IMPLEN DiluCell 10 Automatic Sample Dilution User Manual

IMPLEN-LOGO

IMPLEN DiluCell 10 Automatic Sample Dilution

IMPLEN-DiluCell-10-Automatic-Sample-Dilution-PRODUCT

Specifications

  • Wavelength Range
  • Optical Pathlength
  • Volume Requirement
  • Tolerance
  • Dilution Factor
  • Ordering Information
  • Packaging

Product Overview

  • The DiluCellTM 10 simplifies concentration measurements by reducing the optical pathlength, allowing direct reading of highly concentrated samples while staying within your photometer’s linear range. The Dilution factor is 1:10.

How DiluCellTM Works

  • The Lambert-Beer-Law empirically relates the absorption of light to the properties of the sample. The built-in virtual dilution of DiluCellTM 10 enables a 1:10 dilution in accordance with the Lambert-Beer Law.

INTRODUCTION

Thank You for Choosing DiluCell™

  • This advanced, disposable cuvette is designed to simplify workflows and enhance accuracy for absorbance measurements—especially when working with highly concentrated samples.
  • DiluCell™ offers virtual dilution without the need to physically dilute your samples. The wide wavelength range of 340 nm – 950 nm enables the user to incorporate the DiluCellTM 10 cuvettes for all standard applications.
  • The wide range of applications of our DiluCellTM 10 cuvettes include but is not limited to.
  • Cell Density (OD600) – Monitoring bacterial/yeast/fungal growth
  • Protein Assays – Bradford, BCA, Lowry, etc.
  • Enzyme Kinetics
  • Colorimetric Assays
  • Turbidity/McFarland analyses
  • Please read this manual carefully before use.

Product Overview

  • DiluCell™ 10 simplifies concentration measurements by reducing the optical pathlength, allowing direct reading of highly concentrated samples while staying within your photometer’s linear range.
  • Wavelength range: 340-950 nm
  • Required sample volume: 300 µl
  • Optical pathlength: 1 mm @ 8.5 mm center height
  • Dilution factor: 1:10

How DiluCell™ Works

  • The Lambert-Beer-Law, also known as Beer’s Law, empirically relates the absorption of light to the properties of the sample.

Lambert-Beer Law:IMPLEN-DiluCell-10-Automatic-Sample-Dilution-FIG-1

Where:

  • I0 = incoming light
  • I = outgoing light
  • Ε = extinction coefficient
  • C = concentration
  • D = pathlength (1mm @ 8.5mm center height)IMPLEN-DiluCell-10-Automatic-Sample-Dilution-FIG-2

Built-in Virtual Dilution

  • Traditional cuvettes require manual sample dilution and volumes of at least 1 ml.
  • The innovative design of the DiluCell™ 10 reduces the light path from the standard 10mm to just 1mm, thereby enabling a built-in 1:10 virtual dilution in accordance with the Lambert-Beer Law.
  • This allows you to work with smaller sample volumes (only 300 μl), while eliminating manual dilution steps, saving time, and minimizing the risk of dilution/pipetting errors and cross-contamination.IMPLEN-DiluCell-10-Automatic-Sample-Dilution-FIG-3

Operating Instructions

  1. Correct Insertion
    • Place the DiluCell™ in the cuvette holder with the wide window facing the light beam.
    • Implen devices such as the Implen OD600® and the *NanoPhotometer® product range feature directional arrows to assist with correct alignment.
    • N80 and C40, cuvette-enabled devices.
    • Important: DiluCell™ will not work in any other orientation.
    • Do not mix cuvette types within the same measurement series (e.g., blanks).
  2. Sample Loading
    • The geometry of DiluCell™ supports bubble-free filling.
    • It is, however, still highly recommended to visually check your samples for air bubbles after loading.
    • If bubbles are present, tap gently to eliminate them.
  3. Volume Requirement
    • Use 300 µl of the sample for optimal performance.
  4. Data Handling
    • Simply multiply your absorbance values by 10 post-measurement or set your instrument to apply a 10× dilution factor.

Important Usage Notes

Single Use Only

Reuse is not recommended due to:

  • High cross-contamination risk
  • Bubble formation during refilling
  • Inability to properly decontaminate (especially for biological samples)

Do Not Autoclave

  • Autoclaving may compromise the structure of the cuvette and result in inaccurate readings.

Solvent Compatibility

  • For chemical resistance and solvent use, contact your local Implen representative or visit www.implen.de.

Specifications

  • Parameter DiluCell™ 10
  • Wavelength Range 340 nm – 950 nm
  • Optical Pathlength 1 mm (at 8.5 mm center height)
  • Volume Requirement 300 μl
  • Tolerance ±10%
  • Dilution Factor 1:10
  • Ordering Information DC 10
  • Packaging 96 disposable cuvettes per pack

Contact & Support

Europe, Asia, the South Pacific, North and South America Middle East, and Africa

North and South America

  • Phone: +1 8187486400
  • Fax: +1 8184490416
  • Email: info@implen.com
  • Website: www.implen.com
  • Implen, Inc.
  • Unit 104
  • 31194 La Baya Drive
  • Westlake Village, CA 91362
  • USA

FAQ

Can DiluCellTM be reused?

No, DiluCellTM is designed for single-use only to avoid cross-contamination risks and maintain accuracy.

Can I autoclave DiluCellTM for sterilization?

No, autoclaving may compromise the cuvette's structure and lead to inaccurate readings.

Documents / Resources

PDF thumbnailDiluCell 10 Automatic Sample Dilution
User Manual · DC 10, DiluCell 10 Automatic Sample Dilution, DiluCell 10, Automatic Sample Dilution, Sample Dilution

Ask a Question

Use this section to ask about setup, compatibility, troubleshooting, or anything missing from this manual.

Ask a Question

Ask about setup, compatibility, troubleshooting, or anything missing from this manual. Name and email are optional.