FNIRSI 1013D Oscilloscope
User Manual
General Presentation
- 100MHz analog bandwidth @ 1GSa/s sampling rate (1X = 5MHz, 10X = 100MHz).
- 7-inch full-color TFT LCD display with 800x480 resolution, bright colors, and high contrast.
- Wide measurement voltage range: 1X measures 0-40V, 10X measures 0-400V, 100X measures 0-4000V.
- Measures up to 12 parameters: VPP, VP, Vmax, Vmin, Vavg, Vrms, Frequency, Duty+, Duty-, Time+, Time-, Period.
- Complete triggering functions (single, normal, automatic).
- One-button curve and screenshot editing.
- 1GB built-in storage for up to 1000 screenshots + 1000 curve data sets.
- USB connectivity for computer connection and screenshot sharing.
- Lissajous figures for determining amplitude, frequency, and phase difference of two signal groups.
- Built-in 6000mAh rechargeable lithium battery, providing 4 hours of continuous use at maximum brightness on a full charge.
Notice
The bandwidth of the 1X probe is 5 MHz, and the bandwidth of the 10X probe is 100 MHz. For measurements above 5 MHz, the probe switch must be set to 10X, and the oscilloscope must be set to 10X. Failure to do so will result in significant signal attenuation. The oscilloscope's own probe cable has a capacitance of 100-300 pF, which is substantial for high-frequency signals. The signal is considerably attenuated by the probe before reaching the oscilloscope's input, resulting in an equivalent bandwidth of 5 MHz. Therefore, to match the hundreds of pF of the probe cable, the input is attenuated tenfold (switch in 10X range). A few hundred pF capacitors are used solely for impedance matching. At this point, the bandwidth is 100 MHz. Note that only probes with a bandwidth of 100 MHz or higher can be used.
Screen Instructions
The screen displays various parameters and controls:
- Channel 1 (CH1) and Channel 2 (CH2) Indicators: Indicate the parameters associated with each channel. Clicking these areas displays the respective channel control panel.
- Probe Magnification (1X, 10X, 100X): Settings for the probe magnification for both channels.
- Time Base: Indicates the time duration represented by a large horizontal grid division, determined by the sampling rate. A larger time base means a slower sampling rate, and vice versa.
- Gesture Movement Speed: Adjusts the speed of movement during gesture operations ([Move coarse adjustment] for fast movement, [Move fine adjustment] for slow movement).
- Trigger Indicator: Shows trigger-related parameters. Clicking this area displays the trigger control panel.
- Trigger Channel: Options for CH1 or CH2 triggering.
- Battery Indicator: A green block indicates remaining battery power.
- Main Interface Control Panel Button: Switches between two types of button panels.
- Pause Button: Toggles between run and pause modes.
- Auto Set Button: Automatically identifies the signal and sets the system to optimal parameters for waveform display.
- Time Cursor Button: Activates/deactivates time cursor measurement.
- Voltage Cursor Button: Activates/deactivates voltage cursor measurement.
- Measurement Parameter Selection: Displays the parameter selection panel. Clicking a parameter opens its display.
- One-Click Screenshot Button: Captures the entire screen and saves it to internal memory.
- Waveform Save Button: Saves all waveform data for open channels to internal storage.
- Channel 1 Vertical Sensitivity Adjustment: Buttons to increase or decrease vertical sensitivity for Channel 1.
- Channel 2 Vertical Sensitivity Adjustment: Buttons to increase or decrease vertical sensitivity for Channel 2.
- Trigger Voltage Indicator: Shows the trigger threshold.
- 50% Trigger Button: Sets the trigger voltage to the midpoint of the waveform's amplitude.
- Waveform Data Display: Shows waveform data for Channel 1 and Channel 2.
- Channel 2 Baseline Position Indicator: Indicates the baseline position for Channel 2.
- Channel 1 Baseline Position Indicator: Indicates the baseline position for Channel 1.
- Trigger Position Indicator: Shows the trigger point (X position).
- Trigger Mode Indicator: Displays the trigger mode (Auto, Single, Normal).
- Trigger Edge Indicator: Shows the trigger edge (rising or falling).
Touch Control Panel
The touch control panel allows for direct interaction with the oscilloscope's functions:
- 1: Function Menu Button: Accesses the system function menu and settings.
- 2: Channel 1 Control Panel: Opens the parameter control panel for Channel 1.
- 3: Channel 2 Control Panel: Opens the parameter control panel for Channel 2.
- 4: Movement Speed Button: Switches between fast and slow movement.
- 5: Trigger Control Bar Button: Displays the trigger parameter control bar.
- 6: Trigger Voltage Button: Adjusts the trigger voltage (trigger threshold) by scrolling up or down.
- 7: Time Base Zoom Area (Horizontal): Expands or contracts the waveform horizontally. The area to the right of the center grid line is the key display area.
- 8: Time Base Zoom Area (Horizontal): Expands or contracts the waveform horizontally. The area to the left of the center grid line is the key display area.
- 9: Channel 2 Horizontal Baseline Button: Adjusts the baseline position for Channel 2. The waveform can also be moved by clicking and dragging it. Buttons for left and right waveform movement are also displayed.
- 10: Channel 1 Horizontal Baseline Button: Adjusts the baseline position for Channel 1. The waveform can also be moved by clicking and dragging it. Buttons for left and right waveform movement are also displayed.
Usage Instructions
- Activating/Deactivating Channels: Click the yellow area in the top-left corner to display the channel control panel, then click [ON] or [OFF] to activate or deactivate the channel.
- Activating/Deactivating FFT: Click the yellow area in the top-left corner to display the channel control panel, then click [ON] or [OFF] in the [open FFT] column to activate or deactivate FFT.
- Setting Input Coupling: Click the yellow area in the top-left corner to display the channel control panel, then click [DC] or [AC] in the [Coupling] column to set the input coupling.
- Setting Probe Magnification: Click the yellow area in the top-left corner to display the channel control panel, then click [1], [10], or [100] in the [Probe Mode] column to set the probe magnification.
- Zooming Waveforms: Click the right half of the waveform display area to zoom horizontally (decrease time base). Click [CTRL] and then [V+] to zoom vertically. Click the left half of the waveform display area to zoom out horizontally (increase time base). Click [CTRL] and then [V-] to zoom out vertically.
- Moving Waveforms: Click and drag the waveform or waveform area to move it in any direction.
- Setting Trigger Voltage: Click the green arrow on the right of the screen and drag it up or down to set the trigger voltage.
- Setting Trigger Interface: Click the green area of the T control panel in the top-left corner to display the trigger parameter control panel. Click [Rising] or [Falling] in the [Trigger Mode] column to set the trigger edge.
- Setting Automatic Trigger: Click the green area of the T control panel in the top-left corner to display the trigger parameter control panel. Click [Auto] in the [Trigger Mode] column to set the automatic trigger.
- Setting Single Trigger: Click the green area of the T control panel in the top-left corner to display the trigger parameter control panel. Click [Single] in the [Trigger Mode] column to set the single trigger.
- General Trigger Settings: Click the green area of the T control panel in the top-left corner to display the trigger parameter control panel. Click [Normal] in the [Trigger Mode] column to set the normal trigger.
- Pausing Display: Press the [RUN/STOP] button to pause/stop the system. If this button is not visible, press [CTRL] to switch to the button bar containing it.
- Automatic Waveform Setting: Press the [AUTOSET] button to automatically optimize settings based on signal characteristics. If this button is not available, press [CTRL] to switch to the button bar containing it.
- Slow Scroll Mode: Click the left half of the progress display area to increase the time base. When the time base reaches 100 ms, the system enters slow scroll mode (100 ms - 50 s).
- Cursor Measurement: Press the [Time Cursor] button to manually measure time differences and the [V CURSOR] button to manually measure voltage. If these buttons are not available, press [CTRL] to switch to the button bar containing them.
- Setting Display Parameters: Press the [MEASURS] button to open the parameter control panel and select a parameter to display its details. If this button is not available, press [CTRL] to switch to the button bar containing it.
- Saving Current Screen View: Press the [SAVE PIC] button to capture a screenshot and save it to internal memory. If this button is not available, press [CTRL] to switch to the button bar containing it.
- Saving Current Waveform Data: Press the [SAVE WAVE] button to save all open channel waveform data to internal storage. If this button is not available, press [CTRL] to switch to the button bar containing it.
- Setting Screen Brightness: Navigate to [MENU] > [System Settings] > [Screen Brightness] to adjust brightness (0-100).
- Setting Grid Brightness: Navigate to [MENU] > [System Settings] > [Grid Brightness] to adjust grid brightness (0-100).
- Auto Trigger at 50%: Navigate to [MENU] > [System Settings] > [Always Trigger 50%] to set the auto trigger to 50%.
Troubleshooting Common Issues
- No Power: Ensure the device is charged. Charge for at least half an hour using the original charger.
- No Waveform Display: Check if the pause function is active. Try [Auto Adjust]. Ensure the signal source is outputting a signal and the probe is properly connected.
- Voltage Reading is 0: Adjust vertical sensitivity and time base. Use [AUTOSET] to display a clear waveform.
- Frequency Reading is 0: Ensure the trigger mode is set to Auto. If it remains 0, use [AUTOSET].
- Cycle Reading is 0: Ensure the trigger mode is set to Auto. Adjust the trigger to be within the waveform.
- AC and DC Waveforms are the Same: This can occur with symmetric AC signals or when AC/DC coupling is switched.
- Waveform Jittering: Set the trigger mode to Auto and use [AUTOSET]. Check probe grounding and connection.
- Unstable Waveform: Adjust the trigger voltage. Ensure the signal is properly grounded.
- Cannot Detect Pulse or Digital Logic Signals: Set the trigger mode to Normal or Single, and adjust trigger voltage, time base, and vertical sensitivity.
- No Waveform for Battery or DC Voltage: In DC link mode, adjust vertical sensitivity to display a straight line.
- Charging Issues: Use a 5V 2A charger. Laptop USB ports may provide insufficient power.
- Static Waveform at 50Hz: This is normal for 50Hz signals; the oscilloscope adjusts sampling rate accordingly.
- VPP Reading Lower Than Expected: The displayed value might be RMS. Check the parameter display for RMS voltage.
- Distorted AC Waveform: Household power grids often have harmonics that cause distortion, which is normal.
- Baseline Offset: Perform baseline calibration by navigating to [MENU] > [System Settings] > [Baseline Calibration].
- Signal Attenuation Above 5 MHz: Ensure the probe and oscilloscope are set to 10X for measurements above 5 MHz.
- Green Charging Light Not On: The light indicates charging is in progress. If the battery icon is full, the battery is fully charged.
Common Circuit Testing Methods
Battery Voltage or DC Voltage Measurement
- Set the oscilloscope to automatic trigger mode.
- Set the oscilloscope to the appropriate position (1X or 10X).
- Set the oscilloscope to DC mode.
- Connect the device and probe.
- Verify battery charge or DC voltage.
- Connect the probe to the negative and positive terminals of the battery.
- Press [AUTOSET] to display the DC signal as a straight line.
Crystal Oscillation Measurement
- For crystal oscillation, use 10X probe and oscilloscope settings due to input capacitance.
- Set the oscilloscope to automatic trigger mode.
- Set the oscilloscope to 10X position.
- Set the oscilloscope to AC mode.
- Connect the probe and ensure the crystal oscillator is powered on.
- Connect the probe to the crystal pins.
- Press [AUTOSET] to display the waveform and adjust vertical sensitivity if needed.
PWM Signal Measurement (MOS or IGBT)
- PWM control signals are typically 10-20V. 1X probe is sufficient.
- Set to automatic trigger mode.
- Set oscilloscope to 1X position.
- Set oscilloscope to DC mode.
- Connect probe to MOS tube S and G terminals.
- Press [AUTOSET] to display the PWM waveform and adjust as needed.
Signal Generator Output Measurement
- Signal generator output voltage is typically up to 30V. 1X probe is sufficient.
- Set to automatic trigger mode.
- Set oscilloscope to 1X position.
- Set oscilloscope to DC mode.
- Connect probe to signal generator output terminals.
- Press [AUTOSET] to display the waveform and adjust as needed.
Home Network Voltage Measurement (220V/110V)
- Household electricity is typically 180-260V. Use a 10X probe for measurements up to 400V, or a 100X probe for up to 4000V. A standard 10X high-voltage probe is recommended.
- Set to automatic trigger mode.
- Set oscilloscope to 100X position.
- Set oscilloscope to AC mode.
- Connect probe to household power outlet.
- Press [AUTOSET] to display the waveform and adjust as needed.
Waveform Energy Measurement
- For output voltages below 40V, use 1X settings. For 40-400V, use 10X settings.
- Set to automatic trigger mode.
- Set oscilloscope to the appropriate position (1X or 10X).
- Set oscilloscope to AC mode.
- Connect probe and ensure power supply is on.
- Connect probe to negative and positive terminals of the power supply output.
- Press [AUTOSET] to display the power waveform.
Inverter Output Measurement
- Inverter output voltage is similar to household voltage, with peak-to-peak exceeding 500V. Use 10X or 100X probes accordingly.
- Set to automatic trigger mode.
- Set oscilloscope to 100X position.
- Set oscilloscope to DC mode.
- Connect probe to inverter output terminals.
- Press [AUTOSET] to display the waveform and adjust as needed.
Amplifier or Audio Signal Measurement
- Amplifier output voltage is typically below 40V. 1X probe is sufficient.
- Set to automatic trigger mode.
- Set oscilloscope to 1X position.
- Set oscilloscope to AC mode.
- Connect probe to amplifier output terminals.
- Press [AUTOSET] to display the waveform and adjust as needed.
Automotive Communication Signal / Bus Signal Measurement
- Automotive communication signals are typically below 20V. 1X probe is sufficient.
- Set to normal trigger mode for non-periodic digital signals.
- Set oscilloscope to 1X position.
- Set oscilloscope to AC mode.
- Connect probe to communication line signal terminals.
- Set vertical sensitivity to 50 mV.
- Set time base to 20 uS.
- Press the [50% TRIG] button.
- If a communication signal is present, the oscilloscope will capture it. Adjust time base and trigger voltage if necessary.
Infrared Remote Control Receiver Measurement
- Infrared signals are typically 3-5V. 1X probe is sufficient.
- Set to normal trigger mode.
- Set oscilloscope to 1X position.
- Set oscilloscope to DC mode.
- Connect probe to the infrared receiver's data pin and ground.
- Set vertical sensitivity to 500 mV.
- Set time base to 20 uS.
- Adjust trigger position.
- Send a signal with the remote control to display the waveform.
Amplifier Circuit Measurement with Sensors
- Sensor signals are typically weak (millivolts). Use the amplifier's output.
- Set to automatic trigger mode.
- Set oscilloscope to 1X position.
- Set oscilloscope to DC mode.
- Connect probe to the amplified signal output and ground.
- Set vertical sensitivity to 50 mV.
- Set time base to 500 mS for slow sweep mode.
- Adjust baseline position.
- Reduce vertical sensitivity if the signal appears at the top.
Distributor Information
Sunnysoft s.r.o.
Kovarecká 2390/1a
190 00 Prague 9
Czech Republic
www.sunnysoft.cz
Related Documents
![]() |
FNIRSI-1014D Two-in-One Digital Oscilloscope and Signal Generator Product Instructions The FNIRSI-1014D is a cost-effective two-in-one digital oscilloscope and signal generator, ideal for maintenance and R&D. It features a 1GSa/s real-time sampling rate, 100MHz dual analog bandwidth, comprehensive trigger functions (single, normal, auto), and a built-in DDS signal generator with 14 standard waveforms. The device includes a high-voltage protection module (up to 400V), a 7-inch 800x480 LCD, cursor measurement, and 1GB storage for up to 1000 screen captures and waveform data. It supports Lissajous graphics and FFT viewing for signal analysis and offers convenient USB connectivity for data sharing. |
![]() |
FNIRSI-1014D Dual Channel Oscilloscope & Signal Generator The FNIRSI-1014D is a cost-effective dual-channel desktop oscilloscope and signal generator with 100MHz bandwidth and 1GSa/s sampling rate. It features high-voltage protection, one-button auto-adjustment, comprehensive trigger modes, cursor measurement, 1GB storage for waveforms and screenshots, USB export, and Lissajous graphic display, ideal for R&D and maintenance. |
![]() |
FNIRSI-1014D Two-in-One Digital Oscilloscope and Signal Generator User Manual Comprehensive user manual for the FNIRSI-1014D, a cost-effective dual-channel digital oscilloscope and signal generator. Learn about its 1GSa/s sampling rate, 100MHz bandwidth, trigger functions, DDS signal generator, waveform storage, and various measurement applications for maintenance and R&D. |
![]() |
FNIRSI DPOS350P 4-in-1 Oscilloscope User Guide Comprehensive user guide for the FNIRSI DPOS350P, a 4-in-1 multi-function tablet oscilloscope. Learn about its features, specifications, and operation, including its 350MHz oscilloscope, 50MHz signal generator, spectrum analyzer, and frequency response analyzer capabilities. |
![]() |
FNIRSI-5012H Digital Oscilloscope: 100MHz, 500MS/s Specifications Comprehensive technical specifications and features of the FNIRSI-5012H digital oscilloscope, including its 100MHz bandwidth, 500MS/s sampling rate, 2.4-inch LCD, and advanced measurement capabilities. |
![]() |
FNIRSI-1014D Dual Channel Digital Oscilloscope Product Instructions Comprehensive product instructions for the FNIRSI-1014D, a two-in-one digital oscilloscope and signal generator. This guide details its features, operation, and troubleshooting for maintenance and R&D industries. |
![]() |
FNIRSI DS215H Dual-Channel Oscilloscope and Signal Generator User Manual Comprehensive user manual for the FNIRSI DS215H, a 2-in-1 digital oscilloscope and signal generator with 50 MHz bandwidth and 250 MS/s sampling rate. Features include detailed operation, technical specifications, troubleshooting, and maintenance. |
![]() |
FNIRSI 2C23T Dual Channel Oscilloscope Multimeter Manual User manual for the FNIRSI 2C23T Dual Channel Oscilloscope Multimeter, covering product introduction, panel layout, device parameters, button functions, settings, firmware updates, and testing methods. |