KEITHLEY 2401 Source Meter
Espesifikasyon
- Modèl: 2401
- Firmware Release: B04
- Dat lage: jen 2025
Enstriksyon Itilizasyon Pwodwi
Enstriksyon enstalasyon:
Perform the following steps to install the firmware revision onto the Model 2401 instrument:
- Run the Flash Wizard program.
- Select the appropriate interface at the Communications Select screen and follow any subsequent instructions.
- Ultimately, the program should auto-detect the 2401.
- You will then be asked to specify the firmware file.
- The upgrade proceeds from there. It typically only takes a few minutes.
Contact Keithley if you encounter any problems during the upgrade process.
Upgrade Considerations:
Consider the following table when deciding to upgrade to version B04:
Konsiderasyon | Recalibration Required | Re-qualification Suggested | Èske ou ta dwe ajou? |
---|---|---|---|
From A01 | Non | N/A | Non |
From B01 | Non | Non | Review |
From B02 | Non | Non | Review |
Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-800-833-9200 tek.com/keithley
Enfòmasyon jeneral
Modèl sipòte
This firmware is used on the following Keithley Instruments product model:
- Model 2401 SourceMeter®
Enstriksyon Enstalasyon
Download the following from the Keithley Support Websit
- The Keithley Flash Wizard32, version C12 or higher (if you do not already have this)
- Ouvri zip la files in a folder of your choice and then run the “setup.exe” program.
- Swiv enstriksyon yo pou enstale pwogram nan.
- The firmware revision image file you want to install on your Model 2401 instrument.
Perform the following steps to install the firmware revision onto the Model 2401 instrument:
- Run the Flash Wizard program.
- Select the appropriate interface at the Communications Select screen and follow any subsequent instructions.
- Ultimately the program should auto-detect the 2401.
- You will then be asked to specify the firmware file.
- The upgrade proceeds from there.
It typically only takes a few minutes. Contact Keithley if you have any problems
Upgrade Considerations for 2401
The following table outlines the considerations that should be made when deciding to upgrade to this version or not from any previous version.
Konsiderasyon | Soti nan
A01 |
Soti nan
B01 |
Soti nan
B02 |
Recalibration Required | — | Non | Non |
Re-qualification Suggested | — | Non | Non |
Èske ou ta dwe ajou? | Non | Review | Review |
A Model 2401 instrument with Version A01 firmware is incompatible with Version ‘B’ firmware, such as B03. You should not upgrade A01 to any ‘B’ versions. Contact Keithley Instruments if you require upgrades.
Version B04 Release
Plis paseview
Version D04 is a maintenance-only release to support the new 10-digit serial number format.
Koreksyon kritik
- There are no critical fixes included in this release.
Amelyorasyon
- There are no enhancements included in this release.
Version B03 Release
Plis paseview
Version B03 release fixes issues with RS-232 flash upgrades, RS-232 break-byte functionality, and range change problems.
Koreksyon kritik
AR 66391
Modèl ki afekte yo: 2401
sentòm:
Flash upgrades utilizing RS-232 would fail at all baud rates.
Rezolisyon:
Pwoblèm lan te korije.
Non-critical fixes
AR 67009
Models affected: 2401
sentòm:
When sourcing 100 mA at 2 V compliance and measuring voltage, switching the source range to the 1 A range would cause a problem that would cause the measured value to be incorrect.
Rezolisyon:
Pwoblèm lan te korije.
AR 66390
Modèl ki afekte yo: 2401
sentòm:
When attempting to send a break-byte character to the unit, the pending operation would not be stopped.
Rezolisyon:
Pwoblèm lan te korije.
Version B02 Release
Plis paseview
Version B02 release is to address issues with RS-232 on the new hardware in the Model 2401 SourceMeter®.
Koreksyon kritik
PR 61943
Modèl ki afekte yo: 2401
sentòm:
RS-232 serial communication experiences issues when the baud rate is greater than 9600. Additionally, a baud rate of 300 is not supported on new hardware.
Rezolisyon:
The communication issues when the baud rate is greater than 9600 have been corrected. Removed option of 300 baud rate from unit.
Amelyorasyon
- There are no enhancements included in this release.
Version B01 Release
Plis paseview
The Version B01 release supports new hardware in the Model 2401 SourceMeter®.
Korije kritik
No fixes are included.
Amelyorasyon
- No Enhancements are included.
Version A01 Release
Plis paseview
- The Version A01 release is the initial release of the Model 2401 SourceMeter®.
Pwoblèm li te ye
- There are no known issues with version A01.
FAQ
Q: Can I upgrade from Version A01 to Version B03?
A: No, a Model 2401 instrument with Version A01 firmware is incompatible with Version 'B' firmware. Contact Keithley Instruments for further assistance.
Dokiman / Resous
![]() |
KEITHLEY 2401 Source Meter [pdfGid Itilizatè 2401 Source Meter, 2401, Source Meter, Meter |