KEN-A-VISION T-26001 Vision Scope 2 Stereo Microscope
Product Information
Specifications
- Product Name: Vision Scope 2 Stereo Microscope
- Model: T-22XXX Series T-26001
- Manufacturer: Ken-A-Vision
- Website: www.ken-a-vision.com
Frequently Asked Questions
- What should I do if there are fingerprints or stains on the lens surfaces?
- Clean the lens surfaces with a tissue moistened with a 3:7 mixture of alcohol and ether.
- Can I use organic solutions to clean the plastic surfaces of the microscope?
- No, organic solutions should not be used on plastic surfaces. If necessary, use a neutral detergent.
- Should I disassemble the microscope after use?
- No, it is not recommended to disassemble the microscope as it may cause damage.
- How should I store the microscope after use?
- Cover the microscope with the provided dust-cover and store it in a dry, clean environment to prevent rust.
- How often should I check and clean the microscope?
- The microscope should be checked and cleaned periodically to ensure proper performance.
Before Use
Notice
- Microscope should be stored in a dry and clean environment. Do not expose the microscope to the direct sunlight. Avoid high temperatures and violent vibration.
- This microscope is a precision instrument. Handle with care. Avoid impact and abrupt movement during transportation.
- To keep the image clear, clean any fingerprints or stains on the surfaces of the lens.
- Never turn the left and right focusing knob in the opposite direction at the same time. Microscope damage will occur.
Maintenance
- All lenses must be kept clean. Fine dust on the surface of the lens should be blown off with hand blower or wiped off gently with a soft lens tissue. Fingerprints and oil marks should be cleaned with a tissue moistened with a small amount of a 3:7 mixture of alcohol and ether.
- Never use organic solution to clean the other surfaces (especially the plastic surfaces). If necessary, please use a neutral detergent.
- Once assembled, do not take the microscope apart as this may cause damage to the microscope.
- After use, cover the microscope with the provided dust-cover and store in a dry, clean environment to prevent rust.
- To ensure proper performance, please check and clean the microscope periodically.
Nomenclature
Assembly
Operation
Position the glass stage
- Place the microscope base on a flat, firm surface and position the frosted stage plate on the base. (Figure 1)
Adjust the degree of tightness of the focusing arm
- Check the coarse focus knob on either side of the main support. The coarse focus knob should turn easily.
- If you want to adjust the degree of tightness of the focusing arm, you can hold one of the focusing knobs and turn another one to attain a suitable position. The degree of tightness relies on the direction to be turned. The clockwise direction is tight; counterclockwise is loose. (Figure 2)
Set the specimen slide
- Set the specimen in the center of the stage plate. If necessary, clamp the slide with the clips.
- To operate, plug the small round plug into the back of the base of the microscope and then plug the opposite end into the power outlet.
Adjust focusing knobs and diopter
- Turn the focusing knob and observe the specimen through the right eyepiece until the image of the specimen is clear.
- Observe the specimen through the left eyepiece and adjust the diopter ring 1 until the image is clear. (Figure 3)
Adjust the interpupillary distance
- Adjust the two eyepieces along the direction of the arrow shown in Figure 4 until eyepiece feels comfortable and a single image is seen.
- Use the left rheostat (dimmer knob) to adjust light intensity of the bottom (transmitted) light.
- Use the right rheostat (dimmer knob) to adjust light intensity of upper (incidental) light.
Configuration Chart
T-22XXX Series Configuration
Product Number | T-22001 | T-22011 | T-22021 | T-22041 | T-22051 | T-22061 | |
Magnifications | 10x & 30x | 15x & 45x | 20x & 60x | 20x & 40x | 30x & 60x | 40x & 80x | |
Eyepieces |
2 – 10x | О | О | ||||
2 – 15x | О | О | |||||
2 – 20x | О | О | |||||
Heads /
Objectives |
1x & 3x | О | О | О | |||
2x & 4x | О | О | О |
Note: The items marked “О” are included; others are optional.
T-26001 Series Configuration
Zoom Magnifications | 7x – 45x | 10.5x – 67.5x | 14x – 90x | |
Eyepieces |
2 – 10x | О | ||
2 – 15x | О | |||
2 – 20x | О | |||
Objective Zoom
Range |
0.7x – 4.5x |
О |
О |
О |
Note: The items marked “О” are included; others are optional.
Technical specifications
T-22XXX Series optical specifications
Objective Mag. |
Working Distance (mm) |
Eyepiece | Eyepieces (optional) | ||||
SC6EP10 | SC6EP15 | SC6EP20 | |||||
Mag. |
Objective
field |
Mag. |
Objective
field |
Mag. |
Objective
field |
||
1X |
100 |
10X | 20 | 15X | 15 | 20X | 11 |
2X | 20X | 10 | 30X | 7.5 | 40X | 5 | |
3X |
30X |
6.7 |
45X |
5 |
T-22XXX
SERIESX |
3.3 |
|
4X |
40X |
5 |
T-22XXX
SERIESX |
3.75 |
80X |
2.5 |
Auxiliary objectives for T-22XXX Series
Auxiliary objectives | Magnification | Working distance(mm) |
SC6OB5 | 0.5X | 165 |
SC6OB15 | 1.5X | 45 |
SC6OB20 | 2X | 30 |
- Working distance is fixed regardless of the magnification factor.
- Total magnification = Objective mag. X Auxiliary mag. X Auxiliary mag.
- Diameter of field of view (mm) =
- Photo adaptor mag. = Objective mag. X Auxiliary objective mag. X Photo eyepiece mag.
Configuration specifications of T-22XXX Series
Model |
T-22001 |
T-22041 |
|
Head |
Objective
magnification |
1X、3X | 2X、4X |
Working distance | 100mm | ||
Observation angle | 45° | ||
Interpupillary
distance adjustment |
The linkage between left and right eyepiece tube range of single adjustment: 54-75mm | ||
Diopter adjustment |
Range of single adjustment:±5D | ||
Mount with
auxiliary objectives |
Screw hole: M48*0.75 | ||
Objective | Field of view | φ20mm | |
Main body |
Mount the head | Mount the head in the bracket hole (diameter: φ76mm) | |
Focusing device |
Focus adjustment by turning the focusing knobs.
Range of single adjustment: 10 mm |
||
Glass stage |
Diameter:φ60mm | ||
Clips | Put it on the base from top |
Troubleshooting
Trouble | Cause | Remedy | |||
1.Double images |
Interpupillary distance
correct |
is | not | Adjust the interpupillary
distance |
|
Diopter adjustment
correct |
is | not |
Adjust the diopter |
||
Left and right eyepieces are
different magnifications. |
Mount eyepieces of the
same magnification |
||||
2.Dirt appears in the field of view |
Dirt on the specimen |
Clean the specimen |
|||
Dirt on the surface of eyepiece | Clean the surface | ||||
3.Image is not clear |
Dirt on the surface
objective |
of | the |
Clean the objective |
|
4.Image is not clear while adjusting the focus |
Diopter adjustment
correct |
is | not |
Adjust the diopter |
|
Focus is not correct | Adjust the focus | ||||
5.The focusing knob does not turn smoothly |
The focusing knob is too tight |
Loosen it to a suitable position |
|||
6.The image is obscure because of the head is slipping down by during observation |
The focusing knob is too loose |
Tighten it to a suitable position |
|||
8.Eyes feel tired easily |
Diopter adjustment
correct |
is | not |
Adjust the diopter |
|
Brightness of light is not correct | Adjust the brightness | ||||
The bulb
correctly |
was not | inserted |
Insert it correctly |
Contact
- Ken-A-Vision Mfg. Co., Inc.
- 5615 Raytown Road, Kansas City, MO 64133 USA
- Phone: 816-353-4787
- Fax: 816-358-5072
- www.ken-a-vision.com