Phosphorus, Reactive (Orthophosphate)
Method: 8048
Reagents: Powder Pillows or AccuVac Ampuls
Concentration Range: 0.02 to 2.50 mg/L PO₄³⁻
Scope and Application
For drinking water, wastewater, and seawater analysis.
- USEPA Accepted for reporting for wastewater analyses. Procedure is equivalent to USEPA and Standard Method 4500-P-E for wastewater.
- USEPA Accepted for reporting for drinking water analysis. Procedure is an acceptable version of EPA Method 365.1, approved at 40 CFR part 141 NPDWR compliance monitoring.
- Adapted from Standard Methods for the Examination of Water and Wastewater.
Test Preparation
Instrument-Specific Information
Table 1 and Table 2 provide sample cell and orientation requirements for reagent addition tests, as well as sample cell and adapter requirements for AccuVac Ampul tests. These tables also list the instruments compatible with this test.
Instrument | Sample Cell Orientation | Sample Cell |
---|---|---|
DR 6000, DR 3800, DR 2800, DR 2700, DR 1900, DR 5000, DR 3900 | The fill line is to the right. | 2495402 |
DR 900 | The fill line is toward the user. | |
The orientation mark is toward the user. | 2401906 |
Instrument | Adapter | Sample Cell |
---|---|---|
DR 6000, DR 5000, DR 900, DR 3900 | — | 2427606 |
DR 1900 | LZV846 (A) | |
DR 3800, DR 2800, DR 2700 | 9609900 or 9609800 (C) | LZV584 (C) |
2122800 |
Before Starting
- Install the instrument cap on the DR 900 cell holder before ZERO or READ is pushed.
- For best results, measure the reagent blank value for each new lot of reagent. Replace the sample with deionized water in the test procedure to determine the reagent blank value. Subtract the reagent blank value from the sample results automatically with the reagent blank adjust option.
- Review the Safety Data Sheets (MSDS/SDS) for chemicals used and use recommended personal protective equipment.
- Dispose of reacted solutions according to local, state, and federal regulations. Refer to Safety Data Sheets for disposal information for unused reagents and consult environmental, health, and safety staff for further disposal information.
Items to Collect
Powder Pillows
- PhosVer® 3 Phosphate Reagent powder pillow, 10-mL (1)
- Sample cells (2)
AccuVac Ampuls
- PhosVer® 3 Phosphate Reagent AccuVac® Ampul (1)
- Beaker, 50-mL (1)
- Sample cells (1)
- Stopper for 18-mm tubes and AccuVac Ampuls (1)
Sample Collection and Storage
- Collect samples in clean glass or plastic bottles cleaned with 6 N (1:1) hydrochloric acid and rinsed with deionized water.
- Do not use a detergent containing phosphate to clean sample bottles, as it can contaminate the sample.
- Analyze samples as soon as possible for best results.
- If immediate analysis is not possible, filter and keep samples at or below 6 °C (43 °F) for a maximum of 48 hours.
- Allow sample temperature to increase to room temperature before analysis.
Powder Pillow Procedure
- Start program 490 P React. PP. Refer to Instrument-specific information on page 1 for details on sample cells, adapters, or light shields.
- Prepare the sample: Fill a sample cell with 10 mL of sample.
- Add the contents of one PhosVer 3 Phosphate Reagent Powder Pillow to the cell. A blue color indicates the presence of phosphorus.
- Immediately close the sample cell and shake vigorously for 20-30 seconds.
- Start the instrument timer. A 2-minute reaction time begins. (If the sample was digested using Acid Persulfate digestion, a 10-minute reaction period is necessary.)
- Prepare the blank: Fill a second sample cell with 10 mL of sample.
- When the timer expires, clean the blank sample cell.
- Insert the blank into the cell holder.
- Push ZERO. The display should show 0.00 mg/L PO₄³⁻.
- Clean the prepared sample cell.
- Insert the prepared sample into the cell holder.
- Push READ. Results show in mg/L PO₄³⁻.
AccuVac Ampul Procedure
- Start program 492 P React. PV AV. Refer to Instrument-specific information on page 1 for details on sample cells, adapters, or light shields.
- Prepare the blank: Fill the sample cell with 10 mL of sample.
- Prepare the sample: Collect at least 40 mL of sample in a 50-mL beaker. Fill the AccuVac Ampul with sample, keeping the tip immersed until the ampul fills completely.
- Close the AccuVac Ampul and shake for approximately 30 seconds. Accuracy is not affected by undissolved powder.
- Start the instrument timer. A 2-minute reaction time begins. (If the sample was digested using Acid Persulfate digestion, a 10-minute reaction period is necessary.)
- When the timer expires, clean the blank sample cell.
- Insert the blank into the cell holder.
- Push ZERO. The display shows 0.00 mg/L PO₄³⁻.
- Clean the AccuVac Ampul.
- Insert the prepared sample AccuVac Ampul into the cell holder.
- Push READ. Results show in mg/L PO₄³⁻.
Interferences
Interfering Substance | Interference Level |
---|---|
Aluminum | More than 200 mg/L |
Arsenate | Interferes at any level |
Chromium | More than 100 mg/L |
Copper | More than 10 mg/L |
Hydrogen Sulfide | Interferes at any level |
Iron | More than 100 mg/L |
Nickel | More than 300 mg/L |
Highly buffered samples or extreme sample pH | Can prevent correct pH adjustment by reagents. Sample pretreatment may be necessary. A pH range of 2-10 is recommended. |
Silica | More than 50 mg/L |
Silicate | More than 10 mg/L |
Turbidity or color | Samples with high turbidity can give inconsistent results. Reagent acid may dissolve suspended particles, causing variable orthophosphate desorption. |
Zinc | More than 80 mg/L |
Accuracy Check
Standard Additions Method (Sample Spike)
Use the standard additions method to validate the test procedure, reagents, and instrument, and to identify interferences.
Items to Collect:
- Phosphate standard solution, 50 mg/L PO₄³⁻ ampule
- Ampule breaker
- Pipet, TenSette®, 0.1–1.0 mL and tips
- Mixing cylinders, 25-mL (3)
- Use the test procedure to measure the sample concentration and keep the (unspiked) sample in the instrument.
- Navigate to the Standard Additions option in the instrument menu.
- Select values for standard concentration, sample volume, and spike volumes.
- Open the standard solution.
- Prepare three spiked samples: add 0.1 mL, 0.2 mL, and 0.3 mL of the standard solution to three 10-mL portions of fresh sample using the TenSette pipet. Mix well. (For AccuVac® Ampuls, add 0.2 mL, 0.4 mL, and 0.6 mL of standard solution to three 50-mL portions of fresh sample.)
- Use the test procedure to measure the concentration of each spiked sample, starting with the smallest spike.
- Select Graph to compare expected results with actual results.
Note: If actual results significantly differ from expected results, verify sample volumes and spike measurements. Ensure they match the standard additions menu selections. If results are outside acceptable limits, the sample may contain an interference.
Standard Solution Method
Use the standard solution method to validate the test procedure, reagents, and instrument.
Items to Collect:
- 50 mg/L phosphate standard solution
- 100-mL volumetric flask, Class A
- 4-mL volumetric pipet, Class A and pipet filler safety bulb
- Deionized water
- Prepare a 2.00-mg/L phosphate standard solution: Add 4.00 mL of a 50-mg/L phosphate standard solution to a volumetric flask (or use available mixed parameter standards containing 2.0 mg/L phosphate). Dilute to the mark with deionized water. Mix well. Prepare this solution daily.
- Use the test procedure to measure the concentration of the prepared standard solution.
- Compare the expected result to the actual result.
Note: Factory calibration can be adjusted using the standard adjust option to match the expected standard solution value. This adjustment can improve test accuracy with minor variations in reagents or instruments.
Method Performance
Method performance data is derived from laboratory tests under ideal conditions. Actual results may vary.
Program | Standard | Precision (95% Confidence Interval) | Sensitivity (Concentration change per 0.010 Abs change) |
---|---|---|---|
490 | 2.00 mg/L PO₄³⁻ | 1.98–2.02 mg/L PO₄³⁻ | 0.02 mg/L PO₄³⁻ |
492 | 2.00 mg/L PO₄³⁻ | 1.98–2.02 mg/L PO₄³⁻ | 0.02 mg/L PO₄³⁻ |
Summary of Method
Orthophosphate reacts with molybdate in an acid medium to form a mixed phosphate/molybdate complex. Ascorbic acid reduces this complex, producing an intense molybdenum blue color. Measurement is performed at 880 nm for spectrophotometers (DR 1900: 710 nm) or 610 nm for colorimeters.
Consumables and Replacement Items
Required Reagents
Description | Quantity/Test | Unit | Item no. |
---|---|---|---|
PhosVer® 3 Phosphate Reagent Powder Pillow, 10 mL | 1 | 100/pkg | 2106069 |
PhosVer® 3 Phosphate Reagent AccuVac® Ampul | 1 | 25/pkg | 2508025 |
Required Apparatus
Description | Quantity/Test | Unit | Item no. |
---|---|---|---|
Beaker, 50 mL | 1 | each | 50041H |
Stoppers for 18-mm tubes and AccuVac Ampuls | 2 | 6/pkg | 173106 |
Recommended Standards
Description | Unit | Item no. |
---|---|---|
Phosphate Standard Solution, 10-mL Voluette® Ampule, 50 mg/L as PO₄³⁻ | 16/pkg | 17110 |
Phosphate Standard Solution, 50 mg/L as PO₄³⁻ | 500 mL | 17149 |
Phosphate Standard Solution, 1 mg/L as PO₄³⁻ | 500 mL | 256949 |
Drinking Water Standard, Mixed Parameter, Inorganic for F, NO₃⁻N, PO₄³⁻, SO₄²⁻, COD, SO₄²⁻, TOC | 500 mL | 2833049 |
Wastewater Effluent Standard Solution, Mixed Parameter, for NH₃-N, NO₃-N, PO₄³⁻, COD, SO₄²⁻, TOC | 500 mL | 2833249 |
Water, deionized | 4 L | 27256 |
Optional Reagents and Apparatus
Description | Unit | Item no. |
---|---|---|
AccuVac® Drainer | each | 4103600 |
AccuVac Ampul Snapper® | each | 2405200 |
AccuVac Ampul vials for sample blanks | 25/pkg | 2677925 |
Ampule Breaker, 10-mL Voluette® Ampules | each | 2196800 |
Bottle, sampling, with cap, low density polyethylene, 250 mL | 12/pkg | 2087076 |
Mixing cylinder, graduated, 50 mL | each | 189641 |
Flask, volumetric, Class A, 100 mL, glass | each | 1457442 |
Hydrochloric Acid Solution, 6.0 N (1:1) | 500 mL | 88449 |
Paper, pH, 0-14 pH range | 100/pkg | 2601300 |
Phosphate Treatment Powder Pillows | 100/pkg | 1450199 |
Phosphate Standard Solution, 10 mg/L as PO₄³⁻ | 946 mL | 1420416 |
Phosphate Standard Solution, 15-mg/L as PO₄³⁻ | 100 mL | 1424342 |
Phosphate Standard Solution, 100-mg/L as PO₄³⁻ | 100 mL | 1436832 |
Phosphate Standard Solution, 10-mL ampule, 500 mg/L as PO₄³⁻ | 16/pkg | 1424210 |
Phosphate Standard Solution, 500-mg/L as PO₄³⁻ | 100 mL | 1424232 |
Pipet, TenSette®, 0.1–1.0 mL | each | 1970001 |
Pipet, TenSette®, 1.0–10.0 mL | each | 1970010 |
Pipet tips for TenSette® Pipet, 0.1–1.0 mL | 50/pkg | 2185696 |
Pipet tips for TenSette® Pipet, 0.1–1.0 mL | 1000/pkg | 2185628 |
Pipet tips for TenSette® Pipet, 1.0–10.0 mL | 50/pkg | 2199796 |
Pipet tips for TenSette® Pipet, 1.0–10.0 mL | 250/pkg | 2199725 |
Pipet, volumetric, Class A, 4.00 mL | each | 1451504 |