Bruker helix+ User Manual

Y-STRUCTURE AMINE COUPLING KIT 2

coupling of molecules via primary amines to the green arm of the new Y-structure kit

Dynamic Biosensors GmbH

HK-NYS-NHS-2 v1.0

Key Features

Workflow Overview

3-Step Conjugation Workflow

The workflow is depicted as a sequence starting with a tube labeled 'Crosslinker', leading to a tube labeled 'Ligand', then to a 'proFIRE®' device, and finally to multiple tubes labeled 'Conjugate stock solutions'. The timeline indicates hands-on time < 1 hour, incubation ~ 2 hours, and total time ~ 3 hours.

  1. DNA Modification: The Ligand strand is functionalized with a primary amine reactive NHS.
  2. Ligand Conjugation: The biomolecule (ligand) is added to the functionalized Ligand strand and incubated for at least 1 hour.
  3. Purification: The Ligand strand conjugate is purified using the proFIRE® system. After buffer exchange, the conjugates are aliquoted and stored.
  4. Ready-to-use: The conjugate stock solutions are ready to use in the heliX® biosensor.

Product Description

Order Number: HK-NYS-NHS-2

Table 1. Contents and Storage Information

Material Cap Amount Storage
Y-structure Ligand strand 2 NHS Blue 3 x -20°C
Buffer A [1] Transparent 1 x 1.8 mL -20°C
Buffer C [2] Transparent 3 x 1.8 mL -20°C
Buffer PE40 [3] Transparent 3 x 1.5 mL -20°C
ddH2O Transparent 1.5 mL -20°C
Crosslinker Brown 3 x -20°C
Purification spin column Red 6 x 2-8°C
2.0 mL reaction tubes for purification spin column 6 x RT
Centrifugal filter unit (3 kDa MWCO)[4] 3 x RT
Centrifugation collection tube 3 x RT

For research use only. This product has a limited shelf life, please see expiry date on label.

IMPORTANT: Products may be shipped at different temperatures, but storage should adhere to the guidelines outlined in the Table. The kit contains reagents sufficient for five conjugations of approximately 50-200 µg of biomolecule each. The resin slurry in the purification spin column contains 0.02 % sodium azide.

Additional Materials Required

Material Comments
Benchtop microcentrifuge Required speed range of between 1,000 x g to 13,000 x g
Vortex
1.5 mL reaction tubes
UV-Vis Spectrophotometer (e.g. Nanodrop) For determination of the Ligand strand 2 conjugate's concentration

All necessary solutions and buffers are included in the kit.

Important General Notes

3-Step Conjugation of a Biomolecule to Ligand strand 2

Please read the entire protocol before starting and perform all steps without interruption.

TIP: This protocol can be performed simultaneously for multiple coupling reactions. Avoid using partially purified protein samples or protein samples containing carriers (e.g., BSA).

Before starting, allow the crosslinker to reach room temperature before use.

From now on, Y-structure Ligand strand 2 NHS will be referred to as Ligand strand 2.

I. Nanolever Modification

  1. Dissolve Ligand strand 2 in 40 µL Buffer A prior to use, vortex until all solids are completely dissolved and briefly spin down.
  2. Dissolve the crosslinker (brown cap) by adding 100 µL ddH2O, vortex until all solids are completely dissolved and briefly spin down. IMPORTANT: Always use fresh compound.
  3. Add 10 µL of the freshly prepared linker solution to one Ligand strand 2 aliquot. Discard the remaining linker solution from step 2.
  4. Vortex the reactants for 10 sec, spin down and incubate for 20 minutes at room temperature. IMPORTANT: Do not exceed incubation time or the reaction yield will decrease.
  5. In the meantime, equilibrate two purification spin columns (red cap) for one coupling reaction:
    • a. Remove the column's bottom seal and loosen cap (do not remove cap).
    • b. Place the column in a 2.0 mL reaction tube.
    • c. Centrifuge at 1,500 x g for 1 minute to remove the storage solution.
    • d. Add 400 µL of Buffer C to the column's resin bed. Centrifuge at 1,500 x g for 1 minute to remove buffer.
    • e. Repeat step d and discard the resulting buffer from the reaction tube. The purification spin column should now be in a dry state.
  6. Sample loading:
    • a. Place the columns from step 5 in new 1.5 mL reaction tubes.
    • b. Remove the cap of spin column number 1 and apply the sample from step 4 to the top of the resin bed.
    • c. Centrifuge at 1,500 x g for 2 minutes to collect the sample (flow-through). Discard the Purification spin column after use.
    • d. Remove the cap of spin column number 2 and apply the sample from step c to the resin bed.
    • e. Centrifuge at 1,500 x g for 2 minutes to collect the sample (flow-through). Discard the Purification spin column after use.

II. Ligand Conjugation

  1. Add approx. 100 µg (up to a maximum of 200 µg) of the ligand (concentration approx. 0.5 - 50 mg/mL) to the sample from step 6. For optimal conditions use a volume of approx. 50 µL. EXAMPLE: Adjust protein concentration to 2 mg/mL and use 50 µL for conjugation. IMPORTANT: Ensure the storage buffer of the ligand does not contain any primary amines, e.g., Tris buffers, glycine (please check Important Notes).
  2. Mix the reaction by pipetting up and down and let it react at room temperature for at least 1 hour. IMPORTANT: Do not vortex. If necessary, the reaction can be carried out at 4 °C with a longer reaction time (e.g., overnight).

III. proFIRE® Purification

  1. Perform a purification using the appropriate proFIRE® workflow (please refer to the proFIRE® User Manual). Please make sure that the sample volume is 160 µL.
    • a. If the volume is less than 160 µL, fill the missing volume with Buffer A.
    • b. If the volume exceeds 160 µL, please perform additional 160 µL runs until all the sample is consumed.
  2. Use the Data Viewer software of the proFIRE® to identify which fractions contain pure conjugate. An example chromatogram is shown in Additional Information section: proFIRE® purification of Ligand strand 2 conjugate.
  3. Remove the recommended fractions from the fraction collector. TIP: Do not keep the Ligand strand 2 conjugate for prolonged time in the proFIRE® running buffer. Proceed immediately with the buffer exchange.

IV. Buffer Exchange

  1. Add 500 µL of the first proFIRE® fraction containing the Ligand strand 2 conjugate to the centrifugal filter unit. Centrifuge at 13,000 x g (up to 14,000 x g) for 10 minutes and discard the flow-through.
  2. Add the remaining fractions to the same filter unit and repeat the centrifugation step in order to collect all samples in one tube. (Please check Additional information: Buffer Exchange and Concentration with Centrifugal Filter Units).
  3. Add 350 µL of PE40 (or TE40, HE40) buffer and centrifuge at 13,000 x g for 10 minutes. Discard the flow-through. IMPORTANT: If the protein is not stable in PE40 (or TE40, HE40), please check buffer compatibility with the switchSENSE® compatibility sheet.
  4. Add 350 µL of PE40 (or TE40, HE40) buffer and centrifuge at 13,000 x g for 15 minutes. Discard the flow-through.
  5. To recover the Ligand strand 2 conjugate, place the centrifugal filter unit upside down in a new centrifugal collection tube (provided in the kit). Spin at 1,000 x g for 2 minutes to transfer the sample to the tube.

V. Aliquots and Storage

  1. Measure the absorbance of the Ligand strand 2 conjugate at 260 nm (= A260nm) on a UV-Vis Spectrophotometer (e.g., Nanodrop).
  2. Determine the concentration of the Ligand strand 2 conjugate (c₁) by inserting (A260nm) into the following equation:

    c₁[M] = A260nm / (ε260 * d)

    where d is the path length (usually equal to 1 cm; however, please check the UV-Vis Spectrophotometer user manual) and ε260 is the extinction coefficient of the DNA at 260 nm, equal to 265,000.

  3. For a ready-to-use solution for a biochip functionalization, please adjust the concentration to 500 nM (or up to 1 µM) with PE40 (or TE40, HE40) buffer (including up to 10 % glycerol, if needed) and prepare 20 µL aliquots.
  4. Store between -86 °C and 8 °C, as desired. Stability of the solution is related to the stability of the ligand molecule. IMPORTANT: Before a switchSENSE® interaction measurement, please add the appropriate adapter strand to the conjugate solution.

Additional Information

proFIRE® purification of Ligand strand 2 conjugate

  1. To ensure best results from a measurement, no free Ligand strand 2 should be present on the chip. Therefore, crude Ligand strand 2 conjugates must be purified by ion exchange chromatography prior to measurement. This quality control step gives you additional useful information about your sample purity.
  2. We recommend using the proFIRE® system equipped with an ion exchange column, Buffer A [1] and Buffer B [5], which have the same composition but different salt concentration, allowing peak separation. IMPORTANT: For the Y-structure Ligand strand 2 - NHS kit, please enter 24 as the DNA length (Type 1) when starting the purification program.
  3. In Figure 1, a typical proFIRE® chromatogram of a Ligand strand 2 conjugate purification is depicted, where the peak of the protein-DNA conjugate is separated from the free protein (left) and the free DNA (right). IMPORTANT: The proFIRE® system owns a tailored software for automatic recognition and quantitation of DNA conjugates.
  4. After purification, collect the Ligand strand 2 conjugate fractions (Figure 1: fractions 8-10), concentrate the conjugate, and exchange buffer with your buffer of choice using a Centrifugal filter unit, as described in section II.

Figure 1. proFIRE® chromatogram of a ligand strand conjugate purification. Used buffers: Buffer A [1]; Buffer B [5]. Column: proFIRE® column. Flow: 1 mL/min. Used program: DNA length 24, Type 1.

Buffer Exchange and Concentration with Centrifugal Filter Units

  1. Take one centrifugal filter unit, add the appropriate volume (500 µL) of buffer in the filter device, and cap it.
  2. Place capped filter device into the centrifuge rotor, aligning the cap strap toward the center of the rotor; counterbalance with a similar device.
  3. Spin the device at 13,000 x g (or 14,000 x g) for the given time.
  4. Remove the flow-through and repeat steps 1-3 with 350 µL volume.
  5. Remove the assembled device from the centrifuge and separate the filter device from the microcentrifuge tube.
  6. To recover the conjugate, place the filter device upside down in a clean centrifugal tube, aligning open cap towards the center of the rotor; counterbalance with a similar device. Spin for 2 minutes at 1,000 x g to transfer the sample from the device to the tube.

Compatibility Sheet

Buffer additives

The conjugation of ligands with all available coupling kits can be performed with many different additives. The following list shows all tested ones, but please note that others not listed here may also be successfully used.

Additive Up to Suitability Amine Coupling Suitability Thiol Coupling
EDTA 1 mM
DTT* 1 mM
TCEP 1 mM
Tris** 1 mM
DMSO 2 %
ATP 0.5 mM
MgCl2 2.5 mM
Glycine**
Mannitol 8 %
Glycerol 10 %
Trehalose 8 %
Histidin** 30 mM
Acetonitrile*** 50 %
Trifluoroacetic acid 0.1 %

* thiol-based reducing agents
** contains primary amines
*** caution, may harm the ligand

pH/pI

The pH value for the conjugation buffer may range from pH 5.0 to pH 8.0, depending on the ligand characteristics. When performing a conjugation of proteins with a pI of < 6, please note that using a buffer with lower pH may result in a better yield of conjugate.

Buffer pH Order No Composition
Phosphate-Citrate Buffer pH 5 50 mM buffer salt, 150 mM NaCl
Buffer M pH 6.5 BU-M-150-1 50 mM MES, 150 mM NaCl
Buffer A pH 7.2 BU-P-150-10 50 mM Na2HPO4/NaH2PO4, 150 mM NaCl
Buffer C pH 8.0 BU-C-150-1 50 mM Na2HPO4/NaH2PO4, 150 mM NaCl

Salt concentration

For standard conjugations, 50 mM buffer salt and 150 mM NaCl (monovalent salt) are used. When performing conjugation of strongly charged ligands, make sure that the concentration of NaCl is sufficiently high (up to 400 mM NaCl is recommended). Otherwise, precipitation of DNA may occur. The shielding effect of monovalent sodium cations leads to DNA stabilization through neutralization of the negative charge on the sugar phosphate backbone.

Useful Order Numbers

Table 3. Order Numbers

Product Name Amount Order No
Y-structure Amine coupling kit 2 - Green 3 conjugations HK-NYS-NHS-2
Centrifugal filter unit (10 kDa MWCO) 5 pcs. CF-010-5
10x Buffer A [1] 50 mL (yielding 500 mL) BU-P-150-10
5x Buffer B [5] 50 mL (yielding 250 mL) BU-P-1000-5
1x Buffer C [2] 12 mL BU-C-150-1

Contact

Dynamic Biosensors GmbH
Perchtinger Str. 8/10
81379 Munich
Germany

Bruker Scientific LLC
40 Manning Road, Manning Park
Billerica, MA 01821
USA

Order Information: order.biosensors@bruker.com
Technical Support: support.dbs@bruker.com

www.dynamic-biosensors.com

Instruments and chips are engineered and manufactured in Germany.
©2025 Dynamic Biosensors GmbH
For Research Use Only. Not for use in clinical diagnostic procedures.

[1] Buffer A: 50 mM Na2HPO4/NaH2PO4, 150 mM NaCl, pH 7.2
[2] Buffer C: 50 mM Na2HPO4/NaH2PO4, 150 mM NaCl, pH 8.0
[3] Buffer PE40: 10 mM Na2HPO4/NaH2PO4, 40 mM NaCl, pH 7.4, 0.05 % Tween, 50 µM EDTA, 50 µM EGTA
[4] For conjugation of proteins with a molecular weight higher than 20 kDa: Centrifugal filter units with a MWCO of 10 kDa can be ordered for a faster concentration process (Order No: CF-010-5).
[5] Buffer B: 50 mM Na2HPO4/NaH2PO4, 1 M NaCl, pH 7.2

Models: HK-NYS-NHS-2, heliX plus HK-NYS-NHS-2 Y-Structure Amine Coupling Kit 2, heliX plus HK-NYS-NHS-2, Y-Structure Amine Coupling Kit 2, Amine Coupling Kit 2, Coupling Kit 2, Kit 2

File Info : application/pdf, 13 Pages, 1.84MB

PDF preview unavailable. Download the PDF instead.

HK-NYS-NHS-2

References

HexaPDF version 1.4.0

Related Documents

Preview Bruker Contour X-200 Optical Profiler User Manual
User manual for the Bruker Contour X-200 optical profiler, detailing its capabilities, principles of operation (PSI and VSI), and step-by-step instructions for surface measurements and data analysis.
Preview Bruker TopSpin NMR Software User Manual - Release Letter 4.1.4
Comprehensive user manual and release letter for Bruker's TopSpin NMR software, version 4.1.4, detailing features, bug fixes, and updates for NMR data processing and analysis.
Preview DCMSLink for Bruker Compass/Hystar Troubleshooting Guide
A comprehensive troubleshooting guide for DCMSLink, providing solutions for issues related to Bruker Compass and Hystar software.
Preview Bruker PeakForce QNM User Guide: Operation, Calibration, and Analysis
Comprehensive user guide for Bruker's PeakForce QNM (Quantitative NanoMechanics) Atomic Force Microscopy mode. Learn about operation, calibration procedures, parameter settings, and offline analysis for measuring nano-scale material properties such as modulus, adhesion, deformation, and dissipation.