USER GUIDE

SCC-TC Series Thermocouple Input Modules

Introduction

The SCC-TC Series thermocouple input modules, SCC-TC01 and SCC-TC02, accept input signals from B-, E-, J-, K-, N-, R-, S-, and T-type thermocouples. Each module contains thermistor circuitry powered by 2.5 V, to compensate for cold-junction effects and can detect open thermocouple circuits. Each module has one thermocouple input channel, which consists of a gain of 100 differential amplifier and a dual-pole 2 Hz filter.

? Note: Alternatively, you can use the SCC-TC02 as a 2 Hz lowpass bandwidth ±100 mV input module.

Conventions

The following conventions are used in this guide:

Typography

Bold text: Denotes items that you must select in the software, such as menu items and dialog box options. Bold text also denotes parameter names.

Italic text: Denotes variables, emphasis, a cross-reference, or an introduction to a key concept. Italic text also denotes text that is a placeholder for a word or value that you must supply.

Monospace text:
Denotes text or characters that you should enter from the keyboard, sections of code, programming examples, and syntax examples. This font is also used for the proper names of disk drives, paths, directories, programs, subprograms, subroutines, device names, functions, operations, variables, filenames, and extensions.

Glossary

SC-2345
Refers to both the SC-2345 connector block and the SC-2345 configurable connector.
SCC
Refers to any SCC series signal conditioning module.

What You Need to Get Started

To set up and use the SCC-TC0X, you need the following items:

Hardware

Software

Documentation

You can download needed documents from ni.com/manuals.

Tools

Device Specific Information

For general SCC module installation and signal connection information, and information about the SC-2345 or SC-2350 carriers, refer to the SCC Quick Start Guide.

Installing the Module

⚠️ Caution: Refer to the Read Me First: Safety and Radio-Frequency Interference document before removing equipment covers or connecting/disconnecting any signal wires.

You can plug the SCC-TC0X into any analog input socket on the SCC carrier. The socket you select determines which E/M Series DAQ device channels receive the SCC-TC0X signals.

For single-stage input conditioning, plug the SCC-TC0X into any socket J(X +1), where X is 0 to 7, and connect the input signals to the module as described in the Connecting the Input Signals section.

If you use the SCC-TC0X in a dual-stage configuration, the SCC-TC0X must be the first-stage module. Plug it into any socket J(X+9) and plug the second-stage SCC into socket J(X+1), where X is 0 to 7. Connect the input signals to SCC-TC0X as described in the Connecting the Input Signals section. The SC-2345 connects the output signals of the first-stage SCC to the inputs of the second-stage SCC. An example of dual-stage conditioning is an SCC-TC0X followed by an SCC-LP01 lowpass filter module.

Sockets J9 to J16 are also available for digital input/output (DIO) conditioning or control. Refer to the SC-2345/2350 User Manual for more information on configuring, connecting, and installing SCC modules.

? Note: SCC-TC0X dual-stage configuration is only available in NI-DAQmx 7.1 or later.

Connecting the Input Signals

? Note: The signal names have changed. Refer to ni.com/info and enter 'rdtntg' to confirm the signal names.

The SCC-TC01 has a two-prong uncompensated thermocouple miniconnector that accepts any miniature or subminiature two-prong male thermocouple connector. The SCC-TC02 has a three-position screw-terminal connector that accepts 28 to 16 AWG thermocouple wires. The screw-terminal connector provides a ground connection for shielded thermocouples. Otherwise, the two modules function identically.

To use an SCC-TC01, plug the male thermocouple connector into the module.

The SCC-TC02 accepts up to three signals: TC+, TC−, and GND. TC+ is the positive thermocouple lead and TC− is the negative thermocouple lead. The GND terminal connects to AI GND on the E/M Series DAQ device.

The SCC-TC0X has a 10 MΩ bias resistor connected from the negative thermocouple input to ground. This resistor allows the thermocouple to be ground-referenced or floating without requiring external bias resistors connected to ground.

The amplified thermocouple signal and cold-junction sensor signal are measured by E/M Series DAQ device channel X and channel X+8 respectively, where X is 0 to 7 depending on where you plug the SCC-TC0X. Refer to Figure 1 for SCC-TC0X signal connections.

1 For ANSI color-coded J-type thermocouples, the red wire is negative and the white wire is positive. Refer to the thermocouple data sheet if possible. You can find information about other color-coding schemes in the NI KnowledgeBase at ni.com/support.

Figure 1. SCC-TCOX Signal Connections

Block diagram illustrating the SCC-TCOX module connected to an E/M Series DAQ Device. It shows the thermocouple source input (TC+, TC-, GND), internal circuitry including resistors (10 MΩ, 10 kΩ), a 2-Pole Filter/Buffer Stage, an Offset Calibrator, and a Ref 5V source. The outputs are connected to the DAQ device's AI (X) and AI (X+8) channels, with AI SENSE and AI GND also indicated. The diagram also shows a reference voltage source (LM 4040 2.5 V) for the cold-junction sensor.

For more information about how to configure the SCC-TC0X module using NI-DAQmx, refer to the SCC Quick Start Guide.

Using the TCOX

Scaling Voltage Measurements

? Note: NI-DAQmx includes thermocouple and thermistor conversion utilities that implement the conversions required in steps 3 and 5 of the following procedure. Refer to your software documentation for more information on these utilities.

Your software environment may return only voltage measurements from E/M Series DAQ devices. If so, you must convert voltage measurements to temperature measurements. To make this conversion, complete the following steps:

  1. Measure the thermocouple voltage.
    1. Read the thermocouple channel on the E/M Series DAQ device VERIES [CH(X)].
    2. Calculate the thermocouple voltage by using the following formula:

      $$ V_{TC} = \frac{V_{ERIES}}{100} $$

      where VTC is thermocouple voltage. VERIES is E/M Series DAQ device voltage. This step provides proper scaling for the thermocouple amplifier in the SCC-TC0X.
  2. Measure the reference-junction (cold-junction) temperature.
    1. Read the thermistor voltage [AI (X+8)].
    2. Convert the thermistor voltage to cold-junction temperature using the formula in the Cold-Junction Sensor section.
  3. Calculate the cold-junction compensation voltage by converting the cold-junction temperature from step 2 to a thermocouple voltage. Use the polynomial expressions applicable to the type of thermocouple you are using.
  4. Apply the cold-junction compensation to the thermocouple reading by adding the cold-junction compensation voltage from step 3 to VTC.
  5. Calculate the thermocouple temperature by converting the voltage result from step 4 to a temperature. Use the polynomial expressions applicable to the type of thermocouple you are using. This calculation gives you a linearized temperature measurement.

? Note: Polynomials are from NIST Monograph 175.

Uncompensated Connectors and Accuracy

If you are using an SCC-TC02 with an uncompensated SCC panelette, temperature gradients between the module and the junctions on the panelette affect the accuracy of the measurements.

Detecting Open Thermocouples

The SCC-TC0X contains a 10 MΩ pull-up resistor that connects to +5 V to detect open thermocouples. To determine if you have an open thermocouple, check whether the corresponding E/M Series DAQ device channel is saturated. The pull-up and bias resistors saturate the channel by applying +2.5 V at the input of an open channel. This results in saturation to either the positive or negative rails of the EM Series DAQ device (±10 V).

Errors Due to Open-Thermocouple Detection Circuitry

The open-thermocouple detection circuitry can cause measurement errors. These errors are the results of common-mode voltage at the input of the SCC and current leakage into the signal leads. The 10 MΩ bias resistor in the SCC-TC0X causes this error to be negligible. With the 10 MΩ bias resistor connected to ground and the 10 MΩ pull-up resistor connected to +5 VDC, a current leakage of approximately 0.25 μA (5 V/20 MΩ) flows into the unbroken floating thermocouple. Long thermocouple leads result in larger voltage drops due to lead resistance.

For example, if you have a 24 AWG J-type thermocouple that is 20 ft long, a voltage drop of approximately:

$$ 4 ext{ μV} = (0.145 ext{ Ω/ft} + 0.658 ext{ Ω/ft}) imes 20 ext{ ft} imes 0.25 ext{ μA} $$

can develop in the thermocouple, which corresponds to an error of 0.09 °C. With 10 MΩ pull-up and bias resistors, a common-mode voltage of +2.5 VDC develops if the thermocouple is floating. The common-mode rejection of the SCC-TC0X is sufficiently high, which results in the offset voltage being negligible in most applications.

If your application demands extremely high accuracy, you can eliminate these errors by calibrating the system. Refer to the Calibrating the SCC-TC0X Using a Thermocouple Calibrator section for more information.

Cold-Junction Sensor

The cold-junction sensor voltage output varies from 1.91 V to 0.58 V over a 0 to 55 °C temperature range.

? Note: NI-DAQmx includes thermistor conversion utilities that implement the equations listed below. Refer to your software documentation for more information on these utilities.

You can use the following formulas to convert the cold-junction sensor voltage to cold-junction temperature:

$$ T(°C) = T_K - 273.15 $$

where TK is the temperature in degrees kelvin.

$$ T_K = \frac{1}{a + b(\ln R_T) + c(\ln R_T)^3} $$

where:

$$ R_T = 5,000 \left( \frac{V_{TEMPOUT}}{2.5 - V_{TEMPOUT}} \right) $$

where VTEMPOUT is the output voltage of the cold-junction sensor.

$$ T(°F) = \left[ \frac{T(°C)}{5} \right]^9 + 32 $$

where T(°F) and T(°C) are the temperature readings in degrees Fahrenheit and Celsius, respectively.

? Note: VTEMPOUT varies from 1.91 V (at 0 °C) to 0.58 V (at 55 °C). For best resolution, use the maximum gain for this range on the analog input channel.

For the cold-junction sensor measurement accuracy, refer to the Specifications section. For more information about how to configure the SCC-TC0X using NI-DAQmx, refer to the SCC Quick Start Guide.

Calibrating System Offsets

To calibrate the offset voltage of the SCC-TC0X in the system using the E/M Series DAQ device, complete the following steps. The E/M Series DAQ device must be powered for 30 minutes before you calibrate the offset voltage.

  1. Select the desired channel and gain on the E/M Series DAQ device.
  2. Short the inputs on the SCC-TC0X screw terminals or miniplug.
  3. While acquiring data on the selected channel, use a screwdriver to adjust the potentiometer protruding through the top of the module until you read 0 VDC.

Calibrating the SCC-TC0X Using a Thermocouple Calibrator

To calibrate the system using a thermocouple calibrator, complete the following steps:

  1. Make sure that the thermocouple connections at both the SCC and the thermocouple calibrator are at the same temperature.
  2. Connect the thermocouple calibrator to the SCC-TC0X. For best results, use thermocouple wire of the same length and type that you use in the thermocouple.
  3. Set the thermocouple calibrator to the required calibration temperature.
  4. Measure the calibrator voltage using the SCC-TC0X.
  5. Convert the measured calibrator voltage to a temperature measurement as described in the Scaling Voltage Measurements section.
  6. Adjust the potentiometer on the top of the SCC-TC0X so that the measured temperature is equal to the calibration temperature.

Specifications

These specifications are typical at 25 °C unless otherwise stated.

Analog Input

Amplifier Characteristics

Transfer Characteristics

1 Temperature range is 23 °C ±5 °C.

2 Temperature range is 0 to 50 °C.

Measurement Accuracy

Thermocouple Type Temperature Range (°C) Maximum Inaccuracy (±°C) Typical Inaccuracy (±°C)
B400 to 6002.50.31
600 to 1,80020.13
E-200 to -10021.76
-100 to 60011
600 to 1,0001.50.6
J-200 to -10021.6
-100 to 50010.9
500 to 1,1001.50.72
K-200 to -1002.51.25
-100 to 6001.250.67
600 to 1,20020.45
N-200 to -1002.52
-100 to 1,30020.7
R-50 to 031.24
0 to 10020.82
100 to 1,6001.750.4
S-50 to 031.3
0 to 1,40021
1,400 to 1,6002.51
T-200 to -1002.52
-100 to 40010.9

1 Total system measurement error for operating temperature within ±5 °C of calibration temperature. Includes PCI/AT-MIO-16XE-50 one-year accuracy specification of 0.01% ±412 μV. SCC-TC01/02 accuracy specification of 0.08% ±5 µV, and reference junction measurement accuracy of 0.5 °C. Assumes averaging. Non-averaged, single-point reading has an additional uncertainty (up to ±0.1 °C for J-type thermocouple).

Cold-Junction Sensor (Specifications)

? Note: The accuracy specification includes the combined effects of the temperature sensor accuracy and the temperature difference between the temperature sensor and any thermocouple connector. The temperature sensor accuracy includes component tolerances, temperature drifts, and self-heating effects. It does not include measurement device errors.

Open Thermocouple Detection (Specifications)

Power Requirement

Physical

SCC-TC01 Dimensions

Diagram showing the physical dimensions of the SCC-TC01 module: 1.87 cm (0.74 in.) height, 7.60 cm (2.99 in.) width, and 3.55 cm (1.4 in.) depth.

SCC-TC02 Dimensions

Diagram showing the physical dimensions of the SCC-TC02 module: 1.87 cm (0.74 in.) height, 7.93 cm (3.12 in.) width, and 3.55 cm (1.4 in.) depth.

Maximum Working Voltage

Maximum working voltage refers to the signal voltage plus the common-mode voltage.

Each input must remain within ±12 V of chassis ground.

Measurement Category I

Environmental

Safety

This product is designed to meet the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use:

? Note: For UL and other safety certifications, refer to the product label or visit ni.com/certification, search by model number or product line, and click the appropriate link in the Certification column.

Electromagnetic Compatibility

This product is designed to meet the requirements of the following standards of EMC for electrical equipment for measurement, control, and laboratory use:

? Note: For EMC compliance, operate this device according to product documentation. In addition, all covers and filler panels must be installed.

CE Compliance

This product meets the essential requirements of applicable European Directives, as amended for CE marking, as follows:

? Note: Refer to the Declaration of Conformity (DoC) for this product for any additional regulatory compliance information. To obtain the DoC for this product, visit ni.com/certification, search by model number or product line, and click the appropriate link in the Certification column.

Waste Electrical and Electronic Equipment (WEEE)

EU Customers: At the end of their life cycle, all products must be sent to a WEEE recycling center. For more information about WEEE recycling centers and National Instruments WEEE initiatives, visit ni.com/environment/weee.htm.

I/O Connector Pin Assignments

Figure 4 shows the I/O connector pins on the bottom of the module.

Diagram showing the bottom view of the SCC Module I/O connector with pins labeled 1 through 5, corresponding to Pin 1, Pin 2, PWB Key, Pin 19, and Pin 20.

Figure 4. SCC Module Bottom View

Pin Number Signal
1AI (X)
2
3
4AI (X+8)
5
6DAQ Device AI GND
7
8
9
10
11A GND
12REF 5 V
13+15 V
14-15 V
15
16
17
18
19
20

Table 1 lists the signal connection corresponding to each pin. AI (X) and AI (X+8) are the analog input signal channels of the E/M Series DAQ device. AI GND is the analog input ground signal and is the reference for AI (X) and AI (X+8). A GND is the reference for the ±15 V supplies and REF 5V. AI GND and A GND connect to the SC-2345/2350 at the SCC-PWR connector.

National Instruments, NI, ni.com, and LabVIEW are trademarks of National Instruments Corporation. Refer to the Terms of Use section on ni.com/legal for more information about National Instruments trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National Instruments products, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your CD, or ni.com/patents.

© 2002-2006 National Instruments Corporation. All rights reserved.

PDF preview unavailable. Download the PDF instead.

9ea19291-93d1-44b3-b791-c2ce1c0580fc Acrobat Distiller 7.0.5 (Windows)

Related Documents

Preview Начало работы с системами согласования сигналов National Instruments SCC
Руководство по установке и настройке модулей и шасси National Instruments SCC (SC-2345/2350). Описывает подключение, конфигурацию и использование с NI-DAQmx для согласования сигналов.
Preview National Instruments C Series Analog Input Modules: NI 9201, NI 921x, NI 9221, NI 923x
Overview of National Instruments C Series analog input modules, including NI 9201, NI 921x, NI 9221, and NI 923x. Details features, specifications, system compatibility, and accessories for high-accuracy measurements in CompactRIO and CompactDAQ systems.
Preview NI-9218 Getting Started Guide - National Instruments
A comprehensive guide for the National Instruments NI-9218 module, covering connector types, pinouts, supported measurement types (±16 V, ±65 mV, Full-Bridge, IEPE, ±20 mA, ±60 V, Half-Bridge, Quarter-Bridge), signal conditioning, TEDS support, and filtering.
Preview NI USB-621x User Manual: National Instruments DAQ M Series
Explore the National Instruments USB-621x M Series Data Acquisition (DAQ) devices with this comprehensive user manual. Learn about installation, configuration, analog input/output, digital I/O, counters, and integration with NI-DAQmx software for robust data acquisition and control applications.
Preview National Instruments USB-6001 DAQ Getting Started Guide and Services
A comprehensive guide to installing, connecting, and configuring National Instruments USB-6001 bus-powered USB DAQ devices. Includes software setup, programming basics, troubleshooting, and information on Apex Waves' related services for NI hardware.
Preview NI USB-6001/6002/6003 Quick Start Guide: Low-Cost DAQ USB Devices
A quick start guide for National Instruments USB-6001, USB-6002, and USB-6003 low-cost DAQ USB devices. Learn how to install software, connect hardware, and perform your first measurement using NI MAX.
Preview NI USB-6001/6002/6003 User Guide: Low-Cost DAQ USB Device
Comprehensive user guide for National Instruments USB-6001, USB-6002, and USB-6003 low-cost DAQ USB devices. Covers installation, setup, safety guidelines, hardware features, analog and digital I/O, signal descriptions, and software integration with NI-DAQmx and other development environments.
Preview NI-DAQ User Manual for PC Compatibles - National Instruments
Comprehensive guide to the NI-DAQ software for PC compatibles, version 6.9. Covers data acquisition, software calibration, event messaging, analog and digital I/O, and counter/timer functions. Includes details on hardware support, application tips, and troubleshooting.