Nitrate: Cadmium Reduction Method

Method: 8039
Concentration: 0.3 to 30.0 mg/L NO3-N (HR)
Document ID: DOC316.53.01066
Scope and application: For water, wastewater and seawater.

Test preparation

Instrument-specific information

Table 1 details sample cell and orientation requirements for reagent addition tests, such as powder pillow or bulk reagent tests. Table 2 outlines sample cell and adapter requirements for AccuVac Ampul tests. Both tables list instruments compatible with this test program. To use the tables, select your instrument and find the applicable information.

Table 1: Instrument-specific information for reagent addition

Instrument Sample cell orientation Sample cell
DR 6000
DR 3800
DR 2800
DR 2700
DR 1900
The fill line is to the right. 2495402 (A sample cell with a 10 mL fill line)
DR 5000
DR 3900
The fill line is toward the user.
DR 900 The orientation mark is toward the user. 2401906 (A sample cell with an orientation mark and a 25 mL fill line)

Table 2: Instrument-specific information for AccuVac Ampuls

Instrument Adapter Sample cell
DR 6000
DR 5000
DR 900
DR 3900
LZV846 (A) 2427606 (A sample cell with a 10 mL fill line)
DR 1900
DR 3800
DR 2800
DR 2700
9609900 or 9609800 (C)
LZV584 (C)
2122800 (A sample cell with a 10 mL fill line)

Before starting

Install the instrument cap on the DR 900 cell holder before ZERO or READ is pushed.

For optimal results, measure the reagent blank value for each new lot of reagent. Substitute deionized water for the sample in the test procedure to determine the reagent blank value. The instrument can automatically subtract the reagent blank value from sample results.

This method is technique-sensitive; shaking time and technique influence color development. For the most accurate results, use a standard solution within the test range and repeat the test several times. Adjust shaking time to achieve the expected result and use this adjusted time for sample measurements.

The reagents used in this test contain cadmium. Rinse the sample cell immediately after use to remove all cadmium particles. Collect reacted samples for proper disposal.

A deposit of unoxidized metal may remain at the bottom of the sample cell after reagent dissolution, but it will not affect results.

Review the Safety Data Sheets (MSDS/SDS) for all chemicals used and employ the recommended personal protective equipment.

Dispose of reacted solutions according to local, state, and federal regulations. Consult Safety Data Sheets for disposal information on unused reagents and refer to your facility's environmental, health, and safety staff or local regulatory agencies for further guidance.

Items to collect

Powder pillows

Description Quantity/test Item no.
NitraVer® 5 Nitrate Reagent Powder Pillow, 10-mL 1 2106169
Sample cells (Refer to Instrument-specific information on page 1) 2

Refer to Consumables and replacement items on page 8 for order information.

AccuVac Ampuls

Description Quantity/test Item no.
NitraVer® 5 Nitrate Reagent AccuVac® Ampul 1 2511025
Beaker, 50-mL 1 50041H
Sample cells (Refer to Instrument-specific information on page 1) 1
Stopper for 18-mm tubes and AccuVac Ampuls 1 173106

Refer to Consumables and replacement items on page 8 for order information.

Sample collection and storage

  • Collect samples in clean glass or plastic bottles.
  • Analyze samples as soon as possible for best results.
  • If immediate analysis is not possible, filter samples and store at or below 6 °C (43 °F) for a maximum of 48 hours.
  • To preserve samples for up to 28 days, adjust sample pH to 2 or less with concentrated sulfuric acid (approx. 2 mL per liter) and store at or below 6 °C (43 °F). This preserves both nitrate and nitrite.
  • Allow sample temperature to reach room temperature before analysis.
  • Before analysis, adjust pH to 7 with 5 N sodium hydroxide solution.
  • Correct the test result for dilution caused by volume additions.

Powder pillow procedure

⚠ CAUTION: Hazardous waste exposure. Prepared samples contain cadmium. Refer to the SDS for safe handling and disposal instructions. Obey all local and regional disposal regulations.

  1. Start program 355 N, Nitrate HR PP. For information about sample cells, adapters, or light shields, refer to Instrument-specific information on page 1. Note: Program names may vary between instruments, but the program number remains the same.
  2. Prepare the sample: Fill a sample cell with 10 mL of sample. (Visual: Sample cell being filled with 10 mL of sample).
  3. Add reagent: Add the contents of one NitraVer 5 Nitrate Reagent Powder Pillow. Stopper the sample cell. (Visual: Powder from a pillow being added to the cell, followed by a stopper).
  4. Start reaction timer: Start the instrument timer. A 1-minute reaction time begins. (Visual: Timer displaying [01:00]).
  5. Shake sample: Stopper the sample cell and shake vigorously until the timer expires. Some powder may not dissolve, but this will not affect results. (Visual: Two arrows indicating vigorous shaking).
  6. Start reaction timer: Start the instrument timer. A 5-minute reaction time begins. An amber color indicates the presence of nitrate. (Visual: Timer displaying [05:00]).
  7. Prepare the blank: When the second timer expires, fill a second sample cell with 10 mL of sample. (Visual: Sample cell being filled with 10 mL of sample).
  8. Clean blank cell: Clean the blank sample cell. (Visual: Sample cell being cleaned).
  9. Insert blank: Insert the blank into the cell holder. (Visual: Sample cell being inserted into an instrument cell holder).
  10. Zero instrument: Push [Zero]. The display shows 0.0 mg/L NO3-N.
  11. Clean sample cell: Clean the prepared sample cell. (Visual: Sample cell being cleaned).
  12. Insert sample: Within 1 minute after the timer expires, insert the prepared sample into the cell holder. (Visual: Sample cell being inserted into an instrument cell holder).
  13. Read result: Push [Read]. Results display in mg/L NO3-N.

AccuVac Ampul procedure

⚠ CAUTION: Hazardous waste exposure. Prepared samples contain cadmium. Refer to the SDS for safe handling and disposal instructions. Obey all local and regional disposal regulations.

  1. Start program 361 N, Nitrate HR AV. For information about sample cells, adapters, or light shields, refer to Instrument-specific information on page 1. Note: Program names may vary between instruments, but the program number remains the same.
  2. Prepare the sample: Collect at least 40 mL of sample in a 50-mL beaker. (Visual: 50-mL beaker).
  3. Prepare ampul: Tap the bottom of a NitraVer 5 Nitrate AccuVac Ampul to dislodge the powder. Fill the Ampul with sample, keeping the tip immersed while it fills completely. (Visual: An ampul being tapped, then filled with sample while submerged).
  4. Start reaction timer: Start the instrument timer. A 1-minute reaction time begins. (Visual: Timer displaying [01:00]).
  5. Mix sample: Invert the Ampul 48 to 52 times as the timer counts down. (Visual: A bottle with circular arrows indicating inversion).
  6. Start reaction timer: Start the instrument timer. A 5-minute reaction time begins. Keep the sample still while the timer counts down. An amber color indicates the presence of nitrate. (Visual: Timer displaying [05:00]).
  7. Prepare the blank: When the second timer expires, fill a sample cell with 10 mL of sample. (Visual: Sample cell being filled with 10 mL of sample).
  8. Clean blank cell: Clean the blank sample cell. (Visual: Sample cell being cleaned).
  9. Insert blank: Insert the blank into the cell holder. (Visual: Sample cell being inserted into an instrument cell holder).
  10. Zero instrument: Push [Zero]. The display shows 0.0 mg/L NO3-N.
  11. Clean ampul: Clean the AccuVac Ampul. (Visual: An ampul being cleaned).
  12. Insert sample: Within 1 minute after the timer expires, insert the prepared sample AccuVac Ampul into the cell holder. (Visual: An ampul being inserted into an instrument cell holder).
  13. Read result: Push [Read]. Results display in mg/L NO3-N.

Interferences

Interfering substance Interference level
Chloride Chloride concentrations above 100 mg/L cause low results. The test can be used at high chloride concentrations (seawater) if a calibration is made with standards that have the same chloride concentration as the samples (refer to Seawater calibration on page 6).
Ferric iron Interferes at all levels.
Nitrite Interferes at all levels. Compensate for nitrite interference as follows:
  1. Add 30-g/L Bromine Water dropwise to the sample until a yellow color remains.
  2. Add 1 drop of 30-g/L Phenol Solution to remove the color.
  3. Use the test procedure to measure the concentration of the treated sample. Report results as total nitrate and nitrite.
Highly buffered samples or extreme sample pH Can prevent correct pH adjustment of the sample by the reagents. Sample pre-treatment may be necessary.
Strong oxidizing and reducing substances Interfere at all levels.

Seawater calibration

Chloride concentrations above 100 mg/L can cause low results. To use this method for samples with high chloride concentrations, calibrate the instrument using nitrate standard solutions that match the chloride concentration of the samples. Prepare calibration standards containing chloride and 1.0, 3.0, 5.0, and 10.0 mg/L nitrate (as NO3-N) as follows:

  1. Prepare 1 liter of chloride water with the same chloride concentration as the samples.
    a. Weigh the applicable amount of ACS-grade sodium chloride: (chloride concentration of samples in g/L) x (1.6485) = g of NaCl per liter. Note: 18.8 g/L is the typical chloride concentration of seawater.
    b. Add the sodium chloride to a 1-liter volumetric flask.
    c. Dilute to the mark with deionized water. Mix thoroughly. Use this water as the dilution water for preparing nitrate standard solutions.
  2. Use a pipet to add 1.0, 3.0, 5.0, and 10.0 mL of a 100 mg/L nitrate-nitrogen (NO3-N) standard solution into four separate 100-mL Class A volumetric flasks.
  3. Dilute to the mark with the prepared chloride water. Mix thoroughly.
  4. Complete the test procedure for each standard solution and for the prepared chloride water (as a 0-mg/L standard solution).
  5. Measure the absorbance of the standard solutions and enter a user calibration into the instrument.
  6. Use the user program to measure samples with high chloride concentrations.

Accuracy check

Standard additions method (sample spike)

Use the standard additions method (for applicable instruments) to validate the test procedure, reagents, and instrument, and to identify sample interferences.

Items to collect:

  • 1000 mg/L Nitrate Nitrogen (NO3-N) Standard Solution
  • 100-mL volumetric flask, Class A
  • 25-mL volumetric pipet, Class A and pipet filler
  • Deionized water
  • Pipet, TenSette®, 0.1–1.0 mL and tips

Procedure:

  1. Prepare a 250 mg/L nitrate-nitrogen standard solution:
    a. Use a pipet to add 25 mL of a 1000 mg/L nitrate nitrogen standard solution into a 100-mL volumetric flask.
    b. Dilute to the mark with deionized water. Mix well. Prepare this solution daily.
  2. Use the test procedure to measure the concentration of the sample, then keep the (unspiked) sample in the instrument.
  3. Navigate to the Standard Additions option in the instrument menu.
  4. Select values for standard concentration, sample volume, and spike volumes.
  5. Prepare three spiked samples: use the TenSette pipet to add 0.1 mL, 0.2 mL, and 0.3 mL of the prepared standard solution, respectively, to three 10-mL portions of fresh sample. Mix well. (Note: For AccuVac® Ampuls, add 0.4 mL, 0.8 mL, and 1.2 mL of the prepared standard solution to three 50-mL portions of fresh sample.)
  6. Use the test procedure to measure the concentration of each spiked sample, starting with the smallest spike. Measure each spiked sample in the instrument.
  7. Select Graph to compare expected results to actual results.

Note: If actual results differ significantly from expected results, ensure sample volumes and spikes are measured accurately and match the standard additions menu selections. If results are outside acceptable limits, the sample may contain an interference.

Standard solution method

Use the standard solution method to validate the test procedure, reagents, and instrument.

Items to collect:

  • Nitrate Nitrogen Standard, Solution, 10.0-mg/L NO3-N

Procedure:

  1. Use the test procedure to measure the concentration of the standard solution.
  2. Compare the expected result to the actual result.

Note: Factory calibration can be adjusted using the standard adjust option to match the instrument's expected value for the standard solution. This adjusted calibration is then used for all test results, improving accuracy with slight variations in reagents or instruments.

Method performance

Method performance data was derived from laboratory tests using a spectrophotometer under ideal conditions. Users may experience different results under varying test conditions.

Program Standard Precision (95% Confidence Interval) Sensitivity (Concentration change per 0.010 Abs change)
355 10 mg/L NO3-N 9.3–10.7 mg/L NO3-N 0.3 mg/L at 0 ppm, 0.5 mg/L at 10 ppm, 0.8 mg/L at 30 ppm NO3-N
361 10 mg/L NO3-N 9.3–10.7 mg/L NO3-N 0.5 mg/L at 0 ppm, 0.6 mg/L at 10 ppm, 0.8 mg/L at 30 ppm NO3-N

Summary of method

Cadmium metal reduces nitrate in the sample to nitrite. The nitrite ion reacts in an acidic medium with sulfanilic acid to form an intermediate diazonium salt. This salt couples with gentisic acid to produce an amber-colored solution. The measurement wavelength is 500 nm for spectrophotometers or 520 nm for colorimeters.

Pollution prevention and waste management

Reacted samples contain cadmium and must be disposed of as hazardous waste. Dispose of reacted solutions according to local, state, and federal regulations.

Consumables and replacement items

Required reagents

Description Quantity/test Unit Item no.
NitraVer® 5 Nitrate Reagent Powder Pillow, 10-mL 1 100/pkg 2106169
NitraVer® 5 Nitrate Reagent AccuVac® Ampul 1 25/pkg 2511025

Required apparatus (powder pillows)

Description Quantity/test Unit Item no.
Stopper, Neoprene, solid, size #1 2 12/pkg 1480801
Stoppers for 18-mm tubes and AccuVac Ampuls 2 6/pkg 173106

Required apparatus (AccuVac)

Description Quantity/test Unit Item no.
Beaker, 50-mL 1 each 50041H

Recommended standards

Description Unit Item no.
Nitrate Nitrogen Standard Solution, 10.0-mg/L NO3-N 500 mL 30749
Nitrate Nitrogen Standard Solution 1000-mg/L NO3-N 500 mL 1279249
Wastewater Influent Standard Solution, Mixed Parameter, for NH3-N, NO3-N, PO4, COD, SO4, TOC 500 mL 2833149
Water, deionized 4 L 27256

Optional reagents and apparatus

Description Unit Item no.
Bromine Water, 30 g/L 29 mL 221120
Mixing cylinder, graduated, 50-mL each 2088641
Pipet, TenSette®, 0.1–1.0 mL each 1970001
Pipet, volumetric, Class A, 0.5-mL each 1451534
Pipet, volumetric, Class A, 1.00-mL each 1451535
Pipet, volumetric, Class A, 2-mL each 1451536
Pipet, volumetric, Class A, 3-mL each 1451503
Pipet tips for TenSette® Pipet, 0.1–1.0 mL 50/pkg 2185696
Pipet tips for TenSette® Pipet, 0.1–1.0 mL 1000/pkg 2185628
Phenol Solution, 30-g/L 29 mL 211220
Pipet, volumetric, Class A, 25-mL each 1451540
Pipet filler, safety bulb each 1465100
AccuVac® Ampul Snapper each 2405200
Sodium Hydroxide Solution, 5 N 50 mL 245026
Sulfuric Acid, ACS 500 mL 97949
Flask, volumetric, Class A, 100-mL glass each 1457442

PDF preview unavailable. Download the PDF instead.

32490 324903 Antenna House PDF Output Library 2.6.0 (Windows) AH Formatter V5.1 MR2 (5,1,2010,0409) for Windows

Related Documents

Preview Arsenic Silver Diethyldithiocarbamate Method 8013: A Comprehensive Guide
Detailed instructions for determining arsenic levels in water, wastewater, and seawater using the Silver Diethyldithiocarbamate Method 8013. Covers test preparation, instrument-specific information, sample collection, and the step-by-step test procedure.
Preview Hach PhosVer 3 Method for Reactive Phosphorus Analysis
This document outlines the procedures for analyzing reactive phosphorus (orthophosphate) in water samples using Hach's PhosVer 3 method with either Powder Pillows or AccuVac Ampuls. It details instrument-specific information, test preparation, sample collection, powder pillow and AccuVac ampul procedures, interferences, accuracy checks, method performance, and required reagents and apparatus.
Preview DR 900 User Manual: Portable Colorimeter Guide
Comprehensive user manual for the DR 900 portable colorimeter. Learn about specifications, installation, operation, maintenance, and troubleshooting for this Hach instrument.
Preview Hach DR/Check™ ABS Standards: General Information and Quality Control Instructions
Comprehensive guide to Hach DR/Check™ ABS Standards, detailing general information, storage, alignment procedures, and step-by-step quality control instructions for use with DR/800 and Odyssey spectrophotometers.
Preview Hach Ferrous Iron Test Method 8146: 1,10-Phenanthroline Method
Detailed instructions and specifications for performing the Ferrous Iron test using the 1,10-Phenanthroline Method (Method 8146) with Hach Powder Pillows, covering scope, procedure, accuracy checks, and required materials for water analysis.
Preview Hach DR 3900 Basic User Manual
A comprehensive user manual for the Hach DR 3900 spectrophotometer, covering its specifications, safety information, installation, operation, and troubleshooting. This guide provides essential details for users of this advanced laboratory instrument.
Preview Hach DR 3900 Spectrophotometer Basic User Manual
User manual for the Hach DR 3900 VIS Spectrophotometer, covering specifications, safety information, installation, and operating environment. Learn about its features, including wavelength range, operating modes, and connectivity options.
Preview HACH EZ1008 Colour Analyser: Precision Water Analysis
Detailed overview of the HACH EZ1008 Colour Analyser for online water analysis, featuring precision, automatic features, and flexible configurations. Includes technical specifications and what's included.