Elitech RC-51

Elitech RC-51 Digital Temperature Data Logger User Manual

Model: RC-51 | Brand: Elitech

Introduction

The Elitech RC-51 Digital Temperature Data Logger is a reusable USB device designed for precise temperature monitoring. It records up to 32,000 temperature points and automatically generates PDF reports, making it ideal for various applications including cold chain management, vaccine storage, and industrial processes.

Key Features

Product Overview

Elitech RC-51 Digital Temperature Data Logger
Figure 1: Elitech RC-51 Digital Temperature Data Logger. This image shows the compact, pen-style design of the data logger with its clear cap and LCD screen.
Elitech RC-51 Data Logger Components
Figure 2: Key components of the Elitech RC-51 Data Logger. Labeled parts include the transparent cap, USB port, LCD screen, seal ring, button & bi-color indicator, and temperature sensor.

Setup

The RC-51 data logger is designed for ease of use. Before initial use, it is recommended to configure the device parameters using the Elitechlog software, which can be downloaded from the Elitech website. Parameters such as logging interval, start options, and alarm settings can be customized.

  1. Software Installation: Download and install the Elitechlog software on your computer (Windows or macOS compatible).
  2. Connect Device: Remove the transparent cap to reveal the USB port. Plug the data logger into an available USB port on your computer.
  3. Configure Settings: Open the Elitechlog software. The device should be recognized automatically. Use the software interface to set desired logging parameters, including start mode (button press, timed, or delay), logging interval, and temperature alarm thresholds.
  4. Disconnect Safely: After configuration, safely eject the device from your computer.
Elitech RC-51 generating PDF report
Figure 3: The RC-51 data logger connected to a laptop, illustrating its ability to generate PDF reports without additional software. The image shows the data logger plugged into a USB port, with a sample PDF report displayed on the laptop screen.

Operating Instructions

Starting a Recording Session

Monitoring Data on LCD

A single click on the button allows browsing through various LCD display pages:

Elitech RC-51 LCD Indication Instructions
Figure 4: Visual guide to LCD indications on the Elitech RC-51. This image displays various screens accessible by pressing the button, showing current temperature, record points, record interval, MKT value, average, max, and min temperature values.

Stopping a Recording Session

Recording will stop automatically when the memory is full, or when the device is connected to a computer and a report is generated. If configured, recording can also be stopped by pressing and holding the button on the device.

Generating Reports

To generate a report, simply plug the data logger into a USB port on your computer. The device will automatically create an encrypted PDF report. The Elitechlog software can also be used to export data in PDF, Excel, or TXT formats for more detailed analysis and comparison.

Elitech RC-51 large data storage capacity
Figure 5: Illustration of the RC-51's large data storage capacity, capable of storing up to 32,000 temperature points. This allows for continuous recording over extended periods, such as two years at a one-hour cycle.
Elitechlog software features
Figure 6: Overview of features offered by the Elitechlog software. This includes data query & storage, chart comparison for multiple devices, settable alarm range, FDA 21 CFR Part 11 compliance, email backup, and data export options (PDF & Excel).
Video 1: Elitech RC-51 Pen-Style Design with Waterproof IP67. This video demonstrates the compact, pen-style design and the IP67 waterproof rating of the RC-51 data logger, highlighting its suitability for various environments.

Maintenance

Elitech RC-51 IP65 Protection Class
Figure 7: The RC-51 data logger demonstrating its IP65 protection class, indicating resistance to water splashes and dust. This feature ensures reliability in challenging environments.
Elitech RC-51 Two-year Validation Certificate
Figure 8: Image showing the Elitech RC-51 data logger alongside its two-year validation certificate and user manual. This certificate confirms the device's calibration and performance standards.

Troubleshooting

ProblemPossible CauseSolution
Device not starting recording.Incorrect start mode configuration; battery low; device not activated.Check settings in Elitechlog software. Ensure battery is charged. Press and hold the button to activate if configured for button start.
No PDF report generated.USB connection issue; device memory full; corrupted data.Ensure secure USB connection. Try a different USB port. Check device memory status. Reformat device if data is not critical (this will erase all data).
Inaccurate temperature readings.Sensor obstruction; device placed in extreme conditions; calibration needed.Ensure sensor is clear. Operate within specified temperature range. Consider professional calibration if accuracy is critical.
LCD display not working.Low battery; device malfunction.Charge or replace battery. If problem persists, contact customer support.

Specifications

AttributeValue
Model NumberRC-51
Temperature Measuring Range-22℉~158℉ (-30°C ~+70°C)
Temperature Accuracy±0.5°C
Resolution0.1°C
Data Storage Capacity32,000 points
InterfaceUSB
Battery1 Lithium Metal battery (included)
Product Dimensions0.94 x 0.94 x 5.16 inches
Weight2.82 ounces
Protection ClassIP67
ManufacturerElitech
Elitech RC-51 Specifications and Dimensions
Figure 9: Detailed specifications and dimensions of the Elitech RC-51 data logger, including temperature range, accuracy, and physical measurements.

Warranty and Support

The Elitech RC-51 Digital Temperature Data Logger comes with a two-year validation certificate, ensuring its quality and performance. For any technical assistance, Elitech offers 24/7 US Technician Support via Email and Phone. Please refer to the official Elitech website or the included documentation for specific warranty terms and contact information.

For calibration certificates or further inquiries, please email or message Elitech customer service.

Related Documents - RC-51

Preview Elitech RC-51 Data Logger User Manual
User manual for the Elitech RC-51 temperature data logger, detailing its features, specifications, setup, operation, data logging, and battery replacement procedures.
Preview Elitech RC-51 Multi-use Temperature Data Logger User Manual
Comprehensive user manual for the Elitech RC-51 Multi-use Temperature Data Logger, covering product overview, technical specifications, operation instructions, display menus, report generation, and battery replacement.
Preview Elitech RC-51 Multi-use Temperature Data Logger - Technical Specifications
Explore the Elitech RC-51, a versatile temperature data logger designed for accuracy and reliability. This document provides technical specifications, features, and software download information for the RC-51 model.
Preview Elitech RC-51 Multi-use Temperature Data Logger User Guide and Specifications
Comprehensive guide to the Elitech RC-51 multi-use temperature data logger, covering product overview, technical parameters, usage instructions, data viewing, battery replacement, and PDF report content. Features include a wide temperature range, high accuracy, and large storage capacity for logistics and cold chain monitoring.
Preview Elitech RC-5+ Temperature Data Logger: User Manual and Specifications
This manual provides detailed instructions for using the Elitech RC-5+ USB temperature data logger, covering setup, operation, data download, and analysis with Elitech software. Features and technical specifications are also included.
Preview Elitech RC-5 Quick Start Guide: Temperature Data Logger
Quick start guide for the Elitech RC-5 temperature data logger, covering battery installation, software setup, logger operation, technical specifications, and troubleshooting.