Implement NPOS Software NanoPhotometer
Version History NPOS Software
- Software Version 4.6n build 16350
- Valid from 04.10.2019 last change 27.05.2025
- NPOS Version 4.6n 16350 (28.05.2025)
Improvements
- Automatic System Check on Startup: If the device detects a read-only state on startup, the system now automatically runs a repair check to maintain stability and prevent file system issues.
- Support for WORM Network Server: Our software now supports WORM (Write Once, Read Many) network drives.
- Optimized background checks prevent write operations that could cause file system errors, improving compatibility and ensuring data integrity compliance.
NPOS Version 4.6m 16350 (28.01.2025)
Implemented Functions
N120 single channel: shows always lid 80 in case of ABR
Improvements
- Check all network folders every 5minutesn to keep them alive
- N120: Do not move the horizontal motor with the maintenance stick
- During Linux setup, set eMMC to pSLC mode and enable automatic eMMC BKOPS
Resolved Issues
In the case of Air Bubble Recognition, the corrected wrongly shown lid is 80 / 140.
NPOS Version 4.6L 16350 (19.11.2024)
CFR21 Software
- CFR21: audit trail for backup missing
Improvements
- Add a new screensaver video
- Implement eMMC maintenance
- Fix Japanese translations
- Check all network folders every 5minutesn to keep them alive
NPOS Version 4.6k 16350 (13.06.2024)
Implemented Functions
General (affects all NanoPhotometer® models):
- N/A
NPOS Version 4.6j 16350 (29.04.2024)
Implemented Functions
General (affects all NanoPhotometer® models):
- Nucleic Acids and Concentration Methods: New measurement unit added (nmol/l)
- Automatic Date and Time synchronization via NTP server
- Stored Results: New back button added when IDS/JSON files are opened
CFR21 Software
- Additional CFR21 role added: Viewer – for auditing purposes
- New preference added: Administrators can enable/disable permission for Power Users to add new Users.
Improvements
WiFi functionality improved: Expanded list of SSIDs displayed
Resolved Issues
- Air Bubble Recognition function improved
- HP printer support has been upgraded
NPOS Version 4.6f 16350 (26.04.2023)
Implemented Functions
- N/A
Improvements
- N/A
Resolved Issues
General (affects all NanoPhotometer models)
- The kinetics application improved the duration of the kinetics measurements
- Intermittent disruptions in generating PDF reports have been resolved
NPOS Version 4.6e 16350 (30.01.2023)
Implemented Functions
- N/A
improvements
- N/A
Resolved Issues
Kinetics: Added missing special characters as sample names
NPOS Version 4.6d 16350 (20.01.2023)
Implemented Functions
General (affects all NanoPhotometer models):
- Extend REST API Push Service for general LIMS integration
- NanoPhotometer® N120:
- N120: New multi-channel method: Wavelength
Improvements
General (affects all NanoPhotometer models):
- Make remote access via PageKite configurable via a magic file
- Additional special characters available as sample names
- Kinetics: rounding of Slope value and duration
Updated HP Printer Driver to HPLIP 3.22.10
Resolved Issues
General (affects all NanoPhotometer models):
- Stored Results: handling of dot during file rename
NanoPhotometer® N120
Wavescan: Opening the lid in an 80 measurement
NPOS Version 4.6c 16350 (25.08.2022)
Implemented Functions
NanoPhotometer® N120:
Protein Assay multichannel mode: New assay type Custom
Improvements
General (affects all NanoPhotometer models):
Compatibility for Dymo 550 and 5XL printer
Resolved Issues
General (affects all NanoPhotometer models):
Test dyes removed from preprogrammed dye list
NanoPhotometer® N120
- Windows App: Fix in Protein Assay tab bar
- Warning message: Fix for editing more than 12 sample names
NPOS Version 4.6b 16350 (21.07.2022)
Implemented Functions
N/A
Improvements
N/A
Resolved Issues
NanoPhotometer® N120:
Wrong dilution factors are shown in the graphical user interface after reset.
NPOS Version 4.6a 16350 (20.04.2022)
Implemented Functions
N/A
Improvements
Updated HP printer driver
Resolved Issues
N/A
NPOS Version 4.6 16350 (07.03.2022)
Implemented Functions
N/A
Improvements
N/A
Resolved Issues
- Copy/move data to a USB flash drive with too little storage
- Closing/opening cuvette slider (NP80)
NPOS Version 4.5b 16350 (16.02.2022)
Implemented Functions
- N/A
Improvements
N/A
Resolved Issues
NanoPhotometer® N120:
N120: Configuration Change
NPOS Version 4.5 16350 (27.01.2022)
Implemented Functions
N/A
Improvements
General (affects all NanoPhotometer models):
Available SMB protocols: SMB 3.1, SMB 3.0, SMB 2.1, SMB 2.0, and SMB 1.0 (hardware version C.2)
Resolved Issues
General (affects all NanoPhotometer models):
- Assistance/Selftest: Folder list not shown in the save dialog
- Handling of initialization / selftest with lid open
Protein Assay / Standard Curve method: scaling of the axis
Apps for control devices:
- Mac: OD600 method: selection options for multiplier parameter not shown
- Android Tablet: Toggle switches in Protein Assay and Standard Curve sometimes do not work
- Android Tablet: Stored Results/Methods: Files sometimes not shown
NPOS Version 4.4 16326 (22.12.2021)
Implemented Functions
CFR21 Software:
- Search Option in Audit Trail
- Audit trail available on touchscreen, sorted by year. Saving possible up to 50.000 entries. No direct printing of audit trail in NPOS software.
Improvements
General (affects all NanoPhotometer® models):
- Change of the default dye parameter according to the dye manufacturer
- Improved screensaver with a new movie
- Improved translations
- Improvement of rounding numbers
- New and improved warning messages for easier handling
CFR21 Software
- Improvement of Audit Trail entries (NanoPhotometer booting, opening of Stored
Method and saving in the backup folder/control device)
Resolved Issues
General (affects all NanoPhotometer® models):
- Correct the folder list in the save dialog of Assistance/Selftest
- Correction for remote access in Assistance
- Handling of special characters
- New time setting sometimes does not work in Preferences
- Optimization of load mask
- Protein UV method: No dye selection if the Custom Protein factor is selected
- Protein UV method: No dye selection if OD1 factor is selected
- The result table is sometimes not displayed
- Stored Methods/Results: File directory: folder/files sometimes not displayed
CFR21 Software
- Login is not possible with a password containing some special characters
Apps for control devices
- Smart Phone: Manual Dilution is sometimes not displayed in the parameter
- Tablet: Show print icon only if printer is connected
- Tablet: Resizing of the graph zoom sometimes does not work
- Windows App: N120 sample map sometimes not opening
NPOS Version 4.3e 15670 (18.08.2021)
Implemented Functions
N/A
Improvements
Improvement of USB connection (NanoPhotometer/computer)
Resolved Issues
N/A
NPOS Version 4.3d 15670 (20.04.2021)
Implemented Functions
N/A
Improvements
N/A
Resolved Issues
General (affects all NanoPhotometer® models):
Protein UV method: Calculation of DOL
NPOS Version 4.3c 15670 (30.03.2021)
Implemented Functions
- N/A
Improvements
N/A
Resolved Issues
Concentration alerts
NPOS Version 4.3b 15670 (10.03.2021)
Implemented Functions
General (affects all NanoPhotometer® models):
- Screensaver videos
- Protein Assay and Standard Curve method: remeasurement of standard curves in the Stored
Methods
- Formula editor in the Wavelength method
- New Assistance features: Self-Test, User Manual, and Remote Access
NanoPhotometer® N120
- New multi-channel methods: Protein Assay and Wavescan
- Manual change of Sample IDs in Sample Map
CFR21 Software
- Power User can read the audit trail
Improvements
- General (affects all NanoPhotometer® models):
- New and improved warning messages for easier handling
- Protein UV method: new protein factor Custom (1/ɛ); extended (four) decimal places for Extinction Coefficient
- Protein Assay and Standard Curve method: shows by default three concentrations for curve fit, Regression 2nd Order
- Protein Assay method: default values for Baseline Correction
- Adjusted format for autosave and audit trail file names
Improvement Initialization Test
- Updated HP printer driver
NanoPhotometer® N120
Performance improvement
- Translations for multi-channel methods
- Show .csv files in Stored Methods
CFR21 Software
- New warning message for administrator password
- Additional unique audit trail Entries
Apps for control devices
- Improved saving on smartphone/tablet: data files are saved in the Download folder
- (Android) or NanoPhotometer folder (iOS) for easy access
Resolved Issues
General (affects all NanoPhotometer® models):
- Issues with parameter settings and layout
- Workflow issues for inactive/active blank and sample button, including a warning message
- Auto Sample sometimes not working
- Clickable area not working: alert icons in the table of the Nucleic Acid / Protein UV method
- Show integer numbers for wavelength and time settings
- Keyboard randomly not shown randomly
WiFi status on home screen
- Show the correct WiFi status for connection failure
- Network Folder: issue with permissions; show the correct network status for the wrong network path
- Connection of a static IP address
- The Time Zone dropdown loads too slowly
- The cuvette port sometimes does not close
- Fix of Chinese, Japanese, and Portuguese translation issues
Disable the DNS server
- Overlay graphs are not working after opening the Assistance
- File directory: issues with white frame for folder selection; copying subfolders
- PDF files: layout issues, show correct table columns depending on method, show correct decimal places for all values
- Issues with the results %Transmission for cuvette measurements
- Protein Assay / Standard Curve method: show N/A for the factor for 2nd Order Regression standard curves (Excel / PDF), opened .ids and .json files show saved standard curve with correct values, corrupt autosave file NanoPhotometer® N120:
- Sample Map selection: change of sample loading orientation or switching between single and multi-sample clears Sample Map; selection issues for vertical mode, multiple plates, and a warning message “Position A1 must be selected”
- Sample ID setting on Sample Map: issues with vertical orientation; Sample ID setting in opened .json
- White frame issue on Sample Map/result screens after remeasuring the bad blank
- Hidden Blank: show correct Sample IDs (BLK01…); show correct warning messages on inactive sample button
- Bad blank warning: sort by Sample ID
- Multi-channel result screens: show correct number format onthe concentration grid, correct wavelength range on the graph grid
- The single result overview is empty when swiping screens during measurement.
- Show in the graph the pop-up grid position and not consecutive numbers
- Opened .ids files with selected data saved (multi-channel method): show graphs in grid and Sample IDs in Sample Map
- Sample ID drafts (.csv) – show correct help text in .csv files for exported Sample IDs
- Show correct warning messages in case of an invalid Sample ID definition in the imported .csv file
- Do not stop measurement for imported Sample ID (.csv vertical mode and selection)
- Parameter not shown in the opened .json, navigating over the page indicator
- Bad reaction to the page indicator
- Use the correct lid factor in case of dye measurement inthe Nucleic Acid / Protein UV multichannel method
- Layout issue in Stored Methods/Results after saving in the multichannel method
- Illuminate red LEDs correctly after stopping blank
- Optimize cryo label prints onthe the Dymo printer for multichannel methods
Apps for control devices
Mac computer app
- Layout issues: Auto Sample button, Sample ID pop-up
- Clickable area not working: units dropdown in Nucleic Acids and Protein UV method; alert icons in result view. Saving via the Home button is not a closing method.d Windows computer app:
- Protein Assay / Standard Curve method: toggle switches for replicate selection not working, mouse scrolling issues for result view
- Protein UV multi-channel method: shows the correct value for A260/A280 in the vertical tab bar
- Phone app (iOS and Android): show the correct parameter depending on
- NanoPhotometer® model (opened method and Stored Methods)
- No back button in Stored Results/Stored Methods or opened .ids or .json
- Kinetics save on the home button not working
CFR21 Software
- Full functionality of the CFR21 software without a power user
- The audit trail folder is only accessible via the file server
- The language of the audit trail is English, independent from the language setting
- Use the Login Name and not the User ID in the audit trail
- Layout issues: selected options need to be highlighted (in method color) and greyed out if no selection is possible
- User rights: do not allow the user to change parameters in the loaded .ids files and stored
- Methods (.json); User can only measure defined Sample IDs (multi-channel method) in an open Stored Method (.json);
- Unique login and group names to avoid problems with user rights
- Show the correct file name/path of the .ids PDF and Excel document header
- The screensaver can sometimes not be deactivated / display darkened
- Adding a folder to My NanoPhotometer in the opened .ids file ends in a load mask
- Windows app: forced log off ends in load mask; dye tab in vertical tab bar sometimes not shown in the opened Stored Method
- NPOS Version 4.2h 14900 (29.06.2020)
Implemented Functions
N/A
Improvements
- Activation of SMB2 for new main PCBs
Resolved Issues
General (affects all NanoPhotometer® models):
- File Server: writing permissions on the computer; file server access problems after updating to NPOS 4. X
- Chinese: Date & Time not shown completely in PDF
NPOS Version 4.2g 14900 (31.03.2020)
Implemented Functions
N/A
Improvements
N/A
Resolved Issues
General (affects all NanoPhotometer® models):
Deactivation of screensaver videos (final approval missing)
NPOS Version 4.2f 14900 (24.01.2020)
Implemented Functions
N/A
Improvements
N/A
Resolved Issues
General (affects all NanoPhotometer® models):
Chinese / Japanese language setting: PDF saving and printing are not working for NanoPhotometer® with the new main PCB
NPOS Version 4.2e 14900 (17.12.2019)
Implemented Functions
- N/A
Improvements
General (affects all NanoPhotometer® models):
Initialization test: no measurement with open lid
Resolved Issues
General (affects all NanoPhotometer® models):
Saving issue after save failed on a read-only USB flash drive
NPOS Version 4.2d 14900 (15.11.2019)
Implemented Functions
- N/A
Improvements
N/A
Resolved Issues
NanoPhotometer® N120:
Air Bubble Recognition of multi-channel methods is not working correctly
NPOS Version 4.2c 14900 (07.11.2019)
Implemented Functions
N/A
Improvements
N/A
Resolved Issues
General (affects all NanoPhotometer® models):
Excel of the Wavescan method: shows consistent peak position on all Excel sheets
NPOS Version 4.2b.14900 (29.10.2019)
Implemented Functions
N/A
Improvements
NanoPhotometer® N120:
Improvement ofthe measurement head movement
Resolved Issues
CFR21 Software:
Audit trail: show consecutive IDs (no restart at 1)
NPOS Version 4.2.14900 (04.10.2019)
Implemented Functions
N/A
Improvements
- CFR21 software: second disclaimer for liability added
- Warning Messages: change of ratio alert A260/A280, Confirmation message for deleting files
Resolved Issues
General (affects all NanoPhotometer® models):
- Optimize booting time
- The keyboard sometimes does not open
- Hanging load masks: graph overlay; saving of selected data
- Allow lower and upper case extensions for .ids files
- No Auto Sample function for cuvette mode (NP80)
- Missing text “Selected Time” for time change
- File server: remember user accounts; lock button not working; writing permissions on the computer
- Network Folder: password sometimes lost; warning message for allowed characters for Network Folder Nickname
- Excel: range of x-axis for graph on values sheet
- Cryo label print on Dymo 400 and Dymo 450 Turbo is not working
NanoPhotometer® N120
- Home button not shown after tick box selection
- Layout issues for single result overview
- White frame selection on Sample Map
- Opened .ids files: display of cells without results
- Warning message for selected cells without Sample IDs
- Consistent parameter setting changing between single- and multi-channel options
- Actions for file tree in Sample ID pop-up/import .csv files: copy, delete, and rename
- Excel of single-channel Nucleic Acid / Protein UV: graph sheets are missing
CFR21 Software
- Allow lower and upper case extensions for .lic files (license file)
Software Version 4.6n build 16350
Valid from 04.10.2019 last change 27.05.2025
- Activation of the CFR21 software in a language other than English
- Opened Protein UV Stored Method shows the wrong wavelength parameter
Apps for control devices
- Layout issue on Windows computer: white square instead of x for closing a single result overview
- Concurrency issue opening .ids or .json when the control device is active
- iPhone: saving on the control device
- Android app requests to read ofthe phone state permission
- Android apps: remove the message “Leave this page” on the control device back button when saving data on the control
28. NPOS Version 4.2.14756 (27.08.2019)
Initial release of CFR21 Software
Documents / Resources
![]() |
Implen NPOS Software NanoPhotometer [pdf] User Guide 4.6n, 4.6m, 4.6L, 4.6k, 4.6j, 4.6f, NPOS Software NanoPhotometer, NPOS, Software NanoPhotometer, NanoPhotometer |