Phosphorus, Reactive (Orthophosphate)

Method: 8048
Reagents: Powder Pillows or AccuVac Ampuls
Concentration Range: 0.02 to 2.50 mg/L PO₄³⁻

Scope and Application

For drinking water, wastewater, and seawater analysis.

Test Preparation

Instrument-Specific Information

Table 1 and Table 2 provide sample cell and orientation requirements for reagent addition tests, as well as sample cell and adapter requirements for AccuVac Ampul tests. These tables also list the instruments compatible with this test.

Table 1: Instrument-Specific Information for Reagent Addition
Instrument Sample Cell Orientation Sample Cell
DR 6000, DR 3800, DR 2800, DR 2700, DR 1900, DR 5000, DR 3900 The fill line is to the right. 2495402
DR 900 The fill line is toward the user.
The orientation mark is toward the user. 2401906
Table 2: Instrument-Specific Information for AccuVac Ampuls
Instrument Adapter Sample Cell
DR 6000, DR 5000, DR 900, DR 3900 2427606
DR 1900 LZV846 (A)
DR 3800, DR 2800, DR 2700 9609900 or 9609800 (C) LZV584 (C)
2122800

Before Starting

Items to Collect

Powder Pillows

AccuVac Ampuls

Sample Collection and Storage

Powder Pillow Procedure

  1. Start program 490 P React. PP. Refer to Instrument-specific information on page 1 for details on sample cells, adapters, or light shields.
  2. Prepare the sample: Fill a sample cell with 10 mL of sample.
  3. Add the contents of one PhosVer 3 Phosphate Reagent Powder Pillow to the cell. A blue color indicates the presence of phosphorus.
  4. Immediately close the sample cell and shake vigorously for 20-30 seconds.
  5. Start the instrument timer. A 2-minute reaction time begins. (If the sample was digested using Acid Persulfate digestion, a 10-minute reaction period is necessary.)
  6. Prepare the blank: Fill a second sample cell with 10 mL of sample.
  7. When the timer expires, clean the blank sample cell.
  8. Insert the blank into the cell holder.
  9. Push ZERO. The display should show 0.00 mg/L PO₄³⁻.
  10. Clean the prepared sample cell.
  11. Insert the prepared sample into the cell holder.
  12. Push READ. Results show in mg/L PO₄³⁻.

AccuVac Ampul Procedure

  1. Start program 492 P React. PV AV. Refer to Instrument-specific information on page 1 for details on sample cells, adapters, or light shields.
  2. Prepare the blank: Fill the sample cell with 10 mL of sample.
  3. Prepare the sample: Collect at least 40 mL of sample in a 50-mL beaker. Fill the AccuVac Ampul with sample, keeping the tip immersed until the ampul fills completely.
  4. Close the AccuVac Ampul and shake for approximately 30 seconds. Accuracy is not affected by undissolved powder.
  5. Start the instrument timer. A 2-minute reaction time begins. (If the sample was digested using Acid Persulfate digestion, a 10-minute reaction period is necessary.)
  6. When the timer expires, clean the blank sample cell.
  7. Insert the blank into the cell holder.
  8. Push ZERO. The display shows 0.00 mg/L PO₄³⁻.
  9. Clean the AccuVac Ampul.
  10. Insert the prepared sample AccuVac Ampul into the cell holder.
  11. Push READ. Results show in mg/L PO₄³⁻.

Interferences

Interfering Substances and Levels
Interfering Substance Interference Level
Aluminum More than 200 mg/L
Arsenate Interferes at any level
Chromium More than 100 mg/L
Copper More than 10 mg/L
Hydrogen Sulfide Interferes at any level
Iron More than 100 mg/L
Nickel More than 300 mg/L
Highly buffered samples or extreme sample pH Can prevent correct pH adjustment by reagents. Sample pretreatment may be necessary. A pH range of 2-10 is recommended.
Silica More than 50 mg/L
Silicate More than 10 mg/L
Turbidity or color Samples with high turbidity can give inconsistent results. Reagent acid may dissolve suspended particles, causing variable orthophosphate desorption.
Zinc More than 80 mg/L

Accuracy Check

Standard Additions Method (Sample Spike)

Use the standard additions method to validate the test procedure, reagents, and instrument, and to identify interferences.

Items to Collect:

  1. Use the test procedure to measure the sample concentration and keep the (unspiked) sample in the instrument.
  2. Navigate to the Standard Additions option in the instrument menu.
  3. Select values for standard concentration, sample volume, and spike volumes.
  4. Open the standard solution.
  5. Prepare three spiked samples: add 0.1 mL, 0.2 mL, and 0.3 mL of the standard solution to three 10-mL portions of fresh sample using the TenSette pipet. Mix well. (For AccuVac® Ampuls, add 0.2 mL, 0.4 mL, and 0.6 mL of standard solution to three 50-mL portions of fresh sample.)
  6. Use the test procedure to measure the concentration of each spiked sample, starting with the smallest spike.
  7. Select Graph to compare expected results with actual results.

Note: If actual results significantly differ from expected results, verify sample volumes and spike measurements. Ensure they match the standard additions menu selections. If results are outside acceptable limits, the sample may contain an interference.

Standard Solution Method

Use the standard solution method to validate the test procedure, reagents, and instrument.

Items to Collect:

  1. Prepare a 2.00-mg/L phosphate standard solution: Add 4.00 mL of a 50-mg/L phosphate standard solution to a volumetric flask (or use available mixed parameter standards containing 2.0 mg/L phosphate). Dilute to the mark with deionized water. Mix well. Prepare this solution daily.
  2. Use the test procedure to measure the concentration of the prepared standard solution.
  3. Compare the expected result to the actual result.

Note: Factory calibration can be adjusted using the standard adjust option to match the expected standard solution value. This adjustment can improve test accuracy with minor variations in reagents or instruments.

Method Performance

Method performance data is derived from laboratory tests under ideal conditions. Actual results may vary.

Method Performance Data
Program Standard Precision (95% Confidence Interval) Sensitivity (Concentration change per 0.010 Abs change)
490 2.00 mg/L PO₄³⁻ 1.98–2.02 mg/L PO₄³⁻ 0.02 mg/L PO₄³⁻
492 2.00 mg/L PO₄³⁻ 1.98–2.02 mg/L PO₄³⁻ 0.02 mg/L PO₄³⁻

Summary of Method

Orthophosphate reacts with molybdate in an acid medium to form a mixed phosphate/molybdate complex. Ascorbic acid reduces this complex, producing an intense molybdenum blue color. Measurement is performed at 880 nm for spectrophotometers (DR 1900: 710 nm) or 610 nm for colorimeters.

Consumables and Replacement Items

Required Reagents

Description Quantity/Test Unit Item no.
PhosVer® 3 Phosphate Reagent Powder Pillow, 10 mL 1 100/pkg 2106069
PhosVer® 3 Phosphate Reagent AccuVac® Ampul 1 25/pkg 2508025

Required Apparatus

Description Quantity/Test Unit Item no.
Beaker, 50 mL 1 each 50041H
Stoppers for 18-mm tubes and AccuVac Ampuls 2 6/pkg 173106

Recommended Standards

Description Unit Item no.
Phosphate Standard Solution, 10-mL Voluette® Ampule, 50 mg/L as PO₄³⁻ 16/pkg 17110
Phosphate Standard Solution, 50 mg/L as PO₄³⁻ 500 mL 17149
Phosphate Standard Solution, 1 mg/L as PO₄³⁻ 500 mL 256949
Drinking Water Standard, Mixed Parameter, Inorganic for F, NO₃⁻N, PO₄³⁻, SO₄²⁻, COD, SO₄²⁻, TOC 500 mL 2833049
Wastewater Effluent Standard Solution, Mixed Parameter, for NH₃-N, NO₃-N, PO₄³⁻, COD, SO₄²⁻, TOC 500 mL 2833249
Water, deionized 4 L 27256

Optional Reagents and Apparatus

Description Unit Item no.
AccuVac® Drainer each 4103600
AccuVac Ampul Snapper® each 2405200
AccuVac Ampul vials for sample blanks 25/pkg 2677925
Ampule Breaker, 10-mL Voluette® Ampules each 2196800
Bottle, sampling, with cap, low density polyethylene, 250 mL 12/pkg 2087076
Mixing cylinder, graduated, 50 mL each 189641
Flask, volumetric, Class A, 100 mL, glass each 1457442
Hydrochloric Acid Solution, 6.0 N (1:1) 500 mL 88449
Paper, pH, 0-14 pH range 100/pkg 2601300
Phosphate Treatment Powder Pillows 100/pkg 1450199
Phosphate Standard Solution, 10 mg/L as PO₄³⁻ 946 mL 1420416
Phosphate Standard Solution, 15-mg/L as PO₄³⁻ 100 mL 1424342
Phosphate Standard Solution, 100-mg/L as PO₄³⁻ 100 mL 1436832
Phosphate Standard Solution, 10-mL ampule, 500 mg/L as PO₄³⁻ 16/pkg 1424210
Phosphate Standard Solution, 500-mg/L as PO₄³⁻ 100 mL 1424232
Pipet, TenSette®, 0.1–1.0 mL each 1970001
Pipet, TenSette®, 1.0–10.0 mL each 1970010
Pipet tips for TenSette® Pipet, 0.1–1.0 mL 50/pkg 2185696
Pipet tips for TenSette® Pipet, 0.1–1.0 mL 1000/pkg 2185628
Pipet tips for TenSette® Pipet, 1.0–10.0 mL 50/pkg 2199796
Pipet tips for TenSette® Pipet, 1.0–10.0 mL 250/pkg 2199725
Pipet, volumetric, Class A, 4.00 mL each 1451504

PDF preview unavailable. Download the PDF instead.

DOC316.53.01119 AH XSL Formatter V5.3 MR5 for Windows (x64) : 5.3.6.0 (2012/12/04 12:44:48) Antenna House PDF Output Library 2.6.0 (Windows (x64))

Related Documents

Preview Hach Nitrate Cadmium Reduction Method 8039: Procedure and Guidelines
Comprehensive guide to the Hach Nitrate Cadmium Reduction Method (Method 8039) for determining nitrate levels in water, wastewater, and seawater using NitraVer 5 Powder Pillows and AccuVac Ampuls. Includes detailed procedures, instrument compatibility, interferences, calibration, and safety information.
Preview Arsenic Silver Diethyldithiocarbamate Method 8013: A Comprehensive Guide
Detailed instructions for determining arsenic levels in water, wastewater, and seawater using the Silver Diethyldithiocarbamate Method 8013. Covers test preparation, instrument-specific information, sample collection, and the step-by-step test procedure.
Preview Hach Phosphate, Phosphorus, and Phosphonates Troubleshooting Guide
A comprehensive troubleshooting guide for Hach's phosphate, phosphorus, and phosphonates testing methods, covering principles, chemical reactions, instrumentation, digestion options, and frequently asked questions.
Preview Hach DR/Check™ ABS Standards: General Information and Quality Control Instructions
Comprehensive guide to Hach DR/Check™ ABS Standards, detailing general information, storage, alignment procedures, and step-by-step quality control instructions for use with DR/800 and Odyssey spectrophotometers.
Preview Hach DR 3900 Basic User Manual
A comprehensive user manual for the Hach DR 3900 spectrophotometer, covering its specifications, safety information, installation, operation, and troubleshooting. This guide provides essential details for users of this advanced laboratory instrument.
Preview DR 900 User Manual: Portable Colorimeter Guide
Comprehensive user manual for the DR 900 portable colorimeter. Learn about specifications, installation, operation, maintenance, and troubleshooting for this Hach instrument.
Preview Hach DR 3900 Spectrophotometer Basic User Manual
User manual for the Hach DR 3900 VIS Spectrophotometer, covering specifications, safety information, installation, and operating environment. Learn about its features, including wavelength range, operating modes, and connectivity options.
Preview Hach Ferrous Iron Test Method 8146: 1,10-Phenanthroline Method
Detailed instructions and specifications for performing the Ferrous Iron test using the 1,10-Phenanthroline Method (Method 8146) with Hach Powder Pillows, covering scope, procedure, accuracy checks, and required materials for water analysis.