Cleaning Guide
Braun ThermoScan® PRO 6000
Manufactured by Welch Allyn, Inc.
Thermometer body
NOTE: Place a new probe cover on the thermometer before cleaning.
Approved cleaning solutions:
- 70% isopropyl alcohol or ethyl alcohol
- Hydrogen peroxide (Virox, Oxivir)
- Quaternary ammonium compounds (CaviWipes™, Clinell Universal Wipes, Sanicloth)
- 10% Chlorine bleach solution
Cleaning Steps:
- Apply approved cleaning solution to a wipe or cloth.
- Wipe the thermometer body gently.
- Ensure the thermometer body is dry.
Drying: Wipe dry. Allow 5 minutes drying time. Make sure thermometer body is dry before use. ⏰
Thermometer contacts
Approved cleaning solutions:
- 70% isopropyl alcohol or ethyl alcohol
Cleaning Steps:
- Apply approved cleaning solution to a cotton swab.
- Gently wipe the thermometer contacts.
- Ensure contacts are clean.
Drying: Place the thermometer aside for 1 minute, allowing the contacts to air dry. ⏰
Lens window and probe
Approved cleaning solutions:
- 70% isopropyl alcohol or ethyl alcohol only!
Cleaning Steps:
- Apply approved cleaning solution to a wipe or cloth.
- Gently wipe the probe lens window.
- Ensure the probe lens window is dry.
Drying: Wipe dry. Allow 5 minutes drying time. Make sure probe lens window is dry before use. ⏰
Product Information
This product is manufactured by Welch Allyn, Inc. under license to the 'Braun' trademark. 'Braun' is a registered trademark of Braun GmbH, Kronberg, Germany. ThermoScan is a registered trademark owned by Helen of Troy Limited.
Document Details:
Document Number: 80025920 Ver. D
Revision Date: 2021-03
©2021 Welch Allyn, Inc. All rights reserved.
Hillrom Technical Support
For support, visit: hillrom.com/en-us/about-us/locations/ [info]
Welch Allyn, Inc. is a subsidiary of Hill-Rom Holdings, Inc.
Distributor Information (DocStock)
DocStock Pty Ltd
ABN: 48 115 268 203
Address: 20/7 Anella Ave, Castle Hill, NSW 2154
Phone: (02) 8850 7177
Fax: (02) 8850 7988
Website: www.DocStock.com.au