Braun ThermoScan® PRO 4000 Ear Thermometer
Operation and Technique
Remove the thermometer from the case and attach a new probe cover. The device turns on automatically.
Wait for the "ready" beep and place the probe snugly into the ear canal; correct probe placement is essential for accurate readings.
[Diagram: Cross-section of an ear canal showing probe tip placement near the tympanic membrane, with example temperature readings at different points.]
Press and release the Start Button. The thermometer will beep once. The green ExacTemp light above the Start Button flashes, indicating correct probe placement.
A long beep and a solid green ExacTemp light signals the end of an accurate measurement. Eject the used probe cover. For the next measurement, put on a new, clean probe cover.
If the probe is placed incorrectly or was moved while taking a temperature, the device beeps, the green ExacTemp light goes out and POS (Position Error) displays.
[Display showing "POS"]
Press the [power/memory] button to recall the last temperature.
To change default temperature scale: With the device off, press and hold the [power/memory] button for about 5 seconds. Release the [power/memory] button when desired scale is shown.
What Affects Accuracy
Factor | Affects Accuracy |
---|---|
Poor Probe Placement | Yes |
No Probe Cover | Yes |
Used Probe Cover | Yes |
Ambient Temperature | Yes |
Dirty/Chipped Lens | Yes |
Hearing Aid | Yes |
Lying on Pillow | Yes |
Cerumen (ear wax) | No |
Otitis Media (ear infection) | No |
Tympanostomy tubes | No |
Directions for Opening Battery Door
OLD INSTRUCTIONS
[Diagram: Illustration showing two steps for opening the battery door with arrows indicating movement. Step 1: Push down on the indicated area. Step 2: Slide the cover off.]
NEW INSTRUCTIONS
[Diagram: Illustration showing two steps for opening the battery door with arrows indicating movement. Step 1: Insert a tool (like a pen tip) into the slot and push. Step 2: Lift the battery cover upwards.]
Error Messages and Solutions
No probe cover is attached – Attach new, clean probe cover.
[Icon: Square with dashed border, starburst, and °C symbol]
POS (Position Error): The infrared monitor cannot find temperature equilibrium and allows no measurement. Clear the display by changing the probe cover. Ensure that the positioning of the probe is correct and remains stable.
[Display showing "POS"]
Ambient temperature is not within the allowed operating range 50-104 °F (10-40 °C) or changing too rapidly – Allow the thermometer to remain for 30 minutes in a room where the temperature is between 50 and 104 °F / 10 and 40 °C.
[Icon: Square with dashed border, starburst, and "Err" text]
System Error – Wait 1 minute until the thermometer turns off automatically, then turn on again. If error persists, reset the thermometer by removing the batteries and putting them back in. If error still persists, contact your Welch Allyn representative.
[Display showing "88.8" and memory/power icons]
Battery is low, but thermometer will still operate correctly – Insert new batteries.
[Display showing dashes and °C symbol]
Battery is too low to take correct temperature measurements – Insert new batteries.
[Icon: Square with dashed border and a battery symbol with a starburst]