Information for Patient and Caregiver

smith&nephew

PICO° 7

Single Use Negative Pressure Wound Therapy System Patient and Caregiver User Manual

Introduction

The PICO 7 Single Use Negative Pressure Wound Therapy System is intended for use by or on the direction of a trained and licensed physician in accordance with these instructions for use. Contact a healthcare professional right away if you are unsure as to the correct use of the PICO 7 System.

1. Indications

PICO 7 Single Use Negative Pressure Wound Therapy System is indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of low to moderate levels of exudate and infectious materials. PICO 7 Single Use Negative Pressure Wound Therapy System is suitable for use both in a hospital and homecare setting.

Appropriate wound types include:

When used on closed surgical incisions, PICO 7 Single Use Negative Pressure Wound Therapy System is intended to aid in reducing the incidence of:

2. Contraindications

The PICO 7 System is contraindicated for:

The PICO 7 System should not be used for the purpose of:

3. Description

The PICO 7 System is a device that provides suction to a wound to promote wound healing. When the PICO 7 pump is turned on, air is pulled out of the dressing and excess fluid will start to enter the dressing. The PICO 7 System has visual indicators to let you know when there is an issue. The PICO 7 System does not contain any audible alerts. Dressings should be checked frequently (See Does your dressing look full?). Carry the pump so that the status can be checked frequently. If the air leak, dressing full or battery low indicators illuminate, consult the troubleshooting section.

Visual Indicators:

4. Important Information

Pump Placement Warning

The PICO 7 pump contains a MAGNET. Keep the PICO 7 pump at least 4 inches (10 cm) away from other medical devices at all times. As with all electrical medical equipment, failure to maintain appropriate distance may disrupt the operation of nearby medical devices. See Section 5 Magnet Warning.

5. Warnings

Magnet Warning

The PICO 7 pump contains a MAGNET. As with all electrical medical equipment, failure to maintain appropriate distance may disrupt the operation of nearby medical devices. The PICO 7 pump must be positioned at least 4 inches (10cm) away from other medical devices that could be affected by magnetic interference. These include but are not limited to:

THIS WARNING APPLIES AT ALL TIMES TO ALL USERS. This applies to both Patients and Caregivers. You must keep the PICO 7 pump at least 4 inches (10cm) away from other devices:

Additional Warnings:

5. Warnings (continued)

6. Precautions

7. Technical Information

Storage and Transport Environment

Operating Conditions

Pump Information

8. Activating the PICO 7 pump

  1. Press Orange Button to activate pump.
  2. The green 'OK' indicator and orange 'leak' indicator will start to flash together.
  3. The pump may take up to 65 seconds to establish therapy, during this time you will hear the pump.

If after this time the orange 'leak' indicator continues to flash without the 'OK' indicator, consult the troubleshooting section.

9. How do I know if the PICO 7 pump is working?

Green 'OK' indicator flashes

While the PICO 7 pump is powered on the OK indicator will flash green. The system is functioning correctly and delivering negative pressure. Regularly check the condition of the dressing by referring to Section 12 Does your dressing look full? If dressing looks full or blocked contact a healthcare professional right away.

10. Changing the batteries

If the battery low indicator flashes, the batteries on the PICO 7 pump will require changing. To change the batteries:

  1. Remove the back cover from the PICO 7 pump to access the battery compartment.
  2. Remove the used batteries and dispose of in accordance with local regulations.
  3. Insert new AA batteries in the orientation indicated inside the battery compartment. Following this all four indicators should briefly illuminate.
  4. Replace the back cover.
  5. Press the orange button to restart therapy.

11. Showering and bathing

Light showering is permissible; however, the PICO 7 pump should be disconnected and placed in a safe location where it will not get wet. Before showering and bathing:

While disconnected:

After showering or bathing:

12. Does your dressing look full?

Patients and caregivers should not attempt to change or apply dressings. Dressings should only be changed by a healthcare professional.

If your dressing looks similar to this, you should leave the dressing.

It is acceptable to be left in place.

If your dressing looks similar to either or both of these, you must contact your healthcare professional right away.

Dressing is Full.

Soft port may be blocked with fluid.

13. Troubleshooting

Indicator status What could be the problem? How to solve
All indicators off The pump is in standby. Therapy has been paused. Press the Orange button to resume therapy.
The pump has completed its course of therapy. Pressing the Orange button will not restart therapy. Contact your healthcare professional right away.
The batteries are dead. Replace the batteries (see changing the batteries) then press the Orange button to restart therapy.
Green 'OK' and Orange 'battery low' indicators flash System is functioning properly but the batteries are low. Replace the batteries (see changing the batteries) then press the Orange button to restart therapy.
Orange 'leak' indicator flashes An air leak has been detected. Therapy is not being applied. (Note: the pump will automatically try to restart therapy after 1 hour). 1. Smooth down the edge of the dressings and strips to remove any creases. 2. Ensure that the tube connectors have been twisted together securely. 3. Press the Orange button to restart the pump. 4. If the air leak remains the Orange leak indicator will flash again after approximately 65 seconds (depending on the size of the wound). 5. If the battery low indicator is flashing, replace the batteries (see changing the batteries), then press the orange button to restart therapy.
Orange 'leak' and Orange 'battery low' indicators flash An air leak has been detected and the batteries are low. Therapy is not being applied. (Note: the pump will automatically try to restart therapy after 1 hour). Contact your healthcare professional right away.
Orange 'dressing full' indicator flashes Dressing is full or there is a blockage. Therapy is not being applied. (Note: the pump will automatically try to restart therapy after 1 hour). Contact your healthcare professional right away.
Orange 'dressing full' and orange 'battery low' indicators flash Dressing is full or there is a blockage and the battery is low. Therapy is not being applied. (Note: the pump will automatically try to restart therapy after 1 hour). Contact your healthcare professional right away.
All indicators solidly illuminated A pump error has been detected. The pump can no longer apply therapy. Contact your healthcare professional right away.

14. Glossary of symbols

15. Contact Information

Smith & Nephew Medical Limited, 101 Hessle Road, Hull, HU3 2BN England. *Trade Marks of Smith & Nephew www.smith-nephew.com Smith & Nephew This product may be covered by one or more US patents. See www.smith-nephew.com/patents

UNITED STATES Smith & Nephew, Inc., Smith & Nephew, Inc. 5600 Clearfork Main Street, Suite 600 Fort Worth, TX 76109 Customer Care Center: 1-817-900-4000 Date of issue 04/2022

Models: PICO 7 Patient and Caregiver, PICO 7, Patient and Caregiver, Caregiver

File Info : application/pdf, 4 Pages, 2.62MB

PDF preview unavailable. Download the PDF instead.

35051-pico-7-patientandcaregiver-1674104376612

References

Adobe InDesign 17.1 (Macintosh) Adobe PDF Library 16.0.5

Related Documents

Preview PICO 14 Negative Pressure Wound Therapy System User Manual - Smith & Nephew
User manual for the PICO 14 Single Use Negative Pressure Wound Therapy System by Smith & Nephew, providing instructions for patients and caregivers on indications, contraindications, operation, troubleshooting, and safety. Includes details on system features, warnings, precautions, and technical specifications.
Preview PICO 7 Single Use Negative Pressure Wound Therapy System User Manual | Smith & Nephew
Comprehensive user manual for the Smith & Nephew PICO 7 Single Use Negative Pressure Wound Therapy System, detailing indications, contraindications, warnings, instructions for use, specifications, and troubleshooting for healthcare professionals.
Preview Smith & Nephew PICO 14 Single Use Negative Pressure Wound Therapy System User Manual for Healthcare Professionals
This user manual provides detailed instructions for healthcare professionals on the use of the Smith & Nephew PICO 14 Single Use Negative Pressure Wound Therapy System. It covers product description, indications, contraindications, important information, warnings, precautions, adverse reactions, definitions, instructions for use, troubleshooting, system variants, and specifications.
Preview PICO 7 System Quick Reference Guide - Smith & Nephew
A quick reference guide for the Smith & Nephew PICO 7 system, detailing its features, operation, and troubleshooting.
Preview TRIGEN SURESHOT Distal Targeting System V4.0 User Manual
This user manual provides comprehensive instructions for the Smith & Nephew TRIGEN SURESHOT Distal Targeting System V4.0, an advanced computer-assisted orthopaedic surgery tool designed for precise drill positioning during intramedullary nail implantation procedures using electromagnetic tracking.
Preview VERSAJET III Hydrosurgery System: FDA 510(k) Substantial Equivalence Notification and Product Overview
This document provides an overview of the VERSAJET III Hydrosurgery System by Smith & Nephew, detailing its FDA 510(k) clearance for substantial equivalence. It outlines the device's indications for use, description, technological comparison to its predicate, and regulatory compliance information.
Preview Smith & Nephew SUPER TURBOVAC 90 COBLATION Wands: Product Overview
Explore the Smith & Nephew SUPER TURBOVAC 90 COBLATION Wand family, featuring advanced AMBIENT technology for rapid ablation and clear visualization in surgical procedures. Learn about its key features, specifications, and applications.
Preview Smith & Nephew QUANTUM 2 COBLATION System: Advanced Surgical Ablation Technology
Explore the Smith & Nephew QUANTUM 2 Controller System and its comprehensive range of COBLATION™ Wands for efficient and precise surgical procedures. Learn about COBLATION plasma technology, operating modes, unique waveform features, and detailed wand selection for various arthroscopic applications.