The Metrel MI 3108 EurotestPV is a combined instrument designed for testing photovoltaic (PV) systems and low-voltage installations. It supports all installation tests according to EN 61557 standards and also performs essential tests and measurements on PV systems, including those required by EN 62446. The instrument can measure I-U characteristics, calculate STC values, and measure power and energy on both the DC and AC sides of the inverter. It is built to operate safely in demanding conditions with DC voltages up to 1000 V and currents up to 15 A, featuring a safety probe for secure disconnection.
Key Features and Functions
PV Installations
- DC Side Measurements: Voltage, current, power, energy, open-circuit voltage (Uoc), short-circuit current (Isc), I-V characteristic of PV modules and strings, module illumination, and temperature.
- AC Side Measurements: Voltage, current, power, calculation of efficiencies for PV modules, inverters, and the entire PV system.
- STC Value Calculation: Measures current and voltage values are converted to Standard Test Conditions (STC) based on environmental conditions, enabling comparison of results from different measurement conditions.
- Efficiency Calculation: Utilizes two voltage and two current channels for simultaneous measurement of AC and DC parameters.
- PV Remote Unit: An optional unit for simultaneous measurement of solar irradiance and PV module temperature.
- Graphical Display: Shows the module's I-U curve on the instrument's LCD.
Electrical Installations
- Insulation Resistance: With DC voltage.
- Protective Conductor Continuity: PE continuity testing.
- Installation Impedance: Loop and line impedance measurements.
- Loop Impedance without RCD Tripping: Features a Trip Lock function for impedance testing without triggering RCDs.
- RCD Testing: Supports testing of RCD types AC, A, and B. Includes an RCD Auto function for faster testing.
- Earth Resistance: Supports 3-wire earth resistance measurement.
- AC Current Measurement: Measures load and leakage current (TRMS).
- Voltage and Frequency: TRMS voltage, frequency, and phase sequence measurement.
- Power and Harmonics Analysis: Measures power, energy, and harmonics up to the 11th harmonic for single-phase power.
Data Management and Connectivity
- Memory: Stores up to 1800 test results and 500 graphical results.
- Bluetooth Communication: Connects with Android mobile devices via an optional Bluetooth plug-in.
- Android Application: EuroLink PV app for data management.
- PC Software: EuroLink PRO software for recording, uploading, viewing, analyzing, and printing test results.
Applications
- Testing and fault finding in photovoltaic installations.
- Power and efficiency measurements (AC and DC).
- First and periodic testing of domestic and industrial single and three-phase electrical installations.
Standards Compliance
- Functional: IEC/EN 61557 series, IEC 62446 (photovoltaics), IEC 61829.
- Other Testing Standards: BS 7671, EN 61008, EN 61009, EN 60364-4-41, AS/NZ 3017.
- EMC: EN 61326.
- Safety: EN 61010-1, EN 61010-2-030, EN 61010-031, EN 61010-2-032.
Technical Specifications
PV Installation Measurements
Function | Measuring Range | Accuracy |
---|---|---|
Voltage | 0 VDC ... 999 VDC 0 VAC ... 999 VAC |
±(1% of reading + 2 digits) |
Current | I-V m.: 0 VDC ... 999 VDC Panel m.: 0.0 mA ... 300 ADC Inverter m.: 0.0 mA ... 300 AAC |
±(1.5% of reading + 3 digits) |
Power | I-V m.: 0.00 A ... 15 ADC Panel m.: 0 ... 999 kW I-V m.: 0 ... 14.99 kW |
±(1% of reading + 2 digits) |
I/U Characteristic | 1000 V/15A / 15 kW | ±(1% of reading + 2 digits) |
Irradiance | 0 ... 1750 W/m² | ±(1.5% of reading + 3 digits) |
Temperature | -10 °C ... +85 °C | ±(1% of reading + 2 digits) |
The instrument provides accurate measurement results, fully compliant with technical specifications, for modules with efficiency up to 20%.
Electrical Installation Measurements
Function | Measuring Range | Accuracy |
---|---|---|
Insulation Resistance (EN 61557-2) | U = 50, 100, 250 VDC: R: up to 199.9 MΩ U = 500 VDC, 1 kVDC: R: up to 999 MΩ |
±(5% of reading + 3 digits) |
Resistance, 200 mA (EN 61557-4) | 0.00 Ω ... 1999 Ω | ±(5% of reading + 3 digits) |
Resistance, 7 mA | 0.0 Ω ... 1999 Ω | ±(3% of reading + 3 digits) |
Loop Impedance (EN 61557-3) | 0.0 Ω ... 1999 Ω | ±(5% of reading + 3 digits) |
Short-Circuit Loop Impedance (EN 61557-3) | 0.0 Ω ... 1999 Ω | ±(5% of reading + 5 digits) |
Voltage | 0 VAC ... 550 VAC | ±(2% of reading + 2 digits) |
Frequency | 0.00 Hz ... 499.9 Hz | ±(0.2% of reading + 1 digit) |
Phase Sequence (EN 61557-7) | 1-2-3 or 3-2-1 | - |
RCD Testing (EN 61557-6) | IΔN: 10 mA, 30 mA, 100 mA, 300 mA, 500 mA, 1 A Uc: 0.0 V ... 99.9 V Tripping Time: 0 ms ... 40 ms Tripping Current: 0.2 x IΔN ... 2.2 x IΔN |
(-0% / +15%) of reading ±1 ms ±0.1 x IΔN |
Earth Resistance (EN 61557-5) | 0.00 Ω ... 9999 Ω | ±(5% of reading + 5 digits) |
Energy | 0.000 Wh - 1999 kWh | - |
Harmonics | Up to the 11th | - |
General Data
Parameter | Main Unit | Remote Unit |
---|---|---|
Graphical Display | 128 x 64 pixels with backlight | 128 x 64 pixels with backlight |
Power Supply | 6 x 1.2 V NiMH AA batteries | 6 x 1.2 V NiMH AA batteries |
Overvoltage Category | CAT II/1000 VDC; CAT III / 600 V; CAT IV/300V | - |
Protection Type | Double insulation | - |
COM Port | RS232 and USB | RS232 |
Dimensions | 230 x 103 x 115 mm | 140 x 230 x 80 mm |
Weight | 1.3 kg | 1.0 kg |
PV System Parameters Diagram
A diagram illustrates the connection of PV system components: Solar Module (DC Input) connected to a DC Disconnect, which then connects to the Inverter (DC/AC Inverter). The Inverter's AC Output is connected to the Main consumer unit. Measurements are shown for Uoc, Isc, PmpP, and I-V curve on the DC side, and Power on the AC side. Irradiance (W/m²) and Cell Temperature (°C) are external inputs affecting the Solar Module.
Ordering Information
MI 3108 ST (Standard Kit)
- MI 3108 EurotestPV Instrument
- Carrying bag
- Schuko test lead
- Universal PV measuring lead, 3 x 1.5 m
- Test probe, blue, red, green, black, 4x
- Alligator clip, blue, red, green, black, 4x
- PV Safety Probe
- MC3/4 PV adapters
- Pyranometer
- Temperature probe
- Power adapter + 6 NiMH rechargeable batteries, AA type
- USB cable
- RS-232 - PS/2 cable
- Carrying straps set
- PC SW EuroLink PRO
- Short user manual
- User manual and handbook on CD
- Calibration certificate
MI 3108 PS (PV System Kit)
- PV Remote Unit
- Power adapter + 12 NiMH rechargeable batteries, AA type
- Test lead with probe
- EuroLink PRO Plus license
Optional Accessories
Photo | Order No. | Description |
---|---|---|
A 1378 |
EurotestPV Remote Unit |
|
A 1314 |
Schuko test adapter |
|
A 1401 |
Test lead with 3-wire probe |
|
A 1018 |
Current clamp (low range, leakage current) |
|
A 1391 |
AC/DC Current Clamp |
|
A 1105 |
Barcode reader |
|
A 1431 |
EuroLink Android |
|
A 1436 |
Bluetooth dongle |
|
S 2026 |
Earth resistance measuring set, 3-wire, 20 m |
|
S 2027 |
Earth resistance measuring set, 3-wire, 50 m |
Manufacturer Information
METREL D.O.O.
Measuring and regulatory equipment
Ljubljanska 77, SI-1354 Horjul, Slovenia
T (0)1 75 58 200
metrel@metrel.si
www.metrel.si
Note: Product images in the catalog may differ from the ordered set at the time of delivery. Metrel reserves the right to make technical changes without prior notice.