Broadcom AS20 Encoder

Miniaturization of Energy Harvesting Magnetic Encoder: Compact Designs for Mini Motors

Abstract

Traditional encoders present challenges in space-constrained, weight-sensitive, or high-precision applications due to their size, complexity, and installation requirements. This white paper introduces the miniature energy harvesting multi-turn (EHMT) magnetic encoder, a revolutionary solution designed to overcome these limitations. Combining an exceptionally compact-size and lightweight design with remarkably high accuracy, this encoder enables precise motion feedback where it was previously impossible or impractical. Furthermore, its innovative magnetic sensing principle and simplified mechanical interface ensure easy assembly and integration, significantly reducing installation time and complexity.

Its lightweight rotational magnetic source drastically reduces moment of inertia. This is transformative for applications sensitive to small starting torques, including micro-drives, portable surgical tools, and miniature robotic joints, where every milligram and mili-Newton-meter counts. Combined with its inherent small size and ease of assembly, the EHMT encoder sets a new standard for reliable, precise motion feedback in the most demanding compact applications.

Introduction

The Broadcom® AS20 Encoder offers high performance in a miniature EHMT magnetic encoder. It features an automatic offset and gain calibration across operating temperature. This technology actively nullifies Hall sensor offset drift, ensuring that high accuracy is maintained immediately upon first power-on and across the full operating temperature range, eliminating a major source of system variability.

Unlike conventional counterparts, the magnetic technology demonstrates exceptional immunity to current drift throughout its operational lifetime, providing designers with confidence in stability and precision.

Applications

Core Technology: Automatic Offset and Gain Calibration across Operating Temperature

Traditional magnetic encoders relying on Hall-effect sensors are fundamentally susceptible to offset and gain drift caused by temperature variations. This drift manifests as significant first-power-on accuracy errors (often several percent) when the encoder is powered at a temperature different from its last calibration point. Manual calibration or complex compensation circuits are typically required to mitigate this.

During manufacturing, each EHMT encoder is programmed with the default temperature value of specified operating temperature range. The optimized default offset and gain calibration values are preloaded for four key temperature points (P0-Value D, P1-Value A, P2-Value B, P3-Value C) in the internal memory, spaced at precise 32°C intervals. Point P1 is typically set at a standard room temperature (for example, 25°C).

Memory Settings Illustration of Calibration Values:

Memory Group Factory Default Post calibration
P3 (+64°C) Value C Value C+
P2 (+32°C) Value B Value B+
P1 (Room Temp) Value A Value A+
P0 (-32°C) Value D Value D+

Intelligent Runtime Learning:

During normal operation, when the encoder operates at a temperature close to P0 (P1 – 32°C), the system automatically performs a calibration cycle and stores the optimized offset/gain values for P0. Similarly, when operating near P2 (P1 + 32°C) and P3 (P1 + 64°C), the encoder automatically calibrates and stores values for these points. During operation at room temperature, P1 value also will be updated similar to the three other temperature points.

Calibration Storage: All relearned calibration values (P0, P1, P2, and P3) are stored in the encoder's internal nonvolatile memory.

Automatic Temperature Calibration Flow:

The flow starts with factory default settings. If the operating temperature is close to P0, calibration occurs and stores the new value. If not, it checks for P1, then P2, and finally P3. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next check. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3, calibration occurs. If not, it checks for P2, then P1, and finally P0. If the temperature is close to any of these points, calibration occurs. If the temperature is not close to any defined points, it proceeds to the next step. After checking all points, if no calibration occurred, it proceeds to the next step. If the temperature is close to P3,

PDF preview unavailable. Download the PDF instead.

AS20-M42M-Series-WP100 Acrobat Distiller 25.0 (Windows)

Related Documents

Preview Broadcom AS20-M42M-Pxx Series Absolute Magnetic Encoder SPI 4-Wire Option Software Specification
Technical specification detailing the AS20-M42M-Pxx series absolute magnetic encoder with SPI 4-wire interface, covering communication protocols, data formats, timing, memory mapping, configuration, calibration, and testing procedures.
Preview BCM4390: WICED Wi-Fi IEEE 802.11 b/g/n SoC with Embedded Application Processor Data Sheet
This data sheet provides comprehensive details on the Broadcom BCM4390, a single-chip solution integrating IEEE 802.11 b/g/n MAC/baseband/radio with an ARM Cortex M3 applications processor. It is designed for embedded systems requiring high integration, low power consumption, and flexible form factors, particularly in IoT applications.
Preview Broadcom Emulex Host Bus Adapters Installation Guide Release 14.4
This installation guide (Release 14.4) from Broadcom provides detailed instructions for setting up Emulex Host Bus Adapters (HBAs), including models LPe38100, LPe37100, LPe36000, LPe35000, LPe32000, and LPe31000 series. It covers installation, compatibility, major features, specifications, and regulatory information.
Preview Broadcom White Paper: Predicting Network Congestion with Software to Silicon Solution
Broadcom's white paper details an industry-first software to silicon solution for predicting and mitigating network congestion in the era of 5G, cloud, and SDx. Learn how Broadcom's AIOps platform enhances customer experience through advanced analytics and automation.
Preview Broadcom Zero Trust Architecture (ZTA) Executive Summary
An executive summary of Broadcom's Zero Trust Architecture (ZTA), detailing its principles, application across five pillars, cross-cutting capabilities, and enabling technologies for modern cybersecurity.
Preview Broadcom BCM5419X Quad and Octal Gigabit Ethernet Transceiver with MACsec Product Brief
Product brief detailing the Broadcom BCM5419X family of quad and octal Gigabit Ethernet transceivers with integrated MACsec functionality. Features include IEEE 802.1AE compliance, support for various copper and fiber interfaces, advanced MACsec features like AES/GCM encryption, and energy-efficient operation.
Preview Broadcom Bluetooth Module Variants: BCM20732S, BCM20736S, BCM20737S Similarity Letter
Letter of Similarity from Broadcom Corporation confirming that the BCM20732S, BCM20736S, and BCM20737S Bluetooth modules are variants of an approved product, detailing their differences and similarities.
Preview Broadcom Semiconductor Solutions Terms and Conditions of Sale
Official terms and conditions governing the sale of Broadcom semiconductor products and software licenses, covering applicability, orders, pricing, delivery, warranty, licenses, limitations of liability, and general provisions.