RICHMETERS RM113A Digital Multimeter User Manual

1. Important Safety Information

Please read and understand all safety information and operating instructions before using this multimeter. Failure to follow these instructions may result in electric shock, fire, or damage to the meter.

  • Always ensure the test leads are properly connected and the function switch is set to the correct range before making any measurements.
  • Do not attempt to measure voltages or currents exceeding the maximum rated values for the meter.
  • Exercise extreme caution when working with live circuits. Avoid contact with bare wires or terminals.
  • Never use the meter if it appears damaged or if the test leads are compromised.
  • Replace batteries promptly when the low battery indicator appears to ensure accurate readings.
  • Do not operate the meter in explosive atmospheres or in the presence of flammable gases or dust.
  • Keep hands behind the probe barriers during measurements.

2. Product Overview

2.1 Introduction

The RICHMETERS RM113A is a compact and versatile digital multimeter designed for measuring AC/DC voltage, DC current, resistance, and performing non-contact voltage (NCV) detection, HFE tests, diode tests, and continuity checks. It features a large, backlit LCD display, a built-in flashlight, and magnetic suction for convenient operation.

2.2 Key Features

  • Versatile Electrical Tests: Allows for easy testing of resistance, voltage, and current with a simple switch.
  • Clear Display: Features a large screen with high contrast and full symbol display for easy reading.
  • Versatile Functions: Equipped with a backlight and flashlight for operation in dark environments.
  • Secure Attachment: Magnetic suction function allows the multimeter to stick on metal surfaces securely.
  • One-Handed Operation: Designed for ease of use, facilitating on-the-go measurements.

2.3 Product Components

RICHMETERS RM113A Digital Multimeter with Red and Black Test Leads

Figure 1: RICHMETERS RM113A Digital Multimeter and Test Leads.

The package includes the RICHMETERS RM113A Digital Multimeter and a pair of electronic test probes (one red, one black).

Labeled diagram of RICHMETERS RM113A Digital Multimeter showing NCV Sensor, LCD Display, HFE Testing Input Terminal, Data Hold button, Backlight Key, Protective Shell, Function Knob, and Input Terminals.

Figure 2: Front view of the RM113A with labeled components.

  • NCV Sensor: Detects non-contact voltage.
  • LCD Display: Shows measurement readings and indicators.
  • HFE Testing Input Terminal: Used for transistor HFE measurements.
  • HOLD Button: Freezes the current display reading.
  • Backlight Key: Activates the display backlight and flashlight.
  • Protective Shell: Provides durability and protection.
  • Function Knob: Selects the desired measurement function and range.
  • Input Terminals: Sockets for connecting test leads (COM, VΩmA, 10A).
Back view of RICHMETERS RM113A Digital Multimeter showing magnetic suction strips and test lead storage slots.

Figure 3: Back view of the RM113A, highlighting magnetic suction and test lead storage.

The back of the multimeter features magnetic strips for attaching to metal surfaces and slots for convenient storage of the test leads.

3. Setup

3.1 Battery Installation

The RICHMETERS RM113A requires three (3) AAA batteries (not included) for operation.

  1. Locate the battery compartment cover on the back of the multimeter.
  2. Use a screwdriver to remove the screw securing the battery cover.
  3. Carefully remove the cover.
  4. Insert three AAA batteries, observing the correct polarity (+ and -) as indicated inside the compartment.
  5. Replace the battery cover and secure it with the screw.

3.2 Connecting Test Leads

To prepare for measurements, connect the test leads as follows:

  • Insert the black test lead into the COM (Common) input terminal.
  • For most voltage, resistance, and current measurements (up to 200mA), insert the red test lead into the VΩmA input terminal.
  • For high current measurements (up to 10A), insert the red test lead into the 10A input terminal. Ensure the function knob is set to the appropriate current range.

4. Operating Instructions

4.1 Power On/Off

Rotate the central function knob from the "OFF" position to any desired measurement function to power on the multimeter. To power off, rotate the knob back to the "OFF" position.

4.2 Function Selection

The function knob allows you to select various measurement modes. Ensure the knob is securely clicked into the desired position.

RICHMETERS RM113A Digital Multimeter displaying various function icons including Data Hold, HFE Test, Diode, AC/DC Voltage, Continuity, AC/DC Current, Auto Power Off, Capacitance, Frequency, Flashlight, Resistance, Duty Cycle, and Backlight.

Figure 4: Icons representing the various functions of the RM113A.

4.3 Measuring DC Voltage (V–)

  1. Connect the black test lead to the COM terminal and the red test lead to the VΩmA terminal.
  2. Rotate the function knob to the desired DC Voltage range (e.g., 200V, 600V). If the voltage is unknown, start with the highest range and decrease as needed.
  3. Connect the test probes across the component or circuit to be measured.
  4. Read the voltage value on the LCD display.

4.4 Measuring AC Voltage (V∼)

  1. Connect the black test lead to the COM terminal and the red test lead to the VΩmA terminal.
  2. Rotate the function knob to the desired AC Voltage range (e.g., 200mV, 2V, 20V, 200V, 600V).
  3. Connect the test probes across the AC voltage source.
  4. Read the voltage value on the LCD display.

4.5 Measuring DC Current (A–)

Caution: Never connect the multimeter in parallel with a voltage source when measuring current, as this can damage the meter and the circuit. Always connect in series.

  1. Connect the black test lead to the COM terminal.
  2. For currents up to 200mA, connect the red test lead to the VΩmA terminal. For currents up to 10A, connect the red test lead to the 10A terminal.
  3. Rotate the function knob to the desired DC Current range (e.g., 2mA, 20mA, 200mA, 10A).
  4. Open the circuit where current is to be measured and connect the multimeter in series with the load.
  5. Read the current value on the LCD display.

4.6 Measuring Resistance (Ω)

Caution: Ensure the circuit is de-energized before measuring resistance. Do not measure resistance on live circuits.

  1. Connect the black test lead to the COM terminal and the red test lead to the VΩmA terminal.
  2. Rotate the function knob to the desired Resistance range (e.g., 200Ω, 2KΩ, 200KΩ, 2MΩ, 20MΩ).
  3. Connect the test probes across the component whose resistance is to be measured.
  4. Read the resistance value on the LCD display.

4.7 Non-Contact Voltage (NCV) Detection

The NCV function allows for detection of AC voltage without direct contact with conductors.

  1. Rotate the function knob to the NCV position.
  2. Move the top of the multimeter (where the NCV sensor is located) close to the conductor or outlet.
  3. If AC voltage is detected, the meter will emit an audible beep and the NCV indicator light will illuminate.

4.8 HFE Test (Transistor Test)

This function is used to measure the DC current gain (hFE) of transistors.

  1. Rotate the function knob to the hFE position.
  2. Insert the transistor leads (Emitter, Base, Collector) into the corresponding holes in the HFE testing input terminal, ensuring correct type (NPN or PNP).
  3. Read the hFE value on the LCD display.

4.9 Diode Test

The diode test function checks the forward voltage drop of a diode.

  1. Connect the black test lead to the COM terminal and the red test lead to the VΩmA terminal.
  2. Rotate the function knob to the Diode symbol.
  3. Connect the red probe to the anode and the black probe to the cathode of the diode.
  4. Read the forward voltage drop on the display. Reversing the probes should show an open circuit (OL).

4.10 Continuity Test

The continuity test checks for a complete circuit path.

  1. Connect the black test lead to the COM terminal and the red test lead to the VΩmA terminal.
  2. Rotate the function knob to the Continuity symbol.
  3. Connect the test probes across the circuit or component.
  4. If continuity exists (resistance is low), the meter will emit an audible beep.

4.11 Data Hold Function

Press the "HOLD" button to freeze the current reading on the display. Press it again to release the hold function and resume live readings.

4.12 Backlight and Flashlight

The RM113A features a backlight for the display and a built-in flashlight for working in dimly lit areas.

RICHMETERS RM113A Digital Multimeter with its display backlight and flashlight activated in a dark environment, illuminating the surrounding area.

Figure 5: Multimeter with backlight and flashlight active.

  • Press the Backlight Key (often marked with a sun or light bulb icon) to turn on the display backlight. Press again to turn it off.
  • Long-press the Backlight Key to activate the flashlight. Long-press again to turn it off.

5. Maintenance

5.1 Cleaning

Wipe the meter with a damp cloth and mild detergent. Do not use abrasives or solvents. Ensure the meter is dry before use.

5.2 Battery Replacement

When the low battery indicator appears on the display, replace the batteries as described in Section 3.1. Remove batteries if the meter will not be used for an extended period.

5.3 Storage

Store the multimeter in a cool, dry place, away from direct sunlight and extreme temperatures. Remove batteries if storing for long periods.

6. Troubleshooting

If you encounter issues with your RICHMETERS RM113A, refer to the following table for common problems and solutions.

ProblemPossible CauseSolution
Meter does not power on.Dead or incorrectly installed batteries.Check battery polarity and replace batteries if necessary.
No reading or "OL" displayed.Incorrect function/range selected; open circuit; faulty test leads.Verify function knob setting; check for continuity in the circuit; inspect and replace test leads if damaged.
Inaccurate readings.Low battery; incorrect range; external interference.Replace batteries; select a more appropriate range; move away from strong electromagnetic fields.
Backlight/Flashlight not working.Low battery; button malfunction.Replace batteries. If issue persists, contact customer support.
NCV not detecting voltage.Too far from source; low voltage; shielded wire.Ensure close proximity to unshielded live wire.

7. Technical Specifications

The following table outlines the technical specifications for the RICHMETERS RM113A Digital Multimeter.

ParameterSpecification
Main MaterialPlastics
AltitudeUnder 2000m
Working Temperature0-40℃ (<80%RH, not considered <10℃)
Storage Temperature-10-60℃ (<80%RH, take off battery)
DC Voltage Range200V/600V
DC Voltage Resolution100mV/1V
AC Voltage Range200mV/2V/20V/200V/600V
AC Voltage Resolution0.1mV/1mV/10mV/100mV/1V
DC Current Range2mA/20mA/200mA/10A
DC Current Resolution1uA/10uA/100uA/10mA
Resistance Range200Ω/2KΩ/200KΩ/2MΩ/20MΩ
Resistance Resolution0.1Ω/1Ω/100Ω/1KΩ/10KΩ
Battery Type3 * AAA batteries (NOT included)
Item Weight216g / 7.62 ounces
Item Size147 * 71 * 45 mm / 5.79 * 2.80 * 1.77 inches
Power SourceAC/DC