V+Pro COPD Quick Reference Guide

React Health

Key Features

  • Invasive and Non-invasive modes ⚕️
  • Easily Transportable ?
  • Up to 9 hours of battery life ?
  • Record, Trend and Monitor ?

Contents

  • Recommended settings
  • Predicted body weight to tidal volume charts
  • Flow cycle guide

Example Presets

Example Preset 1: Non-Invasive Ventilation

Setting Value
ModeVol. Targeted-PS
Patient TypeAdult
HumidificationHME or Humidifier
Circuit TypeValveless
Breath Rate12 BPM
Inspiratory Time1.0 Seconds
Tidal Volume7-8 mL/kg (IBW)
Pres. Adj. RateSlow
Pres. Minimum8 cmH2O (above set PEEP)
PEEP5 cmH2O
Flow Trigger3.0 L/min
Flow Cycle465%
Time Cycle1.5 Seconds
Rise Time2 or 3
Apnea Rate10 Backup BPM
Alarms(set Alarms according to your Protocol)
Inspiratory Pressure(High) 35 cmH2O

Example Preset 2: High Flow Therapy

Setting Value
ModeSIMV-Pressure
Patient TypeAdult
High FlowON
Flow230 L/min
Alarms(set Alarms according to your Protocol)
High Insp. Pressure(set Alarms according to your Protocol)
Patient Circuit Discon.(set Alarms according to your Protocol)
High Pressure Delay(set Alarms according to your Protocol)

1 The settings referenced in this guide are suggestions and do not replace physician orders; please consult with patient physician to make sure that all settings are appropriate for the patient.

2 Adjust flow to achieve target SpO2 and/or CO2 targets.

3 Adjust to patient comfort.

4 See Flow Cycle Guide (pg. 4).

Predicted Body Weight (PBW) / Tidal Volume Charts

Males

Height PBW Tidal Volume (mL/kg)
45678
4'10"45.5182227272318363
4'11"47.7191239286334382
5'0"50200250300350400
5'1"52.3209262314366418
5'2"54.6218273328382437
5'3"56.9228285341398455
5'4"59.2237296355414474
5'5"61.5246308369431492
5'6"63.8255319383447510
5'7"66.1264331397463529
5'8"68.4274342410479547
5'9"70.7283354424495566
5'10"73292365438511584
5'11"75.3301377452527602
6'0"77.6310388466543621
6'1"79.9320400479559639
6'2"82.2329411493575658
6'3"84.5338423507592676
6'4"86.8347434521608694
6'5"89.1356446535624713
6'6"91.4366547548640731

Females

Height PBW Tidal Volume (mL/kg)
45678
4'7"34136170204238272
4'8"36.3145182218254290
4'9"38.6154193232270309
4'10"40.9164205245286327
4'11"43.2173216259302346
5'0"45.5182228273319364
5'1"47.8191239287335382
5'2"50.1200251301351401
5'3"52.4210262314367419
5'4"54.7219274328383438
5'5"57228285342399456
5'6"59.3237297356415474
5'7"61.6246308370431493
5'8"63.9256320383447511
5'9"66.2265331397463530
5'10"68.5274343411480548
5'11"70.8283354425496566
6'0"73.1292366439512585
6'1"75.4302377452528603
6'2"77.7311389466544622
6'3"80320400480560640

5 Source: http://www.ardsnet.org/files/pbwtables_2005-02-02.pdf

Flow Cycle Guide

This section illustrates the concept of flow cycling in ventilation.

Diagram Description:

The diagram shows a graph with Pressure (cmH2O) on the vertical axis and Time (sec) on the horizontal axis. A ventilation breath waveform is depicted, starting from a baseline pressure, rising to a peak, and then descending. Two points on the descending phase are marked:

  • One point, labeled "Flow Cycle = 90%", indicates the point where the inspiratory flow has decreased to 90% of its peak value.
  • Another point, labeled "Flow Cycle = 20%", indicates the point where the inspiratory flow has decreased to 20% of its peak value.

A dashed line segment represents the "Time Cycle", indicating a duration on the time axis. The diagram visually explains how flow cycling terminates inspiration based on a percentage of peak flow.

Note: This Quick Reference Guide contains limited information. Please refer to the VOCSN Clinical & Technical Manual for detailed use guidance.

Document Revision: LBL-00333-000 Rev B

Models: LBL-00333-000, LBL-00333-000, V Plus Pro COPD Device, V Plus Pro, COPD Device

PDF preview unavailable. Download the PDF instead.

LBL-00333-000 QRG Volume Targeted Flow Rev B[28017]- V Pro COPD Quick Reference Guide(1) Adobe PDF Library 16.0.3 Adobe InDesign 17.0 (Windows)

Related Documents

Preview React Health Stratus 5 Oxygen Concentrator User Manual
Comprehensive user manual for the React Health Stratus 5 Oxygen Concentrator, detailing safety, operation, maintenance, and specifications for effective use.
Preview React Health Luna G3 CPAP/APAP Patient Quick Set-Up Guide
A quick guide to setting up and using the React Health Luna G3 APAP (LG3600) and Luna G3 CPAP (LG3500) devices, covering button functions, device setup, adjusting settings, starting treatment, and understanding the usage score.
Preview iCodeConnect Quick Start Guide for Physicians - React Health
A quick start guide for physicians on using React Health's iCodeConnect system for seamless sleep report management, including registration, reporting methods, and remote setting changes.
Preview Stratus 5 Filter Guide - React Health
An informational guide detailing the Stratus 5 Sponge Filter (STR1030), HEPA Paper Filter (STR1020), and Bacteria Filter (STR1040) for React Health oxygen concentrators, including maintenance and replacement information.
Preview React Health 3B Luna QR App Instructions: Scan and Sync Sleep Data
Step-by-step guide to download and use the React Health 3B Luna QR App for syncing sleep therapy data from Luna II and Luna G3 devices. Learn how to scan QR codes for seamless data transfer and review.
Preview React Health G3 X: The First Standard Size Foam-Free PAP Device
Introducing the React Health G3 X, the first standard-size, foam-free PAP device in the U.S. Learn about its advanced features like Smart A Custom Pressure Response, a 5-year warranty, and seamless care management for effective sleep therapy.
Preview Luna G3 APAP/CPAP Patient Quick Set-Up Guide | React Health
Learn how to set up and operate your React Health Luna G3 APAP (LG3600) or CPAP (LG3500) device with this quick operation manual. Includes button functions, setup instructions, and usage scoring.
Preview 3B Luna QR App Instructions: Scan Your Sleep Data
Step-by-step guide to downloading and using the 3B Luna QR App for React Health CPAP and BiPAP devices to access and review sleep data.