hygiena BAX-Q7 Molecular Pathogen Detection System
Product Information
Specifications
- Product Name: Yeast and Mold PCR Assay
- Action Levels: 10-50 cfu/g for pooled sample protocol, 25 cfu/g or above for non-pooled sample protocol
- Enrichment Replicates: 3 disrupter tubes for pooled sample protocol
- Sample Homogenate Dilution: 1:10 according to food type
- Incubation Time: 44 hours
- PCR Tablets Hydration Volume: 50 µl lysate from step 12
Product Usage Instructions
- Step 1: Sample Homogenization
- Homogenize the sample in a 1:10 dilution according to the food type.
- Step 2: Determine Sample Volume
- Determine the sample volume to be tested. Refer to the User Guide or table on the back of the reference card for guidance.
- Step 3: Transfer Sample
- Transfer the sample to a disrupter tube.
- Step 4: Incubate Disrupter Tubes
- Incubate the disrupter tubes for 44 hours.
- Step 5: Add DNA Stabilizer
- Add DNA stabilizer to the disrupter tubes.
- Step 6: Agitate in Disrupter Device
- Agitate the disrupter tubes in a disrupter device.
- Step 7: Create a Rack File
- Create a rack file for further processing.
- Step 8: Transfer the Lysis Reagent
- Transfer 12 mL of Protease YM lysis buffer to cluster tubes.
- Step 9: Transfer Disrupted Samples
- Transfer the disrupted samples from the disrupter tubes to the cluster tubes.
- Step 10: Heat Cluster Tubes
- Heat the cluster tubes for 20 seconds.
- Step 11: Cool Cluster Tubes
- Cool the cluster tubes in a cooling block.
- Step 12: PCR Tablet Hydration
- Hydrate PCR tablets with 50 µl lysate from step 12.
- Step 13: Initialize Cycler
- Initialize the cycler for PCR.
- Step 14: Arrange PCR Tubes
- Arrange PCR tubes in the cooling block.
- Step 15: Place Tubes in the Cycler
- Place the PCR tubes in the cycler and run the program.
- Step 16: Review Results
Unload the samples and review the results on the screen. Refer to the User Guide for detailed information on result interpretation.
FAQ:
- Q: What are the action levels for the pooled sample protocol?
- A: The action levels for the pooled sample protocol are 10-50 cfu/g.
- Q: Can the non-pooled sample protocol be used for action levels below 25 cfu/g?
- A: No, the non-pooled sample protocol is recommended for action levels of 25 cfu/g or above.
using instruction
- Homogenize the sample in 1:10 dilution according to the food type.
- Determine the sample volume to be tested.
(See User Guide or table on the back of this reference card.) - Transfer the sample to the disrupter tube.
Pooled sample protocol requires triplicate disrupter tubes. - Incubate disrupter tubes.
- Add DNA stabilizer to disrupter tubes.
DNA stabilizer - Agitate in disrupter device.
- Create a rack file.
- Add protease to YM lysis buffer.
- Transfer lysis reagent made in step 8 to cluster tubes
Lysis reagent mixture can be stored at 2 – 8 °C for up to one week - Transfer disrupted samples to cluster tubes.
*Pooled sample protocol requires pooled volumes from disrupter tubes into 1 cluster tube. - Heat cluster tubes.
- Cool cluster tubes in the cooling block.
* Steps 11 and 12 can also be performed using the Hygiena™ Automated Thermal Block. See the Automated Thermal Block User Guide for details and instructions. - Initialize cycler.
- Arrange PCR tubes in the cooling block.
- Hydrate PCR tablets with 50 μl lysate from step 12.
- Place tubes in the cycler and run the program.
- Unload samples and review results on the screen. See User Guide for details.
Pooled Sample Protocol
This ultra-sensitive protocol uses pooled samples from three disrupter tube enrichment replicates for action levels of 10-50 cfu/g.
If your action level is: |
Then transfer this volume of homogenate to 3 disrupter tubes: | And pool these volumes of disrupted samples for testing: |
10 cfu/g | 400 µL | 7 µL from 3 replicates |
20 cfu/g | 200 µL | 7 µL from 3 replicates |
50 cfu/g | 80 µL | 7 µL from 3 replicates |
Non-Pooled Sample Protocol
This protocol for yeast and mold testing can be used without pooling for action levels of 25 cfu/g or above.
If your action level is: |
Then use this volume of homogenate: |
25 cfu/g | 400 µL |
50 cfu/g | 200 µL |
100 cfu/g | 100 µL |
500 cfu/g | 20 µL |
1000 cfu/g | 10 µL |
Documents / Resources
![]() |
hygiena BAX-Q7 Molecular Pathogen Detection System [pdf] Installation Guide BAX-Q7 Molecular Pathogen Detection System, BAX-Q7, Molecular Pathogen Detection System, Pathogen Detection System, Detection System |
![]() |
hygiena BAX-Q7 Molecular Pathogen Detection System [pdf] Instruction Manual BAX-Q7 Molecular Pathogen Detection System, BAX-Q7, Molecular Pathogen Detection System, Pathogen Detection System, Detection System |