Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using this instrument. This manual will provide you with the necessary information for correct use of this instrument, as well as a precise idea of its versatility. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com or view our contact list at www.hannainst.com.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895, USA.
Remove the tester and accessories from packaging and examine them carefully. For further assistance, please contact your local Hanna Instruments office or email us at tech@hannainst.com.
Each HI9810382 is supplied with a starter kit consisting of:
Note: Save all packing material until you are sure that the tester works correctly. Any damaged or defective item must be returned in its original packing material with the supplied accessories.
Parameter | Range | Notes |
---|---|---|
pH | 0.00 to 12.00 pH | |
mV * | pH/mV conversion | |
Temperature** | 0.0 to 60.0 °C (32.0 to 140.0 °F) | |
pH Resolution | 0.01 or 0.1 pH | |
mV * Resolution | 0.1 or 1 mV | |
Temperature Resolution | 0.1 °C; 0.1 °F | |
pH Accuracy | ±0.05 pH | |
Temperature Accuracy | ±0.5 °C; ±0.9 °F | |
Calibration | Up to three points or four points * | Automatic buffer recognition with Standard buffers Hanna (pH 1.68 *, 4.01, 7.01, 10.01) or NIST (pH 1.68 *, 4.01, 6.86, 9.18) |
Temperature compensation | Automatic (ATC) or Manual (MTC) * | |
Electrode | Body material | Polyvinylidene Fluoride (PVDF) |
Junction | Low Temperature (LT) | |
Reference cell | Open Double, Ag/AgCl | |
Electrolyte | Viscolene | |
Tip / Shape | Conic, Ø 6 x 10 mm (Ø 0.23 x 0.39″) | |
Outer diameter | 13 mm (0.51″) | |
Length | 75 mm (2.95″) | |
Battery type | 3V Lithium – CR2032 | |
Battery life | Approximately 1000 hours (500 hours with Bluetooth enabled) | |
Environment | 0 to 50 °C (32 to 122 °F) | |
Casing | IP65 ingress protection | |
Dimensions / Weight | 51 x 146 x 21 mm (2.0 x 5.7 x 0.8″) / 45 g (1.6 oz.) |
* Available with Hanna Lab App
** Measuring outside the recommended operating temperature range may damage the gel electrolyte and void product warranty.
Note: The tester can display measurements from -2.00 to 16.00 pH. Measurements outside of the pH range will flash. In this case assess the integrity of the tester and the type of measured sample.
HI9810382 is a professional wireless pH tester, part of Hanna Instruments HALO2 family.
The HI9810382 can be used as a stand-alone pH tester or connected to the Hanna Lab App. The Hanna Lab App turns a compatible smart device into a full-featured pH meter. Features include: electrode condition, GLP with time-stamp, live-readings, mV resolution, manual temperature compensation, stability criteria, calibration reminder, pH (mV) and temperature alarms, tester ID, and data sharing.
The food grade PVDF body material is easy to clean and disinfect. Resistant to most chemicals (e.g. solvents, sodium hypochlorite), ultraviolet light, and fungal growth, the PVDF body also has high-abrasion resistance and mechanical strength.
The conical tip allows for easy penetration into solids and semisolids.
The double junction design presents a silver-free electrolyte solution interacting with the sample, making the electrode less susceptible to clogging and guaranteeing a fast response and stable reading.
Built-in temperature sensor at the tip of the pH electrode allows rapid determination of the sample temperature and a high-accuracy temperature reading.
LCD: Displays measurement readings, status indicators, and setup information.
ON/OFF button: Used to turn the tester on/off and navigate menu options.
Maximum immersion level: A marked line on the probe indicating the maximum safe immersion depth.
Recommended immersion: A marked line on the probe indicating the optimal immersion depth for accurate readings.
pH Probe: The sensor for measuring pH and temperature.
Setup button: Located inside the battery compartment, used to enter setup mode.
Battery cover: Covers the battery compartment.
The LCD has several segments:
The display can show values like "8.8.8.8" during initialization or "ATC 25.0 °C" for temperature readings.
Note: Only use specified battery type. Dispose of old battery in accordance with local regulations.
The Setup button is located inside the battery compartment. After Setup configuration, replace the cover.
Option: °C or °F. Press ON/OFF button to select desired temperature unit.
Option: 8, 60 min., or "----" (disabled). Press ON/OFF button to select desired interval. To save battery life, after the selected auto-off interval has elapsed, the tester will automatically turn off.
Option: 2P or 3P. Press ON/OFF button to select between up to two- or up to three-point calibration.
Option: 7.01 pH (Hanna) or 6.86 pH (NIST). Press ON/OFF button to select the calibration buffer set (Hanna or NIST).
Option: 0.01 pH or 0.1 pH. Press the ON/OFF button to select resolution.
Option: On, PAIr, or OFF. Press ON/OFF button to select Bluetooth option at start-up.
Option: dEL PAIr. Press ON/OFF button to delete saved paired device.
Option: SET TIME. Press ON/OFF button to set the date and time.
Options: YEAR, MO, DAY, HOUR, and MIN. Use the Setup button to select the option and press the ON/OFF button to change the selected option.
With "PAIr BLU" or "On BLU" selected in Setup, the Bluetooth icon (?) will blink for up to 45 seconds, indicating the tester is in discoverable mode. Once connected, the icon will stop blinking. If not connected, the icon is not displayed.
Once connected to the application, the ON/OFF button can be pressed to tag the current reading.
For most applications it is recommended to start with pH 7.01 buffer (or pH 6.86).
To restore factory defaults, press and hold the ON/OFF button. "CAL CLR" is displayed.
Note: It is recommended to calibrate the electrode with buffers at the temperature it will be used at.
Connect the tester to the Hanna Lab App and follow the calibration procedure. See the App Help section for the calibration procedure.
To ensure a quick response, the glass tip and junction should be kept hydrated. When not in use, add a few drops of storage solution to the storage cap. If storage solution is not available, pH 4.01 or pH 7.01 buffer can be used.
Note: Do not store the electrode in distilled or deionized water.
Ordering Information | Product Description |
---|---|
HI50016-02 | pH 1.68 buffer solution, 20 mL sachet (25 pcs.) |
HI70004P | pH 4.01 buffer solution, 20 mL sachet (25 pcs.) |
HI70006P | pH 6.86 buffer solution, 20 mL sachet (25 pcs.) |
HI70007P | pH 7.01 buffer solution, 20 mL sachet (25 pcs.) |
HI70009P | pH 9.18 buffer solution, 20 mL sachet (25 pcs.) |
HI70010P | pH 10.01 buffer solution, 20 mL sachet (25 pcs.) |
HI77400P | pH 4.01 & 7.01 buffer solution, 20 mL sachet (10 pcs., 5 each) |
HI770710P | pH 10.01 & 7.01 buffer solution, 20 mL sachet (10 pcs., 5 each) |
Ordering Information | Product Description |
---|---|
HI700601P | Electrode cleaning solution for general use, 20 mL sachet (25 pcs.) |
HI700684P | Electrode cleaning and disinfection solution for bread and dough deposits, 20 mL sachet (25 pcs.) |
Ordering Information | Product Description |
---|---|
HI70300L | Storage solution, 500 mL |
HI70300M | Electrode storage solution, 230 mL |
HI70300S | Electrode storage solution, 30 mL dropper |
HI9072 | Electrode storage solution, 13 mL dropper |
Before using this product, make sure it is entirely suitable for your specific application and for the environment in which it is used. Any variation introduced by the user to the supplied equipment may degrade the tester's performance. For your and the tester's safety do not use or store it in hazardous environments.
HI9810382 is warranted for a period of one year against defects in workmanship and materials when used for its intended purpose and maintained according to instructions. This warranty is limited to repair or replacement free of charge. Damage due to accidents, misuse, tampering, or lack of prescribed maintenance is not covered. If service is required, contact your local Hanna Instruments office. If under warranty, report the model number, date of purchase, serial number, and the nature of the problem. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the tester is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization (RGA) number from the Technical Service department and then send it with shipping costs prepaid. When shipping any product, make sure it is shipped back in compliance with shipping regulations, thoroughly cleaned and free of any chemicals, and is properly packaged for complete protection.
Hanna Instruments reserves the right to modify the design, construction or appearance of its products without advance notice.
All Hanna Instruments conform to the CE European Directives.
The product should not be treated as household waste. Instead, hand it over to the appropriate collection point for the recycling of electrical and electronic equipment, which will conserve natural resources.
This product contains batteries, do not dispose of them with other household waste. Hand them over to the appropriate collection point for recycling. Ensuring proper product and battery disposal prevents potential negative consequences for the environment and human health. For more information, contact your city, your local household waste disposal service, or the place of purchase.
All modules have identical operation. All references to US FCC Rules and Canadian RSS standards on device classification and operation, listed here under BMD-300 Module, apply to all models noted here. Remove the battery cover to check the installed module.
File Info : application/pdf, 12 Pages, 275.46KB
DocumentDocument![]() |
Hanna HI9810362 HALO2 Wireless pH Tester for Meat Instruction Manual Instruction manual for the Hanna HI9810362 HALO2 Wireless pH Tester for Meat, detailing its features, specifications, operation, calibration, maintenance, and troubleshooting. |
![]() |
Hanna HI9810412 HALO2 Wireless pH Tester for Lab Instruction Manual Comprehensive instruction manual for the Hanna HI9810412 HALO2 Wireless pH Tester, detailing its features, operation, calibration, care, and specifications for laboratory use. Includes Bluetooth connectivity and compatibility with the Hanna Lab App. |
![]() |
Hanna Instruments HI9810452 HALO 2 Wireless pH Tester for Meat Instruction Manual This instruction manual provides detailed information on the Hanna Instruments HI9810452 HALO 2 Wireless pH Tester for Meat, covering setup, operation, calibration, maintenance, and troubleshooting for accurate pH measurements in food applications. |
![]() |
Hanna HI9810352 HALO2 Wireless pH Tester for Sushi: Instruction Manual This instruction manual provides detailed information on the Hanna HI9810352 HALO2 Wireless pH Tester for Sushi. It covers product specifications, operational procedures, calibration methods, measurement techniques for sushi rice, and essential care and maintenance guidelines. |
![]() |
Hanna Instruments pHep HI 98107 Pocket-Sized pH Meter The Hanna Instruments pHep HI 98107 is a rugged, pocket-sized pH meter designed for accurate pH measurements. Features include a replaceable electrode, manual 2-point calibration, and a long battery life. |
![]() |
Instrukcja Obsługi Miernika pH i Temperatury HI99121 do Pomiaru w Glebie Instrukcja obsługi dla przenośnego miernika pH i temperatury Hanna Instruments HI99121, przeznaczonego do bezpośredniego pomiaru pH w glebie i roztworach glebowych. Zawiera informacje o funkcjach, obsłudze, kalibracji, konserwacji i specyfikacjach technicznych. |
![]() |
HI98164 Professional pH and Temperature Meter for Yogurt Products The Hanna Instruments HI98164 is a professional portable pH and temperature meter specifically designed for monitoring pH levels in yogurt production. It ensures consistent product quality by accurately measuring pH during the fermentation process, offering advanced features for reliable calibration and data logging. |
![]() |
Hanna Instruments HI6221 Advanced pH & ORP Meter Quick Reference Guide Quick reference guide for the Hanna Instruments HI6221 Advanced pH & ORP Meter, covering key features, getting started, package contents, system overview, and electrode usage guidelines. |