Product Information

Specifications

  • Product Name: Yeast and Mold PCR Assay
  • Action Levels: 10-50 cfu/g for pooled sample protocol, 25 cfu/g or above for non-pooled sample protocol
  • Enrichment Replicates: 3 disrupter tubes for pooled sample protocol
  • Sample Homogenate Dilution: 1:10 according to food type
  • Incubation Time: 44 hours
  • PCR Tablets Hydration Volume: 50 µl lysate from step 12

Product Usage Instructions

  • Step 1: Sample Homogenization
  • Homogenize the sample in a 1:10 dilution according to the food type.
  • Step 2: Determine Sample Volume
  • Determine the sample volume to be tested. Refer to the User Guide or table on the back of the reference card for guidance.
  • Step 3: Transfer Sample
  • Transfer the sample to a disrupter tube.
  • Step 4: Incubate Disrupter Tubes
  • Incubate the disrupter tubes for 44 hours.
  • Step 5: Add DNA Stabilizer
  • Add DNA stabilizer to the disrupter tubes.
  • Step 6: Agitate in Disrupter Device
  • Agitate the disrupter tubes in a disrupter device.
  • Step 7: Create a Rack File
  • Create a rack file for further processing.
  • Step 8: Transfer the Lysis Reagent
  • Transfer 12 mL of Protease YM lysis buffer to cluster tubes.
  • Step 9: Transfer Disrupted Samples
  • Transfer the disrupted samples from the disrupter tubes to the cluster tubes.
  • Step 10: Heat Cluster Tubes
  • Heat the cluster tubes for 20 seconds.
  • Step 11: Cool Cluster Tubes
  • Cool the cluster tubes in a cooling block.
  • Step 12: PCR Tablet Hydration
  • Hydrate PCR tablets with 50 µl lysate from step 12.
  • Step 13: Initialize Cycler
  • Initialize the cycler for PCR.
  • Step 14: Arrange PCR Tubes
  • Arrange PCR tubes in the cooling block.
  • Step 15: Place Tubes in the Cycler
  • Place the PCR tubes in the cycler and run the program.
  • Step 16: Review Results

Unload the samples and review the results on the screen. Refer to the User Guide for detailed information on result interpretation.

FAQ:

  • Q: What are the action levels for the pooled sample protocol?
  • A: The action levels for the pooled sample protocol are 10-50 cfu/g.
  • Q: Can the non-pooled sample protocol be used for action levels below 25 cfu/g?
  • A: No, the non-pooled sample protocol is recommended for action levels of 25 cfu/g or above.

using instruction

  1. Homogenize the sample in 1:10 dilution according to the food type.hygiena-BAX-Q7-Molecular-Pathogen-Detection-System-fig-1
  2. Determine the sample volume to be tested.
    (See User Guide or table on the back of this reference card.)
  3. Transfer the sample to the disrupter tube.hygiena-BAX-Q7-Molecular-Pathogen-Detection-System-fig-2
    Pooled sample protocol requires triplicate disrupter tubes.
  4. Incubate disrupter tubes.hygiena-BAX-Q7-Molecular-Pathogen-Detection-System-fig-3
  5. Add DNA stabilizer to disrupter tubes.hygiena-BAX-Q7-Molecular-Pathogen-Detection-System-fig-3
    DNA stabilizer
  6. Agitate in disrupter device.hygiena-BAX-Q7-Molecular-Pathogen-Detection-System-fig-5
  7. Create a rack file.hygiena-BAX-Q7-Molecular-Pathogen-Detection-System-fig-6
  8. Add protease to YM lysis buffer.hygiena-BAX-Q7-Molecular-Pathogen-Detection-System-fig-18
  9. Transfer lysis reagent made in step 8 to cluster tubeshygiena-BAX-Q7-Molecular-Pathogen-Detection-System-fig-8
    Lysis reagent mixture can be stored at 2 – 8 °C for up to one week
  10. Transfer disrupted samples to cluster tubes.hygiena-BAX-Q7-Molecular-Pathogen-Detection-System-fig-9
    *Pooled sample protocol requires pooled volumes from disrupter tubes into 1 cluster tube.
  11. Heat cluster tubes.hygiena-BAX-Q7-Molecular-Pathogen-Detection-System-fig-10
  12. Cool cluster tubes in the cooling block.hygiena-BAX-Q7-Molecular-Pathogen-Detection-System-fig-11
    * Steps 11 and 12 can also be performed using the Hygiena™ Automated Thermal Block. See the Automated Thermal Block User Guide for details and instructions.
  13. Initialize cycler.hygiena-BAX-Q7-Molecular-Pathogen-Detection-System-fig-12
  14. Arrange PCR tubes in the cooling block.hygie
  15. Hydrate PCR tablets with 50 μl lysate from step 12.hygiena-BAX-Q7-Molecular-Pathogen-Detection-System-fig-14
  16. Place tubes in the cycler and run the program.hygiena-BAX-Q7-Molecular-Pathogen-Detection-System-fig-15
  17. Unload samples and review results on the screen. See User Guide for details.hygiena-BAX-Q7-Molecular-Pathogen-Detection-System-fig-17

Pooled Sample Protocol

This ultra-sensitive protocol uses pooled samples from three disrupter tube enrichment replicates for action levels of 10-50 cfu/g.

 

If your action level is:

Then transfer this volume of homogenate to 3 disrupter tubes: And pool these volumes of disrupted samples for testing:
10 cfu/g 400 µL 7 µL from 3 replicates
20 cfu/g 200 µL 7 µL from 3 replicates
50 cfu/g 80 µL 7 µL from 3 replicates

Non-Pooled Sample Protocol

This protocol for yeast and mold testing can be used without pooling for action levels of 25 cfu/g or above.

 

If your action level is:

 

Then use this volume of homogenate:

25 cfu/g 400 µL
50 cfu/g 200 µL
100 cfu/g 100 µL
500 cfu/g 20 µL
1000 cfu/g 10 µL

 

Documents / Resources

hygiena BAX-Q7 Molecular Pathogen Detection System [pdf] Installation Guide
BAX-Q7 Molecular Pathogen Detection System, BAX-Q7, Molecular Pathogen Detection System, Pathogen Detection System, Detection System
hygiena BAX-Q7 Molecular Pathogen Detection System [pdf] Instruction Manual