The LM5BA 5L Stationary Oxygen Concentrator has 3 filters. For optimal performance and longevity of the unit and to preserve the Manufacturer's Warranty, Rhythm Healthcare provides the following guidance for filter cleaning and replacement.
AIR INTAKE (SPONGE) FILTER
SKU LM5BA-ITF
The Sponge Filter is located on the back of the unit. It should be removed and cleaned on a weekly basis using warm soapy water. Use caution to COMPLETELY dry the filter before replacing. The Sponge Filter should also be replaced between patients. The filter should be replaced if it is torn or damaged.
[Description: A black, rectangular sponge filter with a black strap.]
INLET FILTER (HEPA FILTER)
SKU LM5BA-ILF
The Inlet Filter (HEPA) is located on the back of the unit. Inspect the HEPA Filter in between patients and replace the filter if needed, or every 3 years.
[Description: A grey, rectangular HEPA filter.]
BACTERIAL FILTER
SKU LM5BA-BF
The Bacterial Filter is located inside the LM5BA Concentrator. This filter should be changed during each compressor servicing.
[Description: A white, circular filter with a plastic nozzle.]
[Diagram Description: Orange lines indicate connections from the filter descriptions to specific locations on the rear of the white Rhythm Healthcare LM5BA 5L Stationary Oxygen Concentrator unit.]
[Description: A white, wheeled Rhythm Healthcare LM5BA 5L Stationary Oxygen Concentrator unit.]
Document Reference: RHTHE.0224.V2