PPI LabCon Ultra Multi Purpose Temperature Controller with Recording and PC Software

PPI LabCon Ultra Multi Purpose Temperature Controller with Recording and PC Software

LabCon Ultra

This brief manual is primarily meant for quick reference to wiring connections and parameter searching. For more details on operation and application; please log on to www.ppiindia.net

Level Sub-Level Parameters Range (Default)

SymbolOperator

Temperature Set Value Temp. Setpoint Low Limit  to Temp. Setpoint High Limit (Default : 25.0 °C)
Temperature

Low Deviation Alarm

0.2 to 99.9 °C

(Default : 2.0 °C)

Temperature

High Deviation Alarm

0.2 to 99.9 °C

(Default : 2.0 °C)

SP Edit on Home Screen Enable Disable
Timer Start / Abort Command NA
Time Interval (HH:MM) 0.00 to 500.00

(Default : 0.10)

Zero Offset -50.0 to 50.0°C

(Default : 0.0°C)

Supervisory

Recording

Recording Interval 0 to 250 Minutes (Default : 5 Minutes)
‘Delete Record’ Command NA

Clock & Calendar

Calendar Date (DD/MM/YY) NA
Clock Time (HH:MM:SS) NA

Control Settings

Set Value (Temperature Control Loop) Temp. Setpoint Low Limit to Temp. Setpoint High Limit (Default : 25.0)
Low Limit (for Temp. Control Set Value) -199.9 to Temperature Setpoint High Limit

(Default : 10.0)

High Limit (for Temp. Control Set Value) For RTD : Temperature Setpoint Low Limit to 600.0

For mA/V : Temperature Setpoint Low Limit to 999.9

(Default : 60.0)

Overshoot Inhibit Enable Disable

(Default : Disable)

Cutoff Factor 1.0 to 2.0 (Default : 1.2)
Control PV Strategy MAP-0 PV, Average PV if MAP-0 Fails, Average PV

(Default : MAP-0 PV)

Self Tune NA
Compressor SP 0.0 to 100.0

(Default : 45.0)

Cool Hysteresis 0.1 to 99.9

(Default : 2.0)

Dual Zone
Heat Zone PID Constants

Heat Zone Proportional Band

0.1 to 999.9 °C

(Default : 50.0 °C)

Heat Zone Integral Time 0 to 3600 Sec. (Default : 100 Sec.)
Heat Zone Derivative Time 0 to 600 Sec. (Default : 16 Sec.)
Cool Zone PID Constants

Cool Zone Proportional Band

0.1 to 999.9 °C

(Default : 50.0 °C)

Cool Zone Integral Time 0 to 3600 Sec. (Default : 100 Sec.)
Cool Zone Derivative Time 0 to 600 Sec. (Default : 16 Sec.)
Single Zone
Proportional Band 0.1 to 999.9 °C

(Default : 50.0 °C)

Integral Time 0 to 3600 Sec. (Default : 100 Sec.)
Derivative Time 0 to 600 Sec. (Default : 16 Sec.)
Single Zone & Dual Zone
Output Cycle Time (Sec.) 0.5 to 100.0 Sec. (Default : 10.0 Sec.)
Level Sub-Level Parameters Range (Default)

Supervisory

SMS Alert

GSM Machine ID 1 to 128

(Default : 1)

Reset GSM Module NA

Maintenance

Repair Acknowledge ‘Control Gadget’ NA

Standby Switching

‘Switch Main/Standby Outputs’ NA

Door Lock Access

Lock Position (On / Off) Solenoid ON
Solenoid OFF
(Default : Solenoid OFF)
Password Entry NA

Factory

MAP-0 Sensor

Input Type TC-J, TC-K, TC-P, TC-R, TC-S, TC-B, TC-N, RTD Pt100, 0 to 20 mA, 4 to 20 mA, 0 to 5 V, 0 to 10 V, 1 to 5 V

(Default : RTD Pt100)

Signal Range Low
Input Type Settings Default
0 to 20mA 0.00 to

Signal High

0.00
4 to 20mA 4.00 to Signal High 4.00
0 to 5V 0.000 to Signal High 0.000
0 to 10V 0.00 to Signal High 0.00
1 to 5V 1.000 to Signal High 1.000
Signal Range High
Input Type Settings Default
0 to 20mA Signal Low to 20.00 20.00
4 to 20mA Signal Low to 20.00 20.00
0 to 5V Signal Low to 5.000 5.000
0 to 10V Signal Low to 10.00 10.00
1 to 5V Signal Low to 5.000 5.000
Display Range Low -199.9 to Range High (Default : 0.0)
Display Range High Range Low to 999.9 (Default : 100.0)
Zero Offset -50.0 to 50.0°C

(Default : 0.0°C)

Heat-Cool Select

Mode Continuous OFF, Continuous ON, SP Based ON-OFF, PV Based ON-OFF

(Default : Continuous OFF)

Time Delay 0 to 1000 Sec. (Default : 200 Sec.)
Boundary SP 0.0 to 100.0 (Default : 45.0)
Zone Select Single, Dual (Default : Single l)

Alarm Settings

Inhibit Enable, Disable (Default : Enable)
Low Alarm Deviation High Alarm Deviation 0.2 to 99.9

(Default : 2.0)

Hysteresis 0.1 to 99.9

(Default : 0.2)

Timer

Timer Enable Disable

(Default : Disable)

On Timer END Heater Off Yes / No (Default : Yes)
On Timer END Compressor Off Yes / No (Default : Yes)
Hold Band Type None, Up, Down, Both (Default : None)
Hold Band Value 0.1 to 999.9

(Default : 0.5)

Start Band Value 0 to 999.9

Default : 0.5

Power Up recovery Abort, Restart, Continuous Default : Abort
Level Sub-Level Parameters Range (Default)

Factory

Power Fail

Detection Enable No / Yes (Default : No)
Power Fail Logic Switch Open, Switch Close (Default : Switch Close)

Door Open

Detection Enable No / Yes (Default : No)
Door Open Logic Switch Open, Switch Close (Default : Switch Close)
Alarm Delay (Sec.) 0 to 1000 Sec. (Default : 60 Sec.)

Mapping

Configuration

Select Mapping Inputs 1 MAP (M0),

2 MAP (M0+M1),

3 MAP (M0+M1+M2),

4 MAP (M0+M1+M2+M3), 5 MAP (M0+M1+M2+M3+M4)

(Default : 1 MAP (M0)

Input Settings

Input Type TC-J, TC-K, TC-P, TC-R,

TC-S, TC-B, TC-N, RTD Pt100, 0 to 20 mA, 4 to 20 mA, 0 to 5 V, 0 to 10 V, 1 to 5 V (Default : RTD Pt100)

Signal Range Low
Input Type Settings Default
0 to 20mA 0.00 to Signal High 0.00
4 to 20mA 4.00 to Signal High 4.00
0 to 5V 0.000 to Signal High 0.000
0 to 10V 0.00 to Signal High 0.00
1 to 5V 1.000 to Signal High 1.000
Signal Range High
Input Type Settings Default
0 to 20mA Signal Low to 20.00 20.00
4 to 20mA Signal Low to 20.00 20.00
0 to 5V Signal Low to 5.000 5.000
0 to 10V Signal Low to 10.00 10.00
1 to 5V Signal Low to 5.000 5.000
Display Range Low -199.9 to Range High (Default : 0.0)
Display Range High Range Low to 999.9 (Default : 100.0)

Standby

Fail Detect Time (Min) 0 to 250 Min. (Default : 10 Min.)
Cyclic Time (Hrs.) 0 to 500 Hrs. (Default : 48 Hrs.)
Inhibit Time (Hrs.) 0 to 250 Hrs. (Default : 1 Hrs.)

Control Settings

Heater Output Type (Main & Standby Heater) SSR 0-20 mA
0-4 mA
0-5 V
0-10 V
(Default : SSR)

ELECTRICAL CONNECTIONS

HMI (NEW Version)

HMI (NEW Version)

HMI (NEW Version) 3-Pin Male / Female Connector (5.08 mm pitch)
Supply Voltage : 20 to 28 VDC (24 V Nominal)

HMI (NEW Version) 9 Pin D Type Connector
RS485 Serial Communication with Control Unit & PC
microPLC Electrical Connections

microPLC Electrical Connections

microPLC Mounting

microPLC Mounting

MECHANICAL MOUNTING

New Version

New Version

Dimensions

Overall 204(W) X 145(H) X 34(D), mm
Panel Cutout  192(W) X 138(H), mm
Old Version

Old Version

Dimensions

Overall 204(W) X 145(H) X 44.5(D), mm
Panel Cutout  192(W) X 138(H), mm

Customer Support

101, Diamond Industrial Estate, Navghar,
Vasai Road (E), Dist. Palghar – 401 210.
Sales : 8208199048 / 8208141446
Support : 07498799226 / 08767395333
E: sales@ppiindia.net, support@ppiindia.net

Logo

Documents / Resources

PPI LabCon Ultra Multi Purpose Temperature Controller with Recording and PC Software [pdf] Instruction Manual
LabCon Ultra Multi Purpose Temperature Controller with Recording and PC Software, LabCon, Ultra Multi Purpose Temperature Controller with Recording and PC Software, Temperature Controller with Recording and PC Software, Recording and PC Software, PC Software

References

Leave a comment

Your email address will not be published. Required fields are marked *