Agilent 1290 Infinity Binary Pump

Features, Specifications and Ordering Details

The Agilent 1290 Infinity Binary Pump delivers the industry's widest analytical power range for ultimate chromatographic performance and flexibility. Lowest delay volumes allow you to run ultra fast gradients for narrow-bore applications. Active damping combines innovative pump engineering with tuning algorithms embedded in the firmware for significantly reduced ripples and associated UV noise. Background noise is reduced further by the Agilent Jet Weaver mixer, which is based on multi-layer technology. New ceramic piston material with superior thermal behavior provides robust and reliable performance even at ultra high pressures. An integrated high efficiency degasser and automatic purge valve offer convenient system operation.

Features

Specifications – Agilent 1290 Infinity Binary Pump

SpecificationDetails
Hydraulic systemTwo dual pistons in series, pumps with proprietary servo-controlled variable stroke design and smooth motion control.
Pump resolution step size300 pl step size
Settable flow range0.001–5 mL/min, in 0.001 mL/min increments (– executed in 300 pl/step increments).
Flow precision<=0.07 % RSD or 0.005 min SD, whatever is greater (0.2 – 5.0 mL/min); based on retention time at constant room temperature.
Flow accuracy±1% or 10 µL/min, whatever is greater; pumping degassed H₂O at 10 MPa (100 bar).
Pressure operating range1290 Infinity Binary Pump: Operating range up to 120 MPa (1200 bar) up to 2 mL/min, 80 MPa (800 bar) at 5 mL/min
1290 Infinity Binary Pump VL:
Operating range up to 105 MPa (1050 bar) up to 2 mL/min
Pressure pulsation< 1% amplitude or < 0.5 MPa (5 bar), whatever is greater; at 1 mL/min water.
Compressibility compensationAutomatic, pre-defined, based on mobile phase selection.
Recommended pH-range1.0–12.5, solvents with pH < 2.3 should not contain acid which attack stainless steel.
Gradient formationHigh pressure binary mixing.
Delay volumeJetWeaver V35: < 45 µL
JetWeaver V100: < 75 µL
Note: JetWeaver generally recommended. For applications requiring lowest delay volumes, JetWeaver can be removed out of the flow path. Delay volume is then solely determined by the volume of the connection capillary.
Composition rangeSettable range: 0 - 100%
Recommended range: 1 - 99% or 5 µL/min per channel, whatever is greater.
Composition precision< 0.15 % RSD or 0.01 min SD, whatever is greater (0.2 - 5 mL/min); based on retention time at constant room temperature.
Composition accuracy±0.35% absolute (5 - 95%, 0.2 - 5.0 mL/min, water/caffeine tracer).
Solvent selection valveDefault
Integrated degassing unitNumber of channels: 2 Internal volume per channel: 1.5 mL Materials in contact with solvent: TFE/PDD Copolymer, FEP, PEEK, PPS.
ControlAgilent ChemStation for LC (B.04.02 or above)
EZChrom Elite (3.3.2 SP1 or above)
OpenLAB (3.3.2 SP3)
Masshunter (B.02.01 SP1 or above)
Local controlAgilent Instant Pilot (G4208A) (B.02.08 or above).
CommunicationsController-area network (CAN), RS232C, APG remote: ready, start, stop and shutdown signals, LAN.
Safety and maintenanceExtensive diagnostics, error detection and display through Agilent LabAdvisor, leak detection, safe leak handling, leak output signal for shutdown of the pumping system. Low voltage in major maintenance areas.
GLP featuresEarly maintenance feedback (EMF) for continuous tracking of instrument usage in terms of seal wear and volume of pumped mobile phase with pre-defined and user settable limits and feedback messages. Electronic records of maintenance and errors.
HousingAll materials are recyclable.

Ordering Details – Agilent 1290 Infinity Binary Pump

DescriptionProduct Number
Agilent 1290 Infinity Binary Pump. Maximum. pressure 1200 barG4220A
HPLC Tool KitG4220A#001
Delete Agilent Lab AdvisorG4220A #950
Active seal wash optionG4220A B#030
Agilent 1290 Infinity Binary Pump VL. Maximum. pressure 1050 barG4220B
HPLC Tool KitG4220B #001
Delete Agilent Lab AdvisorG4220B #950
Active seal wash optionG4220B#030
Active seal wash (field upgrade)G1399A

www.agilent.com/chem/1200

© Agilent Technologies, Inc., 2010-2014
Published in USA, June 1, 2014
Publication Number 5990-6097EN

PDF preview unavailable. Download the PDF instead.

agilent-1290-infinity-binary-pump QuarkXPress(R) 8.5 QuarkXPress(R) 8.5

Related Documents

Preview Agilent 1260 Infinity バイナリポンプ ユーザーマニュアル
Agilent 1260 Infinity バイナリポンプの設置、操作、メンテナンス、トラブルシューティング、および仕様に関する詳細情報を提供するユーザーマニュアル。高性能液体クロマトグラフィーシステムに不可欠なコンポーネントについて解説します。
Preview Agilent LC System Best Practices: Technical Guide
Comprehensive guide to best practices for operating and maintaining Agilent Liquid Chromatography (LC) systems, covering housekeeping, daily tasks, power management, and component-specific recommendations for pumps, samplers, degassers, columns, and detectors.
Preview Agilent 1260 Infinity II Binary LC System: Manual and Quick Guide
This manual and quick guide provides comprehensive information on the Agilent 1260 Infinity II Binary LC System, covering product description, features, system components, configuration settings, quick start procedures, parts, consumables, and safety information.
Preview Agilent 1290 Infinity III High-Speed Pumps Service Manual
This service manual provides comprehensive information for the Agilent 1290 Infinity III High-Speed Pumps (G7120A and G7132A), covering product descriptions, features, installation, operation, maintenance, diagnostics, troubleshooting, and parts.
Preview Agilent InfinityLab LC Series Site Preparation Checklist for HPLC Systems
Comprehensive site preparation checklist for Agilent InfinityLab LC Series instruments, detailing laboratory space, environmental conditions, power requirements, solvent and gas needs, and stack configurations to ensure successful and timely installation of HPLC systems.
Preview Agilent InfinityLab LC Series Refractive Index Detectors User Manual
User manual for the Agilent InfinityLab LC Series Refractive Index Detectors, including the 1260 Infinity III (G7162A) and 1290 Infinity III (G7162B) models. This guide covers installation, operation, maintenance, and troubleshooting.
Preview Agilent 1260 Infinity III Binary LC System Manual
Comprehensive guide to the Agilent 1260 Infinity III Binary LC System, covering introduction, installation, configuration, operation, parts, and safety. Designed for advanced liquid chromatography applications.
Preview Agilent 1260 Infinity II Vialsampler Data Sheet
Data sheet for the Agilent 1260 Infinity II Vialsampler, an autosampler designed for routine pharmaceutical tasks, quality control, environmental, and food analyses. Includes product description, features, and detailed physical and performance specifications.