GOYOJO STE-18-E

GOYOJO 18L Autoclave Sterilizer User Manual

Model: STE-18-E

1. Product Overview

The GOYOJO 18L Autoclave Sterilizer is a tabletop Class N high-pressure steam sterilizer designed for efficient sterilization of solid, unwrapped instruments. It features digital controls, fast cycle options, and a secure door lock, making it suitable for various professional settings including tattoo studios, dental offices, beauty salons, and laboratories.

GOYOJO 18L Autoclave Sterilizer main unit

Image 1.1: GOYOJO 18L Autoclave Sterilizer.

Autoclave applications in dental, tattoo, beauty, and lab settings

Image 1.2: Autoclave applications in various professional settings.

2. Important Safety Notes

Adhering to safety guidelines is crucial for the proper and safe operation of the autoclave. Please read and understand all instructions before use.

Important Safety Notes for Autoclave Use

Image 2.1: Important Safety Notes.

  • Use only distilled or deionized water. Tap water can cause scaling and damage the heating element.
  • Ensure the chamber and trays are clean and free of residue before each cycle.
  • Do not overload the chamber. Arrange instruments neatly to allow steam circulation.
  • Check the rubber door gasket regularly. Replace if worn or deformed to maintain airtight sealing.
  • Confirm that the exhaust valve is securely connected to a safe metal drainage container.
  • Do not open the door until the pressure gauge returns to zero.
  • Keep the unit on a stable, flat, and heat-resistant surface with good ventilation.
  • This sterilizer is for solid instruments only (Class N). Not suitable for hollow or porous instruments.
  • Do not place alcohol or other disinfectant liquids, sealed containers, or sharp objects that may damage trays.
  • Disconnect power when not in use or during maintenance.

3. Unboxing and Components

Upon receiving your GOYOJO 18L Autoclave Sterilizer, carefully unbox the unit and verify all components are present and undamaged.

Video 3.1: Unboxing the GOYOJO 18L Autoclave Sterilizer.

Wooden crate packaging for the autoclave

Image 3.1: The autoclave is shipped in a protective wooden crate.

Autoclave components including trays, rack, exhaust tube, power cord, fuse, and handle

Image 3.2: Included components.

Included Accessories:

  • Sterilizer Unit
  • 3 Removable Trays
  • Tray Rack
  • Tray Handle
  • Exhaust Tube
  • Power Cord
  • Spare Fuse

4. Setup

Proper setup ensures safe and effective sterilization.

  1. Placement: Place the autoclave on a stable, flat, and heat-resistant surface with adequate ventilation.
  2. Water Filling: Open the door and pour distilled or deionized water into the chamber. Ensure the water level is above the heating element but below the maximum fill line. Never use tap water.
  3. Load Instruments: Place instruments on the trays, ensuring they are unwrapped and spaced apart to allow proper steam circulation. Use the provided tray handle to insert and remove trays.
  4. Connect Drain Hose: Attach the exhaust tube to the drain outlet and place the other end into a suitable metal drainage container.
  5. Close and Secure Door: Close the door firmly. Rotate the door lock handle clockwise until it is tightly sealed. The digital display should show 'Ld' (Load) when the door is properly closed and locked. If 'do' is displayed, the door is not fully closed.
  6. Connect Power: Plug the power cord into a 110V North American outlet.

Video 4.1: Detailed operation of the GOYOJO 18L Autoclave Sterilizer.

Autoclave with dental and tattoo tools on trays

Image 4.1: Instruments loaded onto trays inside the autoclave.

Power switch on the autoclave

Image 4.2: Power switch.

5. Operating Instructions

Follow these steps to operate the autoclave for sterilization.

  1. Power On: Flip the main power switch to the 'On' position. The LED display will illuminate.
  2. Select Temperature: Use the temperature selection button to choose between 121°C (30 minutes) or 134°C (10 minutes). The indicator light next to the selected temperature will light up. For most applications, 134°C is faster and equally effective.
  3. Start Cycle: Press the start button to begin the sterilization cycle. The display will show the current temperature and pressure.
  4. Monitor Cycle: The autoclave will heat up, build pressure, and maintain the selected temperature for the specified duration. Monitor the pressure gauge to ensure it operates within safe limits.
  5. End of Cycle: Once the sterilization time is complete, the unit will beep and display 'End'.
  6. Pressure Release: Before opening the door, ensure the pressure gauge reads zero. Open the drain valve to release any remaining pressure and steam.
  7. Cooling: Slightly open the door to allow natural cooling and air drying of the instruments. This unit does not have an active drying cycle.
  8. Remove Instruments: Once cooled, use the tray handle to carefully remove the sterilized instruments.

Video 5.1: Operation guide for the 18L autoclave sterilizer.

Digital LED control panel with temperature and cycle indicators

Image 5.1: Digital LED control panel.

LED display control panel on the autoclave

Image 5.2: Close-up of the LED display control.

6. Maintenance

Regular maintenance ensures the longevity and optimal performance of your autoclave.

  • Draining Water: After each use, once the unit has cooled and pressure is released, open the drain valve to empty the water reservoir.
  • Cleaning: Regularly clean the interior of the chamber and trays with a soft cloth and mild, non-abrasive cleaner. Ensure no residue remains.
  • Gasket Inspection: Periodically inspect the door gasket for any signs of wear, cracks, or deformation. A damaged gasket can compromise the seal and sterilization effectiveness. Replace if necessary.
  • Exterior Cleaning: Wipe down the exterior of the unit with a damp cloth. Do not use harsh chemicals.
Exhaust and drain valve on the autoclave

Image 6.1: Exhaust and Drain Valve.

Drain switch on the autoclave

Image 6.2: Drain Switch.

Door gasket on the autoclave

Image 6.3: Door Gasket.

Sealing gasket for the autoclave door

Image 6.4: Sealing Gasket.

7. Troubleshooting

This section provides solutions to common issues you might encounter.

  • Unit Not Powering On: Check the power cord connection and ensure the power switch is in the 'On' position. Verify the spare fuse is correctly installed and not blown.
  • Door Not Sealing/Displaying 'do': Ensure the door is fully closed and the lock handle is rotated clockwise until tight. Inspect the door gasket for any obstructions or damage.
  • No Pressure Build-up: Confirm the drain valve is fully closed. Check the water level in the chamber. Ensure the door is properly sealed.
  • Steam Leakage: Check the door gasket for proper seating and damage. Ensure the door lock is fully engaged.
  • Sterilization Cycle Not Completing: Verify the correct temperature and time settings. Ensure adequate water is in the reservoir.

8. Specifications

Detailed technical specifications for the GOYOJO 18L Autoclave Sterilizer.

Technical parameters and dimensions of the autoclave

Image 8.1: Technical Parameters.

ParameterValue
Model NumberSTE-18-E
Voltage110-130VAC
Power1100 VA
Chamber Size (Inner Diameter/Depth)Φ230 x 360 mm (9.1 x 14.2 in)
Overall Size (W x H x D)560 x 365 x 335 mm (22.0 x 14.4 x 13.2 in)
Net Weight23 kg (50.7 lbs)
Gross Weight28 kg (61.7 lbs)
Sterilization Temperature121°C / 134°C
Sterilization Time (at 121°C)30 minutes
Sterilization Time (at 134°C)10 minutes

9. Warranty & Support

The GOYOJO 18L Autoclave Sterilizer comes with a 1-year warranty.

  • If non-human damage or quality issues occur within 30 days of purchase, please contact us at serenity@goyojo.com for a replacement.
  • For returns, please follow the seller’s return process.
  • When contacting support, provide video, photos, and a detailed problem description. Our team aims to offer a solution within 24 hours.