NOVEX ABBE REFRACTOMETER

Model 98.490

Users Manual

Manufactured by EUROMEX Microscopen B.V., HOLLAND

Website: www.euromex.com

1.0 Introduction

With your purchase of the NOVEX Abbe refractometer, you have chosen a quality product. The NOVEX Abbe refractometer is developed for use in laboratories and in the food industry. The maintenance requirement is limited when using the refractometer in a proper manner. This manual describes the construction of the refractometer, how to use the refractometer, and its maintenance.

2.0 Index

3.0 Construction of the refractometer

The names of the several parts are listed below and are indicated in the picture:

PartDescription
AAdjustable eyepiece
BMain body
CDispersion correction knob
DWater connection (in-out)
EWater connection (out)
FAdjustment knob
GStand foot
HScale window (not visible)
IWater connection (in)
JReflection mirror
KPrimary prism
LLight window
MSecundary prism
NCover light window
OThermometer
PPrism cover

The instrument consists of a main body (B), stand foot (G), and a measuring part consisting of a primary prism (K) and a secondary prism (M). When moving the instrument, always pick it up by its main body.

4.0 Functions of the refractometer

The instrument consists of a main body (B), stand foot (G), and a measuring part comprising a primary prism (K) and a secondary prism (M). When moving the instrument, always pick it up by its main body.

5.0 Preparing the refractometer for use

Remove the instrument from the aluminum case and place it on a flat surface. Remove the black protection cap from the thermometer and screw the thermometer gently into its inlet on the main body, which is situated at the side of the primary prism (K).

6.0 Working with the refractometer

For optimum use, please follow the procedures below.

6.1 Calibrating the refractometer

Before measuring, the instrument should be calibrated with the supplied test piece and immersion fluid, as described below:

  1. Put some drops of the immersion fluid onto the surface of the primary prism (K).
  2. Gently place the test piece into the fluid, with the polished side pointing downwards.
  3. Peep through the eyepiece and turn the eyepiece until the image is sharply focused.
  4. Now turn the adjustment knob F until the cross exactly separates the light and dark image in the field of view.
  5. The scale should now read exactly the same value as mentioned on the test piece. If this is not the case, adjust the scale by turning the set screw (B) with the supplied driver.
Refractometer Scale View

The field of view displays a scale with crosshairs. The scale includes a Brix scale (sugar percentage) ranging from 80 to 90, and an nD scale (refractive index) with values such as 1.490, 1.500, and 1.510. The crosshairs are used for precise alignment between the light and dark portions of the image.

6.2 Measuring liquid samples with the refractometer

  1. Put some drops of the sample fluid onto the primary prism (K), and close the secondary prism (M) by means of the knob (P). Ensure the sample is homogeneous and without air bubbles.
  2. Open the light cap (N) and close the reflection mirror (J).
  3. Peep through the eyepiece (A) and focus the cross line.
  4. Turn the illumination window (H) of the scale until the brightest image is reached.
  5. Now move the cross with knob (F) until it is exactly at the border line of the dark and light image.
  6. Colour shifting in the image can be corrected to "black/white" by means of the dispersion knob (C).
  7. Now a last correction to get the cross exactly on the border line can be made with knob (F).

The refractive index value (nD) can now be read on the bottom part of the scale.

6.3 Measuring the sugar percentage in a liquid

  1. Put some drops of the sample fluid onto the primary prism (K), and close the secondary prism (M) by means of the knob (P). Ensure the sample is homogeneous and without air bubbles.
  2. Open the light cap (N) and close the reflection mirror (J).
  3. Peep through the eyepiece (A) and focus the cross line.
  4. Turn the illumination window (H) of the scale until the brightest image is reached.
  5. Now move the cross with knob (F) until it is exactly at the border line of the dark and light image.
  6. Colour shifting in the image can be corrected to "black/white" by means of the dispersion knob (C).
  7. Now a last correction to get the cross exactly on the border line can be made with knob (F).

The sugar percentage (BRIX) can now be read on the upper part of the scale.

6.4 Measuring clear-transparent solid materials (e.g. glass)

Note! Make sure one of the sides of the sample is perfectly smooth.

  1. Put some drops of the supplied immersion fluid onto the primary prism (K).
  2. Carefully place the specimen with its smooth side into the fluid.
  3. Open the light cap (N) and close the reflection mirror (J).
  4. Peep through the eyepiece (A) and focus the cross line.
  5. Turn the illumination window (H) of the scale until the brightest image is reached.
  6. Now move the cross with knob (F) until it is exactly at the border line of the dark and light image.
  7. Colour shifting in the image can be corrected to "black/white" by means of the dispersion knob (C).

The refractive index can now be read on the bottom part of the scale.

6.5 Measuring non-clear-transparent solid materials (e.g. satinized glass)

Note! Make sure one of the sides of the sample is perfectly smooth.

  1. Put some drops of the supplied immersion fluid onto the primary prism (K).
  2. Carefully place the specimen with its smooth side into the fluid.
  3. Open the reflection mirror (J).
  4. Peep through the eyepiece (A) and focus the cross line.
  5. Turn the illumination window (H) of the scale until the brightest image is reached.
  6. Now move the cross with knob (F) until it is exactly at the border line of the dark and light image.
  7. Colour shifting in the image can be corrected to "black/white" by means of the dispersion knob (C).
  8. Now a last correction to get the cross exactly on the border line can be made with knob (F).

The refractive index can now be read on the bottom part of the scale.

6.6 Establishing the dispersion value DFC

  1. Put some drops of the sample fluid onto the primary prism (K), and close the secondary prism (M) by means of the knob (P). Ensure the sample is homogeneous and without air bubbles.
  2. Open the light cap (N) and close the reflection mirror (J).
  3. Peep through the eyepiece (A) and focus the cross line.
  4. Turn the illumination window (H) of the scale until the brightest image is reached.
  5. Now move the cross with knob (F) until it is exactly at the border line of the dark and light image.
  6. Correct the dispersion with the knob (C) and write down the value "Z" (to read on the side of knob C). If "Z" is higher than 30, write "Z" as a negative number.
  7. Now a last correction to get the cross exactly on the border line can be made with knob (F).
  8. The refractive index can now be read on the bottom part of the scale.
  9. Take the values of "A", "B" and "σ" over from the table on page 9 of this manual with the use of the values "Z" and nD, written down on point 6.
  10. Use the values for the below formula:

DFC = A + σB

6.6a Calculation example 1, measured values with 2 decimals behind the comma:

Measured at 22°C:

nD liquid: 1.3300

Dispersion correction Z : 40.0

Found values A and B in the nD column in table 6.7:

A = 0.02484

B = 0.03304

Found value σ in the Z column in table Z:

σ = -0.500 (negative value for Z is higher as 30)

DFC = A + σB

DFC = 0.02484 + (0.03304 x -0.500)

DFC = 0.00832

6.6b Calculation example 2, measured values with more as 2 decimals behind the comma:

Measured at 20°C:

ND distilled water: 1.3330*

Dispersion correction Z : 41.62*

*In the table the measured values nD 1.3330 and "Z" value 41.62 are not included, therefor calculate "A", "B" and "σ" as follows:

Take the nD value 1.33 (or the corresponding "Z" value) from the table and read the correct value (in this case 5x10⁻⁶ per 0.001) and add it to the value of "A" given for 1.33.

1.3330 - 1.33 = 0.003 so the correcting value is: 3 x 5 x 10⁻⁶ = 0.000015.

The for 1.33 given "A" value is 0.02484 so the corrected "A" value for nD 1.3330 is: 0.02484 + 0.000015 = 0.024825.

Repeat these steps also for the "B" and "σ" values.

Results:

A = 0.024825

B = 0.032983

σ = -0.5716 (negative value for Z is higher as 30)

DFC = A + σB

DFC = 0.024825 + (0.032983 x -0.5716)

DFC = 0.005972

6.7 Table dispersion values

NDA0.001 decimal correction for AB0.001 decimal correction for BZσ0.1 decimal correction for σ
x(10⁻⁶)(10⁻⁹)x(10⁻⁴)Z
1.300000.02499-50.03349-1300.00060
1.310000.02494-50.03336-1610.999159
1.320000.02489-50.03320-1620.995458
1.330000.02484-50.03304-1030.988757
1.340000.02479-50.03285-2040.9781056
1.350000.02474-40.03265-2150.9661255
1.360000.02470-40.03244-2260.9511554
1.370000.02466-50.03221-3470.9341753
1.380000.02461-40.03197-2780.9142052
1.390000.02457-30.03170-2790.8912351
1.400000.02454-40.03143-30100.8665250
1.410000.02450-30.03113-31110.8392749
1.420000.02447-40.03082-32120.8093048
1.430000.02443-30.03050-35130.7773247
1.440000.02440-20.03615-36140.7433446
1.450000.02438-20.02979-38150.7073645
1.460000.02435-20.02941-39160.6693844
1.470000.02433-30.02902-42170.6294043
1.480000.02430-20.02860-43180.5884142
1.490000.02428-10.02817-46190.5454341
1.500000.02427-20.02771-47200.5004540
1.510000.02425-10.02724-49210.4544639
1.520000.02424-10.02675-52220.4074738
1.530000.0242300.02623-54230.3584937
1.540000.0242300.02569-56240.3094936
1.550000.0242300.02513-59250.2595035
1.560000.0242300.02454-61260.2085134
1.570000.02424+10.02393-64270.1565233
1.580000.02425+10.02329-64280.1045232
1.590000.02426+10.02262-67290.0525231
1.600000.02428+20.02192-70300.0005230
1.610000.02430+20.02119-73
1.620000.02433+30.02042-77
1.630000.02437+40.01962-80
1.640000.02442+50.18877-85
1.650000.02447+50.01788-89
1.660000.02453+60.01694-94
1.670000.02461+80.01594-100
1.680000.02470+90.01487-107
1.690000.02480+100.01373-114
1.700000.02493+130.01250-123

6.8 The use of circulating water

For use in so-called "on-line" situations, it can be important to keep the temperature of the sample fluid at a constant level. The refractometer can be linked to a water circulator to keep the prisms of the refractometer at a constant temperature. Therefore, the connection points of the circulation system should be connected to each other by means of flexible rubber hoses (not supplied). Proceed as follows:

  • The water supply should be connected to connection point "I".
  • Connection point "E" should be connected to one of the connection points "D".
  • The last connection point should be used for the discharge.

6.9 Refractive indices and average dispersion values of distilled water

As example dispersion values and nD values corrected at temperatures of 10 - 40°C are given.

Temp in °CRefractive index in nDDispersion value DFCTemp in °CRefractive index in nDDispersion value DFC
101.333690.00600331.331570.00593
111.333640.00600341.331440.00593
121.333580.00599351.331310.00592
131.333520.00599361.331170.00592
141.333460.00599371.331040.00591
151.333390.00599381.330900.00591
161.333310.00598391.330750.00591
171.333240.00598401.330610.00590
181.333160.00598
191.333070.00597
201.332990.00597
211.332900.00597
221.332800.00597
231.332710.00596
241.332610.00596
251.332500.00596
261.332400.00596
271.332290.00595
281.332170.00595
291.332060.00594
301.331940.00594
311.331820.00594
321.331700.00593

7.0 Maintenance and cleaning

Always place the dustcover over the refractometer after use. Clean both prism surfaces with a tissue.

7.1 Cleaning the optics

If the eyepiece or prism surfaces are polluted, they can be cleaned with a lenspaper with a little alcohol. Never use a drop of alcohol directly on the lenses, since it can damage the coating.

Warning

Cleaning cloths containing plastic fibres can damage the coating of the lenses!

7.2 Maintenance of the stand

Dust can be removed with a brush. In case the stand is very dirty, the surface can be cleaned with a non-aggressive cleaning product.

8.0 Refractometer accessories

  • LE.5209: Cold light source EK-0, 12 V 20 W, with single-arm, self-sustaining light guide
  • SL.5208: Spare halogen bulb 12 V 20 W for LE.5209
  • 98.492: Spare thermometer 0-50°C
  • 98.496: Calibrating test piece nD 1.5163
Models: 98.490 ABBE Refractometer, 98.490, ABBE Refractometer, Refractometer

File Info : application/pdf, 12 Pages, 159.83KB

PDF preview unavailable. Download the PDF instead.

9841 2-Bedienungsanleitung f Refraktormeter 98.490 EN

References

AFPL Ghostscript 8.54 PrimoPDF http://www.primopdf.com

Related Documents

Preview Euromex Refractometers: Analog and Digital Product Datasheet
Comprehensive product datasheet for Euromex analog and digital refractometers, detailing specifications, applications in sugar analysis, salinity, battery acid, clinical samples, and gemology. Includes Abbe and gemological refractometer models.
Preview Euromex Digital Refractometer User Manual
User manual for the Euromex Digital Refractometer, providing instructions on specifications, calibration, measurement, maintenance, and troubleshooting for accurate Brix and temperature readings.
Preview Euromex Delphi-X Observer Metall Microscope User Manual
Comprehensive user manual for the Euromex Delphi-X Observer Metall microscope, providing detailed instructions on setup, operation, maintenance, and troubleshooting for observing opaque materials.
Preview Euromex ED.3000 Blue Line Series Digital Microscope User Manual
Comprehensive user manual for the Euromex ED.3000 digital microscope, part of the Blue Line series. Covers handling, maintenance, operation, technical specifications, and interface settings for educational and laboratory use.
Preview Euromex MZ.5000 Digital Macrozoom Microscope User Manual
Comprehensive user manual for the Euromex MZ.5000 digital macrozoom microscope, detailing setup, operation, maintenance, and troubleshooting for Life Sciences applications. Includes component identification and assembly instructions.
Preview Euromex Delphi-X Observer User Manual
This user manual provides detailed instructions for the Euromex Delphi-X Observer microscope, covering its features, safe operation, assembly, maintenance, and troubleshooting for life science applications.
Preview Euromex Delphi-X Inverso DIC Quick Start Manual
A quick start guide for setting up and using the Euromex Delphi-X Inverso microscope with Differential Interference Contrast (DIC) imaging, detailing setup steps and mode switching.
Preview Euromex Delphi-X Observer Metallurgical Microscope User Manual
Comprehensive user manual for the Euromex Delphi-X Observer metallurgical microscope, covering safety, configuration, operation, and troubleshooting.