KEITHLEY 2401 Source Meter

وضاحتون
- ماڊل: 2401
- Firmware Release: B04
- ڇڏڻ جي تاريخ: جون 2025
پيداوار جي استعمال جون هدايتون
انسٽاليشن جون هدايتون:
Perform the following steps to install the firmware revision onto the Model 2401 instrument:
- Run the Flash Wizard program.
- Select the appropriate interface at the Communications Select screen and follow any subsequent instructions.
- Ultimately, the program should auto-detect the 2401.
- You will then be asked to specify the firmware file.
- The upgrade proceeds from there. It typically only takes a few minutes.
Contact Keithley if you encounter any problems during the upgrade process.
Upgrade Considerations:
Consider the following table when deciding to upgrade to version B04:
| غور ڪرڻ | Recalibration Required | Re-qualification Suggested | ڇا توهان کي اپڊيٽ ڪرڻ گهرجي؟ |
|---|---|---|---|
| From A01 | نه | N/A | نه |
| From B01 | نه | نه | Review |
| From B02 | نه | نه | Review |
ڪيٿلي اوزار 28775 اورورا روڊ ڪليلينڊ، اوهائيو 44139 1-800-833-9200 tek.com/keithley
عام معلومات
سپورٽ ماڊلز
This firmware is used on the following Keithley Instruments product model:
- Model 2401 SourceMeter®
تنصيب جون هدايتون
Download the following from the Keithley Support Webسائيٽ
- The Keithley Flash Wizard32, version C12 or higher (if you do not already have this)
- ان زپ ڪريو files in a folder of your choice and then run the “setup.exe” program.
- پروگرام کي انسٽال ڪرڻ جي هدايتن تي عمل ڪريو.
- The firmware revision image file you want to install on your Model 2401 instrument.
Perform the following steps to install the firmware revision onto the Model 2401 instrument:
- Run the Flash Wizard program.
- Select the appropriate interface at the Communications Select screen and follow any subsequent instructions.
- Ultimately the program should auto-detect the 2401.
- You will then be asked to specify the firmware file.
- The upgrade proceeds from there.
It typically only takes a few minutes. Contact Keithley if you have any problems
Upgrade Considerations for 2401
The following table outlines the considerations that should be made when deciding to upgrade to this version or not from any previous version.
| غور ڪرڻ | کان
A01 |
کان
بي 01 |
کان
بي 02 |
| Recalibration Required | - | نه | نه |
| Re-qualification Suggested | - | نه | نه |
| ڇا توهان کي اپڊيٽ ڪرڻ گهرجي؟ | نه | Review | Review |
A Model 2401 instrument with Version A01 firmware is incompatible with Version ‘B’ firmware, such as B03. You should not upgrade A01 to any ‘B’ versions. Contact Keithley Instruments if you require upgrades.
Version B04 Release
مٿانview
Version D04 is a maintenance-only release to support the new 10-digit serial number format.
تنقيدي اصلاحون
- There are no critical fixes included in this release.
واڌايون
- There are no enhancements included in this release.
Version B03 Release
مٿانview
Version B03 release fixes issues with RS-232 flash upgrades, RS-232 break-byte functionality, and range change problems.
تنقيدي اصلاحون
آر 66391
متاثر ٿيل ماڊل: 2401
علامه:
Flash upgrades utilizing RS-232 would fail at all baud rates.
قرارداد:
مسئلو درست ڪيو ويو آهي.
Non-critical fixes
آر 67009
Models affected: 2401
علامه:
When sourcing 100 mA at 2 V compliance and measuring voltage, switching the source range to the 1 A range would cause a problem that would cause the measured value to be incorrect.
قرارداد:
مسئلو درست ڪيو ويو آهي.
آر 66390
متاثر ٿيل ماڊل: 2401
علامه:
When attempting to send a break-byte character to the unit, the pending operation would not be stopped.
قرارداد:
مسئلو درست ڪيو ويو آهي.
Version B02 Release
مٿانview
Version B02 release is to address issues with RS-232 on the new hardware in the Model 2401 SourceMeter®.
تنقيدي اصلاحون
پي آر 61943
متاثر ٿيل ماڊل: 2401
علامه:
RS-232 serial communication experiences issues when the baud rate is greater than 9600. Additionally, a baud rate of 300 is not supported on new hardware.
قرارداد:
The communication issues when the baud rate is greater than 9600 have been corrected. Removed option of 300 baud rate from unit.
واڌايون
- There are no enhancements included in this release.
Version B01 Release
مٿانview
The Version B01 release supports new hardware in the Model 2401 SourceMeter®.
نازڪ اصلاحون
No fixes are included.
واڌايون
- No Enhancements are included.
Version A01 Release
مٿانview
- The Version A01 release is the initial release of the Model 2401 SourceMeter®.
ڄاڻايل مسئلا
- There are no known issues with version A01.
FAQs
Q: Can I upgrade from Version A01 to Version B03?
A: No, a Model 2401 instrument with Version A01 firmware is incompatible with Version 'B' firmware. Contact Keithley Instruments for further assistance.
دستاويز / وسيلا
![]() |
KEITHLEY 2401 Source Meter [pdf] استعمال ڪندڙ ھدايت 2401 Source Meter, 2401, Source Meter, Meter |

