User Manual for BANTE models including: 810, Benchtop Dissolved Oxygen Meter, 810 Benchtop Dissolved Oxygen Meter

User Manual Bante 810 Benchtop Dissolved Oxygen Meter

Bante Instruments

User Manual Bante 810 Benchtop Dissolved Oxygen Meter

This user manual provides a step-by-step guide to help you operate the meter, please carefully read the following instructions before use. Any use outside of ...

3 I Setup Set the barometric pressure coefficient according to the Switching the Meter On and Off Enable Enable Factory Reset 1 2.2 Insert 6-pin connector into the connector socket labeled DO.


File Info : application/pdf, 8 Pages, 1.12MB

PDF preview unavailable. Download the PDF instead.

User Manual Bante 810 Benchtop Dissolved Oxygen Meter
Bante 810 Benchtop Dissolved Oxygen Meter
USER MANUAL
Bante Instruments Inc.

1 I Overview

Introduction
Thank you for selecting the 810 benchtop dissolved oxygen meter. This user manual provides a step-by-step guide to help you operate the meter, please carefully read the following instructions before use. Any use outside of these instructions may invalidate your warranty and cause permanent damage to the meter.
Environmental Conditions
Before unpacking, ensure that current environmental conditions meet the following requirements. · Relative humidity is less than 80% · Ambient temperature between 0°C (32°F) and 50°C (122°F) · No potential electromagnetic interference · No corrosive gas exists
Packing List
The following list describes all components of the meter. If any items are missing or damaged, contact the supplier immediately.
810 meter Electrode arm 9V DC power adapter DO100 dissolved oxygen electrode Membrane cap Electrolyte solution
Meter Overview

Connectors
1 Socket for dissolved oxygen electrode (6-pin mini DIN) 2 Socket for power adapter
Display

Icon

Description

Indicates that the meter is in the measurement mode

Indicates that the meter is in the calibration mode

Indicates that the meter is in the setup mode

Indicates that the measurement is locked
Indicates that the automatic temperature compensation is enabled

Keypad

1 Sensor connections 2 Base plate of electrode arm 3 Display 4 Membrane keypad

Overview I 2

Key

Function

1. Remove the plastic cover from the right side of the electrode arm.

· Switch the meter on or off · Lock or unlock the measurement · Exit the calibration, settings and return to the
measurement mode

Toggle between the % saturation and mg/L (ppm) mode
· Start calibration · Press and hold the key to enter the setup menu

2. Use the screwdriver to tighten the screw moderately.

Set the temperature

Increase value or scroll up through a list of options
Decrease value or scroll down through a list of options

3. Insert the plastic cover to previous position.

Confirm the calibration or displayed option
Installing the Electrode Holder
Take out the electrode arm from the accessory box. The base plate of electrode arm has a circular hole, the electrode arm has a connecting rod. Insert the connecting rod into the circular hole and swivel the electrode arm 90 degrees. The electrode holder is now ready to swing into desired position.

Prior to Use
Filling the Electrolyte Solution
1.1 Take out the dissolved oxygen electrode from the packaging. Unscrew the membrane cap from the bottom of the electrode, rinse the inside and outside with distilled water and blot dry.
1.2 Fill the membrane cap halfway with electrolyte solution.

1.3 Screw membrane cap back onto the electrode. Some electrolyte solution will overflow during this process.
1.4 Check the electrode, ensure that no air bubbles are trapped in the electrolyte solution and membrane is not creased or damaged.

Remove air bubbles

Adjusting the Electrode Arm
After installation, if the electrode arm automatically rises or falls, you are able to adjust the screw until arm locate at any position.

Connecting the Dissolved Oxygen Electrode
2.1 Follow the steps below to place electrode into the left or right side of the electrode arm.

3 I Setup

1 2
3
2.2 Insert 6-pin connector into the connector socket labeled DO. Ensure the connector is fully seated.
2.3 After the connection is completed, DO NOT pull on the cable. Always make sure that the connector is clean and dry.
Connecting the Power Adapter
3.1 Insert the connector of power adapter to the power socket. 3.2 Plug the 9V DC power adapter into the wall outlet.
Switching the Meter On and Off
· Press the Meas key for about 3 seconds and release to switch on the meter.
· Press and hold the Meas key to switch off the meter.
Polarizing the Electrode
Switch on the meter and wait 10 minutes for the electrode to polarize.

Meter Setup
The 810 meter contains an integrated setup menu for customizing the displayed option to meet measurement requirements. The following table describes the functions of each menu item.

Menu Item

Option and Description

Salinity Coefficient Set the salinity compensation coefficient of sample.

0 to 35 g/L (default 0 g/L)

Pressure Coefficient Set the barometric pressure coefficient according to the local altitude (refer to page 4).

450 to 850 mmHg (default 760 mmHg)

Calibration Points Set the number of calibration points.

1 point (default) 2 points

Measurement Unit Set the default measurement unit.

mg/L

Milligrams per liter (default)

ppm

Parts per million

%

Percentage saturation

Degrees Celsius (default)

Degrees Fahrenheit

Auto-Hold If enabled, the meter will automatically sense and lock the measurement endpoint.

Enable Disable (default)

Auto-Power Off If enabled, the meter will automatically switch off if no key is pressed within 8 minutes.

Enable Disable (default)

Factory Reset Reset the meter to factory default settings. Note, the meter must be recalibrated.

Enable Disable (default)

Calibration I 4

Setting the Default Option
1. In the measurement mode, press and hold the key to enter the setup menu.
2. Press the  /  key to select an option, press the Enter key to confirm and switch to the next menu item.
3. Repeat the steps above until the meter returns to the measurement mode.

Dissolved Oxygen Calibration
The 810 meter allows 1 or 2 points dissolved oxygen calibration. If you have selected the single point calibration in the setup menu, we recommend that you perform a 100% saturation calibration in the. air-saturated water. If the 2 points calibration is selected, the zero. oxygen solution needs to be used.
Note: During the calibration and measurement, the temperature sensor on electrode must be immersed in the sample solution completely, the solution should keep 0.3 m/s of minimum flow rate to avoid oxygen starvation at the membrane.

Menu item

Option

Temperature sensor
Single Point Calibration - 100% Saturation
Ensure that you have selected 1 point calibration in the setup menu. 1.1 Press the Cal key, the display shows CAL1/100%

To exit the setup menu without saving changes, press the Meas key.
The following table describes the relationship between the altitude and barometric pressure, make sure to set a compatible parameter before the calibration and measurement.

Altitude (m) kPa

0

101.3

100 100.1

200

98.8

300

97.6

400

96.4

500

95.2

600

94.0

700

92.8

800

91.7

900

90.5

1000 89.4

1100 88.3

1200 87.2

1300 86.1

1400 85.0

1500 84.0

mmHg Altitude (m) kPa

760

1600 82.9

750

1700 81.9

741

1800 80.9

732

1900 79.9

723

2000 78.9

714

2100 77.9

705

2200 76.9

696

2300 76.0

688

2400 75.0

679

2500 74.1

671

2600 73.2

662

2700 72.3

654

2800 71.4

646

2900 70.5

638

3000 69.6

630

3100 68.7

mmHg 622 614 607 599 592 584 577 570 563 556 549 542 536 529 522 515

1.2 Hold the dissolved oxygen electrode in the air at 100% relative humidity or place the electrode into the air-saturated water for about 10 minutes.
Press the Enter key to begin the calibration.
1.3 When the reading has stabilized, the meter will show and return to the measurement mode.

5 I Measurement

2 Point Calibration
Ensure that you have selected 2 points calibration in the setup menu. 2.1 Repeat steps 1.1 and 1.2 above. When the first calibration point.
is completed, the display will show CAL2/0%, the meter prompts. you to continue with second point calibration.
2.2 Place the dissolved oxygen electrode into the zero oxygen solution. for about 10 minutes, press the Enter key to begin the calibration.

Measurement
The 810 meter can be used to measure the water, wastewater, brine and other liquids. If your sample is seawater or water containing large amounts of salt, make sure to set the salinity coefficient before measurement. Some gas and steam such as chloride, sulfur dioxide, sulfureted hydrogen and carbon dioxide can permeate the membrane via diffusion. Their existence will influence the measurements. If the sample contains solvent, grease, sulfide and alga, the membrane will be damaged or eroded.
1. Set the barometric pressure and salinity coefficient in the setup menu.
2. Rinse the electrode with distilled water, place the electrode into the sample solution and stir gently.

When the reading has stabilized, the meter will show . Calibration is completed.
· Performing a percentage saturation calibration will calibrate the corresponding mg/L or ppm concentration value simultaneously.
· To exit the calibration without saving calibrated values, press the Meas key.

3. If the Auto-Hold option in the setup menu is enabled, the meter will automatically sense a stable reading and lock measurement, the icon appears on the display. Press the Meas key to resume measuring. If the option is disabled, the meter will continuously measure and update the readings.

Temperature Calibration
1. Place the dissolved oxygen electrode into a solution with a known accurate temperature.
2. Press and hold the °C key to enter the temperature setting. 3. Press the  /  key to modify the temperature value. 4. Press the Enter key to save.

4. Wait for the measurement to stabilize and record the reading. 5. When all of the samples have been measured, rinse the electrode
with distilled water.

If the meter shows ---- indicating the measurement exceeds the range, remove the electrode from the sample solution immediately.

Appendix I 6

Electrode Maintenance
· Rinse the dissolved oxygen electrode thoroughly with distilled water after use.
· Do not touch the membrane and always keep it is clean and wet. · If you do not use the electrode for long periods, screw off the
membrane cap, rinse the electrode anode, cathode, membrane
cap with distilled water and blot dry. Install the electrode and
store dry.

Cathode Anode

Appendix

Optional Accessories

Order Code Description

DO100

Dissolved oxygen electrode, range: 0 to 20 mg/L

DO-MEM Membrane cap, 2 PCS/set

DO-ES

Electrolyte solution, 480 ml

DCPA-9V

9V DC power adapter, european standard plug

Preparation of Air-Saturated Water
Use an air-pump to blow air into distilled water at least 1 hour, while stirring the solution.

Preparation of Zero Oxygen Solution
Dissolve 500 mg of the sodium sulfate (Na2SO3) reagent and a small amount of cobalt (II) chloride hexahydrate (CoCl2· 6H2O) in the 250 ml distilled water, mix the solution until reagent is completely dissolved.

Meter Specifications
Model Dissolved Oxygen Range Resolution Accuracy % Saturation Range Resolution Accuracy

Bante 810
0.0 to 20.0 mg/L 0.1 mg/L ±0.5 mg/L
0.0 to 200.0% 0.1% ±2.0%

Other Specifications

Calibration Point

1 or 2 points

Temperature Compensation 0 to 40°C (32 to 104°F), automatic

Barometric Pressure Correction 450 to 850 mmHg, manual

Salinity Correction

0 to 35 g/L, manual

Operating Temperature

0 to 50°C (32 to 122°F)

Storage Temperature

0 to 60°C (32 to 140°F)

Relative Humidity

< 80% (non-condensing)

Display

LCD, 135 × 75 mm (5.3 × 2.9 in.)

Power Requirements

9V/400mA DC power adapter

Auto-Off

3 hours after last key pressed

Dimensions

210 (L) × 205 (W) × 75 (H) mm, (8.2 × 8.0 × 2.9 in.)

Weight

1.5 kg (3.3 lb)

Troubleshooting

Fault

Cause and Corrective Action

Dissolved oxygen electrode does not connect to the meter or measured value is out of range.

Drifting erratic Check whether the membrane cap is contaminated

readings

or the electrolyte solution is depleted.

Electrode is broken. Replace the dissolved oxygen electrode.

Disposal
This product is required to comply with the European Union's Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC and may not be disposed of in domestic waste. Please dispose of product in accordance with local regulations at the collecting point specified for electrical and electronic equipment.
Warranty
The warranty period for meter is one year from the date of shipment. Above warranty does not cover the electrode and electrolyte solution. Out of warranty products will be repaired on a charged basis. The warranty on your meter shall not apply to defects resulting from: · Improper or inadequate maintenance by customer · Unauthorized modification or misuse · Operation outside of the environment specifications of the
products. For more information, please contact the supplier.

Office: 4715 Castlewood St., Sugar land, TX 77479, USA Tel: (+1) 346-762-7358 E-mail: banteinstruments@yahoo.com
Factory: A10, No.2066, Laifang Rd., Shanghai 201615, China Tel: (+86) 21-6404-1598 E-mail: banteinstrument@hotmail.com
 www.bante-china.com

The information in this document is subject to change without notice. Copyright © Bante Instruments Inc, 2022. All rights reserved.



References

Microsoft Word 2016