Bante 810 Benchtop Dissolved Oxygen Meter
USER MANUALBANTE 810 Benchtop Dissolved Oxygen Meter - Dissolved

Bante Instruments Inc.

Introduction

Thank you for selecting the 810 benchtop dissolved oxygen meter. This user manual provides a step-by-step guide to help you operate the meter, please carefully read the following instructions before use. Any use outside of these instructions may invalidate your warranty and cause permanent damage to the meter.

Environmental Conditions
Before unpacking, ensure that current environmental conditions meet the following requirements.

  • Relative humidity is less than 80%
  • Ambient temperature between 0° C (32° F) and 50° C (122° F)
  • No potential electromagnetic interference
  • No corrosive gas exists

Packing List

The following list describes all components of the meter. If any items are missing or damaged, contact the supplier immediately.
810 meter
Electrode arm
9V DC power adapter
DO100 dissolved oxygen electrode
Membrane cap
Electrolyte solution

Meter Overview

BANTE 810 Benchtop Dissolved Oxygen Meter - Oxygen

  1. Sensor connections
  2. Base plate of electrode arm
  3. Display
  4. Membrane keypad

ConnectorsBANTE 810 Benchtop Dissolved Oxygen Meter - connect

  1. Socket for dissolved oxygen electrode (6-pin mini DIN)
  2. Socket for power adapter

DisplayBANTE 810 Benchtop Dissolved Oxygen Meter - Display

IconDescription 
BANTE 810 Benchtop Dissolved Oxygen Meter - IconIndicates that the meter is in the measurement mode
BANTE 810 Benchtop Dissolved Oxygen Meter - IndicatesIndicates that the meter is in the calibration mode
BANTE 810 Benchtop Dissolved Oxygen Meter - meterIndicates that the meter is in the setup mode
BANTE 810 Benchtop Dissolved Oxygen Meter - lockedIndicates that the measurement is locked
ACTIndicates that the automatic temperature compensation is enabled

Keypad
BANTE 810 Benchtop Dissolved Oxygen Meter - Dissolv

Key

Function
BANTE 810 Benchtop Dissolved Oxygen Meter - Switch t• Switch the meter on or off
• Lock or unlock the measurement
• Exit the calibration, settings and return to the measurement mode
BANTE 810 Benchtop Dissolved Oxygen Meter - ToggleToggle between the % saturation and mg/L (ppm) mode
BANTE 810 Benchtop Dissolved Oxygen Meter - Start calibration• Start calibration
• Press and hold the key to enter the setup menu
BANTE 810 Benchtop Dissolved Oxygen Meter - Set theSet the temperature
BANTE 810 Benchtop Dissolved Oxygen Meter - IncreaseIncrease value or scroll up through a list of options
BANTE 810 Benchtop Dissolved Oxygen Meter - DecreaseDecrease value or scroll down through a list of options
BANTE 810 Benchtop Dissolved Oxygen Meter - ConfirmConfirm the calibration or displayed option

Installing the Electrode Holder

Take out the electrode arm from the accessory box. The base plate of electrode arm has a circular hole, the electrode arm has a connecting rod. Insert the connecting rod into the circular hole and swivel the electrode arm 90 degrees. The electrode holder is now ready to swing into desired position.
BANTE 810 Benchtop Dissolved Oxygen Meter - Installing

Adjusting the Electrode Arm
After installation, if the electrode arm automatically rises or falls, you are able to adjust the screw until arm locate at any position.

  1. Remove the plastic cover from the right side of the electrode arm.
    BANTE 810 Benchtop Dissolved Oxygen Meter -Remove
  2. Use the screwdriver to tighten the screw moderately.
    BANTE 810 Benchtop Dissolved Oxygen Meter - moderately
  3. Insert the plastic cover to previous position.

Prior to Use

Filling the Electrolyte Solution

  1. Take out the dissolved oxygen electrode from the packaging. Unscrew the membrane cap from the bottom of the electrode, rinse the inside and outside with distilled water and blot dry.
  2. Fill the membrane cap halfway with electrolyte solution.
    BANTE 810 Benchtop Dissolved Oxygen Meter - membrane
  3. Screw membrane cap back onto the electrode. Some electrolyte solution will overflow during this process.
  4. Check the electrode, ensure that no air bubbles are trapped in the electrolyte solution and membrane is not creased or damaged.
    BANTE 810 Benchtop Dissolved Oxygen Meter - air bubbles

Connecting the Dissolved Oxygen Electrode

  1. Follow the steps below to place electrode into the left or right side of the electrode arm.
    BANTE 810 Benchtop Dissolved Oxygen Meter - Insert 6-pin
  2. Insert 6-pin connector into the connector socket labeled DO. Ensure the connector is fully seated.
    BANTE 810 Benchtop Dissolved Oxygen Meter - After the connection
  3. After the connection is completed, DO NOT pull on the cable. Always make sure that the connector is clean and dry.

Connecting the Power Adapter

  1. Insert the connector of power adapter to the power socket.
  2. Plug the 9V DC power adapter into the wall outlet.BANTE 810 Benchtop Dissolved Oxygen Meter - Connecting

Switching the Meter On and Off

  • Press the Meas key for about 3 seconds and release to switch on the meter.
  • Press and hold the Meas key to switch off the meter.BANTE 810 Benchtop Dissolved Oxygen Meter - Press and

Polarizing the Electrode
Switch on the meter and wait 10 minutes for the electrode to polarize.

Meter Setup

The 810 meter contains an integrated setup menu for customizing the displayed option to meet measurement requirements. The following table describes the functions of each menu item.

Menu Item Option and Description 
BANTE 810 Benchtop Dissolved Oxygen Meter - SalinitySalinity Coefficient
Set the salinity compensation coefficient of sample.
OO 0 to 35 g/L (default 0 g/L)
BANTE 810 Benchtop Dissolved Oxygen Meter - PressurePressure Coefficient
Set the barometric pressure coefficient according to the local altitude (refer to page 4).
760   450 to 850 mmHg (default 760 mmHg)
BANTE 810 Benchtop Dissolved Oxygen Meter - CalibrationCalibration Points
Set the number of calibration points.
1 point (default)
2 points
BANTE 810 Benchtop Dissolved Oxygen Meter -MeasurementMeasurement Unit
Set the default measurement unit.
mg/L Milligrams per liter (default)
ppm Parts per million
% Percentage saturation
°C Degrees Celsius (default)
°F Degrees Fahrenheit
BANTE 810 Benchtop Dissolved Oxygen Meter - Auto-HoldAuto-Hold
If enabled, the meter will automatically sense and lock the measurement endpoint.
BANTE 810 Benchtop Dissolved Oxygen Meter - If enabled Enable
BANTE 810 Benchtop Dissolved Oxygen Meter - Disable Disable (default)
BANTE 810 Benchtop Dissolved Oxygen Meter - AutaAuto-Power Off
If enabled, the meter will automatically switch off if no key is pressed within 8 minutes.
BANTE 810 Benchtop Dissolved Oxygen Meter - If enabled  Enable
BANTE 810 Benchtop Dissolved Oxygen Meter - Disable Disable (default)
BANTE 810 Benchtop Dissolved Oxygen Meter - FactoryFactory Reset
Reset the meter to factory default settings.
Note, the meter must be recalibrated.
BANTE 810 Benchtop Dissolved Oxygen Meter - If enabled Enable
BANTE 810 Benchtop Dissolved Oxygen Meter - Disable Disable (default)

Setting the Default Option

  1. In the measurement mode, press and hold the key to enter the setup menu.
  2. Press the Up / Under key to select an option, press the Enter key to confirm and switch to the next menu item.
  3. Repeat the steps above until the meter returns to the measurement mode.
    BANTE 810 Benchtop Dissolved Oxygen Meter - Setting

instructions To exit the setup menu without saving changes, press the Meas key.
The following table describes the relationship between the altitude and barometric pressure, make sure to set a compatible parameter before the calibration and measurement.

Altitude (m)kPa mmHgAltitude (m) kPa mmHg
0101.3760160082.9622
100100.1750170081.9614
20098.8741180080.9607
30097.6732190079.9599
40096.4723200078.9592
50095.2714210077.9584
60094705220076.9577
70092.8696230076570
80091.7688240075563
90090.5679250074.1556
100089.4671260073.2549
110088.3662270072.3542
120087.2654280071.4536
130086.1646290070.5529
140085638300069.6522
150084630310068.7515

Dissolved Oxygen Calibration

The 810 meter allows 1 or 2 points dissolved oxygen calibration. If you have selected the single point calibration in the setup menu, we recommend that you perform a 100% saturation calibration in the. air-saturated water. If the 2 points calibration is selected, the zero. oxygen solution needs to be used.
Note: During the calibration and measurement, the temperature sensor on electrode must be immersed in the sample solution completely, the solution should keep 0.3 m/s of minimum flow rate to avoid oxygen starvation at the membrane.

BANTE 810 Benchtop Dissolved Oxygen Meter -Temperature

Single Point Calibration – 100% Saturation
Ensure that you have selected 1 point calibration in the setup menu.

  1. Press the Cal key, the display shows CAL1/100%BANTE 810 Benchtop Dissolved Oxygen Meter - Single Point
  2. Hold the dissolved oxygen electrode in the air at 100% relative humidity or place the electrode into the air-saturated water for about 10 minutes.
    BANTE 810 Benchtop Dissolved Oxygen Meter -Hold the dissolvedPress the Enter key to begin the calibration.BANTE 810 Benchtop Dissolved Oxygen Meter - Press the Enter
  3. When the reading has stabilized, the meter will show and return to the measurement mode.

2 Point Calibration
Ensure that you have selected 2 points calibration in the setup menu.

  1. Repeat steps 1.1 and 1.2 above. When the first calibration point. is completed, the display will show CAL2/0%, the meter prompts. you to continue with second point calibration.BANTE 810 Benchtop Dissolved Oxygen Meter - calibrationa
  2. Place the dissolved oxygen electrode into the zero oxygen solution. for about 10 minutes, press the Enter key to begin the calibration.

BANTE 810 Benchtop Dissolved Oxygen Meter - calibration

When the reading has stabilized, the meter will show End. Calibration is completed.

instructions • Performing a percentage saturation calibration will calibrate the corresponding mg/L or ppm concentration value simultaneously.
• To exit the calibration without saving calibrated values, press the Meas key.

Temperature Calibration

  1. Place the dissolved oxygen electrode into a solution with a known accurate temperature.
  2. Press and hold the °C key to enter the temperature setting.
  3. Press the Down/UP key to modify the temperature value.
  4. Press the Enter key to save.BANTE 810 Benchtop Dissolved Oxygen Meter - Temperature

Measurement

The 810 meter can be used to measure the water, wastewater, brine and other liquids. If your sample is seawater or water containing large amounts of salt, make sure to set the salinity coefficient before measurement. Some gas and steam such as chloride, sulfur dioxide, sulfureted hydrogen and carbon dioxide can permeate the membrane via diffusion. Their existence will influence the measurements. If the sample contains solvent, grease, sulfide and alga, the membrane will be damaged or eroded.

  1. Set the barometric pressure and salinity coefficient in the setup menu.
  2. Rinse the electrode with distilled water, place the electrode into the sample solution and stir gently.
    BANTE 810 Benchtop Dissolved Oxygen Meter - Rinse the electrode
  3.  If the Auto-Hold option in the setup menu is enabled, the meter will automatically sense a stable reading and lock measurement, theBANTE 810 Benchtop Dissolved Oxygen Meter - locked icon appears on the display. Press the Meas key to resume measuring.
    If the option is disabled, the meter will continuously measure and update the readings.
    BANTE 810 Benchtop Dissolved Oxygen Meter - If the option
  4. Wait for the measurement to stabilize and record the reading.
  5. When all of the samples have been measured, rinse the electrode with distilled water.

instructions If the meter shows —- indicating the measurement exceeds the range, remove the electrode from the sample solution immediately.

Electrode Maintenance

  • Rinse the dissolved oxygen electrode thoroughly with distilled water after use.
  • Do not touch the membrane and always keep it is clean and wet.
  • If you do not use the electrode for long periods, screw off the membrane cap, rinse the electrode anode, cathode, membrane cap with distilled water and blot dry. Install the electrode and store dry.
    BANTE 810 Benchtop Dissolved Oxygen Meter - Cathode

Appendix

Optional Accessories

Order CodeDescription
DO100Dissolved oxygen electrode, range: 0 to 20 mg/L
DO-MEMMembrane cap, 2 PCS/set
DO-ESElectrolyte solution, 480 ml
DCPA-9V9V DC power adapter, european standard plug

Preparation of Air-Saturated Water
Use an air-pump to blow air into distilled water at least 1 hour, while stirring the solution.

Preparation of Zero Oxygen Solution
Dissolve 500 mg of the sodium sulfate (Na2SO3) reagent and a small amount of cobalt (II) chloride hexahydrate (CoCl2 • 6H2O) in the 250 ml distilled water, mix the solution until reagent is completely dissolved.
Meter Specifications

ModelBante 810
Dissolved Oxygen
Range0.0 to 20.0 mg/L
Resolution0.1 mg/L
Accuracy± 0.5 mg/L
% Saturation
Range0.0 to 200.0%
Resolution0.10%
Accuracy± 2.0%
Other Specifications
Calibration Point1 or 2 points
Temperature Compensation0 to 40° C (32 to 104° F), automatic
Barometric Pressure Correction450 to 850 mmHg, manual
Salinity Correction0 to 35 g/L, manual
Operating Temperature0 to 50° C (32 to 122° F)
Storage Temperature0 to 60° C (32 to 140° F)
Relative Humidity< 80% (non-condensing)
DisplayLCD, 135 ×  75 mm (5.3 ×  2.9 in.)
Power Requirements9V/400mA DC power adapter
Auto-Off3 hours after last key pressed
Dimensions210 (L) ×  205 (W) ×  75 (H) mm,
(8.2 ×  8.0 ×  2.9 in.)
Weight1.5 kg (3.3 lb)

Troubleshooting

FaultCause and Corrective Action
—–Dissolved oxygen electrode does not connect to the meter or measured value is out of range.
Drifting erratic readingsCheck whether the membrane cap is contaminated or the electrolyte solution is depleted.
ErrElectrode is broken.
Replace the dissolved oxygen electrode.

Disposal
This product is required to comply with the European Union’s Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC and may not be disposed of in domestic waste. Please dispose of product in accordance with local regulations at the collecting point specified for electrical and electronic equipment.
BANTE 810 Benchtop Dissolved Oxygen Meter - fsss
Warranty
The warranty period for meter is one year from the date of shipment. Above warranty does not cover the electrode and electrolyte solution. Out of warranty products will be repaired on a charged basis. The warranty on your meter shall not apply to defects resulting from:

  • Improper or inadequate maintenance by customer
  • Unauthorized modification or misuse
  • Operation outside of the environment specifications of the products.

For more information, please contact the supplier.

BANTE logo

Office: 4715 Castlewood St., Sugar land, TX 77479, USA
Tel: (+1) 346-762-7358
E-mail: banteinstruments@yahoo.com
Factory: A10, No.2066, Laifang Rd., Shanghai 201615, China
Tel: (+86) 21-6404-1598
E-mail: banteinstrument@hotmail.com
BANTE 810 Benchtop Dissolved Oxygen Meter - bantein www.bante-china.com

BANTE 810 Benchtop Dissolved Oxygen Meter - QRhttp://bante-china.com/enHome/
The information in this document is subject to change without notice.
Copyright © Bante Instruments Inc, 2022. All rights reserved.