Ossila Analytical Balance User Manual

Manual Version: 1.0.A
Product code: LT-BALANCE-ANALYTICAL
Product Version: 1.0
Software Version: 1.0

1. Overview

The Ossila analytical balance allows users to measure out small masses accurately and repeatedly. With our analytical balances being able to measure down to as low as 10 mg the weighing out of the smallest quantities of materials and the counting of small components is possible with this system.

The analytical balance consists of a high precision sensing element with a weighing pan available for placing samples on to. The pan is housed within a shielded draft guarded glass chamber which can be accessed from the top of sides by the user. In addition, the sensitive sensing elements are protected by a draft shield around the pan. Allowing for measurements to be protected from disturbances cause by the external environment.

Our analytical balance come in a range of maximum capacities and have the option for either external calibration or internal calibration depending upon the users needs. The system can also be operated in several measurement modes, including, weighing, density determination, and parts counting. In addition, the analytical balance has an RS232 connector at the back for external communication allowing measurements to be logged on an external computer or printed out.

Users can undertake a measurement by simply zeroing the system via the tare button, samples are then placed onto the pan, the draft shield doors are closed, and the measurement is allowed to stabilise. Users can then adjust the amount of parts or materials they have on the pan until it reaches the desired amount.

Illustration of the Ossila analytical balance, showing its weighing chamber, pan, and digital display.

2. EU Declaration of Conformity (DoC)

Company Name: Ossila BV
Postal Address: Biopartner 3 building, Galileiweg 8
Postcode: 2333 BD Leiden
Country: The Netherlands
Telephone number: +31 (0)71 3322992
Email Address: info@ossila.com

declare that the DoC is issued under our sole responsibility and belongs to the following product:

Product: Analytical Balance (LT-BALANCE-ANALYTICAL-100/120/200/220-I/E)
Serial number: LT-BALANCE-ANALYTICAL-120-E-XXXX

Objects of declaration:

The object of declaration described above is in conformity with the relevant Union harmonisation legislation:

Signed:
Name: Dr James Kingsley
Place: Leiden
Date: 01/07/2024

3. Safety

3.1 Warning

3.2 Use of Equipment

This Analytical Balance is designed to be used as instructed, and in the following environmental conditions:

The analytical balance is supplied with a power adapter and a power cord for the country of purchase (in accordance with European Commission regulations and British Standards). Use of any other electrical power cables or adaptors is not recommended.

3.3 Hazard Icons

The following symbols can be found at various points throughout this manual. Note and read each warning before attempting any associated operations associated with it:

SymbolAssociated Hazard
General warning or caution, which accompanying text will explain
Electrical shock
🌞Inhalation

3.4 General Hazards

Before installing or operating the Analytical balance, there are several health and safety precautions which must be followed and executed to ensure safe installation and operation.

WARNING: Improper handling when operating or servicing this equipment can result in serious injury or death. Read this manual before operating or servicing this equipment.

  1. DANGER: DO NOT use the Analytical balance in the presence of an explosive atmosphere.
  2. WARNING: Emergency Power Disconnect options: Use the power cord as a disconnect method. Ensure that the power outlet for this cord is easily accessible by the user.
  3. CAUTION: Ensure power supply used with the system adheres to local mains voltage before connecting the system.
  4. CAUTION: Make sure the mains cable is free from damage before use of the system.

3.5 Servicing

If servicing is required, please return the unit to Ossila Ltd. The warranty will be invalidated if:

3.6 Health and Safety - Installation

3.7 Health and Safety - Operation

3.8 Health and Safety - Servicing

Servicing should only be performed by an Ossila engineer. Any modification or alteration may damage the equipment, cause injury, or death. It will also avoid your equipment's warranty.

4. Unpacking

4.1 Packing List

The items included with the Analytical Balance are:

4.2 Damage Inspection

Examine the components for any evidence of shipping damage. If damage has occurred, please contact Ossila directly for further action.

5. Specifications

The Analytical Balance specifications are shown in Table 5.1.

Model NumberANALYTICAL-120-IANALYTICAL-100-EANALYTICAL-220-IANALYTICAL-200-E
Capacity120g100g220g200g
Calibration MethodInternalExternalInternalExternal
Readability0.1 mg
Repeatability±0.2 mg
Linearity±0.2 mg
Stabilization Time8s
Warmup Time30-60 minutes
Weighing Unitsg, mg, oz, ct
Piece Counting Reference Quantities5, 10, 20, 50, 100
Piece Counting Minimum Weight10 mg
Weighing Pan SizeØ 80 mm
Net Weight5 kg
Ambient Conditions+10°C to +30°C
20 to 85% Relative Humidity (non-condensing)
Power SupplyDC9V 2.2A/ AC 110-220V 50Hz
Communication InterfaceRS232C (DB9 pin Terminal)
Pin 2 - Receive Data
Pin 3 - Transmit Data
Pin 5 – Signal Ground
Baud Rate1200 / 2400 / 4800 / 9600
Parity8 bits, no parity / 1 stop bit / 1 start bit

6. Installation

6.1 Finding an Installation Location

Before unpacking the instrument, you should identify a suitable location to install the unit. You should look out for the following when finding a suitable location for the analytical balance:

  1. A firm and level surface should be used for placing the balance onto
  2. Avoid places where there may be extremes of heat or large variations in temperature over time. Locating it away from radiators, AC units, or windows is ideal
  3. Avoid locations where large drafts are present, keeping the unit away from doors, windows, fans, or extraction units.
  4. Avoid locations where there may be excessive vibrations, locate the table away from regions where people are walking or from equipment that can produce loud noise and/or vibrations.
  5. Avoid build up of static discharge on or around the equipment as this could potentially damage the sensing element.
  6. Ensure the environment it is placed within is not damp, does not have high humidity, and has few sources of vapours and dust.

6.2 Checking Parts for Damage

Once a suitable location is identified you should open the packaging that the unit comes in and carefully remove each part. You should verify that each part is free of damage and that all items are present. The following items should come with the analytical balance:

  1. Balance
  2. Display Unit
  3. Weighing Pan
  4. Weighing Pan Support
  5. Draft Ring
  6. Bottom Sheet
  7. Power Plug
  8. User Manual

6.3 Removing Transport Screw

Before assembling the unit you should remove the transport securing screw, this is located on the underside of the balance (see Figure 6.1).

Illustration showing the transportation screw located on the underside of the unit.

6.4 Installing Parts

After removing the transportation screw, place the unit in the desired location and install the following components:

  1. Install the bottom sheet on the unit.

    Image showing the bottom sheet installed on the unit.

  2. Place the weighing pan support into the central hole.

    Image showing the weighing pan support placed in the central hole.

  3. Place the weighing pan onto the support.

    Image showing the weighing pan placed onto the support.

  4. Install the draft shield around the pan.

    Image showing the draft shield installed around the weighing pan.

  5. Connect the display unit to the main body.

    Image showing the display unit connected to the main body of the balance.

6.5 Levelling the Balance

Once the balance has had all components installed you should then place it in the position you will be using it and then level it. Using the target spirit level at the rear right of the unit and the two adjustable feet at the rear level the system so that the bubble in the level is centred. This should be checked regularly in case the system has been moved.

6.6 Connecting the Power Supply

Before plugging in the power supply check that the plug is suitable for the mains voltage for the country of use and that the plugs output matches the voltage and current requirements for the system. Once checked plug the system in and turn on the power to the system. Press the ON button. The system should be left for one hour before being used to allow components to reach a steady state temperature. The balance will have a countdown timer start as soon as it is turned on indicating when it is ready to use.

Once ready, the user should undertake a system calibration.

7. Operation

7.1 User Interface

Diagram of the Analytical Balance screen and keypad, showing buttons for Tare, Menu, On, Cal, and Off.

The 'On' button is used to power on and off the system.

The 'Tare' button is used to zero the scales before measuring.

The 'Cal' button is used to perform a calibration of the balance.

The 'Menu' button is used to switch between weighing modes.

The 'Set' button (available on certain models) is used to confirm selections within the weighing modes menu.

The 'Print' button (available on certain models) is used to send data through the RS232C connection to a connected peripheral.

7.2 Settings

Table 7.1 shows all the available settings options for the analytical balance. To enter the settings menu to update any of the values you must first turn the unit on and allow the unit to finish the warm-up countdown. When the unit display the initial zero reading you should then press 'Set' and then 'Print' the display will then show the first menu item C1.

When in this menu you can navigate between the different options by pressing the ‘Tare' key. This will go from C1 to C8 and cycle back to C1 again.

When you have reached the desired settings option, you can change the value by pressing the ‘Print' button. This will cycle through the settings values.

To confirm the new settings value press ‘Tare' this will navigate to the next settings option.

To save the updated values press the ‘On' button, the display will show the word ‘SAVE' to confirm the saved changes press the ‘Tare' button. The unit will return to weighing mode and display the current weight value.

7.3 Calibration

Calibration should be carried out in the following situations:

Users should also consider periodically recalibrating at least once a week to ensure the system remains as accurate as possible.

7.3.1 Internal Calibration

Models with internal calibration have a set of weights within the system which can be measured via an automated process. Here reference weights are lowered onto the weighing pan internally via a motor. The system then performs the necessary calibration process and then the motor lifts the reference weights off the weighing pan. The balance automatically performs the calibration process under the following conditions:

During the automated calibration process the unit should not be moved and the draught shield should remain closed.

In addition to the automated process there is the option to undertake a one-button calibration with internally calibrated scales. To do this ensure that the calibration mode settings value is set to C1 – 0. The user should then press ‘Cal' and the display will show the calibration target weight which should be 100 g. Press ‘Cal' again and the system will undergo the calibration process. This will involve the following:

7.3.2 External Calibration

For systems without internal calibration weights or for users of internal calibration systems who have the calibration mode setting set to C1 – 1 the calibration process requires the use of an external weight. Users should have a certified weight as close to the maximum capacity of the system.

To undertake an external calibration the users should press the ‘Cal' button, the display will then show the word ‘CAL 100' users can press the ‘Tare' button to switch to the 200g calibration process if they wish to calibrate using this weight. Once the desired calibration weight is selected users should then press the 'Cal' button again.

7.4 Basic Operation

7.4.1 Turning On

After plugging the system in the 'On' button icon should appear in the top left corner of the display. To turn the system on press the ‘On’ button, the balance will start up and will enter a warmup mode with a countdown until the system is ready to use.

When the system begins to display the weight, it is ready to use.

7.4.2 Standby Mode

To enter standby mode, press the 'On' button. The display will clear and the ‘On' button icon will remain in the top left corner. When starting up from standby mode the system will not undergo a warmup cycle.

7.4.3 Zeroing the Scale

Before weighing a sample the scales should be zeroed in order to obtain the most accurate weight value to do this ensure that the draft shield is closed and the weight is stable. When the weight readout is constant press the 'Tare' button. The tareing process will take a few seconds and once done the display will read 0.0000g.

7.4.4 Sample Weighing

To weigh a sample the scales must first be zeroed using the tare function. When the scales read zero place the sample of the pan and close the draft shield. You should then wait for the weight readout to be stable, when this occurs a circle icon should appear to the left of the weight readout.

Exceeding the total maximum stated load of the scales should be avoided. If the total weight of the samples exceeds the maximum load the scales will display an ‘E'. If this occurs immediately remove the items from the balance.

7.4.5 Selecting Units

To cycle through available units, press the ‘Mode' button. The order of units is as follows.

7.5 Advanced Weighing Features

7.5.1 Parts Counting

Before performing a parts count an average weight must be determined. Before weighing you should check and set the reference parts value, this can be done via the settings menu and updating the menu item C2 Reference Quantity value (see Section 7.2 Settings for more information.

Once the reference value is selected you will need to set the weighing mode to pieces, this can be done by pressing the ‘Mode' button until the word ‘Pcs' is shown at the top of the display. Tare the scale and then place the reference quantity of parts you wish to count on the scales. Once the measured value is stable press the ‘Set' button to save the reference weight.

Items can now be placed upon the scale and the system will automatically display the number of parts based upon the measured weight and the average weight of the reference parts. You can toggle between displaying the number of parts and the total weight by pressing the 'Mode' button.

7.5.2 Percentage Determination

To enable percentage determination, press the ‘Mode' button until the percentage symbol is shown at the top of the display. Tare the scales and then place the weight which you wish to correspond to 100% of the weight onto the scales. Once the value is stabilised press the ‘Set' button and the display should now show 100%.

Remove the reference weight and place the items you wish to determine the weight percentage of. You can toggle between the displayed percentage and the weight of the sample by pressing the 'Mode' button.

8. Maintenance

8.1 Cleaning

8.2 Repair and Service

Servicing and repairs should be carried out by Ossila. If the unit is faulty, please return it to Ossila. We will promptly quote to repair any faults that occur outside the 2-year warranty period. Parts subject to normal wear and tear.

8.3 Storage Conditions

The Analytical Balance should be kept in dry conditions; away from direct sources of heat or sunlight, and in such a manner as to preserve the working life of the instrument.

9. Troubleshooting

ProblemPossible cause
Display screen is not bright
  • The balance is not switched on
  • The main power cable is not plugged in or faulty.
  • The power supply has been interrupted
The weight value is not stable
  • Draught or air movement
  • Table or floor vibrations
  • Weighing pan in contact with another object
  • Electromagnetic and/or static interference
The weight value is inaccurate
  • Unit has not been correctly zeroed
  • The calibration is no longer valid
  • The balance is no longer level
  • Large temperature variance
  • Insufficient warm up time given
  • Electromagnetic and/or static interference
Models: ANALYTICAL120-I Analytical Balance, ANALYTICAL120-I, Analytical Balance, Balance

File Info : application/pdf, 24 Pages, 953.47KB

PDF preview unavailable. Download the PDF instead.

analytical-balance

References

Microsoft Word for Microsoft 365

Related Documents

Preview Ossila Probe Station User Manual
The Ossila Probe Station User Manual (Model T2009) provides detailed instructions for installation, operation, and maintenance of this advanced scientific instrument. Learn how to perform precise electrical characterization of electronic devices with its high-resolution stage, vacuum chuck, and integrated microscope system. Includes safety guidelines and troubleshooting tips.
Preview Ossila Solar Cell I-V Test System (Automated) User Manual
Comprehensive user manual for the Ossila Solar Cell I-V Test System (Automated) (T2003). Covers installation, operation, measurement types, software settings, troubleshooting, and related products for solar cell characterization.
Preview Ossila Carboxymethylcellulose Sodium (CMC) Safety Data Sheet
Safety Data Sheet for Ossila's Carboxymethylcellulose Sodium (CMC), CAS 9004-32-4. Provides comprehensive information on product identification, hazards, composition, first aid measures, firefighting, handling, storage, exposure controls, physical and chemical properties, stability, toxicology, ecological information, disposal, transport, and regulatory details.
Preview Prussian White Powder Safety Data Sheet (SDS) - Ossila
Safety Data Sheet (SDS) for Prussian White Powder (Sodium manganese hexacyanoferrate) from Ossila, detailing identification, hazards, handling, storage, and regulatory information.
Preview Ossila Polytetrafluoroethylene (PTFE) Safety Data Sheet
Safety Data Sheet for Ossila Polytetrafluoroethylene (PTFE) providing detailed information on identification, hazards, handling, storage, exposure controls, physical and chemical properties, toxicological and ecological data, and regulatory information.
Preview Ossila Polyvinylidene Fluoride (PVDF) Safety Data Sheet
Safety Data Sheet for Ossila's Polyvinylidene Fluoride (PVDF) powder, detailing product identification, hazard classification, first aid measures, handling and storage, exposure controls, physical and chemical properties, stability, toxicological and ecological information, disposal, transport, and regulatory information.
Preview Ossila DCz-BTP Safety Data Sheet (SDS) - Chemical Hazard Information
Comprehensive Safety Data Sheet (SDS) for Ossila DCz-BTP, detailing chemical identification, hazard classification, composition, first aid measures, handling, storage, exposure controls, physical and chemical properties, toxicological and ecological information, disposal, transport, and regulatory compliance.
Preview Ossila p-bPPhenB Safety Data Sheet (SDS) - Chemical Identification and Hazards
Comprehensive Safety Data Sheet (SDS) for Ossila p-bPPhenB (CAS 2135536-38-6), detailing its identification, hazards, composition, first aid, handling, storage, and regulatory information.