PowerPac Basic Power Supply

Instruction Manual

For technical support, call your local Bio-Rad office, or in the U.S., call 1-800-424-6723.

Notice

No part of this publication may be reproduced or transmitted in any form or by any means, electronic or mechanical, including photocopy, recording, or any information storage or retrieval system, without permission in writing from Bio-Rad.

Bio-Rad reserves the right to modify its products and services at any time. This guide is subject to change without notice. Although prepared to ensure accuracy, Bio-Rad assumes no liability for errors or omissions, or for any damage resulting from the application or use of this information.

Bio-Rad is a trademark of Bio-Rad Laboratories, Inc. in certain jurisdictions. All trademarks used herein are the property of their respective owner.

Section 1: Introduction

1.1 Overview

The PowerPac Basic provides constant voltage or constant current to instruments used in electrophoresis. The power supply operates at the values specified for the constant parameter. However, to prevent damage to the electrophoresis cell, the PowerPac Basic provides automatic crossover to constant current or constant voltage, depending on which set value is first reached. When the set limit of the non-constant parameter is reached, and the power capability of the unit is not exceeded, the power supply will switch, making the non-constant parameter the new constant parameter.

Output Specifications:

Figure 1: Front View depicts the front panel of the PowerPac Basic, showing the display, control buttons (STOP, RUN/PAUSE, V/A, CONSTANT), and output jacks.

Figure 2: Rear View shows the rear of the unit, indicating the AC INLET, POWER SWITCH, and FUSE.

1.2 Unpacking

When receiving the power supply, inspect the container for shipping damage. If damage is suspected, file a claim with the carrier before contacting Bio-Rad Laboratories. After unpacking, remove the protective plastic film from the top case. Clean any residue with a soft, damp cloth.

Contents include:

Contact Bio-Rad Laboratories immediately if any part is missing or damaged.

Section 2: Control Features

Figure 3: Front Panel illustrates the controls:

Section 3: Setup and Operation

  1. Turn Power On: Press the power switch on the right side of the unit. The default setting is constant voltage (V), with the LED display showing zero. To view the firmware version, hold the constant parameter key while turning the power switch on.
  2. Connect Electrophoresis Cell(s): Connect the cell(s) to the power supply using the color-coordinated output terminals (red and black). A warning symbol [⚠] indicates high voltages. Power leads must be inserted perpendicular to the case curve.
    • Figure 5: Power leads connected correctly shows proper connection.
    • Figure 6: Power leads connected incorrectly shows improper connection.
  3. Select Constant Parameter: Press the Constant Parameter Key (V/A) to choose between voltage or current. The corresponding LED will illuminate. The display shows zero for the selected parameter.
  4. Enter Constant Parameter Value: Use the scroll keys to set the desired value. Voltage is adjustable from 10 to 300V in 1V increments. Current is adjustable from 4 to 400mA in 1mA increments.
  5. Enter Limit Parameter Value: Use the parameter key to select the limit parameter. The default maximum value (300V or 400mA) is displayed. Set a suitable limit value to prevent excessive power conditions.
  6. Programming a Timed Run: Use the parameter key to select time. Use the scroll key to enter the desired time, up to 999 minutes. If no time is entered, the run continues until stopped. A timed run displays remaining time; an untimed run displays elapsed time.
  7. Power Failure Detection (Optional): Available for timed runs only. To activate, ensure the parameter key is in time mode, enter the desired time, then hold the STOP key for approximately 2 seconds until "Pfd" is displayed. This feature automatically deactivates after a run.
  8. De-activation of Change in Resistance Feature (Optional): For applications with resistance fluctuations, this feature can be deactivated to allow uninterrupted runs. Ensure the parameter key is in current mode, then hold the STOP key for approximately 2 seconds until "dE9" is displayed. Caution: Deactivating this feature may increase the risk of electrical hazard.
  9. Start the Run: Press the RUN/PAUSE key. The run LED will illuminate.
  10. Viewing and Editing During a Run: Press the parameter key to view values. Edit constant parameter and time values (for timed runs) in pause mode. Limit values can also be edited in pause mode. To switch from untimed to timed, stop and reprogram.
  11. Pause Mode: Press the RUN/PAUSE key during a run. The pause LED will light, allowing safe adjustments, editing of all parameters, or changing the constant parameter.
  12. End of Run: When a run completes (timed end or untimed stop), parameter values are preserved, and the timer resets. The run/pause LEDs will be off, indicating no power to output jacks.
  13. Terminating a Timed Run: Press the STOP key. Parameter values are preserved, and the timer resets. The run/pause LEDs will be off.
  14. Powering Down: Press the STOP key before turning the power switch off. Failure to do so may trigger a power failure detection and automatic restart if enabled.

Section 4: Maintenance and Troubleshooting

4.1 Maintenance

The PowerPac Basic requires minimal maintenance. To clean the case, unplug the unit and wipe with a damp cloth.

4.2 Troubleshooting

ProblemCauseSolution
No display/lights/fan1. No AC power.
2. Blown fuse.
3. Power switch cycled rapidly.
1. Check power connection and switch.
2. Replace fuse (see 4.3).
3. Reset unit: turn off, wait 5-10 seconds, turn on.
Repeated blown fusesHardware failureContact Bio-Rad Technical Resources.
Leads from cell not long enough to fit output jacksOutput terminals are recessed; some leads are too short.Use the PowerPac Adaptor for secure connection. (Note: Use of adaptor voids IEC 61010-1 provisions).
E1 error codeNo load detected (Instrument not connected or current < 4mA)Verify electrical connections and buffer levels.
E2 error codeOver current (load current > 400mA)Check for short circuits or excessive load. Remake buffer if needed. Press RUN/PAUSE to resume or any other key to clear.
E3 error codeOver voltage (load voltage > 300V)Turn power supply off, then on to reset. Contact Bio-Rad if persistent.
E5 error codePower failure during timed run with PFD activated.PFD is automatically deactivated after run.
E6 error codePower failure during timed run without PFD activated.Run not completed.
E7 error codePower failure during untimed run or power switch used instead of STOP. PFD not available for untimed runs.Run not completed.
E8 error codeRegulation errorTurn power supply off, then on to reset. Contact Bio-Rad if persistent.
E9 error codeChange in Load Resistance (detects drastic changes indicating potential lead/connection failure)Check/correct resistance issue. Press RUN/PAUSE to resume or any other key to clear. (Feature can be deactivated as per Section 3, Step 8).
E10 error codeUnacceptable value(s) enteredPress any key (except RUN/PAUSE) to clear. Enter valid values and press RUN/PAUSE.
E14 error codeInternal Over VoltagePossible malfunction. Check connections, wires, buffer concentration. Press any key (except RUN/PAUSE) to clear.
E15 error codeInternal Short CircuitPossible malfunction. Check connections, wires, buffer concentration. Press any key (except RUN/PAUSE) to clear.
E17-E19 error codesHardware failureContact Bio-Rad Technical Resources.
E20 error codeInstrument is overheatingCheck fan and vents for obstruction.
E21 error codeRegulation errorTurn power supply off, then on to reset.
E22 error codePower fail while runningPress STOP, then cycle power.
E23 error codeExceeded voltage limit settingPossible malfunction. Check connections, wires, buffer concentration. Press STOP to clear.
E24 error codeExceeded current limit settingPossible malfunction. Check connections, wires, buffer concentration. Press STOP to clear.
E25 error codeExceeded wattage limit settingPossible malfunction. Check connections, wires, buffer concentration. Press STOP to clear.

4.3 Replacing a Fuse

If the unit shows no display, lights, or fan, and is powered on:

  1. Disconnect the power cord.
  2. Use a flat blade screwdriver to release the fuse holder in the power entry module.
  3. Replace the fuse with a 2.5A, 250V, 5 x 20 mm, fast-blow fuse (Bio-Rad catalog number 12007433).
  4. Re-insert the fuse holder.

Note: Repeated fuse blowing indicates hardware failure; contact Bio-Rad for repair.

4.4 Firmware Version Number

To display the firmware version number with the power supply off: Hold down the constant parameter key and turn the power switch on. Release the key when all LEDs light up. The firmware version will appear briefly.

Appendix A: Specifications

1 Fuse specifications apply to products with serial number 041BR300000 and higher.

Appendix B: Warranty and Ordering Information

Warranty

The PowerPac Basic is covered by a standard Bio-Rad Laboratories warranty. Contact your local Bio-Rad representative for details. Bio-Rad will replace defective parts free of charge during the warranty period. Exclusions include defects from improper operation, unauthorized repair/modification, use of non-specified cables/connectors, misuse, or damage from disaster. For service, contact your local Bio-Rad office.

Warranty Information

Ordering Information

Catalog NumberDescription
164-5050PowerPac Basic power supply, 100–240 VAC
165-5061PowerPac Adaptor, qty 1
165-5066PowerPac Adaptor, qty 2
12007433Replacement Fuse, 2.5 A, 250 V, 5 x 20 mm, fast-blow, qty 1

3 Fuse specifications apply to products with serial number 041BR300000 and higher.

PDF preview unavailable. Download the PDF instead.

4006213 Acrobat Distiller 10.0.0 (Windows)

Related Documents

Preview Bio-Rad Mini-PROTEAN II Electrophoresis Cell Instruction Manual
Instruction manual for the Bio-Rad Mini-PROTEAN II Electrophoresis Cell, covering setup, operation, maintenance, troubleshooting, and specifications.
Preview Bio-Rad Protein Blotting Guide: Comprehensive Resource for Transfer and Detection
Explore the Bio-Rad Protein Blotting Guide, a comprehensive resource detailing methods, equipment, and reagents for protein transfer and detection. Includes troubleshooting tips for western blotting and related techniques.
Preview PROTEAN® IEF Cell Instruction Manual
Instruction manual for the PROTEAN® IEF Cell, detailing its features, setup, operation, programming, and maintenance. Includes specifications and safety guidelines.
Preview Bio-Plex Manager Software 6.2 User Guide - Bio-Rad
Comprehensive user guide for Bio-Rad's Bio-Plex Manager Software 6.2, detailing installation, operation, data analysis, and security features for the Bio-Plex Suspension Array System. Learn to manage laboratory assays and ensure compliance with 21 CFR Part 11.
Preview Bio-Rad PR 4100 Absorbance Microplate Reader: Datasheet and Features
Comprehensive product overview of the Bio-Rad PR 4100 Absorbance Microplate Reader, detailing its advanced features for accurate and fast measurements, innovative design, data analysis capabilities, technical specifications, and system integration for ELISA tests.
Preview Bio-Rad Buffer Formulations for Electrophoresis and Sample Preparation
A comprehensive guide to buffer formulations from Bio-Rad Laboratories, including SDS-PAGE, Native PAGE, Tricine, RIPA, PBS, and gel casting buffers with detailed recipes and catalog numbers.
Preview iMark™ マイクロプレートリーダー 取扱説明書 - Bio-Rad
Bio-Rad iMark™ マイクロプレートリーダー取扱説明書。モデル168-1130JA、168-1135JAなどに対応。セットアップ、操作、ソフトウェア、トラブルシューティング、仕様に関する詳細ガイド。
Preview Bio-Rad EP-1 Econo Pump Instruction Manual
Instruction manual for the Bio-Rad EP-1 Econo Pump, covering its features, operation, installation, and troubleshooting.