BIOLAB

OPERATING MANUAL

DRYING OVEN

BODR-404

CE ISO9001 CERTIFIED

www.biolabscientific.com

INDEX

  1. Safety Prompting
  2. Brief Introduction
  3. Operational Instructions
    1. Preparation
    2. Power on
    3. Operation steps
    4. Ways to improve the accuracy of temperature control
    5. Function parameter table
    6. Usage of "Over-temperature Protector" (optional)
  4. Maintenance and Notes
  5. Appendix
    1. Technical Specifications
    2. Troubleshooting
  6. Packing List

1. Safety Prompting

Matters prescribed herein are critically important, which must be practically observed.

! Danger (likely to cause serious property losses or human casualties)

  1. The Product must be reliably grounded (without taking the zero line or neutral line as the earth wire).
  2. Before using, make sure to confirm the compliance of power supply voltage and frequency with the product requirement.
  3. Use a separate power supply socket for the product and confirm the plug and socket are properly grounded.
  4. It is not allowed to pull out or plug in the power supply plug at random without switching off the power when the product is in operation.
  5. It is not allowed to extend or cut short the power supply connection wire of the product at random.
  6. It is prohibited to place in any inflammable, explosive, evaporative and erosive substance for drying or curing.
  7. It is prohibited to touch the door, OBW and ambient surface with the product operating at a temperature above 80°C to avoid any scald.
  8. Do not put hands or other articles into the air inlet/outlet (slot).
  9. It is prohibited to repair without permission and in case of being consigned by the Company for any repair, maintenance shall be carried out by the professionals.

! Warning (Unauthorized repairing may cause damage or injury )

  1. Make sure to read and understand adequately the operating instructions of the Product before operating.
  2. Since 304 stainless steel inner chamber is not acid-resisting, make sure to take corrosion-proof measures and never use any acid media in the case!
  3. Make sure not to directly drag the power cable when pulling out the power supply plug.
  4. In case of any one of the following circumstances, make sure to pull out the power supply plug of the Product:
    1. When replacing the fuse tube.
    2. When checking and repairing the product with fault.
    3. When keeping the product idle for long.
    4. When moving the product.
  5. Do initiate the upper deviation alarm function after power on.

! Attention (likely to affect the useful life and cause the abnormal operation of the product)

  1. The product shall be placed on a hard and firm plane so that it will remain in a level status.
  2. Certain gap shall be kept around the product.
  3. The product must be used under the specified conditions.
  4. Make sure not to open/close the product door by force, otherwise, it may easily cause the door to drop off, damage the product and result in an injury accident.

www.biolabscientific.com

2. Brief Introduction

1. Structure and Functions

This system simulates natural environment of high temperature and is widely used in high temperature adaptability testing (especially on changes to electric performance and mechanical performance) of electronic, electric products and components.

Product Design:

  1. Auto adjustment of temperature can ensure precise temperature control.
  2. Artificial-intelligence controller.
  3. The system contains over temperature, overload and short circuit protection mechanism, which can automatically shut down affected equipment and initiates alert. Safety is guaranteed.
  4. The inner tank is made of stainless steel, which is aesthetic and is easy to clean.
  5. The housing of the system is made of electrostatic sprayed steel with smooth finishing for aesthetic feeling.

Note: This system is tested and used in high-temperature environment. The inner tank, shelf board and door seals carry color changes. This does not affect the functions of the system.

3. Operational Instructions

1. Preparation

Operation Environment:

  1. Ambient Temperature: 5°C~40°C.
  2. Relative humidity is not higher than 85%.
  3. Atmospheric Pressure:(86~106)KPа.
  4. Altitude is not higher than 2000m.
  5. No strong vibration source and strong electro-magnetic fields around.
  6. The product should be placed horizontally and stably in indoor environment without severe dust pollution, direct sunlight and corrosive gas.
  7. Keep sufficient clearance around the system as indicated in the figure. Do not place the system under a fire alarm.
  8. See technical specifications for power voltage (Table 3 of Schedule 1).
  9. Rationally place and adjust the shelf boards. Ensure sufficient upper/lower and surrounding clearance to promote smooth air circulation! (>100mm), Weights on the board should not bend the board.

www.biolabscientific.com

2. Power on

Figure 1 shows the recommended clearance around the drying oven.

  1. Close the door of the box, the handle should be vertically downward.
  2. Turn on the power and the indicator light is on.
  3. The controller enters the working mode after about 4 seconds of self-checking procedure.

Figure 2 illustrates the control panel and indicator lights.

2.4 Indicator light description:

  1. TEMP area (PV): display the measured temperature.
  2. SET area (SV): display the set temperature.
  3. Heating lamp: the lamp is on when there is heating output.
  4. Fan indicator light: on when there is fan output.
  5. TIME: Time display window; display running time or parameter value.

2.5 Button description

  1. SET button: used to modify the set value or enter the internal parameter setting, in the parameter setting state, long press the set button for more than 3 seconds to exit.
  2. Shift key: used to shift the set value, internal parameters and view the ambient temperature.
  3. Decrease key: used to modify the set value, various parameters, or start/stop auto-tuning.
  4. Increase key: used to modify the set value, internal parameters, or view the remaining period (X)
  5. RUN/STOP: Press for 3 seconds to run or stop the controller.

2.6 Check temperature control accuracy

  1. Select a digital thermometer with a resolution of 0.1°C that has passed the verification and is within the verification period as a standard and put it into the product work room, and ensure that the sensor temperature sensor is in the geometric center of the work room.
  2. Choose a point within the temperature control range of the product and set the SV temperature control value. When the PV measurement value is equal to the set value, keep it at a constant temperature (1∼2) hours or so (depending on the product specifications, the constant temperature time may vary), Observe that the difference between the actual measured temperature value of the standard thermometer and the measured value displayed by the controller PV should be ≤±0.5°C.

3. Operation steps

  1. Temperature setting: Press the "SET" button once, the value of the SET window flashes, indicating that the temperature can be set as required, and the required temperature can be set through the "increase", "decrease" and "shift" keys. Press the "SET" key again to return to the standard display mode.
  2. Timing function: Press the "SET" button twice, when the time is set to 0, there is no timing function; when the time is set to 0, the controller has a timing function, press the "SET" button, the TIME value flashes, indicating the time can be set as required. Use the "increase", "decrease" and "shift" keys to set the required time value. When the time is up, the TIME window displays the "END" buzzer, and you can press any key to silence it.

Note: ① Each time you modify a parameter, you need to press the "SET" key to confirm and the modification is effective. ② After setting all the parameters, press the "RUN/STOP" key and wait for about 3 seconds to start running.

4. Ways to improve the accuracy of temperature control

4.1 After the product has been used for a period of time, the temperature control accuracy should be checked according to the method 2.6, if it exceeds ±1°C, it can be corrected according to the following method:

4.1.1 Enter the parameter setting, find the symbol.

Press PK = 4000 × (Standardable value – Measured value PV) / Measured value PV

After the formula is calculated, it shall be modified on the basis of the original PK value at the time of delivery (Note: One correction is not accurate, and the correction can be repeated until it meets).

www.biolabscientific.com

5. Function parameter table

In the standard state, press the SET key and the shift key at the same time for more than 3 seconds, the LCD screen displays the LK code, and you can enter the password setting level interface.

1)When LK is displayed in the output area of the LCD screen, press the plus or minus key or the shift key to make LK=0000, and press the SET key to enter the user parameter level setting.

Symbol Name Setting range Description Factory set value
Pn Work group in operation 0~8 For program control only. To set up the work group for operation of meters. When Pn is set as 8, Group 8 is for fixed value control. Optional
Cy Number of cycle in a period 0~99 Special parameter for program control. When CY is 0, the meter will run between the work group all the time. When CY is not 0, the meter will shut down automatically after CY in the group. Optional
dy Appointment boot selection 0-200::59 0: No appointment; for other values, after pressing the run key to start, it will automatically delay the dy time and then start the machine. Optional
ut UV time 0-200 minute Turn off the UV lamp after ut time, ut=0, turn off the UV lamp manually. Optional
us UV switch 0-1 0: Turn off the UV light 1: Turn on the UV light. Optional

Table 1

2) When LK is displayed in the output area of the LCD screen, set LK=0003, and press the SET key to enter the equipment manufacturer's parameter level setting.The parameter hierarchy menu of the equipment manufacturer is as follows:

Symbol Name Setting range Description Factory set value
tM Setup of maxi temperature permissible by the instrument full range Stop heating beyond maxi temperature and give alarm.
Po Boot mode 0~2 ①When PO =0, after open the power, the controller in a stopped state, by long press star/stop key is up and running. ②When PO =1, after open the power, the controller will be running. When PO =2, running from last power began to run.
AL Setup of alarm 0~100.0 When the temperature exceeds the value of SP+AL, the alarm light is on and the alarm is output (with HOLD function).
Pb Zero adjustment (intercept) -100.0~100.0 When the zero error of the instrument is greater and the full scale error is smaller, the value should be adjusted. As a rule with Pt100 the value is seldom adjusted.
PK Adjustment of full scale (slope) -1000~1000 S When the zero error of the instrument is smaller and the full scale error is greater, the value should be adjusted. PK=4000 × (specified value -actual display value)/actual display value and as a rule with Pt100 the value is adjusted first.
PA Onboard room temperature sensor correction -30-30 When there is an error between the on-board room temperature sensor and the actual situation, adjust the value.
2b The second zero correction -100-100 When the second channel zero error is large and the full scale error is small, adjust this value. Generally, Pt100 rarely adjusts this value.
2K Second channel full scale adjustment -1000~1000 When the zero error of the second channel is small and the full-scale error is large, adjust the value. PK=4000× (mercury thermometer value-display value)/display value, generally Pt100 adjust this value first.

Table 2

*The products have been strictly tested before leaving the factory. When the technical indicators meet the requirements and work normally, no correction is generally required.

6. Usage of "Over-temperature Protector” (optional)

The over temperature protector is a separate protection system. When the temperature controller encounters malfunction and causes temperature out of control and when the temperature in the working room reaches the set value, the overheat protector will cut off the heating and initiate the alarm.

(See Fig. below) working room temperature is lower than set temperature limit The protection system is released after set value and the meter is restored. Cycle repeats until problem is solved.

5.1 Set value is higher or equal to (SV+AL) + (10~15)°C

5.2 See Fig. 6, Use + and — button for over temperature setting on control panel.

To set desired temperature limit.

Example: SV=130°C, AL=10

Set the over temperature to 150°C.

Figure 3 shows a temperature graph illustrating operating temperature versus set temperature limit.

Figure 4 shows the control panel with the ALARM light and SET button.

www.biolabscientific.com

4. Maintenance and Notes

  1. Turn off the power after each use. After testing in the environment with temperature higher than 300°C, first turn off the power and then turn off the electric motor after twenty minutes. Disconnect the three-pin plug after high temperature testing. Be aware of the heat and avoid being burnt when opening the door and taking out articles from the cabinet.
  2. Clean the inside and outside of the system, disconnect the plug and cover the system with plastic dust guard if the system will not be used for a long period.
  3. Clean the testing cabinet with a vacuum cleaner or blow the system with compressed air once a month.
  4. Power on the system to drive away moisture regularly (around once a month) if the system is placed in a high-humidity environment.
  5. Check and calibrate the control precision before reuse or in case of requirements change.
  6. Modification to parameters other than TEMP, TIME, would require consent of out Customer Service Center or should be implemented by authorized personnel.

www.biolabscientific.com

5. Appendix

1. Technical Specifications

This product is manufactured pursuant to Q/TIWY 1—2007

Model Indicator Item BODR-7101 BODR-401 BODR-7102 BODR-402 BODR-7103 BODR-403 BODR-7104 BODR-404
Supply Voltage AC 380V 50/60Hz
Temperature Control Range AH:RT+20~400°C/ BH:RT+20~500°C
Temperature Resolution/Constant Temperature Fluctuation ±0.5°C
Input Power (W) 3250 4050 4050 4900 4900W 6050W 8500W 8500W
Inner Tank Dimensions W×D×H(mm) 350*350*400 450*450*450 600*600*600 980*1000*780
Product Dimensions W×D×H(mm) 890*700*920 990*790*990 1140*950*1140 1324*1263*1770
Shelf board 2 pcs

Table 3

Note: 1. RT: room temperature 2. The above data is based on measurement with a mercury thermometer with 0.1°C resolution in ± the environment of 25°C and 85% relative humidity 0.1°C (the head is placed in the center of the environment).

2. Troubleshooting

Problem Cause Action
No power (Power indicator is not on.) Power outage of socket or poor contact with the plug Repair
Power cord is broken or poorly plugged Repair, replug
power switch is broken (or no turned on) Replace the switch or turn on the switch
Fuse is broken Check the switch, electric motor, heater, temperature controller and other parts for short circuit or 0 insulation resistance, and start the system after repairing.
The screen displays " " Sensor is damaged or connection wires breaks (or fall off) Repair or replace Pt100
Temperature does not grow Timing function is initiated? And set time is up See "timing function”
Temperature controller is broken (no output) HEAT indicator does not light up or 3061 broken, replace
TRIAC has no connectivity Replace the TRIAC (model: BTA16 or BTA26)
Inaccurate temperature control (big static error) or The temperature keeps growing and becomes uncontrollable (The buzzer sounds) Phase loss of three-phase power, or contactor failure Check the three-phase power and switch; turn of the switch again, replace the contactor
Heating tube wires disconnect or short circuit Repair or replace
HEAT indicator does not light up, and the temperature keeps growing SCR is broken, replace it SCR is broken, replace it
The fans are not turned on or damaged (does not rotate) Turn on or replace the fans Turn on or replace the fans
Gap between room temperature and set temperature is too small Minimum temperature controllable is RT + 10°C Minimum temperature controllable is RT + 10°C
Poor Pt100 contact, causing bigger resistance. Reconnect the wires Reconnect the wires
Incorrect settings of Ar and P parameters Reset parameters Reset parameters
Incorrect Pb and Pk Reset Pb and Pk Reset Pb and Pk
A group of heaters or heating plates are damaged (Only applicable to model 9420 or 9620) Replace Replace
Abnormal noise or louder noise Fan bearings is broken or lack of lubricating oil Refill lubricating oil, replace fan
Rubbing back air channel boards Repair (add washer)

Table 4

www.biolabscientific.com

Packing List

Product Name: High-temperature Dry Oven

No Type Item Unit Quantity Remarks
1 Document User Manual 1
2 Document Packing List share 1
3 Accessories Shelf board pcs 2

Packaging list conforms to articles in the package

Packed by: 2 Tested by: 1

Models: BODR-404, BODR-404 Drying Oven, BODR-404, Drying Oven, Oven

File Info : application/pdf, 12 Pages, 448.97KB

PDF preview unavailable. Download the PDF instead.

bodr-404 manual biolab

References

mPDF 8.1.2

Related Documents

Preview BioLAB BONC-101 Natural Convection Oven Operating Manual
Comprehensive operating manual for the BioLAB BONC-101 Natural Convection Oven, detailing its scope of application, technical specifications, structural description, operating methods, precautions, and troubleshooting.
Preview Biolab Natural Convection Oven Operating Manual - BONC-5601, BONC-5602, BONC-5603, BONC-5604
Operating manual for Biolab's Natural Convection Ovens, models BONC-5601, BONC-5602, BONC-5603, and BONC-5604. Covers installation, safety, operation, specifications, maintenance, and troubleshooting.
Preview Biolab Vacuum Oven Operating Manual - BOVA-5801 to BOVA-5805
Comprehensive operating manual for Biolab Vacuum Ovens, models BOVA-5801, BOVA-5802, BOVA-5803, BOVA-5804, and BOVA-5805. Covers installation, safety, operation, maintenance, and troubleshooting.
Preview BioLAB BAVT-2904 Vertical Autoclave Operating Manual
Comprehensive operating manual for the BioLAB BAVT-2904 Vertical Autoclave, covering introduction, structural characteristics, technical parameters, installation, usage, maintenance, and troubleshooting.
Preview BioLAB BLAR-407 Laboratory Refrigerator Operating Manual
Operating manual for the BioLAB BLAR-407 Laboratory Refrigerator, covering safety instructions, installation, operation, maintenance, and troubleshooting.
Preview BioLAB BMGS-104 Magnetic Stirrer Operating Manual
Operating manual for the BioLAB BMGS-104 Magnetic Stirrer, detailing its specifications, preparation, operation, and troubleshooting.
Preview Biolab BHTP-601 Tissue Embedding System Operating Manual
This manual provides comprehensive operating instructions for the Biolab BHTP-601 Tissue Embedding System, covering installation, operation, maintenance, and troubleshooting.
Preview BioLAB BINC-701 Platelet Incubator Operating Manual
Operating manual for the BioLAB BINC-701 Platelet Incubator, covering safety instructions, installation, operation, maintenance, and troubleshooting.