COXO CX235 Series Contra Angle Operation Manual

MODEL: CX235

Professional Dental Manufacturer: COXO

Distributed by: DentalDeal

Please read this operation Manual carefully and file for future reference.

Preface

Thanks for purchasing the COXO dental Contra angle. To fully exert the function of this equipment safely and effectively, please read this instruction carefully before using and keep this manual for reference anytime. Under regulations' request, COXO has committed to the optimization of its products. This may cause some parts of change; when you find the product appears some difference from the description in the manual, please contact the dealers or COXO's after-sales center for details.

Technical Specifications

Website: www.dentaldeal.ca

WARNING

Product Composition: Inner Channel 1:1

Diagrams illustrating different contra angle models:

Cartridge Replacement (Locking Pin)

  1. Loosen the small screw on the curving part, move the head cover, and remove the transmission gear [See Fig. 2.1].
  2. Use the attached wrench to remove the cap [See Fig. 2.2].
  3. Use the bur to push out the cartridge.
  4. Insert the new cartridge into the handpiece head, ensuring the small raised part aligns with the groove inside the front of the head for proper fit.

Fig. 2.1: Illustration of loosening a screw on a handpiece with a screwdriver.

Fig. 2.2: Illustration of removing a cap from a handpiece with a wrench.

Fig. 2.3: Illustration of pushing out a cartridge with a bur.

Fig. 2.4: Illustration of inserting a new cartridge into a handpiece head, aligning a pin with a groove.

Contra Angle and Motor Connections

  1. Straightly insert the motor into the connection of the contra angle until you hear a "cracking" sound.
  2. Hold the bending machine and pull out the motor; it will be separated [See Fig. 3.2].
  3. Align the groove; the contra angle can be inserted into the air motor [See Fig. 3.3].

Fig. 3.1: Shows a Contra angle with generator (CX235-1E) connecting to an Air motor (CX235-3B) with 1:1 drive.

Fig. 3.2: Shows a Contra angle (CX235-1B) connecting to an Air motor (CX235-3B) with 1:1 drive.

Fig. 3.3: Shows an Optic Fiber Contra angle (CX235-1C) connecting to an Optic Fiber/LED Air motor (CX235-3C/CX235-3D) with 1:1 drive.

Fig. 3.4: Close-up illustration of an optic fiber.

Contra Angle Head Technical Parameters

Model Rotate Head BUR
CX235CH-1/2 max.40,000rpm for CA burs Φ2.35
CX235CH-3/4 max.40,000rpm for CA burs Φ2.35
CX235CH-6/7 max.40,000rpm for use files
CX235CH-13 max.30,000rpm for CA burs Φ2.35
CX235CH-15/16 max.3,000rpm for CA burs Φ2.35
CX235CH-17 max.40,000rpm for CA burs Φ2.35

Fig. 1.1, 1.2, 1.3: Illustrations of "360° reciprocating rotate head" movement.

Fig. 1.4: Illustration showing a handpiece head with a bur, indicating a length of 10.7mm.

Special Head Movement Technical Parameters

Model Rotate Head BUR
CX235CH-10 90° reciprocating for CA burs Φ2.35/prophylaxis brush
CX235CH-11 Up and down movement for CA burs Φ2.35/the crevices knife
CX235CH-12 60° reciprocating for use files
CX235CH-9 360° reciprocating for CA burs Φ2.35/implant files
CX235CH-14 360° reciprocating for CA burs Φ2.35/implant files

Fig. 1.5: Illustration of "90° reciprocating rotate head" movement.

Fig. 1.6: Illustration of "Up and down movement".

Fig. 1.7: Illustration of "60° reciprocating rotate head" movement.

Fig. 1.8, 1.9: Illustrations showing internal and external spray nozzles with irrigation tubes and water outlets.

Product Composition: 1:1 Direct Drive

Diagrams illustrating various 1:1 direct drive contra angle models:

Fig. 2.1 to 2.5: Visual representations of the aforementioned models.

Fig. 3.1: Shows dimensions Φ2.35 and Φ19.6 for a handpiece.

Fig. 3.2: Shows dimensions Φ1.6 and Φ19.6 for a handpiece.

External contra angle low speed handpiece: 1:1 Direct Drive.

Burs: For CA burs Φ2.35; for FG burs Φ1.6.

Series: CX235C1 Series; CX235C2 Series.

Mounting of Bur (Locking Pin)

  1. Open the latch chuck, insert the bur, ensuring the quadrate part of the bur aligns with the proper part in the cartridge, then turn the latch chuck back.
  2. When removing the bur, open the latch chuck, then take out the bur.

Fig. 3.3: Illustration of inserting a bur using a locking pin mechanism.

Mounting of Bur (Push Button)

The fixed method of internal contra low speed handpiece is push button. Push the button to the end, and then the bur can be ejected from the head.

Fig. 3.4: Illustration of inserting a bur using a push button mechanism.

Product Composition: 1:1 Direct Drive (Continued)

Diagrams illustrating various 1:1 direct drive contra angle models:

Fig. 2.6 to 2.8: Visual representations of the aforementioned models.

Fig. 3.1: Shows dimensions Φ2.35 and Φ19.6 for a handpiece.

Fig. 3.2: Shows dimensions Φ1.6 and Φ19.6 for a handpiece.

Mounting of Bur (Locking Pin)

  1. Open the latch chuck, insert the bur, be sure the quadrate part of the bur aligns the proper part in the cartridge, then turn the latch chuck back.
  2. When remove the bur, open the latch chuck, then take out the bur.

Fig. 9.1: Illustration of inserting a bur using a locking pin mechanism.

Fig. 9.2: Illustration of inserting a bur using a push button mechanism.

Fig. 9.3: Illustration of inserting a bur using a push button mechanism.

Fig. 9.4: Illustration of inserting a bur using a push button mechanism.

Fig. 9.5: Illustration of inserting a bur using a push button mechanism.

Product Composition: 4:1 Reduction

Diagrams illustrating various 4:1 reduction contra angle models:

All models are labeled with "4:1 reduction".

Fig. 4.1 to 4.6: Visual representations of the aforementioned models.

Mounting of Bur (Locking Pin)

  1. Open the latch chuck, insert the bur, be sure the quadrate part of the bur aligns the proper part in the cartridge, then turn the latch chuck back.
  2. When remove the bur, open the latch chuck, then take out the bur.

Fig. 5.1: Illustration of inserting a bur using a locking pin mechanism.

Mounting of Bur (Push Button)

The fixed method of internal contra low speed handpiece is push button. Push the button to the end, and then the bur can be ejected from the head.

Fig. 5.2: Illustration of inserting a bur using a push button mechanism.

Fig. 5.3: Illustration of inserting a bur using a push button mechanism.

Fig. 5.4: Illustration of inserting a bur using a push button mechanism.

3. Prophylaxis

Mounting and Removing the Rubber Cup/Brush

Mounting

  1. Insert the rubber cup/brush into a screw hole of the handpiece head.
  2. Turn the rubber cup/brush in a clockwise direction to fix the screw shank.

Removing

Perform a reverse procedure of Mounting. Turn the Rubber cup/brush in the "Loosen" direction.

Fig. 6.1: Illustration of mounting a rubber cup/brush onto a handpiece head.

Fig. 6.2: Illustration of a mounted rubber cup/brush.

Product Composition: 10:1 Reduction

CX235C5-12: Features a "push button", "hand use file", and "10:1 reduction hand".

Fig. 7: Visual representation of the CX235C5-12 model.

Mounting and Removing File

⚠️ NOTE:

  1. Do not connect or disconnect the handpiece until the drive motor has completely stopped.
  2. After the file is locked in place, lightly pull out the file to make sure the file is locked.

Fig. 7.1: Illustration of inserting hand use files into a handpiece with a push button mechanism.

Product Composition: 16:1 Reduction

Diagrams illustrating various 16:1 reduction contra angle models:

All models are labeled with "16:1 reduction".

Fig. 8.1 to 8.8: Visual representations of the aforementioned models.

Mounting of Bur (Locking Pin)

  1. Open the latch chuck, insert the bur, be sure the quadrate part of the bur aligns the proper part in the cartridge, then turn the latch chuck back.
  2. When remove the bur, open the latch chuck, then take out the bur.

Fig. 9.1: Illustration of inserting a bur using a locking pin mechanism.

Mounting of Bur (Push Button)

The fixed method of internal contra low speed handpiece is push button. Push the button to the end, and then the bur can be ejected from the head.

Fig. 9.2: Illustration of inserting a bur using a push button mechanism.

Fig. 9.3: Illustration of inserting a bur using a push button mechanism.

Fig. 9.4: Illustration of inserting a bur using a push button mechanism.

Fig. 9.5: Illustration of inserting a bur using a push button mechanism.

Product Composition: 20:1 Reduction

Diagrams illustrating various 20:1 reduction contra angle models:

All models are labeled with "20:1 reduction".

Fig. 10.1 to 10.4: Visual representations of the aforementioned models.

Mounting and Removing Bur (Locking Pin)

Locking pin

  1. Open the latch chuck, insert the bur, be sure the quadrate part of the bur aligns the proper part in the cartridge, then turn the latch chuck back.
  2. When remove the bur, open the latch chuck, then take out the bur.

⚠️ NOTE:

  1. Do not connect or disconnect the handpiece until the drive motor has completely stopped.
  2. After the file is locked in place, lightly pull out the bur to make sure the bur is locked.

Fig. 11.1: Illustration of inserting a bur using a locking pin mechanism.

Fig. 11.2: Illustration of inserting a bur using a locking pin mechanism.

Fig. 11.3: Illustration of inserting a bur using a push button mechanism.

Mounting the "Y" type water tube

Three irrigation methods are available: External, Internal, or both External and Internal, depending on the tool and operation procedure.

(1) External Spray Nozzle

Connect the Irrigation Tube to the External Spray Nozzle firmly.

Fig. 12.1: Illustration showing connection of irrigation tube to an external spray nozzle.

(2) Internal Spray Nozzle

Only a drill with internal irrigation is used.

  1. Connect the Irrigation Tube to the Internal Spray Nozzle firmly.
  2. Put the Internal Spray Nozzle into the slot on the head.

Fig. 12.2: Illustration showing connection of irrigation tube to an internal spray nozzle.

Fig. 12.3: Illustration showing connection of irrigation tube to an internal spray nozzle.

(3) External and Internal Irrigation together (locking pin)

Only a drill with internal irrigation is used.

  1. Connect the ends of the Y-Connector to the External Spray Nozzle and Internal Spray Nozzles as detailed in procedures (1) and (2).
  2. Connect the Irrigation Tube to the Y-Connector firmly.

Fig. 12.4: Illustration showing connection of a "Y" Connector and irrigation tube to external and internal spray nozzles.

(4) External and Internal Irrigation together (push button)

Only a drill with internal irrigation is used.

  1. Connect the ends of the Y-Connector to the External Spray Nozzle and Internal Spray Nozzles as detailed in procedures (1) and (2).
  2. Connect the Irrigation Tube to the Y-Connector firmly.

Fig. 12.5 to 12.7: Illustrations showing connection of irrigation tube and "Y" Connector for push button type.

Fig. 12.8: Illustration showing a handpiece with connected irrigation tubes and Y-connector.

Cartridge Replacement (Locking Pin)

  1. Loosen the curving lock section and remove the head from the sheath.
  2. Pull out the drive gear assembly from the head [See Fig. 13.1].
  3. Cap the wrench on the head's screw and turn counter-clockwise. Unscrew and separate the cover [See Fig. 13.2].
  4. Swing the new cartridge, with its pin aligned in the groove of the head.
  5. Reassemble in the reverse order.

Fig. 13.1: Illustration of removing the drive gear assembly.

Fig. 13.2: Illustration of unscrewing a cover with a wrench.

Fig. 13.3: Illustration of a cartridge with a pin and slit.

Fig. 13.4: Illustration of removing a cartridge.

1.2. Removal of the cartridge (CX235CH-11)

  1. Loosen the curving lock section and remove the head from the sheath.
  2. Pull out the drive gear assembly from the head [See Fig. 14.1].
  3. Directly remove the cartridge [See Fig. 14.4].

Fig. 14.1: Illustration of removing the drive gear assembly.

Fig. 14.2: Illustration of a screwdriver being used.

Fig. 14.3: Illustration of a cartridge with a slit.

Fig. 14.4: Illustration of direct cartridge removal.

Fig. 14.5, 14.6: Illustrations showing different handpiece models for cartridge removal.

2. Push button Cartridge Replacement

  1. Loosen the small screw on the curving part, move the head cover, and remove the transmission gear [See Fig. 15.1].
  2. Use the attached wrench to remove the cap [See Fig. 15.2].
  3. Use the bur to push out the cartridge.
  4. Insert the new cartridge into the handpiece head, ensuring the small raised part aligns with the groove inside the front of the head for proper fit.

Fig. 15.1: Illustration of loosening a screw and removing a head cover.

Fig. 15.2: Illustration of removing a cap with a wrench.

Fig. 15.3: Illustration of a cap being removed.

Fig. 15.4: Illustration of inserting a new cartridge.

Contra Angle and Motor Connections

  1. Straightly insert the motor into the connection of the contra angle until you hear a "cracking" sound.
  2. Hold the bending machine and pull out the motor; it will be separated.

Fig. 16.1: Illustration showing separation of a handpiece from a motor.

Fig. 16.2: Illustration showing connection of a handpiece to a motor.

Lubrication

After daily use, spray lubricant for 1 to 2 seconds by inserting the lubricant nozzle all the way into the drive air tube.

If work is heavy, lubrication should be implemented at noon and afternoon after using every day.

Lubricate the head, gear, and inside [See Fig. 16.3]. When lubricating the head, please open the latch chuck.

Fig. 16.3: Illustration showing lubricant being sprayed into the drive air tube.

Caution

Fig. 18: Illustration showing lubricant being sprayed from a can into a handpiece using an E spray nozzle.

Clean-head system

  1. Scrub dirt and debris from the handpiece.
  2. Stop the handpiece and wipe it dry. If the dirt cannot be removed from the hole, remove the cartridge and clean by brush.

Fig. 17: Illustration showing cleaning a handpiece head with a brush.

⚠️ Notice! Use only clean water to maintain the clean head system.

Cleaning-bulb

Wipe clean the Cellular optical bulb entry point with an alcohol-immersed cotton swab. Remove all debris and oil.

Sterilization and Cleaning

Fig. 18.2: Illustration showing a handpiece in an autoclave pouch with "135°C SSS" symbol.

Fig. 18.3: Illustrations showing heat sterilization of tubing contra angles at 135°C.

⚠️ NOTE: After treatment or before autoclave, please do the lubrication.

Fig. 18.4: Illustration showing cleaning a handpiece under running water with a brush.

Fig. 18.5: Illustration showing lubricant being sprayed onto a handpiece.

Fig. 18.6: Illustration showing a handpiece in a beaker with a cross symbol (indicating not to immerse).

Fig. 18.7: Illustration showing a handpiece in a "STERILE" pouch.

Fig. 18.8: Illustration showing temperature and pressure symbols (135°C, 275° F).

Trouble Shooting

Trouble Possible Cause Solution
Body of straight head or contra angle rotates during motor running O-ring on the nose of motor wears Change o-ring
Handpiece fail to rotate Contra angle chucking with bur but latch plate fail to be fixed in the proper position. Dirt get into the handpiece. Clean and lubricate, move it by hand

COXO's service center can offer technical assistance to you.

Warranty

COXO grants the user a 12 months guarantee for its complete product range, except ball bearing (3 months guarantee) from the date of invoice issued. Maintenance over the term of guarantee will be at the customer's charge. COXO will not be responsible for damage or injury resulting from:

The technical specifications and dimensions in these instructions are given only as a guide. They may not be subject to any claim.

Standard Symbols

On the instrument label appear standard symbols as follows:

Transportation and deposited condition

Environmental temperature range: -40°C to +50°C; Relative humidity ≤80%; The scope of air pressure: 500hPa to 1060hPa.

Symbols indicate:

Website: www.dentaldeal.ca

Models: CX235 Contra Angle, CX235, Contra Angle, Angle

File Info : application/pdf, 33 Pages, 6.68MB

PDF preview unavailable. Download the PDF instead.

COXO CX235 Series Contra Angle User Manual(5) Corel PDF Engine Version 16.1.0.843

Related Documents

Preview COXO CX235 Serie Håndstykker - Betjeningsmanual
Omfattende betjeningsmanual for COXO CX235 serien af dentale håndstykker, der dækker installation, brug, vedligeholdelse, fejlfinding og sikkerhedsprocedurer for professionelle tandplejere.
Preview COXO CX235 Contra Angle Operation Manual - Dental Handpiece Guide
Comprehensive operation manual for the COXO CX235 Contra Angle dental handpiece, covering specifications, usage, maintenance, sterilization, and troubleshooting.
Preview COXO PT Master Airflow Dental Scaler and Air Polisher User Guide
Comprehensive user guide for the COXO PT Master Airflow Dental Scaler and Air Polisher, detailing operation, maintenance, and safety features for dental professionals. Learn about its heated water function and optimal usage for dental hygiene.
Preview COXO CX207 High-Speed Dental Turbine Handpiece User Manual
This manual provides detailed instructions for the safe and effective operation, maintenance, and troubleshooting of the COXO CX207 series high-speed dental turbine handpiece. It covers technical specifications, assembly, usage guidelines, and warranty information.
Preview COXO C-Smart Nova Endo Motor User Manual - Dental Equipment Guide
Comprehensive user manual for the COXO C-Smart Nova dental endo motor. Learn about safety, operation, maintenance, troubleshooting, and EMC compliance for this professional dental instrument.
Preview COXO C-SMART-I PILOT Endo Motors User Manual
User manual for the COXO C-SMART-I PILOT Endo Motors, detailing its features, operation, safety precautions, and technical specifications for dental professionals.
Preview User Manual: COXO Air Abrasion System with Spray (CA-1)
This user manual provides detailed instructions for the COXO Air Abrasion System with Spray (CA-1), covering its operation, maintenance, cleaning, disinfection, sterilization, troubleshooting, and technical specifications. It is intended for use by qualified dentists.
Preview COXO SC-Pro Dental Endo Motor and Apex Locator User Manual
Comprehensive user manual for the COXO SC-Pro dental endodontic motor and apex locator, detailing its features, operation, and maintenance. Covers file selection, apex locator functions, multi-function mode, battery management, and troubleshooting.