Tektronix A622 100 Amp AC/DC Current Probe

Instructions and Specifications

Contacting Tektronix

Tektronix, Inc.
14150 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA

For product information, sales, service, and technical support:

Important Safety Information

This manual contains information and warnings that must be followed by the user for safe operation and to keep the product in a safe condition. To safely perform service on this product, refer to the Service safety summary that follows the General safety summary.

General Safety Summary

Use the product only as specified. Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. Carefully read all instructions. Retain these instructions for future reference.

To Avoid Fire or Personal Injury

Do Not Operate Without Covers

Do not operate this product with covers or panels removed, or with the case open. Hazardous voltage exposure is possible.

Avoid Exposed Circuitry

Do not touch exposed connections and components when power is present.

Replace Batteries Properly

Replace batteries only with the specified type and rating.

Do Not Operate in Wet/Damp Conditions

Be aware that condensation may occur if a unit is moved from a cold to a warm environment.

Do Not Operate in an Explosive Atmosphere

Keep product surfaces clean and dry. Remove input signals before cleaning.

Probes and Test Leads

Connect and Disconnect Properly

Connect the probe output to the measurement product before connecting the probe to the circuit under test. Connect the probe reference lead to the circuit under test before connecting the probe input. Disconnect the probe input and reference lead from the circuit under test before disconnecting the probe from the measurement product.

Do not connect a current probe to any wire that carries voltages or frequencies above the current probe voltage rating.

Inspect the Probe and Accessories

Before each use, inspect probe and accessories for damage (cuts, tears, or defects). Do not use if damaged.

Service Safety Summary

This section contains additional information required to safely perform service. Only qualified personnel should perform service procedures. Read this and the General safety summary before performing any service.

To Avoid Electric Shock

Disconnect Power

To avoid electric shock, switch off product power and disconnect the power cord from the mains power before removing covers, panels, or opening the case for servicing.

Use Care When Servicing With Power On

Dangerous voltages or currents may exist. Remove battery before removing protective panels, soldering, or replacing components.

Terms in this Manual

Terms on the Product

Symbols on the Product

When a symbol is marked on the product, consult the manual to find out the nature of potential hazards and actions to avoid them. This symbol may also refer the user to ratings in the manual.

The following symbols may appear on the product:

Compliance Information

This section lists the safety and environmental standards with which the instrument complies. This product is intended for use by professionals and trained personnel only; it is not designed for use in households or by children.

Questions about compliance information may be directed to:

Tektronix, Inc.
PO Box 500, MS 19-045
Beaverton, OR 97077, USA
tek.com

Safety Compliance

EU Declaration of Conformity – Low Voltage

Compliance demonstrated to Low Voltage Directive 2014/35/EU:

U.S. Nationally Recognized Testing Laboratory Listing

Canadian Certification

Additional Compliances

Equipment Type

Test and measuring equipment.

Safety Class

Class 2

Pollution Degree Description

A measure of contaminants that could occur in the environment around and within a product. Products should be used only in the environment for which they are rated.

Pollution Degree Rating

Pollution Degree 2 (as defined in IEC 61010-1). Rated for indoor, dry location use only.

IP Rating

IP20 (as defined in IEC 60529).

Measurement and Overvoltage Category Descriptions

Measurement terminals may be rated for measuring mains voltages from one or more of the following categories:

Note: Only mains power supply circuits have an overvoltage category rating. Only measurement circuits have a measurement category rating.

Environmental Compliance

This section provides information about the environmental impact of the product.

Product End-of-Life Handling

Observe the following guidelines when recycling an instrument or component:

Equipment Recycling

Production required extraction and use of natural resources. Equipment may contain harmful substances. Recycle this product in an appropriate system to ensure materials are reused or recycled.

This symbol [WEEE/Battery symbol] indicates compliance with EU requirements (Directives 2012/19/EU and 2006/66/EC). For recycling options, check the Tektronix Web site (www.tek.com/productrecycling).

Getting Started

Overview

The A622 current probe enables a general purpose oscilloscope to display AC and DC current signals up to 100 amps Peak (70 A RMS). It can also make AC and DC measurements with a multimeter using the recommended BNC-to-banana plug adapter.

Figure 1: A622 AC/DC Current Probe

The probe consists of a body containing the electronics, jaws to clamp around a conductor, a handle, an OFF/Range switch, a Zero adjustment control, a current flow symbol indicating polarity, and a battery compartment.

Controls and Indicators

Control/Indicator Description
Current flow symbol The arrow shows the polarity convention for measuring current flowing from positive to negative.
Zero adjustment Rotate to adjust the probe output to zero when no current is present. Can also offset a DC signal component. Zeroing is not needed for AC measurements unless the instrument cannot isolate a DC component.
OFF/Range switch Slide from OFF to either the 10 mV/A or 100 mV/A range. The probe powers on when a range is selected, and the green battery indicator lights. If the indicator does not light, see Battery Notes.
Battery indicator The green LED lights when the probe is turned on. For more information, see Battery Notes.
Overload indicator The red LED lights if the measured signal exceeds the selected range capacity. Switch to 10 mV/A if possible, or remove the probe from the circuit.

Operating Basics

Before using the probe, the batteries must be installed (see Battery Installation).

Warning: Personal injury or damage to the probe can result if you clamp the probe onto circuits with voltages greater than 600 VAC. Do not clamp onto circuits with higher voltages. Always connect the probe to the instrument before clamping onto the circuit under test.

  1. Connect the probe BNC connector to the oscilloscope input. Set the oscilloscope channel vertical coupling to DC volts and the vertical deflection to 0.1 V/div.
  2. Move the OFF/Range switch to the 10 mV/A or 100 mV/A position to power on the probe. The A622 has a green LED power/battery indicator. If the LED does not light, replace the battery.
  3. Use the ZERO adjustment to zero or offset the probe output.
  4. Connect the probe to the circuit by opening the jaws and clamping around the conductor.
  5. Note: Clamping around both the "hot" and neutral wires may yield a zero reading.

    (Remember to unclamp the probe from the conductor before disconnecting it from your meter or instrument.)

Figure 2: Connecting the A622 Current Probe

The figure shows the probe jaws clamping around a conductor, with connections to the instrument.

  1. Adjust the probe and channel as necessary for a clear signal view. Set the channel to DC volts to see both AC and DC currents; set to AC to see AC current only.

The current drawn by some devices differs significantly. While RMS current may be low, momentary peaks can be high. The following figures show the difference between line current drawn by a resistive load and a motor controller.

Figure 3: Typical Current Waveforms

This figure displays two waveforms: one for a resistive load (sinusoidal) and one for a motor controller (pulsed). The Y-axis represents current, and the X-axis represents time/frequency.

Using with a Multimeter

If using the A622 with a multimeter, connect the probe with the BNC-to-banana adapter. Connect the black lead to the meter COM, and the red lead to the VΩ input.

Increasing Measurement Sensitivity

To increase sensitivity, loop additional turns of the wire under test through the jaws. The sensitivity is multiplied by the number of loops. For example: 10 mV/A × 4 turns = 40 mV/A.

Figure 4: Increasing the Sensitivity

The figure illustrates a wire being passed through the probe jaws multiple times to increase measurement sensitivity.

CAUTION: To prevent damage to the probes, do not force the jaws closed. If you cannot close the jaws around the conductor(s), either reduce the number of conductors or measure on a smaller conductor.

Maintenance

Cleaning

To clean the probe body, use a soft cloth dampened in mild detergent and water. To clean the core, open the jaw and clean exposed core surfaces with a cotton swab dampened with isopropyl alcohol. Lubricate jaw mating surfaces with a light oil.

Do not clean with solvents or abrasives. Do not immerse the probe.

Battery Notes

The A622 uses a single 9 V battery. As the battery drains, significant gain errors can occur. The green LED remains lit until a low battery voltage of 6.5 V is reached. If gain errors are detected, replace the battery with a fresh one.

Battery Installation

  1. Remove the probe from the circuit.
  2. Open the battery compartment by loosening the captive screw and sliding the cover off.
  3. While observing polarity, attach the battery to the battery connector.
  4. Replace the cover and lightly tighten the screw.

Figure 5: A622 Battery Compartment

The figure shows the battery compartment with the battery and screw.

Specifications

These characteristics apply to an A622 AC/DC Current Probe installed on a Tektronix TDS 220 oscilloscope, after 20 minutes warm-up and within specified environmental limits.

Table 2: Electrical Characteristics

Characteristic Description
Current Ranges 10/100 mV/A
DC Accuracy, typical ±3% ± 50 mA at 100 mV/A (50 mA to 10 A peak range)
±4% ± 50 mA at 10 mV/A (500 mA to 40 A peak range)
Gain versus frequency, typical Figure 6 on page 13
Maximum Working Current Table 3 on page 12
Maximum Working Voltage 1 Table 3 on page 12
30 VRMS, 42 VPK, 60 VDC. For voltages above these limits, use insulated conductors only.
Maximum Float Voltage Table 3 on page 12
Frequency Range DC to 100 kHz (-3 dB)
Battery Type and Life, typical 9V NEDA 1604A, IEC 6LR61 40 hours minimum (1 each)
DC signal linearity, typical Figure 8 on page 14
Phase shift, typical Figure 9 on page 14

Table 3: Voltage and Current Ratings

Rating Maximum working current (A) Maximum working voltage (V) Maximum floating
Range 10 mV/A Range 100 mV/A
DC 100 2 10 600 600
DC + peak AC 100 2 10 600 600
AC peak 100 10 600 600
AC peak-peak 200 20 1200 600
RMS CAT III 70.7 7.07 600 600
RMS CAT II 70.7 7.07 600 600
RMS CAT I 70.7 7.07 600 600

1 An insulated conductor is any conductor surrounded by insulating material capable of isolating the voltage. Lacquer coatings on transformer windings may not provide sufficient insulation and can be easily nicked or damaged.

2 See frequency derating information in Figure 7 on page 13.

Table 4: Physical Characteristics

Characteristic Description
Dimensions 231 mm x 36 mm x 67 mm (9.09 x 1.42 x 2.64 inches)
Maximum Conductor Size 11.8 mm (0.46 inches)
Cable Length 2 m (6.6 feet)
Weight 330 g (12 oz) with battery

Table 5: Environmental Characteristics

Characteristic Description
Temperature Operating: 0 °C to +50 °C (+32 °F to +122°F)
Storage: -20 °C to +80°C (-4 °F to +176°F)
Humidity 0 °C to 40 °C, 95% humidity
40 °C to 50 °C, 45% humidity
Pollution Degree 2

Graphs

Figure 6: Gain versus frequency at 1 A peak, typical

This graph shows the gain in dB versus frequency (Hz) for the 10 mV/A and 100 mV/A ranges, illustrating how gain changes with frequency.

Figure 7: Maximum current versus frequency

This graph shows maximum current (peak and RMS) versus frequency (Hz), indicating the probe's current handling capability at different frequencies.

Figure 8: DC signal linearity in the 10 mV/A range, typical

This graph shows the percentage of uncertainty versus amperes (A) for DC signal linearity in the 10 mV/A range, comparing typical performance to specifications.

Figure 9: Phase versus frequency at 1 A peak, typical

This graph shows phase shift in degrees versus frequency (Hz) for the 10 mV/A and 100 mV/A ranges, illustrating phase response.

Preparation for Shipment

If the original packaging is unfit for use or not available, use the following packaging guidelines:

  1. Use a sturdy shipping carton with inside dimensions at least one inch greater than the probe dimensions.
  2. Put the probe into a plastic bag or wrap to protect it from dampness.
  3. Place the probe into the box and stabilize it with light packing material.
  4. Seal the carton with shipping tape.

Replaceable Parts

The A622 AC/DC Current Probe is shipped with the following items:

Recommended accessory for use with DMMs:

The A622 does not have any user-repairable assemblies. If you should have trouble with your probe, contact your local Tektronix Service Center or representative for help.

Models: A622, 100 Amp AC and DC Current Probe

File Info : application/pdf, 15 Pages, 327.30KB

PDF preview unavailable. Download the PDF instead.

TEKTRONIX A622 (01)

References

AH XSL Formatter V7.0 MR4 for Windows (x64) : 7.0.5.46841 (2020-08-31T10:02 09) Antenna House PDF Output Library 7.0.1604

Related Documents

Preview Tektronix TCP0030A AC/DC Current Probe User Manual
User manual for the Tektronix TCP0030A, a 120 MHz, 30 Amp AC/DC Current Probe. This document provides safety information, operating instructions, specifications, and maintenance details for the probe.
Preview Tektronix P5102 100 MHz High-Voltage Probe Instruction Manual
Comprehensive instruction manual for the Tektronix P5102 100 MHz High-Voltage Probe, covering safety, operation, adjustment, verification, maintenance, and specifications. Includes detailed guidance for high-voltage measurements.
Preview Tektronix P6105A 100 MHz 10X Passive Probe Instruction Manual
Comprehensive instruction manual for the Tektronix P6105A 100 MHz 10X passive probe, detailing its features, specifications, installation, operation, safety, and maintenance.
Preview Tektronix TPP0250 and TPP0500B 10X Passive Probes: Instructions and Specifications
User instructions and technical specifications for Tektronix TPP0250 (250 MHz) and TPP0500B (500 MHz) 10X passive oscilloscope probes, covering setup, operation, accessories, and safety.
Preview Tektronix P6006 Probe Instruction Manual
Instruction manual for the Tektronix P6006 passive probe, detailing its specifications, characteristics, compensation procedures, derating curves, and replaceable parts.
Preview P6339A 500 MHz Buffered Passive Probe Instruction Manual
Comprehensive instruction manual for the Tektronix P6339A 500 MHz Buffered Passive Probe, detailing its features, operation, specifications, safety precautions, adjustment procedures, and maintenance.
Preview Tektronix TPH1000 Probe Holder: Instructions and Features
Comprehensive instructions and overview of the Tektronix TPH1000 Probe Holder, detailing its features, probe grip compatibility, dimensions, usage guidelines, and safety precautions for oscilloscope probe accessories.
Preview Tektronix Type 317 Oscilloscope Instruction Manual
This comprehensive instruction manual provides detailed characteristics, operating instructions, and maintenance procedures for the Tektronix Type 317 Oscilloscope. Learn about its vertical and horizontal deflection systems, triggering modes, and mechanical specifications for effective use and upkeep.