Product Information

Specifications

  • Product Name: Vision Scope 2 Stereo Microscope
  • Model: T-22XXX Series T-26001
  • Manufacturer: Ken-A-Vision
  • Website: www.ken-a-vision.com

Frequently Asked Questions

  • What should I do if there are fingerprints or stains on the lens surfaces?
    • Clean the lens surfaces with a tissue moistened with a 3:7 mixture of alcohol and ether.
  • Can I use organic solutions to clean the plastic surfaces of the microscope?
    • No, organic solutions should not be used on plastic surfaces. If necessary, use a neutral detergent.
  • Should I disassemble the microscope after use?
    • No, it is not recommended to disassemble the microscope as it may cause damage.
  • How should I store the microscope after use?
    • Cover the microscope with the provided dust-cover and store it in a dry, clean environment to prevent rust.
  • How often should I check and clean the microscope?
    • The microscope should be checked and cleaned periodically to ensure proper performance.

Before Use

Notice

  1. Microscope should be stored in a dry and clean environment. Do not expose the microscope to the direct sunlight. Avoid high temperatures and violent vibration.
  2. This microscope is a precision instrument. Handle with care. Avoid impact and abrupt movement during transportation.
  3. To keep the image clear, clean any fingerprints or stains on the surfaces of the lens.
  4. Never turn the left and right focusing knob in the opposite direction at the same time. Microscope damage will occur.

Maintenance

  1. All lenses must be kept clean. Fine dust on the surface of the lens should be blown off with hand blower or wiped off gently with a soft lens tissue. Fingerprints and oil marks should be cleaned with a tissue moistened with a small amount of a 3:7 mixture of alcohol and ether.
  2. Never use organic solution to clean the other surfaces (especially the plastic surfaces). If necessary, please use a neutral detergent.
  3. Once assembled, do not take the microscope apart as this may cause damage to the microscope.
  4. After use, cover the microscope with the provided dust-cover and store in a dry, clean environment to prevent rust.
  5. To ensure proper performance, please check and clean the microscope periodically.

Nomenclature

KEN-A-VISION-T-26001-Vision-Scope-2-Stereo-Microscope-fig-1

Assembly

KEN-A-VISION-T-26001-Vision-Scope-2-Stereo-Microscope-fig-2

Operation

Position the glass stage

KEN-A-VISION-T-26001-Vision-Scope-2-Stereo-Microscope-fig-3

  1. Place the microscope base on a flat, firm surface and position the frosted stage plate on the base. (Figure 1)

Adjust the degree of tightness of the focusing arm

KEN-A-VISION-T-26001-Vision-Scope-2-Stereo-Microscope-fig-4

  1. Check the coarse focus knob on either side of the main support. The coarse focus knob should turn easily.
  2. If you want to adjust the degree of tightness of the focusing arm, you can hold one of the focusing knobs and turn another one to attain a suitable position. The degree of tightness relies on the direction to be turned. The clockwise direction is tight; counterclockwise is loose. (Figure 2)

Set the specimen slide

  1. Set the specimen in the center of the stage plate. If necessary, clamp the slide with the clips.
  2. To operate, plug the small round plug into the back of the base of the microscope and then plug the opposite end into the power outlet.

Adjust focusing knobs and diopter

KEN-A-VISION-T-26001-Vision-Scope-2-Stereo-Microscope-fig-5

  1. Turn the focusing knob and observe the specimen through the right eyepiece until the image of the specimen is clear.
  2. Observe the specimen through the left eyepiece and adjust the diopter ring 1 until the image is clear. (Figure 3)

Adjust the interpupillary distance

KEN-A-VISION-T-26001-Vision-Scope-2-Stereo-Microscope-fig-6

  1. Adjust the two eyepieces along the direction of the arrow shown in Figure 4 until eyepiece feels comfortable and a single image is seen.
  2. Use the left rheostat (dimmer knob) to adjust light intensity of the bottom (transmitted) light.
  3. Use the right rheostat (dimmer knob) to adjust light intensity of upper (incidental) light.

Configuration Chart

T-22XXX Series Configuration

Product Number   T-22001 T-22011 T-22021 T-22041 T-22051 T-22061
Magnifications   10x & 30x 15x & 45x 20x & 60x 20x & 40x 30x & 60x 40x & 80x
 

Eyepieces

2 – 10x О     О    
2 – 15x   О     О  
2 – 20x     О     О
Heads /

Objectives

1x & 3x О О О      
2x & 4x       О О О

Note: The items marked “О” are included; others are optional.

T-26001 Series Configuration

Zoom Magnifications   7x – 45x 10.5x – 67.5x 14x – 90x
 

Eyepieces

2 – 10x О    
2 – 15x   О  
2 – 20x     О
Objective Zoom

Range

 

0.7x – 4.5x

 

О

 

О

 

О

Note: The items marked “О” are included; others are optional.

Technical specifications

T-22XXX Series optical specifications

 

Objective Mag.

 

Working Distance (mm)

Eyepiece Eyepieces     (optional)
SC6EP10 SC6EP15 SC6EP20
 

Mag.

Objective

field

 

Mag.

Objective

field

 

Mag.

Objective

field

1X  

 

 

100

10X 20 15X 15 20X 11
2X 20X 10 30X 7.5 40X 5
 

3X

 

30X

 

6.7

 

45X

 

5

T-22XXX

SERIESX

 

3.3

 

4X

 

40X

 

5

T-22XXX

SERIESX

 

3.75

 

80X

 

2.5

Auxiliary objectives for T-22XXX Series

Auxiliary objectives Magnification Working distance(mm)
SC6OB5 0.5X 165
SC6OB15 1.5X 45
SC6OB20 2X 30
  • Working distance is fixed regardless of the magnification factor.
  • Total magnification = Objective mag. X Auxiliary mag. X Auxiliary mag.
  • Diameter of field of view (mm) =KEN-A-VISION-T-26001-Vision-Scope-2-Stereo-Microscope-fig-7
  • Photo adaptor mag. = Objective mag. X Auxiliary objective mag. X Photo eyepiece mag.

Configuration specifications of T-22XXX Series

Model  

T-22001

 

T-22041

 

 

 

 

 

 

 

Head

Objective

magnification

1X、3X 2X、4X
Working distance 100mm
Observation angle 45°
Interpupillary

distance adjustment

The linkage between left and right eyepiece tube range of single adjustment:     54-75mm
 

Diopter adjustment

Range of single adjustment:±5D
   
   
Mount with

auxiliary objectives

Screw hole: M48*0.75
Objective Field of view φ20mm
 

 

 

Main body

Mount the head Mount the head in the bracket hole (diameter: φ76mm)
 

Focusing device

Focus adjustment by turning the focusing knobs.

Range of single adjustment: 10 mm

 

Glass stage

Diameter:φ60mm
Clips Put it on the base from top

Troubleshooting

Trouble Cause Remedy
 

 

 

1.Double images

Interpupillary distance

correct

is not Adjust the interpupillary

distance

Diopter      adjustment

correct

is not  

Adjust the diopter

Left and right eyepieces are

different magnifications.

Mount eyepieces of the

same magnification

 

2.Dirt appears in the field of view

Dirt on the specimen  

Clean the specimen

Dirt on the surface of eyepiece Clean the surface
 

3.Image is not clear

Dirt on the surface

objective

of the  

Clean the objective

 

4.Image is not clear while adjusting the focus

Diopter      adjustment

correct

is not  

Adjust the diopter

Focus is not correct Adjust the focus
 

5.The focusing knob does not turn     smoothly

 

 

The focusing knob is too tight

 

Loosen it to a suitable position

 

6.The image is obscure because of the head  is slipping down by during observation

 

 

 

The focusing knob is too loose

 

 

 

Tighten it to a suitable position

 

 

 

8.Eyes feel tired easily

Diopter      adjustment

correct

is not  

Adjust the diopter

Brightness of light is not correct Adjust the brightness
The bulb

correctly

was not inserted  

Insert it correctly

Contact