Braun ThermoScan® PRO 4000 Ear Thermometer

Operation and Technique

  1. Remove the thermometer from the case and attach a new probe cover. The device turns on automatically.

  2. Wait for the "ready" beep and place the probe snugly into the ear canal; correct probe placement is essential for accurate readings.

    [Diagram: Cross-section of an ear canal showing probe tip placement near the tympanic membrane, with example temperature readings at different points.]

  3. Press and release the Start Button. The thermometer will beep once. The green ExacTemp light above the Start Button flashes, indicating correct probe placement.

  4. A long beep and a solid green ExacTemp light signals the end of an accurate measurement. Eject the used probe cover. For the next measurement, put on a new, clean probe cover.

  5. If the probe is placed incorrectly or was moved while taking a temperature, the device beeps, the green ExacTemp light goes out and POS (Position Error) displays.

    [Display showing "POS"]

  6. Press the [power/memory] button to recall the last temperature.

  7. To change default temperature scale: With the device off, press and hold the [power/memory] button for about 5 seconds. Release the [power/memory] button when desired scale is shown.

What Affects Accuracy

FactorAffects Accuracy
Poor Probe PlacementYes
No Probe CoverYes
Used Probe CoverYes
Ambient TemperatureYes
Dirty/Chipped LensYes
Hearing AidYes
Lying on PillowYes
Cerumen (ear wax)No
Otitis Media (ear infection)No
Tympanostomy tubesNo

Directions for Opening Battery Door

OLD INSTRUCTIONS

[Diagram: Illustration showing two steps for opening the battery door with arrows indicating movement. Step 1: Push down on the indicated area. Step 2: Slide the cover off.]

NEW INSTRUCTIONS

[Diagram: Illustration showing two steps for opening the battery door with arrows indicating movement. Step 1: Insert a tool (like a pen tip) into the slot and push. Step 2: Lift the battery cover upwards.]

Error Messages and Solutions

  • No probe cover is attached – Attach new, clean probe cover.

    [Icon: Square with dashed border, starburst, and °C symbol]

  • POS (Position Error): The infrared monitor cannot find temperature equilibrium and allows no measurement. Clear the display by changing the probe cover. Ensure that the positioning of the probe is correct and remains stable.

    [Display showing "POS"]

  • Ambient temperature is not within the allowed operating range 50-104 °F (10-40 °C) or changing too rapidly – Allow the thermometer to remain for 30 minutes in a room where the temperature is between 50 and 104 °F / 10 and 40 °C.

    [Icon: Square with dashed border, starburst, and "Err" text]

  • System Error – Wait 1 minute until the thermometer turns off automatically, then turn on again. If error persists, reset the thermometer by removing the batteries and putting them back in. If error still persists, contact your Welch Allyn representative.

    [Display showing "88.8" and memory/power icons]

  • Battery is low, but thermometer will still operate correctly – Insert new batteries.

    [Display showing dashes and °C symbol]

  • Battery is too low to take correct temperature measurements – Insert new batteries.

    [Icon: Square with dashed border and a battery symbol with a starburst]

Models: ThermoScan PRO 4000, Ear Thermometer

File Info : application/pdf, 1 Pages, 494.36KB

PDF preview unavailable. Download the PDF instead.

80010876LITPDF Adobe InDesign CS5.5 (7.5) Adobe PDF Library 9.9

Related Documents

Preview Braun ThermoScan PRO 4000 Ear Thermometer User Manual
Comprehensive user manual for the Braun ThermoScan PRO 4000 ear thermometer, detailing its features, usage instructions, troubleshooting, technical specifications, and care guidelines.
Preview Braun ThermoScan PRO 6000 Ear Thermometer: Directions for Use and User Manual
This document provides comprehensive directions for use and operational guidance for the Braun ThermoScan PRO 6000 ear thermometer, a clinical-grade device for professional medical environments. Learn about its features, setup, usage, maintenance, and troubleshooting.
Preview Braun ThermoScan PRO 4000: Guide to Taking Accurate Ear Temperatures
Learn how to use the Braun ThermoScan PRO 4000 ear thermometer for precise temperature readings. Includes step-by-step instructions, tips for accuracy, and factors affecting measurements.
Preview Braun ThermoScan PRO 4000 Ear Thermometer User Manual
Comprehensive user manual for the Braun ThermoScan PRO 4000 ear thermometer, covering how to use, temperature ranges, troubleshooting, care, and specifications. Features accurate, fast, and safe temperature readings.
Preview Braun ThermoScan PRO 4000 Ear Thermometer Manual
Comprehensive user manual for the Braun ThermoScan PRO 4000 ear thermometer by Welch Allyn. Includes instructions for use, troubleshooting, product specifications, and warranty information.
Preview Braun ThermoScan PRO 4000 Ear Thermometer User Manual
Comprehensive user manual for the Braun ThermoScan PRO 4000 ear thermometer, detailing usage instructions, product features, troubleshooting, specifications, care, and warranty information.
Preview Braun ThermoScan PRO 6000 Ear Thermometer: Directions for Use and Technical Guide
Comprehensive guide for the Braun ThermoScan PRO 6000 ear thermometer, covering usage instructions, features like PerfecTemp™ and ExacTemp™, maintenance, troubleshooting, specifications, and warranty information.
Preview Braun ThermoScan PRO 6000 Ear Thermometer Cleaning Guide
Official cleaning guide for the Braun ThermoScan PRO 6000 ear thermometer, detailing approved cleaning solutions and step-by-step instructions for the thermometer body, contacts, and lens window.