DYNALABS-LOGO

DynaLabs DYN-I-8000 Triaxial MEMS Gyroscopes

DynaLabs DYN-I-8000-Triaxial-MEMS-Gyroscopes-FIG- (2)

Product Support

If at any time you have questions or problems with the DYN-I-8000 sensors, please contact a Dynalabs engineer at:

Warranty

Our products are warranted against defective materials and workmanship for one year. Defects arising from user errors are not covered by the warranty.

Copyright

All copyrights of this manual belonging to Dynalabs products are reserved. It cannot be reproduced without written consent.

Disclaimer

Dynalabs Ltd. provides this publication “as is” without warranty of any kind, express or implied, including but not limited to, the implied warranties of merchantability or fitness for a particular purpose. This document is subject to change without notice, and should not be construed as a commitment or representation by Dynalabs Ltd.
This publication may contain inaccuracies or typographical errors. Dynalabs Ltd. will periodically update the material for inclusion in new editions. Changes and improvements to the product described in this manual may be made at any time.

Introduction

Dynalabs Inertial Measurement Units are based on triaxial accelerometers and triaxial gyroscopes that are integrated in a single housing. The inertial measurement unit (IMU) is based on proven microelectromechanical systems (MEMS) accelerometers and gyroscopes for detecting the smallest linear accelerations and angular rates. Dynalabs IMUs enable separate analog voltage outputs for all 6 degrees of freedom (DOF). IMUs enable power supply voltage from 6 to 20 VDC. Dynalabs IMUs feature a lightweight, reliable aluminum housing with protection class IP68 and have cable with configurable length and connectors.

DYN-I-8000 sensors offer the following options;

  • Custom Cable Length
  • Custom Housing Material
  • Custom Connector
  • Base plate

General Information

Unpacking and Inspection

Dynalabs products provide adequate protection for undamaged products to be transported. Document the damages that occur indirectly during the transport and contact the customer representative.

System Components

The DYN-I-8000 has the following components:

  • MEMS Sensor
  • Calibration Certificate
  • Product Manual

Specifications

Table 1 Gyroscopes Specifications Datasheet

Full scale angular velocity 

(°/s)

 

±75

 

±150

 

±300

 

±900

 

Frequency range

 

(Hz)

 

0-150

 

0-150

 

0-150

 

0-150

Non-linearity (full scale) 

(%)

 

0.06

 

0.06

 

0.06

 

0.06

 

Noise (in band)

 

(°/s /√Hz)

 

0.0075

 

0.0075

 

0.0075

 

0.0075

Scale factor (nominal) 

(V/°/s)

 

0.012

 

0.006

 

0.003

 

0.001

Scale factor var. over. temp. 

(%)

 

0.5

 

0.5

 

0.5

 

0.5

Bias variation with temp. 

(°/s)

 

± 1

 

± 2

 

± 3

 

± 4

Table 2: Accecelerometers Specifications Datasheet

Full scale acceceleration 

(g)

 

±2

 

±5

 

±10

 

±30

 

±50

 

±100

 

±200

Frequency range (±5%)(Hz) 

150

 

150

 

300

 

600

 

750

 

750

 

1000

Non-linearity (full scale)(%) 

0.1

 

0.1

 

0.1

 

0.1

 

0.1

 

0.1

 

0.1

 

Noise (in band)

(μg/√ Hz) 

9

 

21

 

40

 

100

 

180

 

340

 

680

Scale factor (nominal)(mV/ g) 

1,350

 

540

 

270

 

90

 

54

 

27

 

13.5

Scale factor temp. coeff.(ppm/° C) 

120

 

120

 

120

 

120

 

120

 

120

 

120

Bias temperature coeff.(mg/° C) 

± 0.2

 

± 0.5

 

± 1

 

± 3

 

± 5

 

± 10

 

± 20

Environmental

Table 3: Environmental Specifications datasheet

Protection LevelIP 68
Operating Voltage6 V – 20 V
Operating Temperature-40 °C to +100 °C
Operating Current Consumption mA<50 mA
IsolationCase isolated

Physical

Table 4: Physical Specifications datasheet

Sensing ElementMEMS Capacitive – Vibrating ring
Housing MaterialAluminum or Steel
Connector (Optional)D-Sub 9 or 15 pin, Lemo, Binder
MountingAdhesive or screw mount
Base plate (Optional)Aluminum or Steel
 

Weight (without cable)

27.5g (aluminum)

69.5 g (steel)

Outline Drawing

The dimensional properties of DYN-I-8000 sensors are given belowDynaLabs DYN-I-8000-Triaxial-MEMS-Gyroscopes-FIG- (3)

Operation and Installation

General

The general sensor connector configuration is given below;

Cable Code/Pin Configuration

  • Red: V + Power supply voltage +6 to +20 VDC
  • Black: Ground Power GND
  • X-Axis: Yellow: Signal(+) Positive, analog output voltage signal for differential mode.
  • Purple: Signal(-) Negative, analog output voltage signal for differential mode.
  • Y-Axis : Blue: Signal(+) Positive, analog output voltage signal for differential mode.
  • Green: Signal(-) Negative, analog output voltage signal for differential mode.
  • Z -Axis: White: Signal(+) Positive, analog output voltage signal for differential mode.
  • Orange: Signal(-) Negative, analog output voltage signal for differential mode

Cable Code/Pin Configuration:

  • Red: V + Power supply voltage +5 V to +20 VDC.
  • Black: Ground Power GND
  • RX: Brown: Gyro Signal X-Axis: Analog output voltage signal for single-ended mode.
  • RY : Grey: Gyro Signal Y-Axis: Analog output voltage signal for single-ended mode.
  • RZ: Pink: Gyro Signal Z-Axis: Analog output voltage signal for single-ended mode.
  • White- Brown: NC Not connected.

WARNING

  • Never connect the power supply and/or the power ground to yellow, purple, blue, green, white, orange, brown, grey, and/or pink cables.
  • Never connect the power supply to the power ground. Always use a clean power source and check the voltage range.

Sensor Static Calibration Verification

  • Using gravity for the accelerometer of the 8000 series IMU, voltage values are measured in the + and – gravity directions, providing a value of ±1 g. The measurement should be made as follows;
  • When the acceleration value of 8000 series IMU’s accelerometers is entered into the data acquisition system, the sensor shows +1 g with the effect of gravity, which is in the direction of the axis to be calibrated.DynaLabs DYN-I-8000-Triaxial-MEMS-Gyroscopes-FIG- (4)
  • When the sensor is positioned in the opposite direction to the axis to be calibrated, the arrow shows -1g as shown below under the effect of gravity.
  • Using gravity, the voltage values that provide 1 g in the + and – directions are measured and compared with the catalog value. The calibration value should be close to the catalog value with 10% tolerance. Sensor catalog sensitivity values are given in Table 2.

Declaration of Conformity

This declaration of conformity is issued under the sole responsibility of the manufacturer. The product(s) are developed, produced and tested according to following EC- directives:

  • 2014/35/EU – Low Voltage Directive (LVD)
  • 2006/42/EU – Machinery Safety Directive
  • 2015/863/EU – RoHS Directive

Applied standards:

  • EN 61010-1:2010
  • EN ISO 12100:2010
  • MIL-STD-810-H-2019 (Test Methods: 501.7 – High Temperature, 502.7 – Low Temperature, 514.8 – Vibration, 516.8 – Shock)

DYNALABS MÜHENDİSLİK SANAYİ TİCARET LİMİTED ŞİRKETİ declares that above mentioned products meet all the requirements of the above mentioned standards and regulations.
Murat Aykan, Technical Manager Ankara, 15.07.2021

Documents / Resources

DynaLabs DYN-I-8000 Triaxial MEMS Gyroscopes [pdf] Instruction Manual
DYN-I-8000 Triaxial MEMS Gyroscopes, DYN-I-8000, DYN-I-8000 Gyroscopes, Triaxial MEMS Gyroscopes, Triaxial MEMS, Triaxial Gyroscopes, MEMS Gyroscopes, Gyroscopes

References

Leave a comment

Your email address will not be published. Required fields are marked *