Urgent Medical Device Removal: Accu-Chek Guide System

May 13, 2024

URGENT MEDICAL DEVICE REMOVAL

Accu-Chek® Guide system: Meters showing incorrect measurement unit (mmol/L rather than mg/dL)

This Urgent Medical Device Removal (UMDR) only applies to the following products:

Affected Meter Serial NumberProduct DescriptionProduct REF NumberProduct Device Identifier (UDI / GTIN)Product Lot Number
92339920445Accu-Chek® Guide (SC) Kit0845307100100365702729100207482
92339955415Accu-Chek® Guide (SC) Kit0845307100100365702729100207482
92340116052Accu-Chek® Guide (SC) Kit0845307100100365702729100207519
92340117408Accu-Chek® Guide (SC) Kit0845307100100365702729100207519
92340120006Accu-Chek® Guide (SC) Kit0845307100100365702729100207519
92340120057Accu-Chek® Guide (SC) Kit0845307100100365702729100207519
92339920116Accu-Chek® Guide (SC) Kit0845307100100365702729100207526
92339094787Accu-Chek® Guide (SC) Kit0845307100100365702729100207545
92339744998Accu-Chek® Guide (SC) Kit0845307100100365702729100207454

Issue:

Roche Diabetes Care has received three customer complaints, including one US complaint, in which the meter was reading in the incorrect unit of measure. The complaints were reported to FDA as quality issues via the Medical Device Reporting program. The resulting Roche complaint investigations indicated that the above Accu-Chek Guide systems were configured incorrectly, and thus show the results in mmol/L rather than in the labeled measuring unit, mg/dL. This becomes visible once a blood glucose measurement is taken and the user could recognize that the result is displayed mmol/L rather than mg/dL. Roche Diabetes Care has conducted an in-depth evaluation of the underlying root cause of the above-described issue and has corrected the issue to prevent recurrence.

Risk to Health:

This misconfiguration may result in customers perceiving wrong units of measure, and/or interpreting a result as approximately 18 times lower than the actual blood glucose. In the worst case, these misleading displayed results, in the wrong unit of measure, may trigger an inappropriate treatment decision (consumption of carbohydrates) with potentially severe consequences.

Reason for the Notification:

Investigations have revealed that the nine meters listed above were configured incorrectly and thus show the results in a different than labeled measuring unit. This notification is to inform customers of the issue and to instruct customers to discontinue use of the Accu-Chek Guide blood glucose meters with the serial numbers listed above and contact Accu-Chek Customer Care at 1-800-858-8072 to arrange for return and replacement.

Accu-Chek Guide System Visual Description

The Accu-Chek Guide blood glucose monitoring system includes the Accu-Chek Guide meter, Accu-Chek Softclix lancing device, and Accu-Chek FastClix lancets. The meter features a backlit display showing the time and blood glucose reading (e.g., "106 mg/dL"), and has buttons for operation. The system is designed for single patient use.

Actions Required by End Users

  • Please check your meter to see if the display shows the appropriate measuring unit (mg/dL).
  • Please compare the serial number of your meter with the listed affected serial numbers to determine if it is impacted by this issue.

Affected Serial Numbers:

  • 92339920445
  • 92339955415
  • 92340116052
  • 92340117408
  • 92340120006
  • 92340120057
  • 92339920116
  • 92339094787
  • 92339744998
  • Please stop using your Accu-Chek Guide meter if you notice that your product shows an incorrect measuring unit or the serial number is one of the listed impacted products.
  • Please visit http://accu-chek.com/notices/UMDR-24-001 and enter your serial number. The form will provide instructions on how to proceed.
  • Alternatively, contact Roche Diabetes Care Accu-Chek Customer Care at 1-800-858-8072 (Monday - Friday from 8:00 a.m. ET - 8:00 p.m. ET).
  • Please keep this UMDR for future reference.

Actions Required for Distributors/Retailers

  • Read this UMDR (24-001).
  • Please check meters remaining in your stock for impacted serial numbers.
Affected Meter Serial NumberProduct DescriptionProduct REF NumberProduct Device Identifier (UDI / GTIN)Product Lot Number
92339920445Accu-Chek® Guide (SC) Kit0845307100100365702729100207482
92339955415Accu-Chek® Guide (SC) Kit0845307100100365702729100207482
92340116052Accu-Chek® Guide (SC) Kit0845307100100365702729100207519
92340117408Accu-Chek® Guide (SC) Kit0845307100100365702729100207519
92340120006Accu-Chek® Guide (SC) Kit0845307100100365702729100207519
92340120057Accu-Chek® Guide (SC) Kit0845307100100365702729100207519
92339920116Accu-Chek® Guide (SC) Kit0845307100100365702729100207526
92339094787Accu-Chek® Guide (SC) Kit0845307100100365702729100207545
92339744998Accu-Chek® Guide (SC) Kit0845307100100365702729100207454
  • If you identify an Accu-Chek Guide blood glucose meter with one of the serial numbers noted above, remove it from your inventory and contact Roche Diabetes Care Accu-Chek Customer Care at 1-800-858-8072 (Monday – Friday from 8:00 a.m. ET - 8:00 p.m. ET) to arrange for return and replacement.
  • If you have distributed the products noted above to other suppliers, provide a copy of this UMDR to those suppliers.
  • If you have distributed products of the lots noted above to consumer customers, make the UMDR available to those customers.
  • Complete all sections of the enclosed distributor faxback form (24-001) and fax or email it according to the instructions on the form.
  • Contact your Roche Account Manager or your distributor if you have questions regarding the information in this UMDR.
  • Please keep this UMDR for future reference.

Contact Information and Reporting

For questions regarding this UMDR or other questions related to Roche Diabetes Care products, please contact your Roche Account Manager or contact Accu-Chek Customer Care at 1-800-858-8072.

ACCU-CHEK GUIDE and ROCHE DIABETES CARE are trademarks of Roche. ART-10333570001

This notification is being made with the knowledge of the Food and Drug Administration (FDA). You may also report adverse events or quality problems experienced with the use of this product to the FDA's MedWatch Adverse Events Reporting Program online at https://www.fda.gov/Safety/MedWatch/HowToReport/default.htm (form available to fax or mail), or call the FDA 1-800-FDA-1088.

Models: 08453071001 Urgent Medical Device Removal, 08453071001, Urgent Medical Device Removal, Medical Device Removal, Device Removal, Removal

File Info : application/pdf, 4 Pages, 305.42KB

PDF preview unavailable. Download the PDF instead.

UMDR 24-001 ART-10333570001

References

atkinsc3 PDF generator

Related Documents

Preview Accu-Chek Instant & mySugr Quick Reference Guide: Pairing, Target Range, and Logging
A concise guide to pairing your Accu-Chek Instant blood glucose meter with the mySugr app, changing target ranges, and logging blood glucose readings, carbs, insulin, and activities. This document provides essential instructions for managing diabetes with Accu-Chek and mySugr.
Preview ACCU-CHEK Mobile Wireless Adapter: Instructions for Use and Setup
A guide to attaching, pairing, transferring data, and replacing the battery for the ACCU-CHEK Mobile Wireless Adapter. Includes safety warnings and troubleshooting tips.
Preview ACCU-CHEK ADVANTAGE CONTROL SOLUTION CA Safety Data Sheet
Safety Data Sheet for ACCU-CHEK ADVANTAGE CONTROL SOLUTION CA, providing comprehensive information on product identification, hazard assessment, composition, first aid measures, firefighting, accidental release, handling and storage, exposure controls, physical and chemical properties, stability and reactivity, toxicological information, ecological information, disposal considerations, transport information, and regulatory information.
Preview cobas h 232 Quick Reference Guide | Roche Diagnostics
Concise guide for the Roche cobas h 232 point-of-care testing system, covering essential operational steps and sample results. Includes proBNP testing information.
Preview Roche v-TAC Standalone Software User Guide
Comprehensive user guide for the Roche v-TAC Standalone software, version 1.5. This guide provides detailed information on installation, operation, safety precautions, troubleshooting, and technical specifications for the v-TAC Standalone software used in in vitro diagnostics.
Preview Roche Diagnostics Indianapolis Campus Map and Directory
Detailed campus map and directory for Roche Diagnostics North America and Diabetes Care North America headquarters in Indianapolis, including a legend of facilities and driving directions to an off-campus site.
Preview Richeal Cline Biography and Roche Diagnostics Overview
Profile of Richeal Cline, Head of Diagnostics Global Operations and Global Head of Quality & Regulatory at Roche Diagnostics, along with an overview of Roche's leadership in healthcare.
Preview Roche Whole Blood Collection Tube - Instructions for Use and Technical Information
Comprehensive guide for the Roche Whole Blood Collection Tube, detailing its intended use, required materials, handling precautions, storage guidelines, step-by-step instructions, symbol interpretations, and manufacturer information.