iCAP 6000 Series Pre-Installation Manual

Version 4.2

Part Number: 8499 400 30001

Published: 2007 Thermo Fisher Scientific. All rights reserved.

Registration No. 441506

Thermo Fisher Scientific
SOLAAR House
19 Mercers Row,
Cambridge CB5 8BZ,
United Kingdom.
Telephone +44 (0) 1223 347400,
Fax +44 (0) 1223 347402

Website: http://www.thermofisher.com

Chapter 1: Introduction

This manual is designed to help ensure that your iCAP 6000 Series ICP-OES Spectrometer will be installed efficiently and will be able to meet your requirements quickly and completely. It will minimise installation time if the services and facilities detailed here are available before the equipment is delivered.

Properly installed and maintained, your system will provide you with years of reliable service. Please inform your local Thermo Fisher organisation what aspects of support are most important to you and a tailored agreement can be provided to prolong the life of your instrument.

This manual describes the environment and the resources required at the installation site of the iCAP 6000 Series ICP-OES Spectrometer and associated equipment. Safety requirements for the installation are also detailed.

WARNING: The installation of all services must comply with the appropriate rules and regulations required by the local authorities responsible for those services in the workplace, at the installation site. An installation engineer is not responsible for the fitting, or compliance of the facilities, or services.

The choice of an operating site for the instrument will be influenced by local considerations, for example, ease of access and availability of electrical power. Logical planning of the installation can save both time and money. The objective of this manual is to provide information that will enable the best site to be chosen and to highlight the essential requirements.

If further information or advice is required, contact your local Thermo Fisher organisation. A list of Thermo Fisher organisations and agents can be found on http://www.thermofisher.com.

Installation of the system will include some familiarisation training; however, comprehensive method development and software training will require additional training that should be ordered separately.

Chapter 2: System Accessories

Required:

Optional:

Please request site requirement guides for any purchased accessories.

Chapter 3: Transportation

The spectrometer is supplied with all compatible accessories that have been purchased to meet the customer's system requirements. Items may be shipped in separate packages.

WARNING: All items should be transported to the installation site on pallets, in their original packaging and the right way up. Under no circumstances may any package be moved without its pallet unless under the direct control of an authorised Thermo Fisher service engineer.

Proposed routes from the customer receiving area to the installation site must be checked for suitability. The largest pallet size should be used to calculate clearances.

WARNING: The boxed Spectrometer should be moved using a mechanical lift. Five people should be used if it has to be moved by hand taking appropriate precautions. Care should be taken in doorways, corridors or when lowering to avoid trapping fingers.

Package Size

The instrument will be supplied inside especially designed packaging. The dimensions are:

Table 3-1. Packaged Dimensions and Weight
Width Depth Height Weight
980 mm (38.6 in) 900 mm (35.4 in) 920 mm (36.2 in) 95 Kg (210 Ib)

Transportation to some destinations may necessitate additional packaging, for example a packing crate. Dimensions and exact specifications may vary.

Short Term Storage

To maintain the instrument in serviceable condition and to comply with the conditions of warranty, ensure that storage of each item is maintained within the stated parameters detailed for the Working Area Requirements below. Make sure that the packages are stored the correct way up.

Note: Do not open the Packages without the permission of the Thermo Fisher engineer.

When the packages are opened prior to installation, the contents will be checked against the packing list(s).

Transit Damage

Before accepting delivery of any equipment, the packages should be inspected for signs of obvious damage. The nature of the damage should be noted on the delivery notice and signed by the carrier's representative.

Within the time stated in the Thermo Fisher terms and conditions, an inspection should be made for concealed damage. The local Thermo Fisher organisation should be advised of any damage in writing and, on receipt of specific instructions, the customer should return the equipment complete and in the original packing material.

Warranty

Note that the instrument is warranted against defects in material and workmanship for a period of 12 months from installation, or 15 months from shipment, whichever comes first. This warranty does not cover damage sustained as a result of improper storage by the customer prior to installation, or resale to a third party.

Note: Warranty is not transferable to a third party without Thermo Fisher approval.

Chapter 4: Working Area Requirements

The choice of site will be influenced by the dimensions and weights of the spectrometer and accessories. Other factors are the environment and the availability of electricity, water and gas supplies, as well as the need for a suitable ventilation system to dispose of the exhaust gases. All of these factors are covered in the following sections.

Pre-installation visits can be made on request, but may be chargeable.

Introduction

Prior to installation, make sure that the proposed area is compatible with the conditions specified. The laboratory must offer dry, even temperature and dust-free conditions, with no possibility of condensation forming. Sample preparation activities and corrosive materials should be located in a separate room to avoid problems due to corrosive fumes.

A comprehensive ‘Risk Assessment’ should be carried out that is specific to the handling of solvents, samples, and sample preparation.

Particular consideration should be taken to avoid direct sunlight, proximity to heat sources, draughts, and vibration. Do not locate the system where sudden changes in temperature can occur, for example, near a door or window. Care should be taken with the location of items such as air conditioning vents and heating vents.

Location

The instruments are designed for use on a normal laboratory bench. Ideally, the instrument should be placed on a moveable bench with 0.5 meters of access behind the instrument. The mounting surface must be level, and the instruments must not be placed on any type of cushioning as this could block ventilation.

The mounting arrangements should be capable of supporting the weight of the spectrometer and its accessories. Make sure that the working surface is sufficiently rigid to prevent vibration, as this may affect the optical alignment of the spectrometer and accessories.

To avoid the possibility of liquid ingress into the top of the spectrometer, the location should ensure it is not possible to store sample or other liquids directly above the instrumentation.

Organic or volatile solvents should not be stored, even for a short time, near the instrument.

Do not position the equipment so that it is difficult to operate the extraction, electrical supply, cooling water, purge gas, and plasma gas controls.

Environmental Requirements

The atmospheric temperature requirement is 15 to 35°C (60 to 95°F); the temperature should not change by more than 2°C per hour.

Atmospheric humidity should be 20 to 80% m/v for an ambient temperature between 15 and 30°C, and 20 to 60% m/v for an ambient temperature between 30 and 35°C. Atmospheric conditions must be non-condensing. The instrument room should be at a positive pressure with respect to rooms with a corrosive atmosphere.

Warning: Argon is an asphyxiant; if the extraction is turned off, sufficient air must be supplied to prevent the concentration of Argon reaching a harmful level.

Electrical

The spectrometer will require an electrical supply at 200 to 240 VAC, 4 KVA, 50/60 Hz. The spectrometer will be supplied with a 2.5m mains cord.

For customers in the USA and Canada, the mains cord supplied will terminate in a NEMA L6-20P Twist&Lock plug. (Coding is black (X – AC high), white (Y – AC low), green (G – earth/ground).

Figure 4-1. NEMA L6-20P plug for USA/Canada. This plug features three pins: two for AC power (typically black for high voltage, white for low voltage) and one for ground (green).

For customers in the rest of the world, the mains lead is shipped with no terminating plug, but with wires ready for a suitable 20 A plug. Coding is blue (neutral), brown (live), green/yellow (earth/ground).

Connection to the laboratory supply should be with an appropriately rated plug conforming to local electrical guidelines and local requirements. It must be possible to electrically isolate the instrumentation.

Additional standard mains sockets will be required for the Data Station PC, and for an autosampler, chiller, printer, or any other additional accessory required. Each electrical outlet must have an effective earth/ground connection. This protection must not be negated by the use of an extension cable without a protective earth conductor.

Spectrometer Size

Table 4-2. Spectrometer Dimensions and Weight
Width Depth Height Weight
832 mm (32.7 in) 744 mm (29.3 in) 589 mm (23.2 in) 85.5 Kg (188 lbs)

Figure 4-2. iCAP 6000 Series Front View. Dimensions provided are 832 mm (width) x 588.2 mm (height).

Figure 4-3. iCAP 6000 Series Top View. Dimensions provided are 831.2 mm (width) x 743.1 mm (depth).

Gas Requirements

The spectrometer will require argon at 6 bar (87psi) (minimum quality of 99.998% pure with less than 10ppm water and less than 10ppm oxygen - used for operation and optical path purge). The maximum flow requirement will be 25 l/min during installation.

For purging the optical path, nitrogen at 5.5 bar (80psi) (minimum quality of 99.995% pure with less than 10ppm water and less than 10ppm oxygen) can be used, instead of argon. The maximum flow requirement will be 15 l/min during installation. For optimum performance, a quality of 99.998% pure nitrogen is recommended.

If an additional gas accessory is required (optional), clean oil-free air is required at a pressure of 2.5 bar (35psi) and a flow of approx 100ml.min⁻¹.

The instrument will be supplied with gas tubing terminating in 6mm outside diameter gas connections (contamination-free, refrigeration-grade copper for the purge and plastic for the plasma gas).

Connection to the laboratory gas supply should be within 1 metre of the instrument. A regulator should be close to the connection.

Care should be taken to ensure that laboratory gas lines do not contaminate the gases used for the iCAP.

If liquid argon is not used, it is recommended that banks of cylinders be used with switch-over valves so that the gas supplies can be used continuously.

Warning: Nitrogen and argon gas may cause asphyxiation at high concentrations; appropriate care should be taken.

PLEASE NOTE: GAS REQUIREMENTS ARE NOT THE SAME AS FOR PREVIOUS THERMO FISHER ICP SPECTROMETERS.

Water Temperature Control

A recirculating chiller should be used to remove waste heat from the spectrometer. The water temperature should be set to 5°C below the ambient temperature.

The water used in the chiller must be distilled water containing 5% by volume of the correct inhibitor (included with maintenance kit which can be ordered from your local Thermo Fisher organisation).

Temperature variation must be less than 0.2°C per hour.

The flow requirement of the spectrometer is 5 l/min and a cooling capacity in excess of 900 W.

A suitable chiller (ThermoFlex 900) can be supplied from your local Thermo Fisher organisation.

Requirements for ThermoFlex 900:

The spectrometer will be supplied with plastic connecting tubing that has a 12mm outside diameter.

Connection to the water chiller should be within 3 meters of the instrument, or insulated tubing must be used.

Waste Storage

The analysis of a sample, by ICP-OES, usually involves production of a fine mist from a liquid sample. Waste will be produced that could be corrosive and toxic, or an organic solvent. An appropriate container is required that is solvent-proof, shatterproof, and vented away from the instrument. Ensure the waste container does not constitute a spill or trip hazard. It may be necessary to neutralise waste to prevent any toxin build-up. Appropriate facilities should be provided for the disposal of any waste, which should be disposed of following local procedures and regulations.

Spectrometer Fume Extraction

This instrumentation is designed for operation in clean air conditions. The laboratory must be free of all contaminants that could have a degrading effect on the instrument components. Dust, acid, and organic vapours must be excluded from the work area. The warranty will be void if the equipment is operated in substandard conditions.

WARNING: The spectrometer must never be operated without an effective fume extraction system attached to the torch compartment chimney.

Hot fumes, which may be corrosive and toxic, are discharged from the instrument chimney during operation. To ensure a safe working environment and safe removal of waste combustion products, an effective extraction system must be installed, this should include appropriate filtering of hazardous toxic fumes.

The extraction must be capable of exceeding the minimum required velocity, measured in the centre of the extraction tube at the spectrometer end, while disconnected from the instrument, as shown in Figure 4-4 below. The extraction tube has an inlet diameter of 125 mm (5 in).

An iCAP 6000 Duo instrument will require an extraction of 10 m/s (33 ft/s; 22.4 mph).

An iCAP 6000 Radial instrument will require an extraction of 5 m/s (16 ft/s; 11.2 mph).

The extraction flow must be adjustable to 50% of the maximum to allow optimization during installation. This can be achieved with a butterfly valve, or an adjustable gate as shown in Figure 4-4 below. The extraction system must not be affected by external weather conditions, or other uses the system may be used for.

Figure 4-4. iCAP 6000 Extraction System. This diagram illustrates the fume extraction setup, showing the spectrometer connected via an extraction hose to an exhaust system, with a note to measure flow rate at the instrument end of the extraction system.

PLEASE NOTE: EXTRACTION REQUIREMENTS ARE NOT THE SAME AS FOR PREVIOUS THERMO FISHER ICP SPECTROMETERS.

Sample Fume Extraction

Additional, separate, extraction should be considered if significant numbers of volatile, or acidic, samples are left in the proximity of the instrumentation, even those in an autosampler.

Chapter 5: Communications Interface

On the Spectrometer

An Ethernet connection is fitted for Data Station communication. An optional accessory control PCB may be fitted for additional communication requirements for some accessories. The use of each connection is specified in the user manuals for each accessory.

On the Data Station

The Data Station requirements will normally be met by a PC meeting the minimum specification:

A suitable PC is available from your local Thermo Fisher organisation.

The installation engineer is not responsible for network and customer-specific set-up of the PC.

Chapter 6: Pre-Installation Checklist

This checklist is to be filled in by the customer to confirm that all relevant factors concerned with the installation have been considered and dealt with. On completion of the list, a copy should be forwarded to the local Thermo Fisher organisation with a request for installation to be carried out. Failure to send the checklist may result in delays with the instrument installation; any delays may incur extra charge.

Item Required Completed
Location
Table dimensions for the spectrometer 900 x 600 mm
Table dimension for the auto sampler (right side of the machine) 600 x 550 mm
Extraction
Tube diameter for extraction 125 mm
Extraction capability 10 m/s (Duo)
5 m/s (Radial)
Adjustable valve in the extraction tube Yes
Electrical supply
Plug 200-240 VAC, 32A single phase 4000W, 50Hz Yes
Standard mains plug 200-240 VAC for each accessory Yes
Argon Gas
Regulator valve in the wall 0 -10 bars (0 – 100 PSI) Yes
Gas union dimension at the exit of the regulator valve 6 mm (1/4 inch adapter supplied with instrument)
Maximum distance between the instrument and the regulator valve 2 m
Purity of argon (Gas or Liquid) 99.995% O₂<10 ppm H₂O<10 ppm
Pressure required at the input of spectrometer 6 bars (87 Psi)
Optic nitrogen purge (Optional)
Regulator valve in wall 0 – 10 bars (0 – 100 Psi) Yes
Gas union dimension at the exit of the regulator valve 6 mm (1/4 inch adapter supplied with instrument)
Maximum distance between the instrument and the regulator valve 2 m
Purity of nitrogen (Gas or Liquid) 99.995% O₂<10 ppm H₂O<10 ppm
Pressure required at the input of spectrometer 6 bar (87 Psi)
Waste
Waste container, a minimum of 5 l for liquids waste Yes
Auto Sampler rinse and waste
Minimum 1l container for the rinse supply station and waste 2 needed
Chiller (if not bought with the instrument)
Maximum distance from the spectrometer 3 m
Pipe supply O.D. dimension 12 mm
Cooling capability 900 W
Flow rate 5 l/min
Chiller inhibitor Yes
PC and printer (if not bought with the machine)
512 Mb RAM, 40 Gb HDD, 2 Ethernet ports, 1 serial port, USB ports, monitor, ink jet printer Yes
Windows XP Professional (Service pack 2) Yes

Caution: The instrument installation will not start before Thermo Fisher receives this document back completed. In the case of false information causing the installation to be stopped, a travel cost may be charged.

I certify the exactness of the information supplied in this document.

Name: _________________________ Position: _________________________

Signature: _________________________ Company stamp: _________________________

Date: _________________________

Chapter 7: Special Agreements

This section should detail any special agreements, which have been arranged with your local Thermo Fisher organisation.

PDF preview unavailable. Download the PDF instead.

iCAP 6000 Spectrometers Pre-Installation Manual v4 2 GPL Ghostscript 8.15

Related Documents

Preview iCAP PRO Series ICP-OES Installation Procedure
This document provides guidelines for installing Thermo Scientific iCAP PRO Series ICP-OES systems, detailing hardware and software setup, performance verification, and customer familiarization.
Preview Thermo Fisher Scientific iCAP Q Operating Manual
Comprehensive operating manual for the Thermo Fisher Scientific iCAP Q mass spectrometer, covering system description, setup, operation, troubleshooting, and maintenance.
Preview Nicolet Summit LITE FTIR Spectrometer Site and Safety Guide
This guide provides essential site preparation, safety precautions, and handling instructions for the Nicolet Summit LITE FTIR Spectrometer, ensuring safe operation and optimal performance.
Preview Thermo Fisher Nicolet iS5 User Guide - Setup and Operation
User guide for the Thermo Fisher Nicolet iS5 instrument, providing detailed instructions for setup and operation.
Preview Thermo Scientific Nicolet iS20 Spectrometer: Getting Started Guide
A comprehensive guide to getting started with the Thermo Scientific Nicolet iS20 FT-IR spectrometer, covering features, operation, accessories, software, and maintenance.
Preview Thermo Scientific Environmental Contaminant Analysis Compendium: Methods from Sample to Compliance
Explore Thermo Fisher Scientific's comprehensive solutions for environmental contaminant analysis. This compendium details methods for organic, inorganic, trace elemental, and emerging contaminants in water, air, and soil using advanced chromatography and mass spectrometry techniques.
Preview Thermo Compound Discoverer User Guide - Software Version 3.0
Comprehensive user guide for Thermo Fisher Scientific's Compound Discoverer software (v3.0). Learn data processing, compound identification, workflow creation, and results analysis for mass spectrometry research.
Preview ARL 9800 Series User Manual - Thermo Fisher Scientific
This user manual provides comprehensive instructions and information for operating the ARL 9800 Series, a sophisticated X-ray fluorescence spectrometer system from Thermo Fisher Scientific. It covers instrument description, sample preparation, analysis principles, instrument calibration, routine analysis, and troubleshooting.