NAVAC NX1 NX1V Nexus Digital Manifold Gauge User Manual
Safety Information
Use the instrument only as specified in this manual. Otherwise, the protection provided by the instrument may be impaired.
To avoid personal injury or death, follow these guidelines:
- Most governments and legal authorities require that HVAC technicians be trained and certified in the safe and proper operation of HVAC tools, such as this instrument. Since this tool may be connected to many types of equipment through a limitless combination of hoses and fittings, proper training is the most important element of using this tool safely.
- Read the entire Users Manual before using the instrument.
- Use the instrument only as described in the Users Manual, otherwise the protection provided by the equipment may be impaired.
- Do not use the instrument if it is damaged. Before you use the instrument, inspect the case. Look for cracks or loose components.
- The instrument contains no internal user-serviceable parts. Do not open the instrument.
- Do not use the instrument if it operates abnormally. Protection may be impaired. When in doubt, have the instrument serviced.
- Do not operate the instrument around explosive gas, vapor, or dust.
- The refrigerant database in this unit may include refrigerants classified as flammable. If such refrigerants are selected, the operator may need additional certifications and/or training. Consult your government and legal authority and comply fully with all requirements.
- Always wear eye and skin protection when working with refrigerants. Escaping refrigerant vapors will present a freezing danger. Do not direct refrigerant vapors venting from hoses towards the skin.
- Do not allow pressures beyond the specifications listed in this manual.
⚠️ Warning This product operates under high pressure. Follow all safety guidelines regarding refrigerant handling including wearing Personal Protective Equipment such as safety glasses, and gloves.
NX1 Nexus Digital Manifold Gauge
Product Overview
The NX1 Nexus Digital Manifold Gauge features a 5V DC Charger Port (USB Type C), Indicator Light, Control Keys, Low Side Valve (Black), High Side Valve (Orange), Temperature Probe Holder, Sight Glass, and 1/4" SAE Hose Connection-Ports. It also includes a Foldable Hook on the rear and a Display.
Technical Specifications
Model | NX1 | NX1V |
---|---|---|
Accessory | 2 Temperature Probes, Charging Cable, Carry Case, NMV1S Micron Gauge in additional to the NX1 Kit | |
72 Refrigerant Types | R-11, R-113, R-114, R-115, R-116, R-12, R-123, R-1233zd, R-1234yf, R-1234ze, R-124, R-125, R-13, R-134a, R-22, R-23, R-236fa, R-245fa, R-290, R-32, R-401A, R-401B, R-402A, R-402B, R-403B, R-404A, R-406A, R-407A, R-407B, R-407C, R-408A, R-409A, R-410A, R-413A, R-414A, R-414B, R-416A, R-417A, R-417C, R-420A, R-421A, R-421B, R-422A, R-422B, R-422C, R-422D, R-424A, R-427A, R-428A, R-434A, R-437A, R-438A, R-448A, R-449A, R-450A, R-452A, R-452B, R-453A, R-454B, R-458A, R-500, R-501, R-502, R-503, R-507A, R-508B, R-513A, R-600, R-600A, R-601, R-601A, R-744* | |
Max. Over Pressure | Maximum pressure: 600 Psi (4200 kPa) | |
Pressure Unit | 1015.3 psi (70 bar) | |
Vacuum Unit | bar, Mpa, -Kpa, psi, kgf/cm² | |
Temperature Unit | Pa, micron, mBar, mmHg | |
Pressure Scale | °C, °F | |
Resolution | -1~50 bar, -0.1~5 Mpa, -100~5000Kpa, -14.5~725psi, -1.02~51kgf/cm² | |
Temperature Resolution | 0.01 bar, 0.001Mpa, 1Kpa, 0.1psi, 0.01kgf/cm² | |
Battery | 32~104°F (0~40°C) | |
Connection | 3000 mAh Li-polymer | |
Sensor | 7/16" UNF | |
Unit Dimension | Digital Sensor | |
Unit Weight | 7.7"x2.8"x9" |
Indicator Light
Status | Indicator | Description |
---|---|---|
Always on/ Flashing | Power on | Power off |
Bluetooth is not connected | / | |
Bluetooth is connected | / | |
Device power is low | Charging | |
Fully charged |
Icons
Power display:
- T1 - low battery of Thermometer -LP
- T2 - low battery of Thermometer -HP
- VG - low battery of Vacuum Gauge
- Bluetooth connection identification; 0/1/2/3/4 indicates the number of Bluetooth connections, 0 indicates no connection.
Auto-Shutdown
- Pressure value stays at 0, no action on buttons, automatic shutdown after 30 minutes;
- Pressure sensor malfunction, no action on buttons, automatic shutdown after 10 minutes;
- Low battery, automatic shutdown.
Function Introduction
- On/Off key: Press and hold for 1s to start the machine, press and hold for 2s to shut it down.
- MODE Mode Selection Key: Manifold Gauge Mode, Leak Test Mode, Evacuation Mode and Decay Test Mode.
- SET Setting Key: Press and hold for 2s to enter the parameter setting.
- ▲ Up Key: Moves the parameter upwards.
- ▼ Down Key: Moves the parameter downwards.
- ENTER Confirm Key: Setup interface to confirm parameters; Start/Stop in Leak Test mode.
Calibration
Press and hold for 2 seconds to zero out the pressure.
Modes of Operation
Bluetooth Pairing
- To pair the low pressure side, press and hold ENTER + ▲ for 3 seconds.
- To pair the high pressure side, press and hold ENTER + ▼ for 3 seconds.
- To pair the vacuum gauge, click "Mode" button to switch NX1 to Evacuation Mode, then press and hold ENTER + for 3 seconds.
- In Evacuation Mode, Press and hold ENTER + ▼ for 3 seconds.
Parameter Setting
In Manifold Gauge mode, Leak Test mode, Evacuation mode, press and hold "SET" key for 2s, enter the parameter setting interface, press ▼ or ▲ to cycle and select the parameter content (refrigerant setting, press and hold to scroll through refrigerant type), press "ENTER" will switch to the next parameter setting, when all the parameter setting is completed, the screen will return to the original interface.
Factory Refrigerant Type: R-11, R-12, R-13, R-22...... (Total 72 types)
Pressure unit: MPa, bar, kgf/cm², psi, Kpa.
Temperature unit: °C, °F
Vacuum unit: Pa, micron, mBar, mmHg
Sound: ON/OFF
Backlight: 1min/3min/10 min/OFF (the number indicates that the backlight turns on for a few minutes after operation without buttons, and OFF indicates that it turns off)
Manifold Gauge Mode
VSAT: Vapor saturation temperature.
LSAT: Liquid saturation temperature.
T1 SLT: Suction line temperature.
T2 LLT: Liquid line temperature.
SH: Superheat value SH = T1 SLT-VSAT.
SC: Subcooling value SC = LSAT-T2 LLT.
T1-T2: temperature difference.
Dial scale (in Mpa):
Pressure values at the high and low pressure ends are displayed as gauge pressure. (Gauge pressure = measured pressure - atmospheric pressure)
When the pressure value is less than 0, the pressure unit automatically switches to inHg.
Leak Test Mode
Screen Display Introduction
Initial Pressure Value, Real-time Pressure Value, Leak Test Time, TIMER, dP (Pressure Difference).
Operation
Note: the system needs to be connected to the high side of the manifold gauge.
- Press the ENTER (show button as original) key to start the leak test. The bottom of the screen displays "START", the right side of the screen displays the real-time pressure value, the leak test timer starts, and the value of "dP" starts calculating.
- Press the ENTER key again, the bottom of the screen shows "END", the leak test timer stops and the final pressure difference is displayed at "dP";
- Press the ENTER key again to Reset.
Evacuation Mode
Screen Display Introduction
Vacuum Gauge Connection Status, Vacuum Unit, Evacuation Time, TIMER, Real-time Vacuum Value, Vacuum Target Value.
Evacuation value display range is 0-9999;
- 10000-19999 micron is displayed as "1E4";
- 20000-29999 micron is displayed as "2E4";
- ≥30000 micron is displayed as "....".
Vacuum Gauge Bluetooth Pairing
The first time you connect the vacuum gauge NMV1S, you need to do the Bluetooth pairing manually:
At this time, the Bluetooth logo "CONNECT NMV1S" of the digital display meter is flashing, long press ENTER + for more than 3 seconds to enter the Bluetooth pairing interface of the vacuum gauge, is lit in sequence, indicating that the device is searching for Bluetooth. If the screen shows "", indicating that the Bluetooth has been successfully connected, and then it will return to the interface of the Evacuation Mode in 2s, and the Bluetooth logo "CONNECT NMV1S" will disappear.
Notes:
- (1) When in Bluetooth pairing, no other operation can be performed for 30 seconds.
- (2) The digital gauge will automatically connect to the paired vacuum gauge when it is in vacuum mode.
Decay Test Mode
Screen Display
DECAY TEST, MICRON, Real-time Pressure Values, TIMER, DECAY TARGET, Evacuation Target Value.
Vacuum Attenuation Mode Parameter Settings
Long press the SET key for two seconds to enter the vacuum attenuation parameter setting interface. There are 4 kinds of setting parameters, the order are: vacuum unit, "TARGET", "DECAY" and "TIMER". When the parameter is selected, the corresponding icon and value will flash, short press the ▼ or ▲ key to select the value cyclically, press the "ENTER" key to lock and save, the parameter will display constant light and automatically switch to the next parameter setting.
Function
When connected to the vacuum gauge, the digital display will receive the "TARGET", "DECAY" and "TIMER" setting values from the vacuum gauge and display them in the corresponding positions on the screen.
- (1) Start vacuum extraction, when the vacuum level of the measuring device is less than the "Target" setting;
- (2) The vacuum pumping is stopped and when the vacuum value is greater than the "Target", the vacuum decay mode is entered and the "TIMER" starts timing;
- (3) If the vacuum value does not exceed the setting value of "DECAY" within the time set by "TIMER", it will display "PASS" (Pressure Holding Successful), otherwise it will display "FAIL".
Exploded View & Spare Parts List
Exploded View
The exploded view shows the components of the NAVAC NX1/NX1V manifold gauge, including the main body, valves, display, circuit board, battery, and various covers and knobs.
Spare Parts List
No. | Item | Qty |
---|---|---|
1 | Master Clamp Handle | 1 |
2 | LCD | 1 |
3 | Toothed Block | 1 |
4 | Temperature Sensor | 1 |
5 | O-ring | 1 |
6 | Cover | 1 |
7 | Toothed Block | 1 |
8 | Spring | 1 |
9 | Nut | 1 |
10 | Sub Clamp Handle | 1 |
11 | Battery Cover Plate | 1 |
12 | AAA Battery | 2 |
13 | Retaining Stud | 1 |
14 | Battery Shrapnel | 1 |
15 | PCBA | 1 |
16 | Button | 1 |
NX1 Refrigerant Update Via MyNAVAC App
Download the myNAVAC APP from the App Store or Google Play. Connect your NX1 to the app via Bluetooth. Navigate to Settings > Refrigerant Update. Confirm and select refrigerants.
Refrigerant update conditions:
- Both gauge pressures should be 0 and the vacuum reading should also be "-----".
- Unit power is greater than 50%.
- Not in setup screen.
When the APP is updating with the unit, the unit refrigerant model display area displays "Loading...". When the update is successful, "Complete" is displayed.
After successful update, it displays "Complete" and returns to the original interface automatically, and the refrigerant model number displays ".......". If the refrigerant data update fails, "Failed" will be displayed, and after 3 seconds, it will automatically return to the original interface and keep the original refrigerant data.
NMV1S Overview
Instrument Overview
The NMV1S is a vacuum gauge with a display area, operation key, and battery charge port.
Function Introduction
1. Operation Key
- 1.1.: Power on / off key. Press 1 second to turn on, press 2 seconds to turn off.
- 1.2.: Down key. In setting mode, adjust settings. In decay testing mode, set timer and leakage rate.
- 1.3. ▲: Up key. In setting mode, adjust settings. In decay testing mode, set timer and leakage rate.
- 1.4. SET: Parameter setting key. Interface display "○○○" press SET and hold for 2s to enter the parameter setting interface.
2. Screen Display
- 2.1. SET: Parameter setting mode.
- 2.2. 0000: Power indicator.
- 2.3. 01: Bluetooth connection indicator; 0/1 indicates the number of Bluetooth connections.
- 2.4. Evac Target: minimum target vacuum; (Vacuum value expected to be reached by the evacuation system)
- 2.5. Decayed target: the maximum target vacuum degree which should be greater than the target vacuum value;
- 2.6. Decay time: pressure maintaining time. (Triggered only when the minimum and maximum target vacuum are not setting to "off");
- 2.7. Decay test: enter the decay testing procedure;
- 2.8. Rate: leakage rate, leakage rate unit: / min.
Technical Specifications
Model | NMV1S |
---|---|
Location | Indoor use |
Altitude | ≤5000m |
Humidity | ≤75% R.H. |
Intended Use | Pressure Measurement |
Maximum Overload Pressure | 50 psi / 3.4 bar |
Range | 0~25000 microns, 0~3333.3 Pa, 0~33.3 mBar, 0~25 mmHg, |
Resolution | 0.01 micron (<10 microns), 0.01Pa (<10 Pa), 0.001 mBar, 0.001 mmHg |
Accuracy | 50~1000 microns:± 5% of reading(at 68°F) |
Operating Temperature | 32~104°F(0~40°C), 10~90%RH |
Battery | 2000 mAh Li-polymer |
Charge Parameter | Max. 5V 2A |
Unit | Pa, mBar, mmHg, microns |
Sensor | 1/4" SAE |
Connections | Pirani sensor |
Weight | 6.2 oz |
Indicator Light:
Indicate | Status |
---|---|
Low Battery | |
Powered on, Bluetooth is not connected | |
Bluetooth is connected |
Display Range
Vacuum display range for vacuum pumping: 0-25000 microns (3333.3 Pa, 33.3 mbar, 25 mmHg). Over range display "----" .
Automatic Shutdown
When the vacuum gauge has a reading displayed, the automatic shutdown program will not be triggered for 2 hours; The automatic shutdown program will be triggered only when the interface displays atmospheric pressure "----", and without any operation on the interface.
Operation
1. Start, Shut off
Press 1 second to turn on, press 2 seconds to turn off.
2. Parameter Setting Interface
2.1. The vacuum display area displays "----", long press" SET "2 seconds to enter the parameter setting interface.
2.2. Vacuum unit setting:
Press ▲▽ to select unit, press " SET " to lock and switch to next parameter setting.
2.3. "Evac target" setting: (minimum target vacuum)
Press, to select the flashing value, then press" SET "to lock and stay on, then switch to the next "delayed target" setting.
2.4. Delayed target setting: (maximum target vacuum)
Press ▲ to select, the value is higher than the value set by "evac target" or "off". If "off" is selected, the "delay time" pressure maintaining time will default to "off", and skip directly to enter the sound setting.
2.5. Pressure maintaining time setting (delay time)
After the "delayed target" setting is completed, if the values of 3&4 are not "off", press" SET enter to the setting of "delayed time". Press ▲▽ to select. Then press" SET to lock and stay on, then switch to the next parameter setting.
2.6. Sound setting "beep"
Press ▲▽ to select, press" SET "to lock and then enter the next parameter "backlight" setting.
2.7. Backlight setting "B-L"
Press ▲ to select, press" SET "to lock, it will return to the standby interface automatically.
3. Evacuation and Pressure Maintaining Interface I
(Both evac target and Delayed target are not set to off)
3.1. After the vacuum pumping starts, the actual measured vacuum reading keeps getting smaller. When it reaches the "evac target" setting value, the value such as "XXX" keeps flashing. Press (any key) to turn off flashing and stop the prompt.
3.2. When the actual measured vacuum reading is less than the "evac target" value and rises to this reading, the pressure maintaining time starts to count from 0. Until exiting this interface or complete pressure maintaining. (if the vacuum is less than the evac target value, time will be paused until the collected number is greater than evac target. Then continue timing. Press ▲ or ▼ to select evac target, delayed target, delayed time and rate.
3.3. When do decay testing, If the vacuum reading does not exceed the set value of delayed target. It displays "pass" until you press (any key) to stop flashing, exit the prompt and return to the main standby interface.
3.4. When do decay testing, if the vacuum reading gets larger, reaches and exceeds the set value of delayed target, the buzzer will alarm, and the word "FAIL" will flash until you press any key to stop flashing and exit the prompt, and return to the standby main interface.
3.5. If pressure holding time is set to "off", alarm will start after the vacuum value exceeds decayed target.
4. Evacuation and Pressure Maintaining Interface II
(Evac target and Delayed target are set to off)
4.1. Evac target is set to off and delayed target is not set to off when the real-time vacuum reduces, the prompt of delayed target value will not be triggered. The prompt of "fail" and "alarm tone" will be triggered only when the vacuum value increases to the delayed target value.
4.2. Evac target is set to non off and delayed target is set to off when the target vacuum reduces, the "alarm tone" beep of evac target value will be triggered. When the vacuum value increases, the "fail" alarm will not be triggered.
4.3. Both evaluation target and delayed target are set to off all alarm prompts will not be triggered.
5. Zero setting calibration
Under the atmosphere, press and hold + ✔ at same time for more than 2 seconds until the vacuum display area displays "------- ", the full calibration is completed.
Warning
WARNING: Read all safety warning and instructions. Failure to follow the warnings and instructions may result in electric shock, fire and/or serious injury.
- Do not dismantle, open or shred rechargeable Li-polymer battery pack.
- Do not expose battery pack to heat or fire. Avoid storage in direct sunlight.
- Do not short-circuit a battery pack.
- When battery pack is not in use, keep it away from other metal objects, like paper clips, coins, keys, nails, screws or other small metal objects, that can make a connection from one terminal to another, Shorting the battery terminals together may cause burns or a fire.
- Do not subject battery pack to mechanical shock.
- In the event of a battery leaking, do not allow the liquid to come in contact with the skin or eyes. If contact has been made, wash the affected area with copious amounts of water and seek medical advice.
- Seek medical advice immediately if a battery pack has been swallowed.
- Do not use any battery pack which is not designed for use with the equipment.
- Use only the battery pack in the application for which it was intended.
- Do not use a battery pack or appliance that is damaged or modified. Damaged or modified batteries may exhibit unpredictable behavior resulting in fire, explosion or risk of injury.
- Keep battery pack out of the reach of children.
- Always purchase the correct battery for the equipment.
- Keep battery pack clean and dry.
- Wipe the battery pack terminals with a clean dry cloth if they become dirty.
- Dispose of properly.
- Follow all charging instructions and do not charge the battery pack or appliance outside of the temperature range specified in the instructions. Charging improperly or at temperatures outside of the specified range may damage the battery and increase the risk of fire.
- Instructions regarding battery charging. information regarding ambient temperature range for battery use and storage, and the recommended ambient temperature range for charging system during charging.
NMV1S Technical Specifications
Model | NMV1S |
---|---|
Operating Temperature | Charge: 32~104°F(0~40°C) Discharge: 14~140°F(-10~60°C) |
Storage Temperature | 1 month: -4~140°F(-20~60°C) 3 month: -4~113°F(-20~45°C) 1 year: -4~77°F (-20~25°C) |
18. Rechargeable battery pack need to be charged before use. Always refer to the equipment manual for proper charging instructions. Charge only with USB (Type C) not exceed 5V-2A.
NMV1S Screen Display
The screen displays Power indicator, APO (Automatic shutdown), Bluetooth status, and Temperature.
Indicator Light
Power on status: Green light blinks 3 times.
Rest state:
- (1) Entering or exiting function setting, green light flashes 3 times fast.
- (2) Normal function and successful Bluetooth connection, green light flashes slowly.
- (3) Normal function but Bluetooth is not connected, yellow light flashes slowly.
- (4) Low battery or hardware failure, red light flashes slowly.
Alarm
Over range:
- (1) When the measured temperature is greater than 125°C and less than 150°C, the temperature value is displayed in red and the buzzer sounds continuously at a frequency of 0.5S.
- (12) When the measuring temperature is less than -40°C or more than 150°C, the display will show "OL" and the buzzer will sound for a long time.
Hardware failure:
- (1) When the temperature sensor fails, "ERR.-" is displayed and the buzzer sounds continuously for 1 second.
- (2) When Bluetooth is faulty, the buzzer sounds continuously for 1 second.
Bluetooth Connection
The device will keep the Bluetooth broadcast status after powering on; if the device is not connected, " * " is blinking constantly; If the device is connected, " * " is always on.
Automatic Shutdown
When "APO" is turned on, the unit automatically shuts down if no key is pressed for 2 hours.
Key Function
- Switch on/off: Press and hold the key for 2 seconds to turn on the power, press and hold the key for 3s to turn off the power after turning on the power.
- Turn off and light up the screen: under the non-function selection state, press the key to switch back and forth between off and light up the screen.
- Function Selection:
- (1) Quickly double-click the button to enter the function setting interface, the setting order is Auto Power Off, Bluetooth Switch and Temperature Unit in order.
- (2) Click the button to select function.
Guide for App Download & Connection
iOS
Search for "myNAVAC" in the app market.
Android
Search Google Mall for "myNAVAC".
Log In
Account Login: Internet connection is required, all data records are stored in the backend server. Visitor Mode: No internet connection is required, all data records are stored locally on the cell phone.
Page Introduction
The page introduction illustrates the process of connecting devices, viewing job listings, accessing device pages, and utilizing functions like chart data logging and test reports.
Basic Functions
Link Device
Turn on your phone's Bluetooth. In the main interface, tap the plus button in the lower right corner to search available devices. (Note: If you are using the NX1 with a vacuum gauge and temperature probes, please connect the vacuum gauge and temperature probes to the NX1 first, and then connect the NX1 to the APP).
Find the target device you need to connect to, click the plus sign in the upper right corner and wait for the connection to be established.
If the connection is successfully established, it will automatically jump to the main interface, click the device icon to enter the device interface.
Data logging
Click " " in the bottom navigation bar to turn on the data logging function. Click Start to record data.
Click again on the bottom navigation bar " ", to save or restart the record. Then select a Job you wish to save to.
View Test Report
Click " " in the bottom navigation bar on the device page to view the test report on the current device.
On the main page, click "Job" in the bottom navigation bar to view the test reports saved in the workbook.
On the main page, click "Test Report" on the bottom navigation bar to view the test reports of all devices.
Share Test Reports
Press the "Share" button at the bottom of the test report, a pop-up window will prompt you to select the the file type and interval for the test report format.
After determining the the file type and interval, a data table will be automatically generated, click " " in the upper right corner to share the test report.
Related Documents
![]() |
NAVAC N2DX Flex-X Manifold Gauge User Manual User manual for the NAVAC N2DX Flex-X Manifold Gauge with HD Color Graphic Interface. Covers safety information, product overview, technical specifications, screen display, indicator icons, pressure calibration, device standby/off, button functions, and an exploded view. |
![]() |
NAVAC Summer Slam Promotion: Deals on HVAC Tools Discover NAVAC's Summer Slam Promotion, offering free tools with qualifying purchases of PowerT3 tools, NEXUS gauges, and recovery units. Valid from August 1st to September 30th, 2025. |
![]() |
NAVAC Flex72 N2D72 Digital Manifold Gauge User Manual User manual for the NAVAC Flex72 Manifold Gauge with Digital Display (N2D72). Provides comprehensive guidance on setup, operation, maintenance, and troubleshooting for HVAC professionals. |
![]() |
NAVAC Counter Day at Geary Pacific Supply Join Geary Pacific Supply for a NAVAC Counter Day event featuring NAVAC vacuum pumps, recovery units, manifold gauges, and tube benders. Learn about the latest HVAC tools and technology. |
![]() |
NAVAC N2A4A & N2A4B Manifold Gauge User Manual User manual for NAVAC N2A4A and N2A4B manifold gauges, covering safety guidelines, parts, specifications, technical parameters, operation instructions for pressure testing and system evacuation, and maintenance. |
![]() |
NAVAC N2A4 Parts Diagram and Component List Detailed parts diagram and component list for the NAVAC N2A4. Includes part numbers, descriptions, and a visual breakdown of the assembly. |
![]() |
N4A4H Parts Diagram and Component List Detailed parts diagram and component list for the N4A4H, including part numbers and descriptions for each component. |
![]() |
NAVAC NR7 Recovery Unit User Manual Comprehensive user manual for the NAVAC NR7 Recovery Unit, detailing safety measures, operating instructions, specifications, and troubleshooting for HVAC and refrigeration system service. |