WB-T Transparent Water Bath

User Manual

Preface

Thank you for purchasing our Heating Mantle. Users should read this manual carefully, follow the instructions and procedures, and beware of all the preventive measures when using this instrument.

Service

If help is needed, you can always contact your dealer or Labbox via www.labbox.com. Please, provide the customer service representative with the following information:

Warranty

This instrument is guaranteed to be free from defects in materials and workmanship under normal use and service, for a period of 12 months from the date of invoice. The warranty is extended only to the original purchaser. It shall not apply to any product or parts that have been damaged due to improper installation, improper connections, misuse, accident, or abnormal conditions of operation. For claim under the warranty, please contact your supplier.

1. Safety Instructions

When working, wear the necessary personal protective equipment ?️ to avoid the risk of:

2. Product Description

2.1. Summary

This product is suitable for distillation, concentration, drying and thermostatic heating in medical units, universities and schools, scientific research units, and laboratories of industrial and mining enterprises like chemical painting and dyeing enterprises or pharmaceutical enterprises.

2.2. Structure Features

2.3. Cleaning and Maintenance

2.4. Main Technical Parameters

Function Parameter Value
Function Voltage (V) 200-240 / 100-120
Power (W) 800
Sensor Pt100
Temperature Motion (°C) ±0.5
Temperature Range (°C) 0~100
Temperature Sensitivity (°C) ≤±1
Time Range 0~9999 min or hours
Measurement error < 0.3%
Structure Shelves Height (mm) Distance from bottom 100mm
Shelves Holes Diameter 5 X104mm
Inner Chamber Size (L*W*H) 280 x 220 x150 mm
Exterior Size (L*W*H) 400 x 260 x 190 mm
Packing Size (L*W*H) 450 x 350 x270 mm
N.W. (KG) 3.5
G.W. (KG) 5
Working conditions Environmental temperature (°C) 5~40
Relative humidity RH < 85%

2.5. Control Monitor

Indicators

Keys

Screen

3. Operation

  1. Place the water bath on a stable surface that can support the weight of the device and leave 30 cm clear on each side.
  2. Fill the tank to 2/3 of the total volume.
  3. Connect to a power supply that is consistent with the voltage requirement of this instrument. Make sure it is properly earthed .
  4. Switch the device on. The power supply indicator will light up.
  5. Set the desired temperature (see instructions below). The temperature controller will display the temperature inside the working chamber.
  6. If desired, set the time (see instructions below).

Note: The thermostatic bath cannot be used with deionized, demineralized, or distilled water as it will not conduct electricity properly.

4. Temperature and Time Setting

4.1. No Timer Function

  1. Press the SET button to enter the setting state SP and the temperature set value will show.
  2. Use to modify the value.
  3. Press SET to enter; the controller will return to the normal display and the value will be saved.

4.2. Timer Function

  1. Press SET to enter the setting state SP and the temperature set value will show.
  2. Press SET again; the indicator SP and the time set value will show.
  3. Use to modify the value.
  4. Press SET to return to the initial screen. The value will be saved automatically.

When the time is set to 0, the controller will run continuously until manually stopped. Otherwise, when the timer is set, the lower window will display the temperature set value or the running time depending on the ndt parameter configuration (see later). When the time is over, the lower window will display "End" and the alarm will go off and ring for 60s. Stop it by pressing any button. Press the RST button for 3s to restart the controller.

5. Self-tuning of PID

When the temperature control effect is not ideal, a system self-tuning can be carried out where the equipment will automatically configure its PID parameters. To do that, follow the next instructions:

6. Internal Parameters Setting

6.1. How to Enter

To adjust the internal parameters, you must enter the required password. To do that, follow these instructions:

  1. Press SET for 3s Lc; the controller will display the password prompt Lc.
  2. Use to choose the required value (see the password values below).
  3. Press SET to enter; the controller will enter the internal parameters setting state.

To return to the regular state at any time, press SET for 3s.

6.2. Parameter Tables

INDICATOR NAME INSTRUCTIONS OF THE FUNCTION (RANGE) DEFAULT VALUE
Lc Password The password to modify the parameters in this table is Lc = 3 0
ALH Over-temp alarm If SV > (SP + ALH), the ALM light turns on. The buzzer sounds and the heating output turns off. (0~100.0°C) 20.0
ALL Under-temp alarm If SV < (SP – ALL), the ALM light flashes, the buzzer sounds. (0~100.0°C) 0
T Control cycle The temperature control cycle. (1~60S) 5
P Proportional band Adjustment of proportional function. (1~400.0°C) 35.0
I Integration time Adjustment of integration function. (1~2000S) 200
d Differential time Adjustment of differential function. (0~1000S) 200
Pb Zero point adjustment When the error is comparatively larger, this value might need adjustment. Pb = actual value – measure value (-12.0~12.0°C) 0
PK Large value adjustment Correction of errors caused by sensor high temperature values. PK = 1000 × [ (actual value - measured value) / measured value ] (-999~999) 0
INDICATOR NAME INSTRUCTIONS OF THE FUNCTION (RANGE) DEFAULT VALUE
Lc Password key The password to modify the parameters in this table is Lc = 9 0
ndA Temp alarm mode 0: With over-temp alarm only.
1: With over-temp alarm and under-temp alarm at the same time.
(0~1) 0
ndt Timer mode 0: No timer function.
1: With timer function, the lower window displays the running time after the measured temperature reaches to the set value.
2: With timer function, the lower window always displays the setting value.
(0~2) 1
Hn Timer unit 0: Minutes
1: Hours
(0~1) 0
EH Timer end mode 0: Continue to maintain the constant temperature when the running time is over.
1: Stop the temperature control when the running time is over.
(0~1) 0
oPn Auxiliary parameter 0: ON
1: OFF
(0~1) 0
nP Maximum power Percentage of maximum power heating output. (0~100%) 100
Co Off point If SV > (SP + Co), stop the heating output forcibly. (0~100.0°C) 50.0
SPL Minimum set point Adjustment of the minimum temperature. (0~0) 0
SPH Maximum set point Adjustment of the maximum temperature. (0~400.0) 300.0

7. Fault Analysis

FAILURE CAUSE HANDLING METHOD
No power supply Bad contact between plug and socket
The fuse is burnt
Replace the plug or socket tube
Replace the fuse with one with the same specifications
No temperature rise The temperature controller is broken
The sensor is broken
The set temperature is lower than water temperature
Replace the instrument
Replace the sensor
Reset the temperature
There is a big difference between display temp. and actual temp. The heating pipe is burnt
The temperature controller
The temperature sensor is broken.
Replace the heating pipe
Replace the temperature controller
Replace the temperature sensor

8. Wiring Diagrams

The following diagrams illustrate the wiring configurations for different models:

Environmental Information (WEEE Directives)

This section provides information regarding the disposal of electronic waste according to European Union directives. Devices marked with a symbol of a crossed-out wheelie bin (?️) should not be disposed of as municipal waste. Users in the EU have the option to return WEEE (Waste Electrical and Electronic Equipment) to distributors or manufacturers for disposal after purchasing a new product. Illegal disposal of electrical and electronic equipment is subject to administrative fines.

Note: The original document contains specific information for Spain, France, and Italy regarding these directives.

Models: WB-T Transparent Water Bath, WB-T, Transparent Water Bath, Water Bath, Bath

File Info : application/pdf, 11 Pages, 1.34MB

PDF preview unavailable. Download the PDF instead.

MANU EN 1357

References

Microsoft: Print To PDF

Related Documents

Preview HMM1 Heating Mantle, Multi-Row User Manual | Ibx Instruments
Comprehensive user manual for the Ibx Instruments HMM1 multi-row heating mantle, covering safety instructions, operation, maintenance, and troubleshooting.
Preview Ibx S01 Compact Magnetic Stirrer User Manual
Comprehensive user manual for the Ibx S01 Compact Magnetic Stirrer without heating, covering safety instructions, operation, maintenance, specifications, and troubleshooting.