Analog Chlorine Sensors
For free and total chlorine
The electrochemical WTW chlorine sensors can be applied for measurements in drinking water or swimming pools. Connectable via MIQ/IDS module to the IQ Sensor NET.
We would like to inform you about the application range on our website.
Sensor Overview and Specifications
Three sensor models are presented visually, each depicted as a cylindrical probe with a protruding tip. Key dimensions are indicated: overall length approximately 200 mm, with a tip section of approximately 25 mm.
Model | FCML 412-M12-2 | FCML 412-M12-20 | TCML 412-M12-2 |
---|---|---|---|
Measuring principle | Amperometric | Amperometric | Amperometric |
Measured value | Free chlorine | Free chlorine | Total chlorine |
Measuring range | 0.005 … 2.000 mg/l Cl² | 0.05 … 20.00 mg/l Cl² | 0.005 … 2.000 mg/l Cl² |
Accuracy | <1 % | <1 % at 4 mg/l, >3 % at 16 mg/l | <2 % |
Response time | t90 approx. 120 s | ||
Minimum flow rate | Recommended minimum flow rate in flow cell D 19: 15 l/h, or alternatively in flow cell D-CL: > 30 l/h | ||
Temperature measurement | 0 °C … 45 °C | ||
Temperature compensation | Automatically via integrated sensor | ||
pH range | 4 … 9 | 4 … 9 | 4 … 12 |
Polarization time | Approx. 1 hour after new installation or change of electrolyte | ||
Calibration method | 1-point-calibration (according to DPD method as reference) | ||
Pressure resistance | 3 bar | ||
Electrical connection | M12 plug* | ||
Certifications | CE, UKCA | ||
Mechanical | |||
Shaft: | PVC | PVC | PVC |
Membrane cap: | PVC | PVC | PVC |
Working electrode: | Gold | Gold | Gold |
Reference electrode: | Ag/AgCl | Ag/AgCl | Ag/AgCl |
Cable connection: | Polyamid | Polyamid | Polyamid |
Protection rate: | IP64 | IP64 | IP64 |
Weight | Approx. 0.5 kg | ||
Warranty | 1/2 year with intended use |
* Sensor with M12 plug for connection to the IQ Sensor NET, please order adapter ADA CI/IDS separately.
Product Models and Descriptions
Model | Description | Order No. |
---|---|---|
FCML 412-M12-2 | pH value independent chlorine sensor based on the electrochemical principle, suitable for online measurement of free chlorine in drinking water and swimming pool water. Measuring range: 0-2 mg/l, pH range 4-9 (please order adapter to connect with IQ Sensor Net module MIQ/IDS separately) | 201189 |
FCML 412-M12-20 | Like FDML 412-M12-2, but with measuring range: 0-20 mg/l (please order adapter to connect with IQ Sensor Net module MIQ/IDS separately) | 201194 |
TCML 412 M12-2 | pH value independent chlorine sensor based on the electrochemical principle, suitable for online measurement of total chlorine in drinking water and swimming pool water. Measuring range: 0-2 mg/l, pH range 4-12 (please order adapter to connect with IQ Sensor Net module MIQ/IDS separately) | 201197 |
Manufacturer Information
Xylem Analytics Germany Sales GmbH & Co. KG
Am Achalaich 11, 82362 Weilheim, Germany
Phone: +49 881 1830
Fax: +49 881 183-420
Email: Info.XAGS@xylem.com
Website: www.xylemanalytics.com
All names are registered tradenames or trademarks of Xylem Inc. or one of its subsidiaries. Technical changes reserved. © 2023 Xylem Analytics Germany Sales GmbH & Co. KG.