AEMC Amperimetric Probes Datasheet: Series MR, MN, SR, MF, and Oscilloscope Models

AEMC Amperimetric Probes

Image of AEMC MR415, MR416, MR525, and MR526 Amperimetric Probes.

Series MR Amperimetric Probes

Models MR415, MR416, MR525, and MR526

Hall-effect and general-purpose amperimetric probes, for use with digital multimeters, data loggers, and power analyzers.

Specifications

MODELMR415MR416MR525MR526
ELECTRICAL CHARACTERISTICS
Measurement Range0.5 to 400 ACA, 600 ACC0.5 to 40 ACA, 60 ACC;
0.5 to 400 ACA, 600 ACC
0.5 to 1000 ACA,
1400 ACC
0.5 to 100 ACA, 150 ACC;
0.5 to 1000 ACA, 1400 ACC
Output Signal1 mV/A10 mV/A, 1 mV/A1 mV/A10 mV/A, 1 mV/A
Frequency RangeDC to 30kHz (at -3 dB) (depends on current value)
Phase Shift (at 50/60 Hz)≤1.5° at 400 A
<2.2° at 40 A
≤1.5° at 400 A
<2.2° at 40 A
≤1.5° at 400 A
<2.2° at 40 A
≤1.5° at 400 A
<2.2° at 40 A
Load Impedance>1 MΩ and ≤100 pF
Overload3000 ACC or 1000 ACA continuous at <1 kHz
ZeroingAutomatic on all ranges
Power SourceOne 9V alkaline battery or 5 VDC with micro USB Type B
Battery Life50 hours typical
Low Battery IndicatorGreen LED flashing
Overload IndicatorRed LED flashing (indicates measurement exceeds selected range)
Output Terminal1.5 m (5 ft) cable with double insulation and two 4 mm safety banana terminals
MECHANICAL CHARACTERISTICS
Max. conductor sizeCables: (1) 30 mm (1.18 in.) or (2) 24 mm (0.94 in.)
Busbars: (1) 50 x 10 mm (1.97 x 0.39 in.) or (2) 31.5 x 10 mm (1.23 x 0.39 in.) or (3) 25 x 8 mm (0.98 x 0.31 in.)
Cables: (1) 39 mm (1.5 in.) or (2) 25.4 mm (1 in.)
Busbars: (1) 50 x 12.5 mm (1.97 x 0.49 in.) or (2) 25 x 5 mm (0.98 x 0.2 in.) or (1) 31.5 x 10 mm (1.24 x 0.30 in.) or (3) 25 x 8 mm (0.98 x 0.31 in.)
Dimensions224 x 97 x 44 mm (8.82 x 3.82 x 1.73 in.)236.5 x 97 x 44 mm (9.31 x 3.82 x 1.73 in.)
Weight (with battery)440 g (0.98 lbs)520 g (1.15 lbs)
ENVIRONMENTAL CHARACTERISTICS
Operating Temperature-10 to 55°C (14° to 131°F)
Storage Temperature-40 to 80°C (-40° to 176°F)
Relative Humidity0 to 85% with linear decrease above 35°C
SAFETY
ProtectionIP40
Electromagnetic CompatibilityEN 61326-1
Electrical SafetyIEC 61010-1, EN 61010-2-32, Pollution Degree 2, 600 V CAT III

Key Features

  • ? The clamp shape allows it to be placed around small cables or busbars.
  • ? Battery powered or via an external 5V source using the micro USB connector.
  • Measurements up to 1000 ACA and 1400 ACC (model dependent).
  • ⚙️ DC zeroing function.
  • ⚙️ Auto-power-off function that can be enabled or disabled.
  • ? Millivolt output compatible with most instruments.
  • ? Battery life up to 50 hours.
  • 600 V CAT III safety rating.

Catalog Numbers

CATALOG NUMBERDESCRIPTION
1200.80Amperimetric probe CA/CC model MR415 (400 ACA/600 ACC, 1 mV/A, cable with 4 mm banana connectors) Replacement for model MR410
1200.82Amperimetric probe CA/CC model MR416 (40 ACA/60 ACC, 10 mV/A and 400 ACA/600 ACC, 1 mV/A, cable with 4 mm banana connectors) Replacement for model MR411
1200.81Amperimetric probe CA/CC model MR525 (1000 ACA/1400 ACA, 1 mV/A, cable with 4 mm banana connectors) Replacement for model MR520
1200.83Amperimetric probe CA/CC model MR526 (100 ACA/150 ACC, 10 mV/A and 1000 ACA/ 1400 ACC, 1 mV/A, cable with 4 mm banana connectors) Replacement for model MR521

General Use Amperimetric Probes (MN Series)

Image of various AEMC MN Series Amperimetric Probes.

SERIESMODELRATIOMEASUREMENT RANGE
AC
DCOUTPUT SIGNAL
CURRENT
VOLTAGEPHASE SHIFT**MAX. CONDUCTOR SIZE
CABLE
BUS BAROUTPUT CONNECTIONCATALOG NUMBER
MN012 to 150 A1 mA/A*10 mm (0.39 in.)N/A2129.17
MN021000:150 mA to 100 A
50 mA to 90 A
2129.20
MN032 to 100 A1 mV/A2129.18
MN055 MACA to 10 A
1 to 100 A
1 mV/mA
1 mV/A
Cable2129.19
MN1031 mA to 10 A
1 to 100 A
1 mV/mA
1 mV/A
1031.02
MN1061000:12 to 150 A1 mA/A<10°12 mm (0.47 in.)1031.17
MN1141 mA to 10 A100 mV/A<8°2110.71
MN18550 mA to 120 A1 mA/A<3.5°Connector100.185
MN2131000:10.5 to 240 A1 mA/A*2115.75
MN2510.1 to 240 A1 mV/A<2.5°Cable2115.77
MN2530.1 to 24 A
0.1 to 240 A
10 mV/A
100 mV/A
N/A2115.79
MN25510 mV/A2115.81
MN2910.5 to 240 A100 mV/AN/A2115.84
MN30710 mA to 12 A100 mV/A2116.23
MN3121000:10.1 to 200 A1 mA/A*<2.5°19.8 mm (0.78 in.)Connector2116.24
MN3131 mA/A*Cable2116.25
MN3520.1 to 150 A10 mV/AConnector2116.26
MN3530.01 to 2.4 A
0.1 to 200 A
1000 mV/A
10 mV/A
<3°2116.27
MN3732116.28
MN3750.1 to 10 A100 mV/A<1.5°2115.41
MN3795mA to 6 A
0.1 to 120 A
200 mV/A
10 mV/A
Cable2153.01
SL20610 mA to 1.5 A
50 mA to 60 A
10 mA to 2 A
50 mA to 80 A
1 mV/mACA/CC
10 mV/ACA/CC
<1°11.8 mm (0.46 in.)1201.45
MD3011000:12 to 500 A1 mVcc/ACAN/A30 mm
(1.18 in.)
63 x 5 mm
(2.48 x
0.20 in.)
1201.07
MD3051 to 600 A1 mA/A*<1°2 x 500 kcmil1201.36

*Protection for open secondary output
**Phase shift indicated at maximum load
Notes: MN103, MN106, MN114, and MN185 models are not CE marked. Probes from the MN200 and MN300 series have UL certification except for model MN379. Consult factory for NIST calibration pricing.

AmpFlex® and Miniflex® Probes

Image of AEMC AmpFlex and Miniflex probes.

SERIESMODELINDEXMEASUREMENT RANGE
AC
DCOUTPUT SIGNAL
CURRENT
VOLTAGEPHASE SHIFT**MAX. CONDUCTOR SIZE
CABLE
BUS BAROUTPUT CONNECTIONCATALOG NUMBER
MF 300-6-2-10-HF30 A / 300 A100 mV/A / 10 mV/A70 mm (2.95 in.)2126.84
MF 300-10-2-10-HF45 mm (1.77 in.)2126.83
MA1143 A / 30 A / 300 A / 3000 A1 mV/mA / 100 mV/A
10 mV/A / 1 mV/A
101 mm (4 in.)2153.41
300-24-2-1030 A / 300 A100 mV/A / 10 mV/A190 mm (8 in.)2112.88
1000-24-1-11000 A1 mV/A2112.39
1000-24-2-1100 A / 1000 A10 mV/A / 1 mV/A290 mm (11 in.)2129.98
1000-36-2-12113.00
3000-24-1-13000 A1 mV/A190 mm (8 in.)2112.46
3000-36-1-1290 mm (11 in.)2112.48
3000-24-2-1300 A10 mV/A190 mm (8 in.)2113.05
3000-36-2-13000 A1 mV/A290 mm (11 in.)2112.00
3000-48-2-1390 mm (15 in.)2112.01
6000-36-2-0.1600 A / 6000 A1 mV/A / 0.1 mV/A290 mm (11 in.)2113.21
30000-24-2-0.13000 A / 30000 A2113.33
24-3001300 ACA / 3000 ACA10 mV/A / 1 mV/A
(4000 mV peak max.)
190 mm (8 in.)2120.81

Note: Consult factory for calibration pricing.

Oscilloscope Probes with BNC Connectors

Image of AEMC Oscilloscope Probes with BNC connectors.

MODELMEASUREMENT RANGE
AC
DCOUTPUT SIGNAL
VOLTAGE
PHASE SHIFT**MAX. CONDUCTOR SIZE
CABLE
BUS BAROUTPUT CONNECTION
SL261100 mA to 10 A
1 to 100 A
100 mV/A
10 mV/A
<1.5°11.8 mm (0.46 in.)N/A2 m (6.6 ft) cable with BNC
MN2610.1 to 24 A
0.5 to 240 A
100 mV/A
10 mV/A
<2.5°19.8 mm (0.78 in.)
SR6610.1 to 12 A
0.1 to 120 A
1 to 1200 A
100 mV/A
10 mV/A
1 mV/A
<1°52 mm (2.05 in.)50 x 5 mm (1.96 x 0.19 in.)
MN251T0.5 to 240 A1 mV/A<2.5°20 mm (0.78 in.)20 mm (0.78 in.)3 m (10 ft) cable with BNC
MN379T0.005 to 6 A
0.1 to 120 A
200 mV/A
10 mV/A
<4°
<2°
MH600.5 to 100 A10 mV/A<1°26 mm (1.02 in.)N/A
MR4170.5 to 40 A
0.5 to 400 A
0.5 to 60 A
0.5 to 600 A
10 mV/A
1 mV/A
<2.2°
<1.5°
30 mm (1.18 in.)2 busbars of 31.5 x 10 mm (1.24 x 0.39 in.)2 m (6.6 ft) cable with BNC
MR5270.5 to 100 A
0.5 to 1000 A
0.5 to 150 A
0.5 to 1400 A
39 mm (1.53 in.)2 busbars of 50 x 5 mm (1.96 x 0.19 in.)

*Phase shift indicated at maximum load.
Note: All probes comply with 600 V CAT III safety rating and CE marking. Not all models have UL certification; please consult factory.

Output Terminals

Images illustrating different output terminal types.

Cable with BNC

2 m (6.5 ft) coaxial cable with a BNC connector, rated for 600 Vrms.

Connectors

Standard 4 mm safety banana connectors.

Cables

Reinforced 1.5 m (5 ft) cable with 4 mm banana terminals.

Protected Banana Terminals

Two 4 mm banana terminals with standard 19 mm (3/4 in.) spacing.

Safety and Certifications

All probes comply with the 600 V CAT III safety rating and CE marking. Not all models have UL certification; please consult factory for details.

PDF preview unavailable. Download the PDF instead.

1200-81-SP Adobe InDesign 16.4 (Windows) Adobe PDF Library 16.0

Related Documents

Preview AEMC MN Series AC Current Probes: Specifications and Selection Guide
Discover the AEMC MN Series of compact AC current probes, ideal for measuring AC currents in low-power secondary transformers and industrial applications. This guide details specifications, features, and selection charts for models like MN01, MN02, MN05, and MN09, alongside other AEMC probe series.
Preview AEMC MiniFlex HF Flexible AC Current Probes User Manual
User manual for AEMC MiniFlex HF series flexible AC current probes (MF 300-10-2-10 HF, MF 3000-14-1-1 HF), detailing features, specifications, operation, maintenance, and warranty.
Preview AEMC Instruments MD305 AC Current Probe User Manual and Specifications
Detailed user manual and technical specifications for the AEMC Instruments MD305 AC Current Probe. Covers description, safety warnings, electrical and mechanical specifications, operation, maintenance, and warranty information for industrial electrical measurements.
Preview AEMC MR6292 AC/DC Current Probe User Manual and Specifications
Comprehensive user manual for the AEMC MR6292 AC/DC Current Probe, detailing its features, specifications, operation, maintenance, and warranty. Designed for use with the AEMC Micro-Ohmmeter Model 6292.
Preview AEMC MN103 AC Current Probe User Manual
User manual for the AEMC MN103 AC Current Probe, detailing its description, safety warnings, electrical and mechanical specifications, operation, maintenance, repair, and warranty. Learn how to measure leakage and low AC currents with this instrument.
Preview AEMC Current Probes and Sensors for Power Quality Meters - User Manual
Comprehensive user manual for AEMC Instruments' range of current probes and flexible sensors, including models like AmpFlex, MiniFlex, J93, MN93, MN193, MR193, SL261, and SR193. Covers features, specifications, operation, maintenance, and warranty.