Product Functions

TEC103 is primarily used for temperature measurement and control in optical components, such as lasers, detectors, and small sample chambers.

SenseFuture-TEC103L-Single-Channel-Temperature-Controller-FIG (1)

Product Features

  • Thermal measurement sensitivity of 0.1 mK, long-term drift (over 24 hours) less than 1 mK.
  • Temperature control stability of ±0.001°C, is suitable for most scenarios, including the stringent temperature control requirements for semiconductor lasers.
  • Optional bipolar or unipolar output.
  • Capable of limiting the maximum rate of temperature change.
  • Supports NTC (Negative Temperature Coefficient) thermistor temperature sensors.
  • Chip-level design, facilitating integration into circuit board designs.
  • Features overheat protection for the circuit board, ensuring reliable performance.
  • Allows direct parameter setting via the display control module, with settings retained in memory after power loss, simplifying production operations.
  • Provides a comprehensive set of serial port control commands, offering an open platform for customization and integration.

Product Parameters

Table1 Basic Parameters of TEC103

 

PARAMETERS

MODEL  

UNIT

TEC103L TEC103 TEC130

(Pending Launch)

24-hour Temperature Measurement Stability

(with the matched thermistor)

 

<0.001@20℃

 

<0.001@20℃

 

°C

Temperature drift caused by ambient temperature 0.0001 0.0001 °C/°C
Optimal Temperature Control Stability

(related to the overall system)

 

±0.01

 

±0.001

 

±0.001

 

°C

Temperature Change Limit Setting Range 0.01~2.5 0.01~2.5 °C/s
 

Temperature Setting Method

 

UART

UART

Analog Voltage: 1V =10kΩ

 
Power Supply Voltage (Short-term Maximum Voltage: 28V)  

7~24

 

7~24

 

V

Output Polarity Bipolar、Unipolar Bipolar、Unipolar V
Number of Channels 1 1  
Maximum Output Voltage ±90%Vin (Settable) ±90%Vin (Settable)  
Output Current Range 0~±3 0~±30 A
Ambient Temperature -55~60 -55~60 °C
Ambient Humidity 0~98 0~98 %RH
Thermal Dissipation Requirements No Additional Thermal Dissipation Needed Within Rated Operating Range  
Circuit Board Overheat Protection Yes  
Power Loss Memory Yes  
PID Parameters User Adjustable  
Size 46.5*39.0*9.6 —— mm
Weight ≈30 —— g

Interface Introduction

SenseFuture-TEC103L-Single-Channel-Temperature-Controller-FIG (2)

Pin Number Pin Name Pin Type  

Pin Definition (High Level: 3.3V, Low Level: 0V)

 

1

 

GND

 

Input

 

Power Input Negative Pole (Low Current).

 

 

2

 

 

STATE

 

 

Output

Temperature Control Status Output. High Level: Temperature control is functioning normally (temperature control error < 0.01°C). Low Level: Temperature control anomaly detected (temperature control error ≥ 0.01°C).

The temperature control standard of 0.01°C can be set.

 

3

 

ENABLE

 

Input

Output Enable Pin. High Level (Default): Enables temperature control output. Low Level: Disables temperature control output.
 

4

 

TX2

 

Output

Serial Port 2 Receiver, TTL Level, used for connecting to the Screen Display Control Module.
 

5

 

RX2

 

Input

Serial Port 2 Receiver, TTL Level, used for connecting to the Screen Display Control Module.
 

6

 

VCC

 

Output

3.3V Output, intended for connection to the screen display control module and not recommended for other uses.
 

7

 

TX1

 

Output

Serial Port 1 Receive End, TTL level, used for connecting to PC control software. Data bits: 8 bits, Stop bits: 1 bit, Parity: None, Baud Rate: 38400.
 

8

 

RX1

 

Input

Serial Port 1 Receive End, TTL level, used for connecting to PC control software. Data bits: 8 bits, Stop bits: 1 bit, Parity: None, Baud Rate: 38400.
 

9

 

NTC-

 

Input

Thermistor (NTC) Interface, compatible with different resistance values of NTC thermistors, with wiring polarity not required.
 

10

 

NTC+

 

Input

Thermistor (NTC) Interface, compatible with different resistance values of NTC thermistors, with wiring polarity not required.
 

11

 

GND

 

Input

 

Power Input Negative Pole (High Current).

 

12

 

GND

 

Input

 

Power Input Negative Pole (High Current).

 

13

 

Vin

 

Input

 

Power Input Positive Pole, with an input voltage range of 7 to 24V.

 

14

 

Vin

 

Input

 

Power Input Positive Pole, with an input voltage range of 7 to 24V.

 

15

 

TEC-

 

Output

The negative terminal of the temperature control current output is usually connected to the negative terminal of the Thermoelectric Cooler (TEC).
 

16

 

TEC-

 

Output

The negative terminal of the temperature control current output is usually connected to the negative terminal of the Thermoelectric Cooler (TEC).
 

17

 

TEC+

 

Output

The positive terminal of the temperature control current output is typically connected to the positive terminal of the Thermoelectric Cooler (TEC).
 

18

 

TEC+

 

Output

The positive terminal of the temperature control current output is typically connected to the positive terminal of the Thermoelectric Cooler (TEC).

SenseFuture-TEC103L-Single-Channel-Temperature-Controller-FIG (3)

Dimensional Drawing

SenseFuture-TEC103L-Single-Channel-Temperature-Controller-FIG (4)SenseFuture-TEC103L-Single-Channel-Temperature-Controller-FIG (5)

Computer Software

(Communication Protocol Refer to Attachment)

SenseFuture-TEC103L-Single-Channel-Temperature-Controller-FIG (6)Download:

https://drive.google.com/file/d/1-2Ruffh7yyJPImV5w0OIUB0lyTqhbeRD/view?usp=sharing

Instructional Video

YOUTUBE【SenseFuture】±0.001℃ Temperature Controller (TEC103 Series) Instructions for Use —— DFB Laser Temperature Control
https://www.youtube.com/watch?v=exZvXJUNZ1c

Selection Guide

Table3 Temperature Controllers Selection Guide

SenseFuture-TEC103L-Single-Channel-Temperature-Controller-FIG (7)

Temperature Measurement Range and Sensitivity of Temperature Sensors Compatible with

TEC103/207/215

 

Sensitivity

NTC (500k B4250) NTC (100k B3950) NTC (10k B3950) NTC (1k B3470)  

PT1000

 

PT100

≤±0.001℃ 60~300℃ 25~210℃ -20~150℃ -60~70℃ -200~800℃ ——
≤±0.01℃ 300~470℃ 210~350℃ 150~200℃ 70~110℃ —— -200~800℃
≤±0.1℃ 470~550℃ 350~500℃ 200~290℃ 110~180℃ —— ——

Temperature Measurement Range and Sensitivity of Temperature Sensors Compatible with TEC103L/207L/215L

 

Sensitivity

NTC (500k B4250) NTC (100k B3950) NTC (10k B3950) NTC (1k B3470)  

PT1000

 

PT100

≤±0.01℃ 60~400℃ 25~290℃ -20~180℃ -60~100℃ -200~800℃ ——
≤±0.1℃ 400~550℃ 290~430℃ 180~280℃ 100~130℃ —— -200~800℃
≤±1℃ —— 430~550℃ —— 130~180℃ —— ——

Customized Temperature Control System Services

We offer complete temperature control solutions, providing custom temperature control systems for institutions such as the National Institute of Metrology of China, the Anhui Institute of Optics and Fine Mechanics, Nanjing University, and Shenzhen University.

For customized temperature control systems, please contact our technical support at +86 191 2054 5883(WhatsApp ID same as phone number)

Attachment 1. Typical Application Cases

01 DFB Semiconductor Laser Temperature Control Case Study

  • Temperature Control Object Information: A domestically manufactured Distributed Feedback (DFB) laser diode operating at a wavelength of 1370nm and a power output of 10mW.
  • Temperature Sensor Specification: The laser module incorporates an NTC 10K B3950 thermistor internally.
  • Heating/Cooling Device: The laser features an integrated thermoelectric cooler (TEC) capable of 1.5A at 2.6V.
  • Temperature Controller Brand and Model: SenseFuture™ TEC103.
  • Target Temperature: 25°C.
  • Temperature Controller Settings: Power supply voltage is 12V, with a maximum output voltage percentage set at 20% (i.e., 12V × 20% = 2.4V); PID parameters configured as P = 200, I = 100, D = 0, with a positive hysteresis duty cycle of 0.005%, and a negative hysteresis duty cycle also at 0.005%.
  • Measured Results: The actual temperature stability achieved is ±0.0005°C after 5 hours of testing under ambient conditions of 25±1.5°C, and ±0.0005°C maintained over a 24-hour period, again within an ambient range of 25±1.5°C.
    • (Need a specific solution? Please consult technical support for quotation at +86 191 2054 5883)SenseFuture-TEC103L-Single-Channel-Temperature-Controller-FIG (8)SenseFuture-TEC103L-Single-Channel-Temperature-Controller-FIG (9)

ICL Semiconductor Laser Temperature Control Case Study

  • Performance is similar to 01, with specific details to be shared upon update.
    • (Need a specific solution? Please consult technical support for quotation at +86 191 2054 5883)

LD Laser Diode Temperature Control Case Study

  • Performance is similar to 01, with specific details to be shared upon update.
    • (Need a specific solution? Please consult technical support for quotation at +86 191 2054 5883)

QCL Temperature Control Case Study

  • Temperature Control Object Details: QCL (Quantum Cascade Laser) with a wavelength of 4332nm and a power output of 100mW.
  • Temperature Sensor: An internal NTC 10K B3950 thermistor built into the laser.
  • Heating/Cooling Device: An integrated thermoelectric cooler (TEC) within the laser operating at 7V.
  • Temperature Controller Brand and Model: SenseFuture™ TEC103.
  • Target Temperature: 47°C.
  • Temperature Controller Settings: Supply voltage is 12V, with a maximum output voltage setting of 20% (corresponding to 12V x 20% = 2.4V), PID parameters configured as P = 5000, I = 500, and D = 0.
  • Actual Test Results: Achieved temperature stability was ±0.001°C over a 1-hour test period.
    • (Need a specific solution? Please consult technical support for quotation at +86 191 2054 5883)SenseFuture-TEC103L-Single-Channel-Temperature-Controller-FIG (10)

MCT Detector Temperature Control Case Study

SenseFuture-TEC103L-Single-Channel-Temperature-Controller-FIG (11)

  • Temperature Control Object Information: MCT Detector from brand VIGO.
  • Temperature Sensor: Built-in NTC 2K B3950 thermistor inside the detector.
  • Heating/Cooling Element: Integrated thermoelectric cooler (TEC) within the detector rated at 1V and 100mA.
  • Temperature Controller Brand and Model: SenseFuture™ TEC103.
  • Target Temperature: 25°C.
  • Temperature Controller Settings: Power supply voltage is 9V with a maximum output voltage percentage of 3% (which translates to 9V × 3% = 0.27V), PID parameters set to P = 15, I = 5, and D = 0.
  • Measured Results: Achieved temperature stability of ±0.0025°C over a 14-hour test period.
    • (Need a specific solution? Please consult technical support for quotation at +86 191 2054 5883)SenseFuture-TEC103L-Single-Channel-Temperature-Controller-FIG (12)

Partners

  • Universities and Research InstitutesSenseFuture-TEC103L-Single-Channel-Temperature-Controller-FIG (13)
  • Optical Instrument Technology CompanySenseFuture-TEC103L-Single-Channel-Temperature-Controller-FIG (14)

CONTACT INFORMATION

Download

SenseFuture-TEC103L-Single-Channel-Temperature-Controller-FIG (15)

Documents / Resources

SenseFuture TEC103L Single Channel Temperature Controller [pdf] Installation Guide
TEC103L, TEC103, TEC103L Single Channel Temperature Controller, TEC103L, Single Channel Temperature Controller, Channel Temperature Controller, Temperature Controller, Controller

References

Leave a comment

Your email address will not be published. Required fields are marked *