LABMAN logoLMTB200 Turbidity Meter
User GuideLABMAN LMTB200 Turbidity Meter

Salient Features

  • Turbidity meter with TSS measurement mode, selectable four turbidity units
  • Turbidity meter is equipped with a 4.5″ TFT display.
  • 2 to 7 points calibration using the Formazin Standards.
  • Selectable four turbidity units, including the NTU, FNU, EBC and ASBC.
  • TSS conversion factor ensures accurate measurement for Total Suspended Solids.
  • Continuous measurement mode that can used for indexing and matching the sample vials.
  • Hold function freezes current reading for easy viewing and recording.
  • Password protection prevents the unauthorized calibration and settings.
  • Help message as a operational guide that helps user quickly begin using the meter.
  • System menu allows setting the 9 parameters, including the date and time, resolution, etc.
  • Reset feature automatically resumes all settings back to factory default settings.

SPECIFICATIONS

MODEL LMTB200
Make LABMAN
Measurement Range 0-2000 NTU, 0-500 EBC, 0-9999 ASBC,
Resolution 0.01 (0-100 NTU), 0.1 (100-999 NTU), 1(999^2000 NTU)
Accuracy ±2% of reading for 0-500 NTU, ±3% of reading for 50172000 NTU
Calibration Points 2 to7 points ( 0.02, 10.00, 200, 500, 1000, 1500, 2000 NTU )
TSS Range deponding on the TSS conversion factor
TSS Accuracy ±3% of reading
Light Source Infrared-emitting diode (850 nm wavelength)
Detector Silicon Photodiode
Stray Light < 0.02 NTU
Sample Vials 60 (H) x 25 (Dia)mm
Memory Stores up to 200 data sets
Display 4.5″ TFT display
Operating Temperature 0∼50°C
Power Requirements DC12V/2A, using AC adapters, 220VAC/50Hz
Dimensions LxWxH (mm) & Weight 250 x 177 x 96 mm & 1.2kg

* Specifications are subject to change without notice.

Catalogue 2021-22
HSN Code: 90275090
STANDARD ACCESSORIES:
Empty Vial – 1no, Calibration Standards – 3
(200NTU, SOONTU, 1000NTU)
Cloth, 12V Power Adaptor, Test Cer tificate

Documents / Resources

LABMAN LMTB200 Turbidity Meter [pdf] User Guide
LMTB200 Turbidity Meter, LMTB200, Turbidity Meter, Meter

References

Leave a comment

Your email address will not be published. Required fields are marked *