Rice Lake 150-10-5 Digital Physician Scale

Technical Manual

This technical manual provides comprehensive information for the Rice Lake 150-10-5 Digital Physician Scale. The scale is designed for accurate, reliable, and repeatable weight measurements, featuring a built-in height rod, a handle, and two heavy-duty rear wheels for portability.

Manuals are available from Rice Lake Weighing Systems at www.ricelake.com/manuals. Warranty information is available at www.ricelake.com/warranties.

1.0 Introduction

The 150-10-5 digital physician scale is efficiently designed to provide accurate, reliable and repeatable weight measurements. It is equipped with a built-in height rod, a handle and two heavy-duty rear wheels for easy portability.

1.1 FCC Compliance

United States
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.

Canada
This digital apparatus does not exceed the Class A limits for radio noise emissions from digital apparatus set out in the Radio Interference Regulations of the Canadian Department of Communications.

1.2 Disposal

Product Disposal
The product must be brought to appropriate separate waste collection centers at the end of its life cycle. Proper separate collection to recycle the product helps prevent possible negative effects on the environment and to health, and promotes the recycling of the materials. Users who dispose of the product illegally shall face administrative sanctions as provided by law.

Battery Disposal
Dispose of batteries at appropriate waste collection centers at the end of their life cycle in accordance with local laws and regulations. Batteries and rechargeable batteries may contain harmful substances that should not be disposed of in household waste. Batteries may contain harmful substances including but not limited to: cadmium (Cd), lithium (Li), mercury (Hg) or lead (Pb). Users who dispose of batteries illegally shall face administrative sanctions as provided by law.

WARNING: Risk of fire and explosion. Do not burn, crush, disassemble or short-circuit lithium batteries.

1.3 Safety

Safety Definitions:

  • DANGER: Indicates an imminently hazardous situation that, if not avoided, will result in death or serious injury. Includes hazards that are exposed when guards are removed.
  • WARNING: Indicates a potentially hazardous situation that, if not avoided, could result in serious injury or death. Includes hazards that are exposed when guards are removed.
  • CAUTION: Indicates a potentially hazardous situation that, if not avoided, could result in minor or moderate injury.
  • IMPORTANT: Indicates information about procedures that, if not observed, could result in damage to equipment or corruption to and loss of data.

General Safety
Do not operate or work on this equipment unless this manual has been read and all instructions are understood. Failure to follow the instructions or heed the warnings could result in injury or death. Contact any Rice Lake Weighing Systems dealer for replacement manuals.

WARNING
Failure to heed could result in serious injury or death.

  • Ensure every individual who operates or works with this unit has read and understands all safety information.
  • Do not transport the scale while someone is on the scale.
  • Do not allow minors (children) or inexperienced persons to operate this scale.
  • Do not use in the presence of flammable materials.
  • Do not use this product if any of the components are loose or cracked.
  • Do not use near water.
  • Do not use the scale on slippery surfaces, such as a wet floor.
  • Do not use this scale when a person's body or feet are wet, such as after taking a bath.
  • Do not place fingers into slots or possible pinch points.
  • To avoid cross contamination, the scale should be cleaned regularly.
  • Prior to cleaning, make sure the scale is disconnected from the power source.
  • People with disabilities, or who are physically frail, should always be assisted by another person when using this scale.

IMPORTANT

  • Do not drop the scale or subject it to violent shocks.
  • Do not jump on the scale.
  • For accurate weighing, the scale must be placed on a flat, stable surface.
  • Operating at voltages and frequencies other than specified could damage the equipment.
  • Avoid contact with excessive moisture.
  • Do not make alterations or modifications to the scale.
  • Rice Lake Weighing Systems offers optional AC adapters; utilizing an adapter not supplied by Rice Lake Weighing Systems voids all warranties and approvals.
  • Weight exceeding the maximum capacity may damage the scale.

2.0 Assembly

2.1 Unpacking

Place the carton on a hard, level surface for unpacking, preferably in the area where it will be used. Carefully lift the scale out of the packaging material; lifting it by the scale base. If parts were damaged in shipment, notify Rice Lake Weighing Systems and the shipper immediately.

IMPORTANT: The scale base and scale column are connected by a cable and requires great care when removing from the box so that the cabling does not get damaged.

Parts contained in the shipping box include:

  • Scale (the base and attached column)
  • Height rod
  • White box labeled Parts Inside which contains the following:
    • Wheels (2)
    • Wheel hinge (1)
    • Pan head screws (8)
    • AA batteries (6)

2.1.1 Repacking Scale

All component parts have been wrapped in plastic. Retain the packaging for use in the event that the scale must be returned or moved. It must be properly packed with sufficient packing materials. Whenever possible, use the original shipping carton when shipping the scale back.

IMPORTANT: Damage caused by improper packaging is not covered by the warranty.

2.2 Scale Assembly

Use the following steps to set up the scale.

NOTE: A Phillips head screwdriver will be needed for installation.

  1. Place the scale on a hard, level surface for the most accurate weighments.
  2. Thread the excess cable through the lower column to eliminate possible pinching during installation.
  3. Place the lower column on the base making sure not to pinch the cable.
  4. Insert the screws through the bottom of the scale base to secure the column.
  5. Attach wheels to the wheel assembly rod.
  6. Attach the wheel assembly to the scale with a screw, tighten securely.
  7. Ensure that the cable is extending through the top of the lower column.
  8. Plug the cable from the lower column into the upper column.
  9. Place the upper column onto the lower column, aligning the screw holes. Ensure the cable does not get pinched.
  10. Secure with two screws.

2.3 Height Rod Installation

Use the following steps to install the height rod assembly.

  1. The height rod comes in two pieces. Connect the two pieces together, ensuring the push buttons are popped out and fully engaged to lock the two height rod pieces together.
  2. Slide the assembled height rod from the top of the column into the slot located on the front of the scale.

2.4 Inserting Batteries

The six AA batteries that come with the scale offer an average of 25 hours of continuous use.

To install the batteries:

  1. Turn thumbscrew counterclockwise then remove battery cover.
  2. Insert batteries into the battery chamber.
  3. Put the cover in place and turn the thumbscrew clockwise to secure.

2.5 AC Power Connections

Use the optional 120 VAC adapter or 230 VAC adapter when power is available. The optional AC power adapter plugs into the back of the indicator as shown below. Rice Lake Weighing Systems offers optional AC adapters; utilizing an adapter not supplied by Rice Lake Weighing Systems voids all warranties.

  • RS-232 Connection
  • Connect the optional AC power adapter.

2.6 Leveling the Scale

Use bubble level to check for level and adjust feet as needed. On a flat surface, adjust the scale feet until the bubble indicates that the scale is level.

3.0 Operation

This section describes the front panel and includes procedures for operation of the scale.

Key Descriptions

The unit has nine front panel keys. Key functions are described in the table below.

IMPORTANT: The front panel keys are very sensitive, so only a gentle press is required.

KeyNameFunction
On/OffOn/OffPowers the scale on or off.
Print LB/KGPrint LB/KGSends data out from the RS-232 port. Allows to toggle between kilograms and pounds providing that it is enabled in Configuration mode. Cannot toggle while in the BMI mode.
Zero+0+Only functions if the current weight is stable and less than 2% of the capacity of the scale. Anything over 2% requires a recalibration.
Hold ReleaseHold ReleaseDisplays most current weight value on the display and holds that value when the patient is off the scale. A second press releases the weight value. Not active while in BMI mode.
BMIBMIPressing the BMI key enables access to the BMI (Body Mass Index) mode (defaults when scale is turned on). The patient is gets on the scale, weight stabilizes and press the BMI key. The display then asks for the patient height to calculate out the patient BMI. When using the BMI function, the display will be looking for a height entry. Pressing Clear changes this entry back to 190.0 cm (default) or 5 ft, 7.5 in. Once BMI is displayed, pressing the Clear key exits BMI.
CLEARCLEAR
ENTERENTERUsed to accept height in BMI mode; accepts the value of the parameter last entered and moves to the next stage. Pressing and holding Enter during startup will display ID. This is the first setup on entering into configuration mode.
Up/Down ArrowsAdjusts height input (0.5 in/0.5 cm) while in BMI mode.

3.1 Weighing

Use the following steps to weigh a patient.

  1. Press On/Off to turn on the scale. 0.0 displays along with ZERO on the upper display.
  2. Have the patient step on the scale. The patient's weight is displayed, the LOCK annunciator is on and the indicator beeps to indicate the end of the weighing process.
  3. Press Print LB/KG to change the display from lb to kg and vice-versa.
  4. Press and hold On/Off until OFF displays to turn off the scale.

3.2 Hold/Release Function

Use the following steps to use the Hold/Release function:

  1. Press On/Off to turn on the scale. 0.0 displays along with ZERO on the upper display.
  2. Have the patient step on the scale.
  3. Press Hold Release once the patient's weight stabilizes. The patient's weight and the HOLD and LOCK annunciators remain on the display when the patient is off the scale.
  4. Press Hold Release again to return the scale to zero.

NOTE: Pressing Hold Release prior to the patient getting on the scale will also hold the weight display.

3.3 Using the Body Mass Index (BMI) Function

Use the following steps in determining the BMI.

LB Mode

  1. Ensure the scale is at zero.
  2. Have the patient step onto the scale.
  3. Have the person step on the scale to obtain a weight. The LOCK annunciator appears on the display.
  4. Press BMI. The BMI and FT/IN annunciators light and a default height value of 5 feet and 7.5 inches (5-07.5) flashes.
  5. Use the Up/Down Arrows to adjust the height value, and press ENTER.
  6. The BMI value and BMI annunciator are shown on the display. Press CLEAR to return to weighing mode and the BMI function will be turned off.

KG Mode

  1. Ensure that the scale is at zero.
  2. Have the patient step on the scale to obtain a weight. The LOCK annunciator appears on the display.
  3. Press BMI. The BMI and CM annunciators appear and a default height value of 170.0 cm (170.0) flashes.
  4. Use the Up/Down Arrows to adjust the height value.
  5. Press ENTER.
  6. The BMI value and BMI annunciator appear on the display. Press CLEAR to return to weighing mode and the BMI function.

3.4 Troubleshooting

Refer to the following table to check and correct any failure before contacting service personnel.

SymptomPossible CauseCorrective Action
Scale does not turn onDead batteriesReplace batteries or connect to AC power
Faulty electrical outletUse a different electrical outlet
Bad power supplyReplace adapter
Questionable weight or the scale does not zeroExternal object is interfering with the scaleRemove the interfering object from the scale
Display did not show 0.0 before weighingHelp the patient off the scale, zero the scale and begin the weighing process again
Scale is not placed on a level floorEnsure scale is level and begin the weighing process again
Scale is out of calibrationCheck the weight with a certified calibration weight
The display shows a STOP messageThe load on the scale exceeds the capacity of the scaleRemove the excess weight and use the scale according to manufacture specifications
The display shows LO Bat messageThe battery is lowReplace batteries
The display shows E and Err messages as detailed belowE06Identifier - ADCAD too high
E07ADCAD too low
E10OverloadScale has been overloaded. Remove load from scale
E4LBATBattery low, but still usable- one bar left on indicator display
E4UBATBattery low and unstable - no bars left on indicator display
E11CALCalibration Error - recalibrate scale
Err 1Load cell cable may be plugged into wrong connection portEnsure cable is connected to the load cell connection port. Note: Load cell connection point is located underneath the curved plastic cover of the indicator. Remove four back retaining screws, remove curved back cover to access load cell connection point.
Err 2Low saturation state (low A/D)The load cell is not connected properly; Check the cables and mechanical connections; if the problem persists, replace the set of load cells
Err 3High saturation state (high A/D)See Err 2
Err 6Unstable weight; Cannot calibrateCheck the load cell mechanical surroundings and ensure nothing is contacting the load cell and that the cables are properly welded
Err 7Scale isn't movingMake sure feet are installed on the scale. Turn the feet all the way in and then back them out three full turns, then level the scale
SATDamaged load cell cableReplace load cell cable
Load cell cable may be plugged into wrong connection portEnsure cable is connected to the load cell connection port. Note: Load cell connection point is located underneath the curved plastic cover of the indicator. Remove four back retaining screws, remove curved back cover to access load cell connection point.

3.5 Using the Height Rod

The height rod can be extended to accommodate patients of different heights and measures from 27.5 to 82 x 1/8 inch (70 to 208 x 1 mm). It is comprised of a stationary outer sleeve and an inner rod that slides up and stays in place once extended. Measurements are shown in inches and centimeters.

  1. To extend the length of the height rod, hold the white latch located next to the indicator display and pull it up vertically.
  2. Raise the headpiece until it is perpendicular to the height rod and snaps into place.
  3. Raise the rod until the person can easily stand without touching the headpiece.
  4. Have the patient step onto the scale with their head level.
  5. With a patient on the scale, lower the entire height rod (not just the headpiece), until the headpiece just touches the top of the patient's head.

NOTE: The headpiece should remain level, not slanted up or down.

  • Inaccurate Headpiece Reading
  • Inaccurate Headpiece Reading

6. For patients taller than 43 in (109 cm) use the measurements at the read line (inner rod).

For patients shorter than 43 in (109 cm), use the outer sleeve measurements (below the read line).

4.0 Configuration

Options and parameter setup are done through the scale configuration. Access to the setup switch is located inside the battery cover (shown below).

4.1 General Navigation

Use the buttons on the front panel to navigate through the menus and parameters.

4.1.1 Change Parameters

  • Press BMI to scroll through the menus and/or parameters
  • Press ENTER to enter a displayed menu and/or parameter
  • Press Up/Down Arrows or to scroll through values
  • Press ENTER to save the displayed selection and move to the next parameter

4.1.2 Enter Numbers

  • Press ENTER to enter parameter value
  • Press Up/Down Arrows or to increment/decrement numbers
  • Press BMI to move to the next digit
  • Press ENTER to save value and move to the next parameter

4.1.3 Save and Return to Main Menu

  • When a parameters selection/value is correct, press ENTER. The next parameter displays
  • When all parameters selections/values are correct, SAVE displays
  • Press ENTER. DONE displays.
  • Press ENTER to save settings and return to weigh mode.

4.2 Configuration Mode

Use the following steps to enter into Configuration mode.

  1. Ensure the scale is turned off.
  2. Turn the scale on by simultaneously pressing On/Off and ENTER. Continue to hold both keys until Id appears. The unit cycles through its startup function and continues to display the software version.
  3. Access the setup switch located in the back of the scale to enter the setup parameters for the scale. Use a small paper clip, small screwdriver or other similar object to press the setup switch.
  4. Once the setup switch is pressed, PROG displays.
  5. Scale can be configured using a series of menus accessed through the front panel when the scale is in Setup mode.
  6. Press BMI to advance to the desired menu.
  7. Press ENTER and advance in the manual to the related menu selection for further instructions.

4.3 Programming Mode Menu

Various parameters can be set while in Programming mode.

Table 4-1. Configuration Mode Menu

ParameterDescriptionChoicesSteps
FULLFull capacity of the scaleValue (capacity of scale)The display toggles between a numeric value and FULL; If you do not want to change the value, press the BMI key to move to the next setting; Example: from FULL to LOAD. If you want to change the value, use the following steps: 1. Press ENTER key to change value starting with the right most digit. 2. Use the Up/Down arrow keys to increment/decrement numbers. 3. Press BMI key to move to the left. 4. Use the Up/Down arrow keys to increment/decrement numbers. 5. Press BMI key again to move to the left 6. Use the Up/Down arrow keys to increment/decrement numbers. 7. When done press ENTER key to move to the next parameter (LOAD).
LOADThis is the amount of weight applied during calibration; Can also be changed in the calibration menuValue (200 lb)The display toggles between a numeric value and LOAD; If you do not want to change this value, press the BMI key to move to the next setting; Example: from LOAD to ASTART; If you want to change the value, use the following steps: 1. Press ENTER to change value starting with the right most digit. 2. Use the Up/Down arrow keys to increment/decrement numbers. 3. Press BMI key to move to the left. 4. Use the Up/Down arrow keys to increment/decrement numbers. 5. When done, press ENTER key to move to next parameter (G-CAL).
G-CALGravity calibration; sets gravity for original calibration00000The display toggles between 00000 and G-CAL. Press the BMI key to move to the next setting. If you want to change the value, use the following steps; 1. Press ENTER to change the value starting with the right most digit. 2. Use the Up/Down arrow keys to increment/decrement numbers 3. Press BMI key to move to the left 4. Use the Up/Down arrow key to increment/decrement numbers 5. When done, press ENTER to move to the new parameter (G-LOC)
G-LOCGravity location; set gravity for location of us00000The display toggles between 00000 and G-LOC. Press the BMI key to move to the next setting. If you want to change the value, use the following steps; 1. Press ENTER to change the value starting with the right most digit. 2. Use the Up/Down arrow keys to increment/decrement numbers 3. Press BMI key to move to the left 4. Use the Up/Down arrow key to increment/decrement numbers 5. When done, press ENTER to move to the new parameter (ASTART)
ASTRTWeight process start limit — Maximum (full capacity)/10; Determine when the weight algorithm starts (when the "----" is displayed), below this value the scale will show live weightValue (2.0)The display toggles between a numeric value and ASTART; If you do not want to change this value, press the BMI key to move to the next setting; Example: from ASTART to ARW; If you want to change the value, use the following steps: 1. Press ENTER key to change value starting with the right most digit. 2. Use the Up/Down arrow keys to increment/decrement numbers. 3. Press BMI key to move to the left. 4. Use the Up/Down arrow keys to increment/decrement numbers. 5. When done, press ENTER key to move to the next parameter (ARW).
ARWAuto Reweigh - Restarts the weight algorithm if the weight changed by more than this value.Value (4.0 lb)The display toggles between a numeric value and ARW; If you do not want to change this value, press the BMI key to move to the next setting; Example: from ARW to SAL; If you want to change the value, use the following steps; 1. Press ENTER key to change value starting with the right most digit. 2. Use the Up/Down arrow keys to increment/decrement numbers. 3. Press BMI key to move to the left. 4. Use the Up/Down arrow keys to increment/decrement numbers. 5. When done, press ENTER key to move to the next parameter (SAL).
SALSemi Auto Live —This value is the interval between weight displays during the algorithm processValue (0.5)The display toggles between a numeric value and SAL; If you do not want to change this value, press the BMI key to move to the next setting; Example: from SAL to ROUND; If you want to change the value, use the following steps: 1. Press ENTER key to change value starting with the right most digit. 2. Use the Up/Down arrow keys to increment/decrement numbers. 3. When done, press ENTER key to move to the next parameter (ROUND).
ROUNDScale Resolution0.2
0.5
1.0
2.0
5.0
10.0
20.0
The display toggles between a numeric value and ROUND; The decimal point location is set to the DISP parameter display decimal point location; If you do not want to change this value, press the BMI key to move to the next setting; Example: from ROUND to DISP; If you want to change the value, use the following steps: 1. Press ENTER key to change value. 2. Press the Up/Down arrow keys to change the available parameters. 3. When done, press ENTER key to move to the next parameter (DISP).
DISP0.1
0.0000
0.000
0.00
The display toggles between a numeric value and DISP; If you do not want to change this value, press the BMI key to move to the next setting; Example: from DISP to BAUD; If you want to change the value, use the following steps: 1. Press ENTER key to change value. 2. Use the Up/Down arrow keys to change the available parameters. 3. When done, press ENTER key to move to the next parameter (BAUD).
BAUDBaud rate96
48
1152
576
384
288
192
144
Indicator display illustrates first two digits of baud rate only; The display toggles between a numeric value and baud; If you do not want to change this value, press the BMI key to move to the next setting; Example: from BAUD to ATOL; If you want to change the value, use the following steps 1. Press ENTER key to change value: 2. Use the Up/Down arrow keys to change the available parameters. 3. When done, press ENTER key to move to the next parameter (ATOL).
ATOLAlgorithm initial tolerance - Maximum value is 255. Values above 255 will not let you proceed and will return to the previous value.Value (10)The display toggles between a numeric value and ATOL; If you do not want to change this value, press the BMI key to move to the next setting; Example: from ATOL to ALEN; If you want to change the value, use the following steps: 1. Press ENTER key to change value starting with the right most digit. 2. Use the Up/Down arrow keys to increment/decrement numbers. 3. Press the BMI key to move to the left. 4. When done, press ENTER key to move to the next parameter (ALEN).
ALENAlgorithm initial exponent - Maximum value 10. Values above 10, will not let you proceed and will return to the previous value.Value (8)The display toggles between a numeric value and ALEN; If you do not want to change this value, press the BMI key to move to the next setting; If you want to change the value, use the following steps: 1. Press ENTER key to change value starting with the right most digit. 2. Use the Up/Down arrow keys to increment/decrement numbers. 3. When done, press ENTER key to move to the next parameter (ATOUT).
ATOUTAlgorithm maximal exponent — Maximum value is 15. Values above 15, will not let you proceed and will return to the previous value.Value (10)The display toggles between a numeric value and ALEN; If you do not want to change this value, press the BMI key to move to the next setting; If you want to change the value, use the following steps: 1. Press ENTER key to change value starting with the right most digit. 2. Use the Up/Down arrow keys to increment/decrement numbers. 3. When done, press ENTER key to move to the next parameter (ATOUT).
TOFFAuto off timer - Measured in minutes; 0 = always on; Maximum is 9 minutes; When using an external power supply, this parameter is irrelevant5
4
3
2
1
0
9
8
7
6
The display toggles between a numeric value and ATOUT; If you do not want to change this value, press the BMI key to move to the next setting; If you want to change the value, use the following steps: 1. Press ENTER key to change value starting with the right most digit. 2. Use the Up/Down arrow keys to increment/decrement numbers. 3. Press BMI key to move to the left. 4. Use the Up/Down arrow keys to increment/decrement numbers. 5. When done, press ENTER key to move to the next parameter (TOFF).
SPACENumber of new lines after print7The display toggles between a numeric value and TOFF; If you do not want to change this value, press the BMI key to move to the next setting; Press the ENTER key to move to the next parameter; (UNITS); If you want to change the value, use the following steps; 1. Press the ENTER key to change values. 2. Use the Up/Down arrow keys to change the available parameters. 3. When done, press ENTER key to move to the next parameter (UNITS).
UNITSUnits - Selects the unit of measure; It can be either Kg/Lb, Kg only or Lb onlyKG/LB
KG
LB
The display toggles between a numeric value and SPACE; If you do not want to change this value, press the BMI key to move to the next setting: Press the ENTER key to move to the next parameter; (UNITS); If you want to change the value, use the following steps; 1. Press the ENTER key to change values. 2. Use the Up/Down arrow keys to change the available parameters. 3. When done, press ENTER key to move to the next parameter (UNITS). The display toggles between unit of measurements and UNIT; If you do not want to change this value, press the BMI key to move to the next setting; Example: from UNITS to OP; If you want to change the value, use the following steps: 1. Press the ENTER key to change values. 2. Press the Up/Down arrow keys to change the value. 3. When done, press the ENTER key to move to the next parameter (OP).
OPBinary options: OPO - Live weighing options (0=disable, 1=enable) OP1 - Communication protocol (0=ESC, 1=maintenance) OP2 - BMI menu (0=disable, 1=enable) OP3-RTC power (0=disable, 1=enable) OP4 - Semi-Auto-Live — (0=disable, 1=enable) OP5 - Full calculation — (0=spatial, 1=multiply by (0=disable, 1=enable OP6 - Tare - (0=disable, 1=enable OP7 - Bat type - (0=dry batteries, 1=rechargeable batteries) OP8 - OIML mode - (0=disable, 1=enableOPO
OP1
OP2
OP3
OP4
OP5
OP6
OP7
OP8
The display toggles between a binary option and POO-0; If you do not want to change this value, press the BMI key to move to the next settings If you want to change the value, use the following steps: 1. Press ENTER to change parameters. 2. Use the Up/Down arrow keys to select the display value you want to change. 3. Press BMI key to move the flashing cursor a. Use the Up/Down arrows to change the value. b. Press the BMI key to move the flashing cursor. 4. Press ENTER key to save all of the display parameters. SAVE appears on the display. 5. Press ENTER key again and DONE appears indicating that you are now done entering all of the parameters of the scale. Press the BMI key to cycle back to the top level menu choices, ie: PROG/CALIB/DEF/DONE

4.4 Default Menu

The default menu is used to return the scale back to its factory settings and is shown below.

  1. Press ENTER. The display shows a default value of NO.
  2. To change to YES, use Up/Down Arrows to adjust the value.
  3. Press ENTER and the display shows DONE.
  4. Press BMI to return to PROG/CALIB/DEF/DONE upper level menu.

NOTE: Selecting YES and pressing ENTER will reset to factory defaults settings without changing the calibration and will return you to Weigh mode.

4.5 Scale Calibration

Use the following steps to calibrate the scale.

  1. Press On/Off and ENTER simultaneously to power on the scale.
  2. The unit cycles through its startup function and continues to display the software version. Continue to hold both keys until Id appears.
  3. Access the setup switch located in the back of the scale to enter the setup parameters for the scale. Use a small paper clip, small screwdriver or other similar object to press the setup switch on the back of the indicator (under the cover).
  4. Prog displays. Press BMI to toggle along the parameter menu.
  5. Calib displays and enter the calibration parameters.
  6. Press ENTER and a numeric value is displayed which represents the amount of weight that is used for calibration.
  7. Lb will be flashing. To switch between lb and kg, press Print LB/KG or Up/Down Arrows. Once a unit is selected, press ENTER and the right most digit will be flashing.
  8. To change the calibration load value, press Up/Down Arrows or to increment/decrement the flashing digit.
  9. Use BMI to move the flashing digit to the left or right.
  10. Once all the digits have been entered, press ENTER and Clear displays.
  11. Make sure the scale platform is clear of weight and press ENTER again then ===== displays.
  12. A request to put the chosen load on the platform is displayed by Put xxx.xx.
  13. Put the chosen weight on the platform and press ENTER ===== displays then Save displays.
  14. Press ENTER again and the display indicates Done.
  15. Press BMI three times to exit back out to the top level Done parameter.
  16. Press ENTER to return to Weigh mode.
  17. To exit calibration without changing zero or span existing calibration, press CLEAR, then press BMI.

4.6 Test Menu

To access the TEST menu, use the following steps.

  1. Simultaneously press On/Off and ENTER to power on the scale until ID flashes.
  2. Press ENTER. Press ENTER to access the test menu.
  3. Continue to press BMI to scroll through the various menu items.
  4. Once complete, press BMI until Done displays.
  5. Press ENTER to start the weighing process.

Table 4-2. Test Menu

ParameterChoiceDescription
RIDValueInternal ID number
VERValueDisplays the current software version
VALUEValueDisplays the actual value
A2DValueDisplays the actual raw counts of the scale
BATValueDisplays the current battery level

5.0 Communication

The unit comes with an RS-232 port that enables weight data to be transmitted to other equipment, such as a computer or printer. The RS-232 cable with DB-9 connector (PN 100719) is available from Rice Lake Weighing Systems. That connection is shown in USB Connection section.

The RS-232 parameters are:

  • 9600 baud (selectable in the programming mode)
  • 8 data bits
  • 1 stop bit
  • No parity
  • No handshaking

There are three methods of communication:

  • Push-button keypad print
  • Standard remote protocol
  • Escape protocol

5.1 Push-button Keypad Print

With a stable, in-range weight, press and hold Print LB/KG for at least three seconds, or until the scale emits two quick beeps.

NOTE: If the scale does not beep after five seconds, release Print LB/KG as the weight was either in motion or out of range.

If displaying weight and not BMI, the scale will send out the following 21 character string:

xxxxxxxxxuummmmm

Table 5-1. Print Format Tokens

TokenDescription
XXXXXXXXXWeight with decimal point and “-” sign
Space
uuUnit - Ib or kg
mmmmmMode - gross or net
Carriage return
Line feed (moves cursor down to the next line)

60.1 KG= -60.1KG

In BMI mode (displaying the BMI value), the scale will send out the following data:

PATIENT WEIGHT 60.1 KG
PATIENT HEIGHT 170.0 CM
PATIENT BMI 20.8

Example in KG:

-60.1KG
-170.0CM
<20.8

Example in LB:

132.4LB
-5-07.5FT
<20.4

In case of under weight or over weight, the word Under or Over will be sent correspondingly.

5.2 Communication Protocols

The scale has two communication protocols, escape and maintenance protocol.

5.2.1 Escape Protocol

An escape protocol is where the escape (0X1B or ASCII 27) is used to indicate that there is a command following. On the PC side there must be a listener created by the vendor that will interpret this protocol. This listener must also take care of all the issues regarding data integrity to make sure that the data that was sent and received is valid.

Two examples include:

  • Scale initiated communication
  • PC initiated communication

The escape protocol commands table shows (below) what can be sent across communications lines.

Table 5-2. Escape Protocol Commands

PC InitiatedESC Value
Request current values/settingsR
DiagnosticsA
Send scale control messagesC
PC InitiatedESC Value
Send single readingR
Send diagnostic responseR

ESC characters that will be used is shown below.

Table 5-3. ESC Characters

NameESC CharacterESC Value with ParametersDescription
ReadingRRTells PC the scale is sending a reading; immediately following this is the value that is sent Example: NmE
WeightWWnnn.nThe patient weight (Example: W02000 means 200.0). If scale is overloaded or under loaded, 999.99 is returned
HeightHHnnn.nPatient height
BMIBBnn.nPatient BMI
UnitsNNcIndicates the units the values have been taken (m=metric, c=constitutional).
End of Packet (EOP)EEIndicates the end of the command has been reached.
Diagnostics (request)AAcccA request for a diagnostic test on certain parts of the scale (like battery life, load cells).
Diagnostics (response)ZZcccThe response of the diagnostics done on the scale; values include error codes to indicate an issue, or all zeros (Z000) to indicate the scale is performing properly
Control (set a value)CCccc=cSets the value of the scale's global settings Example:

Table 5-4. Scale Global Values and Identifiers

Name of ControlIdentifierUnit
Unit of Measure (metric or constitutional)UOMc (m or c)

Figure 5-1. Escape Protocol Examples

Figure 5-2. Diagnose Battery Examples

5.2.2 Maintenance Protocol

Maintenance protocol commands are listed below.

Table 5-5. Maintenance Protocol Commands

CommandDefinition
RReboot
VFirmware ID + development version
WCurrent weight
ACurrent AD
ZZero the scale
FShow flash values (used for the first flash process)
LUSB On/Off (not available on USB communication

5.3 USB Connection

The scale has the capability of connecting to a Windows® computer (PC) using a USB cable (not included) and a terminal emulation program. A terminal emulation program allows the transfer of data between the scale and PC using a serial port.

  • RS-232 Connection
  • USB Connection

NOTE: Apple® and Macintosh® computers are unable to transfer the necessary data to the scale. Only use a PC for data transfer.

Connecting software and downloads should always be addressed by the IT department for safety reasons and can vary depending on what type of computer platform is being used.

NOTE: Consult the IT department if driver protections are preventing the use of the USB driver. Driver protections may need to be temporarily disabled on Windows 10 or later computers to allow for the installation of the USB driver.

  1. Connect the scale's indicator to a PC using a USB-Type B to USB-Type A cable (not included).
  2. Turn the indicator on.

NOTE: In most cases, the PC should find the driver and automatically configure the driver when the scale is plugged into a USB port.

  1. Open a terminal emulation program, such as Advanced Serial Port Terminal, pUtty or Hercules (used in this example).
  2. Connect to the serial port assigned by the PC (COM5 in example). This can be found in Device Manager. Once selected, press Open.

Figure 5-4. Connecting to a Serial Port

5. With weight on the scale, press and hold the Print button on the indicator for three seconds. The patient's weight is sent to the PC.

Figure 5-5. Patient Weight Displayed

6.0 Maintenance

The following section provides instructions for maintaining and cleaning the unit.

IMPORTANT: Do not immerse the scale in cleaning or other liquid solutions.

Do not use Isopropyl alcohol or other solutions to clean the indicator display surface.

6.1 Basic Maintenance

Before the first use of the scale and after periods of non-use, check the scale for proper operation and function. If the scale does not operate correctly, contact a qualified service personnel.

Go through the following steps for basic maintenance.

  • Check the overall appearance of the entire scale for any obvious signs of damage
  • Inspect the condition of the AC power adapter cord for cracking, fraying or for broken or bent prongs

6.2 Cleaning

Proper care and cleaning is essential to ensure a long life of accurate and effective operation. Before beginning the cleaning process, disconnect the scale from the AC power source.

  • Clean all external surfaces with a clean, damp cloth or tissue. Mild soap and water solution may be used. Dry with a clean soft cloth
  • Do not immerse the scale into cleaning or other liquid solutions
  • Do not use Isopropyl alcohol or other solutions to clean the display surface

7.0 Specifications

  • Power: 120 VAC-9VDC-60Hz / 230 VAC-9VDC-50Hz
  • Battery Type: 6 AA size Alkaline batteries
  • Battery Use: 25 hours continuous use; Automatic power-off can be configured
  • Data Communications: RS-232 with RJ-45 jack; Selectable baud rate, default - 9600; 8 bits; No parity; 1 stop bit; No handshaking
  • Environmental:
    • Operating Temperature: 50°F to 104°F (14°C to 40°C)
    • Storage Temperature: 32°F to 158°F (0°C to 70°C)
    • Humidity: 85% relative humidity
  • Capacity: 550 lb x 0.2 lb (250 kg x 0.1 kg)
  • Dimensions:
    • Scale Platform: 14.7 in W x 14.7 in L x 3 in H (37.5 cm W x 37.5 cm L x 7.6 cm H)
    • Overall Height: 54 in (137.2 cm)
Models: 150-10-5 Digital Physician Scale, 150-10-5, Digital Physician Scale, Physician Scale, Scale

File Info : application/pdf, 32 Pages, 1.88MB

PDF preview unavailable. Download the PDF instead.

m 126178 150-10-5 tech enus revh

References

Related Documents

Preview Rice Lake 150-10-5 Digital Physician Scale Operation Manual
This operation manual provides detailed instructions for the Rice Lake 150-10-5 Digital Physician Scale, covering assembly, safe operation, BMI calculation, troubleshooting, maintenance, and technical specifications. Learn how to use the scale effectively and ensure its longevity.
Preview Rice Lake 150-10-8 Digital Physician Scale Operation Manual
Operation manual for the Rice Lake 150-10-8 Digital Physician Scale, covering assembly, operation, maintenance, troubleshooting, and specifications.
Preview Rice Lake 150-10-5 Digital Physician Scale Technical Manual
This technical manual provides comprehensive instructions for the Rice Lake 150-10-5 Digital Physician Scale, covering assembly, operation, configuration, communication, maintenance, and specifications. It details how to set up, use, and troubleshoot the scale for accurate weight and BMI measurements.
Preview Rice Lake 150-10-6 Digital Physician Scale Technical Manual
This technical manual provides comprehensive instructions for the Rice Lake 150-10-6 Digital Physician Scale, covering assembly, operation, configuration, communication, maintenance, and specifications.
Preview Rice Lake 150-10-5 Digital Physician Scale Operation Manual
This operation manual provides detailed instructions for the Rice Lake 150-10-5 Digital Physician Scale, covering setup, operation, maintenance, and troubleshooting.
Preview Rice Lake 150-10-6 Digital Physician Scale Operation Manual
Operation manual for the Rice Lake 150-10-6 Digital Physician Scale, covering setup, operation, maintenance, and troubleshooting.
Preview Rice Lake 150-10-6 Digital Physician Scale Operation Manual
Comprehensive operation manual for the Rice Lake 150-10-6 Digital Physician Scale. Includes detailed instructions on assembly, operation, weighing, BMI calculation, communication, maintenance, and specifications. Essential guide for healthcare professionals.
Preview Rice Lake 150-10-7 Digital Physician Scale Operation Manual
Operation manual for the Rice Lake 150-10-7 Digital Physician Scale, covering setup, operation, maintenance, and specifications.