ED.3000 Blue Line Series Digital Microscope

User Manual

Brand: Euromex Microscopes Holland

A description of the Euromex ED.3000 digital microscope: a white and blue laboratory/educational microscope featuring a prominent 9-inch display screen, a microscope head with an objective lens, a focusing knob, an LED ring light, and a base with a working stage. It is designed for excellent price-performance ratio.

Introduction

Thank you for purchasing the Euromex ED.3000. These microscopes are developed for use in schools and laboratories, with specific attention to production methods resulting in an excellent price-performance ratio. Please read this manual carefully before using this product to ensure correct and safe usage. The content of this manual is subject to change without notice. The appearance of the actual product may differ from the models described. Not all equipment mentioned may be part of the purchased set. All optics are anti-fungus treated and anti-reflection coated for maximum light throughput.

Notes on Handling and Safety

Handle with care

This product is a high-quality optical instrument. Handle it with care, avoiding sudden shocks and impacts, as even small impacts can affect the precision of the objective.

Handling the LED

Always disconnect the power cord and allow the system approximately 35 minutes to cool down before handling the LED bulb and power unit to avoid burns. Never touch the LED with bare hands. Dirt or fingerprints on the LED can reduce its lifespan and result in uneven illumination. Use only Euromex original replacement LEDs; using other products may cause malfunctions and void the warranty. The power unit gets hot during use; never touch it while in operation and allow it to cool down for approximately 35 minutes.

Dirt on the lenses

Dirt on or inside optical components (eyepieces, lenses, etc.) negatively affects image quality. Always try to prevent your microscope from getting dirty by using the dust cover. Prevent leaving fingerprints on the lenses and regularly clean the outer surface of the lenses. Cleaning optical components is a delicate matter; refer to specific cleaning instructions in this manual.

Environment, Storage, and Use

This is a precision instrument and should be used in a proper environment for optimal use. Install the product indoors on a stable, vibration-free, and level surface. Do not place the product in direct sunlight. The ambient temperature should be between 5°C to +40°C, and relative humidity should be a maximum of 80% at 31°C, decreasing linearly to 50% at 40°C. Although the system is anti-mold treated, installation in a hot, humid location may still result in mold or condensation on lenses, impairing performance or causing malfunctions. Never turn the right and left focus knobs in opposite directions simultaneously or turn the coarse focus knob past its farthest point, as this will damage the microscope. Never use undue force when turning the knobs. Ensure the microscope system can dissipate heat. Keep the microscope approximately 15 cm clear from walls and obstructions. Never turn the microscope on when the dust cover is in place or when items are placed on the microscope. Keep flammable fluids, fabric, etc., well out of the way.

Disconnect power

Always disconnect your microscope from power before performing any maintenance, cleaning, assembling, or replacing LEDs to prevent electric shocks.

Prevent contact with water and other fluids

Never allow water or other fluids to come into contact with your microscope, as this can cause short-circuiting, malfunction, and damage to your system.

Moving and Assembling

The ED.3000 microscope is a relatively heavy system. Consider this when moving and installing it. Always lift the microscope by holding the main body and base. Never lift or move the microscope by its focusing knobs, stage, or head. If needed, move the microscope with two persons instead of one.

Component

The Euromex ED.3000 microscope system includes the following key components:

Maintenance

All lens surfaces should be kept clean. They can be lightly wiped with a cotton cloth to remove dust. For dirt or fingerprints, gently wipe with a neutral detergent or Euromex cleaning agent. Do not use organic substances. Optical parts should only be cleaned with dedicated cleaning agents. Do not disassemble any parts of the microscope. When the microscope is not in use, use a dust cover.

Operation

Operating environment

Ensure the operating environment is within room temperature (-5°C to +40°C) and relative humidity is between 45% to 85%.

Choose the working stage

The ED.3000 comes with a black/white stage as an accessory. To use it, loosen the fixing screw, remove the glass table, and replace it with the black/white stage. Tighten the fixing screw. Typically, the white side of the stage is used upward for dark samples to increase contrast. If the sample is light-colored, use the black side upward, which requires incident light.

Operating steps

  1. Place the microscope on a stable table.
  2. Turn the 9" display screen from a horizontal to a vertical position.
  3. Tear off the protective film from the screen.
  4. Insert an SD card if you need to take photos or record videos.
  5. Connect the mouse to the USB 2.0 interface.
  6. Connect the power adapter (100V-240V/12V 1A).
  7. Turn the ON/OFF switch for the microscope and display screen to "ON".
  8. Adjust the light intensity for incident and transmitted light if needed using the dimmer controls.
  9. Place the sample in the center of the working stage.
  10. Rotate the lens adjustment ring to the minimum position (1x).
  11. Turn the focusing knob slowly until a clear image appears on the display screen.
  12. If magnification needs to be changed, rotate the lens adjustment ring; it can be changed from 1x to 4x.

Technical Data

ParameterSpecification
Objective Lens1x~4x
Reduction lens0.3x
Fixed working distance90.5 mm
Image sensor1/3", 3 MP
White balanceauto or one push white balance
Resolution1024 x 600/30 fps
Pixel size2.75 x 2.75 μm
Signal/noise ratioabove 65 DB
Optical zoom8.1x-32.4x
Digital zoom1x-10x
Functioncapture and take videos
OutputUSB2.0 interface
Power adapter100V-240V/12V 1A
IlluminationIncident: 21 x 20 mA LED ring light; Transmitted: 7x 20 mA LED. Both adjustable.

Calculation of total magnification: Objective magnification x (25.4x size of monitor in inch diagonal line of CMOS) x magnification of reduced lens.

Remarks: A: 25.4 x size of monitor (mm): 1 inch = 25.4 mm. B: Diagonal line of CCD: 1/3" CMOS: 6 mm, 1/2" CMOS: 8 mm, 2/3" CMOS: 11 mm. Example: For 1x objective magnification, 0.3x reduction lens, 1/3" CCD, and 9" LCD, the total magnification is: 1x (25.4x 9/6) x 0.3 = 11.43x.

Function Menu and Interface Settings

Move the mouse to the top right of the display to access the function menu. Move the mouse to the right side of the display to bring up the interface setup menu.

Main Features

Interface Setting Options

The interface settings allow for detailed control over the image display and capture:

Image Adjustment Controls:

Image Manipulation Tools:

Grid and Line Tools:

Models: ED.3000 Blue Line Series Microscope, ED.3000, Blue Line Series Microscope

PDF preview unavailable. Download the PDF instead.

ED.3000-manual-academy EN Adobe Acrobat 21.5 Adobe Acrobat 21.5 Image Conversion Plug-in

Related Documents

Preview Euromex Delphi-X Observer User Manual
This user manual provides detailed instructions for the Euromex Delphi-X Observer microscope, covering its features, safe operation, assembly, maintenance, and troubleshooting for life science applications.
Preview Euromex Delphi-X Observer Metallurgical Microscope User Manual
Comprehensive user manual for the Euromex Delphi-X Observer metallurgical microscope, covering safety, configuration, operation, and troubleshooting.
Preview Euromex Oxion Inverso Fluo Microscope User Manual
Comprehensive user manual for the Euromex Oxion Inverso Fluo inverted fluorescence microscope, detailing setup, operation, maintenance, and troubleshooting for optimal performance.
Preview Euromex NexiusZoom Stereo Microscopes: Features, Specifications, and Accessories
Explore the Euromex NexiusZoom series of stereo microscopes, offering high precision, advanced optics, and versatile configurations for materials and life sciences. Details on ESD safe models, various stands, and comprehensive accessories.
Preview Euromex MZ.5000 Software System User Manual
Comprehensive user manual for the Euromex MZ.5000 software system, detailing its features, functions, measurement tools, calibration, and customization options for video integrated microscopes.
Preview Euromex VC.3042 UHD-4K Camera User Manual - Features, Configuration, and Operation
A comprehensive user manual for the Euromex VC.3042 UHD-4K microscope camera. This document details the camera's specifications, software integration (ImageFocusAlpha), various connection configurations (HDMI, USB, LAN, WLAN), user interface functions, measurement tools, and settings.
Preview Euromex PB.5275 Cleaning Kit: How to Keep Your Microscope Optics Clean
A step-by-step guide on how to effectively clean microscope optics using the Euromex PB.5275 cleaning kit, ensuring optimal image quality and longevity. Covers lens cleaning, air blowing, and fluid application.
Preview Euromex MZ.5000 Digital Macrozoom Microscope User Manual
Comprehensive user manual for the Euromex MZ.5000 digital macrozoom microscope, detailing setup, operation, maintenance, and troubleshooting for Life Sciences applications. Includes component identification and assembly instructions.