Apogee Instruments Extended Range PFD Sensor

Owner's Manual

Models SQ-622 and SQ-625

Rev: 28-Oct-2020

Apogee Instruments, Inc. | 721 West 1800 North, Logan, Utah 84321, USA | TEL: (435) 792-4700 | FAX: (435) 787-8268 | WEB: APOGEEINSTRUMENTS.COM

Copyright © 2021 Apogee Instruments, Inc.

Certificate of Compliance

EU Declaration of Conformity

This declaration of conformity is issued under the sole responsibility of the manufacturer:

Apogee Instruments, Inc.
721 W 1800 N
Logan, Utah 84321
USA

for the following product(s):

Models: SQ-622, SQ-625
Type: Extended Range PFD Sensor

The object of the declaration described above is in conformity with the relevant Union harmonization legislation:

  • 2014/30/EU: Electromagnetic Compatibility (EMC) Directive
  • 2011/65/EU: Restriction of Hazardous Substances (RoHS 2) Directive
  • 2015/863/EU: Amending Annex II to Directive 2011/65/EU (RoHS 3)

Standards referenced during compliance assessment:

  • EN 61326-1:2013: Electrical equipment for measurement, control and laboratory use – EMC requirements
  • EN 50581:2012: Technical documentation for the assessment of electrical and electronic products with respect to the restriction of hazardous substances

Please be advised that based on the information available to us from our raw material suppliers, the products manufactured by us do not contain, as intentional additives, any of the restricted materials including lead (see note below), mercury, cadmium, hexavalent chromium, polybrominated biphenyls (PBB), polybrominated diphenyls (PBDE), bis(2-ethylhexyl) phthalate (DEHP), butyl benzyl phthalate (BBP), dibutyl phthalate (DBP), and diisobutyl phthalate (DIBP). However, please note that articles containing greater than 0.1% lead concentration are RoHS 3 compliant using exemption 6c.

Further note that Apogee Instruments does not specifically run any analysis on our raw materials or end products for the presence of these substances, but rely on the information provided to us by our material suppliers.

Signed for and on behalf of:
Apogee Instruments, January 2021

Bruce Bugbee
President
Apogee Instruments, Inc.

Introduction

Radiation that drives photosynthesis is called photosynthetically active radiation (PAR) and is typically defined as total radiation across a range of 400 to 700 nm. PAR is almost universally quantified as photosynthetic photon flux density (PPFD) in units of micromoles per square meter per second (µmol m⁻² s⁻¹, equal to microEinsteins per square meter per second) summed from 400 to 700 nm (total number of photons from 400 to 700 nm). However, ultraviolet and far red photons outside the defined PAR range of 400-700 nm can also contribute to photosynthesis and influence plant responses (e.g., flowering).

Sensors that measure PPFD are often called quantum sensors due to the quantized nature of radiation. A quantum refers to the minimum quantity of radiation, one photon, involved in physical interactions (e.g., absorption by photosynthetic pigments). In other words, one photon is a single quantum of radiation. Sensors that function like traditional quantum sensors, but measure a wider range of wavelengths can be thought of as an ‘extended range’ quantum sensor.

Typical applications of traditional quantum sensors include incoming PPFD measurement over plant canopies in outdoor environments or in greenhouses and growth chambers, and reflected or under-canopy (transmitted) PPFD measurement in the same environments. The Extended Range PFD Sensor detailed in this manual uses a detector that is sensitive to radiation up to about 1100 nm, well beyond the range of wavelengths that influence photosynthesis and plant responses. This means this particular sensor should only be used for photon flux density measurements under LEDs.

Apogee Instruments SQ-600 series Extended Range PFD Sensors consist of a cast acrylic diffuser (filter), photodiode, and signal processing circuitry mounted in an anodized aluminum housing. A cable to connect the sensor to a measurement device is also included. SQ-600 series Extended Range PFD Sensors are designed for continuous photon flux density measurements in indoor environments under LEDs. SQ-620 models output an analog signal that is directly proportional to photon flux density. The analog signal from the sensor is directly proportional to radiation incident on a planar surface (does not have to be horizontal), where the radiation emanates from all angles of a hemisphere.

Sensor Models

This manual covers the amplified voltage output Extended Range PFD Sensors, models SQ-622 and SQ-625 (in bold below). Additional models are covered in their respective manuals.

ModelSignal
SQ-620Self-powered
SQ-6220-2.5 V
SQ-6244-20 mA
SQ-6250-5 V
SQ-626USB
SQ-627SDI-12
SQ-628Modbus

Sensor Description: The sensor is a compact, cylindrical device made of anodized aluminum with a diffuser on top. It features labels indicating "Extended Range Quantum", "SQ-622 SN: 1001", "CE", and "Apogee Instruments, Inc. Made in USA".

Sensor model number and serial number are located on the bottom of the sensor. If the manufacturing date of a specific sensor is required, please contact Apogee Instruments with the serial number of the sensor.

Specifications

SQ-622-SSSQ-625-SS
Power Supply5 to 24 V DC5.5 to 24 V DC
Current Draw12 V is 57 µA
Sensitivity0.625 mV per µmol m⁻² s⁻¹1.25 mV per µmol m⁻² s⁻¹
Calibration Factor1.6 µmol m⁻² s⁻¹ per mV0.8 µmol m⁻² s⁻¹ per mV
Calibration Uncertainty± 5% (see Calibration Traceability below)
Output Range0 to 2.5 V0 to 5 V
Measurement Range0 to 4000 µmol m⁻² s⁻¹
Measurement RepeatabilityLess than 0.5%
Long-term Drift (Non-stability)Less than 2% per year
Non-linearityLess than 1% (up to 4000 µmol m⁻² s⁻¹)
Response TimeLess than 1 ms
Field of View180°
Spectral Range340 to 1040 nm ± 5 nm (wavelengths where response is greater than 50%; see Spectral Response below)
Directional (Cosine) Response± 2 % at 45° zenith angle, ± 5 % at 75° zenith angle (see Directional Response below)
Azimuth ErrorLess than 0.5%
Tilt ErrorLess than 0.5%
Temperature Response-0.11 ± 0.04% per C
Uncertainty in Daily TotalLess than 5%
HousingAnodized aluminum body with acrylic diffuser
IP RatingIP68
Operating Environment-40 to 70 C; 0 to 100 % relative humidity; can be submerged in water up to depths of 30 m
Dimensions30.5 mm diameter, 37 mm height
Mass (with 5 m of cable)140 g
Cable5 m of two conductor, shielded, twisted-pair wire; TPR jacket; pigtail lead wires; stainless steel (316), M8 connector
Warranty4 years against defects in materials and workmanship

Calibration Traceability: Apogee Instruments SQ-600 series PFD sensors are calibrated through side-by-side comparison to the mean of four transfer standard PFD sensors under a reference lamp. The transfer standard PFD sensors are recalibrated with a quartz halogen lamp traceable to the National Institute of Standards and Technology (NIST).

Spectral Response: A line graph shows the mean spectral response measurements of six replicate Apogee SQ-600 series Extended Range PFD Sensors. Spectral response measurements were made at 10 nm increments across a wavelength range of 300 to 1100 nm. The graph displays relative response to photons on the y-axis (0.0 to 1.1) versus wavelength in nanometers on the x-axis (300 to 1100 nm). The response peaks at approximately 1.0 between 400 nm and 950 nm, then declines. Dashed vertical lines are present at 700 nm and 950 nm.

Cosine Response: A diagram illustrates the directional, or cosine, response, defined as the measurement error at a specific angle of radiation incidence. The diagram shows that for light incident at 0° zenith angle, the error is minimal. At 45° zenith angle, the error is approximately ±2%, and at 75° zenith angle, the error increases to approximately ±5%. This indicates that the sensor's accuracy is highest for light directly overhead and decreases for light incident at oblique angles.

Deployment and Installation

Mount the sensor to a solid surface using the provided nylon mounting screw. To accurately measure photon flux density incident on a horizontal surface, the sensor must be level. An Apogee Instruments model AL-100 leveling plate is recommended for this purpose. To facilitate mounting on a cross arm, an Apogee Instruments model AL-120 mounting bracket is recommended.

Mounting Accessories: Images show the Apogee Instruments AL-100 leveling plate (a flat circular plate with mounting holes) and the AL-120 mounting bracket (a clamp-style bracket for cross arm mounting). Both are depicted with a sensor attached via a nylon screw (10-32 x 3/8).

To minimize azimuth error, the sensor should be mounted with the cable pointing toward true north in the northern hemisphere or true south in the southern hemisphere. Azimuth error is typically less than 0.5%, but it is easy to minimize by proper cable orientation.

Azimuth Orientation: Diagrams illustrate correct sensor orientation. In the Northern Hemisphere, the cable should point North (N). In the Southern Hemisphere, the cable should point South (S) to minimize azimuth error.

In addition to orienting the cable to point toward the nearest pole, the sensor should also be mounted such that obstructions (e.g., weather station tripod/tower or other instrumentation) do not shade the sensor. Once mounted, the blue cap should be removed from the sensor. The blue cap can be used as a protective covering for the sensor when it is not in use.

Cable Connectors

Apogee started offering cable connectors on some bare-lead sensors in March 2018 to simplify the process of removing sensors from weather stations for calibration (the entire cable does not have to be removed from the station and shipped with the sensor).

The ruggedized M8 connectors are rated IP68, made of corrosion-resistant marine-grade stainless-steel, and designed for extended use in harsh environmental conditions.

Connector Attachment: Cable connectors are attached directly to the sensor head.

Instructions:

Pins and Wiring Colors: All Apogee connectors have six pins, but not all pins are used for every sensor. There may also be unused wire colors inside the cable. To simplify datalogger connection, unused pigtail lead colors are removed at the datalogger end of the cable.

If a replacement cable is required, please contact Apogee directly to ensure ordering the proper pigtail configuration.

Alignment: When reconnecting a sensor, arrows on the connector jacket and an aligning notch ensure proper orientation.

Connector Alignment: An image shows the inside of the M8 connector, highlighting the pins and a reference notch that ensures proper alignment before tightening.

Disconnection for extended periods: When disconnecting the sensor for an extended period of time from a station, protect the remaining half of the connector still on the station from water and dirt with electrical tape or other method.

Calibration Shipment: When sending sensors in for calibration, only send the sensor head.

Tightening: Connectors are designed to be firmly finger-tightened only. There is an o-ring inside the connector that can be overly compressed if a wrench is used. Pay attention to thread alignment to avoid cross-threading. When fully tightened, 1-2 threads may still be visible.

WARNING: Do not tighten the connector by twisting the black cable or sensor head, only twist the metal connector (yellow arrows).

Finger-tighten firmly: An image shows the sensor head with the cable and a warning graphic with yellow arrows pointing to the metal connector, emphasizing to twist only the connector.

Operation and Measurement

Connect the sensor to a measurement device (meter, datalogger, controller) capable of measuring and displaying or recording a millivolt signal (an input measurement range of approximately 0-2.5 V (SQ-622) or 0-5 V (SQ-625) is required to cover the entire range of PFD from the sun). To maximize measurement resolution and signal-to-noise ratio, the input range of the measurement device should closely match the output range of the extended range PFD sensor. DO NOT connect the sensor to a power source greater than 24 V DC.

Wiring for SQ-622 and SQ-625:

A diagram shows the sensor cable with four wires:

  • White: Positive (signal from sensor)
  • Red: Input power (SQ-622 5-24 V DC, SQ-625 5.5-24 V DC)
  • Black: Ground (for sensor signal and input power)
  • Clear: Shield/Ground

Sensor Calibration

Apogee SQ-622 and SQ-625 Extended Range PFD Sensors have standard calibration factors of exactly:

  • SQ-622: 1.6 µmol m⁻² s⁻¹ per mV
  • SQ-625: 0.8 µmol m⁻² s⁻¹ per mV

Multiply the calibration factor by the measured mV signal to convert sensor output to photon flux density in units of µmol m⁻² s⁻¹:

Calibration Factor (0.8 µmol m⁻² s⁻¹ per mV) * Sensor Output Signal (mV) = Photon Flux Density (µmol m⁻² s⁻¹)

Example Calculation:

Calibration Factor (0.8 µmol m⁻² s⁻¹ per mV) * 2500 mV = 2000 µmol m⁻² s⁻¹

LED Output Example: An illustration shows the sensor head with five colored wires, each representing an output signal corresponding to a specific photon flux density. For example, an output signal of 2500 mV (for the 0-5 V option) or 1250 mV (for the 0-2.5 V option) corresponds to a photon flux density of 2000 µmol m⁻² s⁻¹ when multiplied by the respective calibration factor.

  • SQ-622 Sensor Output: 1250 mV
  • Conversion Factor: 1.6 µmol m⁻² s⁻¹ per mV
  • SQ-625 Sensor Output: 2500 mV
  • Conversion Factor: 0.8 µmol m⁻² s⁻¹ per mV

Maintenance and Recalibration

Dust or organic deposits are best removed using water or window cleaner and a soft cloth or cotton swab. Salt deposits should be dissolved with vinegar and removed with a soft cloth or cotton swab.

Blocking of the optical path between the target and detector can cause low readings. Occasionally, accumulated materials on the diffuser of the upward-looking radiometer and in the apertures of the downward-looking radiometer can block the optical path in three common ways:

  1. Moisture or debris on the diffuser (upward-looking) or in the apertures (downward-looking).
  2. Dust during periods of low rainfall.
  3. Salt deposit accumulation from evaporation of sea spray or sprinkler irrigation water.

Apogee Instruments upward-looking sensors have a domed diffuser and housing for improved self-cleaning from rainfall, but active cleaning may be necessary. Dust or organic deposits are best removed using water, or window cleaner, and a soft cloth or cotton swab. Salt deposits should be dissolved with vinegar and removed with a cloth or cotton swab. Salt deposits cannot be removed with solvents such as alcohol or acetone. Use only gentle pressure when cleaning the diffuser with a cotton swab or soft cloth, to avoid scratching the outer surface. The solvent should be allowed to do the cleaning, not mechanical force. Never use an abrasive material or cleaner on the diffuser.

It is recommended that two-band radiometers be recalibrated every two years. See the Apogee webpage for details regarding return of sensors for recalibration (https://www.apogeeinstruments.com/tech-support-recalibration-repairs/).

Troubleshooting and Customer Support

Independent Verification of Functionality

Apogee SQ-622 and SQ-625 Extended Range PFD Sensors provide an amplified voltage output that is proportional to incident photon flux density for the 340-1040 nm wavelength range. A quick and easy check of sensor functionality can be determined using a DC power supply and a voltmeter. Power the sensor with a DC voltage by connecting the positive voltage signal to the red wire from the sensor and the negative (or common) to the black wire from the sensor. Use the voltmeter to measure across the white wire (signal output) and black wire (signal ground). Direct the sensor head toward a light source and verify the sensor provides a signal. Increase and decrease the distance from the sensor head to the light source to verify that the signal changes proportionally (decreasing signal with increasing distance and increasing signal with decreasing distance). Blocking all radiation from the sensor should force the sensor signal to zero.

Compatible Measurement Devices (Dataloggers/Controllers/Meters)

SQ-600 Extended Range PFD Sensors are calibrated with a standard calibration factor of 1.6 µmol m⁻² s⁻¹ per mV (SQ-622) or 0.8 µmol m⁻² s⁻¹ per mV (SQ-625), yielding a sensitivity of 0.6 mV per µmol m⁻² s⁻¹ (SQ-622) or 1.3 mV per µmol m⁻² s⁻¹ (SQ-625). Thus, a compatible measurement device (e.g., datalogger or controller) should have resolution of at least 0.6 mV (SQ-622) or 1.3 mV (SQ-625) in order to provide photon flux density resolution of 1 µmol m⁻² s⁻¹.

An example datalogger program for Campbell Scientific dataloggers can be found on the Apogee webpage at https://www.apogeeinstruments.com/downloads/#datalogger.

Cable Length

When the sensor is connected to a measurement device with high input impedance, sensor output signals are not changed by shortening the cable or splicing on additional cable in the field. Tests have shown that if the input impedance of the measurements device is greater than 1 mega-ohm there is negligible effect on the calibration, even after adding up to 100 m of cable. All Apogee sensors use shielded, twisted pair cable to minimize electromagnetic interference. For best measurements, the shield wire must be connected to an earth ground. This is particularly important when using the sensor with long lead lengths in electromagnetically noisy environments.

Modifying Cable Length: See Apogee webpage for details on how to extend sensor cable length: (https://www.apogeeinstruments.com/how-to-make-a-weatherproof-cable-splice/).

Return and Warranty Policy

Return Policy

Apogee Instruments will accept returns within 30 days of purchase as long as the product is in new condition (to be determined by Apogee). Returns are subject to a 10% restocking fee.

Warranty Policy

What is Covered: All products manufactured by Apogee Instruments are warranted to be free from defects in materials and craftsmanship for a period of four (4) years from the date of shipment from our factory. To be considered for warranty coverage an item must be evaluated by Apogee.

Products not manufactured by Apogee (spectroradiometers, chlorophyll content meters, EE08-SS probes) are covered for a period of one (1) year.

What is Not Covered: The customer is responsible for all costs associated with the removal, reinstallation, and shipping of suspected warranty items to our factory.

The warranty does not cover equipment that has been damaged due to the following conditions:

  1. Improper installation or abuse.
  2. Operation of the instrument outside of its specified operating range.
  3. Natural occurrences such as lightning, fire, etc.
  4. Unauthorized modification.
  5. Improper or unauthorized repair.

Please note that nominal accuracy drift is normal over time. Routine recalibration of sensors/meters is considered part of proper maintenance and is not covered under warranty.

Who is Covered: This warranty covers the original purchaser of the product or other party who may own it during the warranty period.

What Apogee Will Do: At no charge Apogee will:

  1. Either repair or replace (at our discretion) the item under warranty.
  2. Ship the item back to the customer by the carrier of our choice.

Different or expedited shipping methods will be at the customer's expense.

How To Return An Item:

  1. Please do not send any products back to Apogee Instruments until you have received a Return Merchandise Authorization (RMA) number from our technical support department by submitting an online RMA form at https://www.apogeeinstruments.com/tech-support-recalibration-repairs/. We will use your RMA number for tracking of the service item. Call (435) 245-8012 or email techsupport@apogeeinstruments.com with questions.
  2. For warranty evaluations, send all RMA sensors and meters back in the following condition: Clean the sensor's exterior and cord. Do not modify the sensors or wires, including splicing, cutting wire leads, etc. If a connector has been attached to the cable end, please include the mating connector – otherwise the sensor connector will be removed in order to complete the repair/recalibration. Note: When sending back sensors for routine calibration that have Apogee's standard stainless-steel connectors, you only need to send the sensor with the 30 cm section of cable and one-half of the connector. We have mating connectors at our factory that can be used for calibrating the sensor.
  3. Please write the RMA number on the outside of the shipping container.
  4. Return the item with freight pre-paid and fully insured to our factory address shown below. We are not responsible for any costs associated with the transportation of products across international borders.
    Apogee Instruments, Inc.
    721 West 1800 North Logan, UT
    84321, USA
  5. Upon receipt, Apogee Instruments will determine the cause of failure. If the product is found to be defective in terms of operation to the published specifications due to a failure of product materials or craftsmanship, Apogee Instruments will repair or replace the items free of charge. If it is determined that your product is not covered under warranty, you will be informed and given an estimated repair/replacement cost.

Products Beyond the Warranty Period: For issues with sensors beyond the warranty period, please contact Apogee at techsupport@apogeeinstruments.com to discuss repair or replacement options.

Other Terms

The available remedy of defects under this warranty is for the repair or replacement of the original product, and Apogee Instruments is not responsible for any direct, indirect, incidental, or consequential damages, including but not limited to loss of income, loss of revenue, loss of profit, loss of data, loss of wages, loss of time, loss of sales, accruement of debts or expenses, injury to personal property, or injury to any person or any other type of damage or loss.

This limited warranty and any disputes arising out of or in connection with this limited warranty ("Disputes") shall be governed by the laws of the State of Utah, USA, excluding conflicts of law principles and excluding the Convention for the International Sale of Goods. The courts located in the State of Utah, USA, shall have exclusive jurisdiction over any Disputes.

This limited warranty gives you specific legal rights, and you may also have other rights, which vary from state to state and jurisdiction to jurisdiction, and which shall not be affected by this limited warranty. This warranty extends only to you and cannot by transferred or assigned. If any provision of this limited warranty is unlawful, void or unenforceable, that provision shall be deemed severable and shall not affect any remaining provisions. In case of any inconsistency between the English and other versions of this limited warranty, the English version shall prevail.

This warranty cannot be changed, assumed, or amended by any other person or agreement.

APOGEE INSTRUMENTS, INC. | 721 WEST 1800 NORTH, LOGAN, UTAH 84321, USA TEL: (435) 792-4700 | FAX: (435) 787-8268 | WEB: APOGEEINSTRUMENTS.COM

Copyright © 2021 Apogee Instruments, Inc.

Models: SQ-622, SQ-625, Extended Range PFD Sensor, Range PFD Sensor, Extended PFD Sensor, SQ-622, Sensor

File Info : application/pdf, 15 Pages, 708.27KB

PDF preview unavailable. Download the PDF instead.

SQ-622-625-manual

References

Microsoft Word for Microsoft 365

Related Documents

Preview Apogee SQ-500 Quantum Sensor Owner's Manual
Comprehensive owner's manual for the Apogee Instruments SQ-500 Quantum Sensor, detailing specifications, deployment, operation, maintenance, and troubleshooting for measuring Photosynthetically Active Radiation (PAR).
Preview Apogee Instruments MQ-510 Underwater Quantum Meter Owner's Manual
Owner's Manual for the Apogee Instruments MQ-510 Underwater Quantum Meter, detailing its specifications, operation, maintenance, troubleshooting, and warranty information.
Preview Apogee Instruments SE-100 Photometric Sensor Owner's Manual
Owner's manual for the Apogee Instruments SE-100 Photometric Sensor, covering specifications, deployment, operation, maintenance, and troubleshooting.
Preview Apogee SQ-521 Quantum Sensor with METER ZENTRA System: Installation and Operation Guide
This guide details the installation and operation of Apogee SQ-521 Full-Spectrum Quantum Sensors with METER ZENTRA series data loggers. It covers sensor setup, connection, data interpretation, and troubleshooting for accurate photosynthetic photon flux density (PPFD) measurements.
Preview Apogee SP-422 Pyranometer Owner's Manual
Owner's manual for the Apogee SP-422 pyranometer, detailing its specifications, installation, operation, maintenance, troubleshooting, and warranty. Covers Modbus RTU communication for solar radiation measurement.
Preview Apogee Instruments µCache AT-100 Owner's Manual
Comprehensive owner's manual for the Apogee Instruments µCache AT-100 Bluetooth® Memory Module, detailing its features, specifications, setup, operation, maintenance, troubleshooting, and warranty information.
Preview Apogee Instruments MP-100/MP-200 Solar Radiation Meter Owner's Manual
Comprehensive owner's manual for Apogee Instruments MP-100 and MP-200 solar radiation meters, detailing specifications, deployment, operation, maintenance, troubleshooting, and calibration.
Preview Apogee Instruments Pyranometers: Silicon-Cell and Thermopile Series Datasheet
Detailed specifications and features for Apogee Instruments' silicon-cell and thermopile pyranometers, including models SP-100, SP-200, SP-400, SP-510, SP-610, SP-522, and their digital output options (USB, SDI-12, Modbus).