.uk www.sealey.co.uk

Models: TM102, TM102 Professional Auto Ranging Digital Multimeter, Professional Auto Ranging Digital Multimeter, Auto Ranging Digital Multimeter, Ranging Digital Multimeter, Digital Multimeter, Multimeter

PDF preview unavailable. Download the PDF instead.

TM102 Instructions 20240319 (4) DFC1594824 Adobe PDF Library 17.0 Adobe InDesign 19.3 (Windows)

Related Documents

Preview Sealey MM102 Professional Auto-Ranging Digital Multimeter User Manual and Instructions
Detailed user manual and operating instructions for the Sealey MM102 Professional Auto-Ranging Digital Multimeter, including safety precautions, specifications, and maintenance guides.
Preview Sealey 6-24V Auto Probe PP1.V4: User Manual and Operating Instructions
A comprehensive guide to the Sealey 6-24V Auto Probe (Model PP1.V4). This manual covers essential safety precautions, detailed features, step-by-step operation, various testing procedures for automotive electrical systems (polarity, continuity, component activation, trailer lamps, ground contacts, short circuits), and jumper lead functionality. It also includes environmental protection and warranty information.
Preview Sealey 6-24V Auto Probe Plus PPX.V4 User Manual
Comprehensive instructions for the Sealey 6-24V Auto Probe Plus (Model No. PPX.V4), covering safety, features, operation, testing, and troubleshooting for automotive electrical systems.
Preview Sealey VS921 IR & RF Key Fob Tester: Instructions and Specifications
Comprehensive instructions and specifications for the Sealey VS921 IR & RF Key Fob Tester. Learn how to use, assemble, and troubleshoot this diagnostic tool for key fobs and remote controls, including safety guidelines and environmental information.
Preview Sealey TA131 Relay Tester: User Manual and Specifications
Comprehensive user manual for the Sealey TA131 Relay Tester, covering safety, operation, specifications, and maintenance for electro-mechanical relays in modern vehicles.
Preview Sealey AK1998.V3 80-1000V Non-Contact Voltage Detector User Manual & Safety Guide
User manual for the Sealey AK1998.V3 non-contact voltage detector. Covers safety, operation, specifications, battery replacement, maintenance, and environmental disposal for AC voltage detection up to 1000V.
Preview Sealey VS270 Multi Voltage Glow Plug Tester: Instructions and Operation
A comprehensive guide to the Sealey VS270 Multi Voltage Glow Plug Tester, detailing safety precautions, product introduction, operational procedures including initialization and testing methods, and maintenance guidelines for modern diesel engines.
Preview Sealey AK1998.V3 80-1000V Non-Contact Voltage Detector User Manual
Comprehensive user manual for the Sealey AK1998.V3 80-1000V Non-Contact Voltage Detector. Includes safety instructions, operation guide, specifications, and maintenance tips.