1. Introduction
This manual provides essential instructions for the safe and efficient operation, setup, maintenance, and troubleshooting of the Radwag MA 50.R Moisture Analyzer. Please read this manual thoroughly before using the device to ensure proper functionality and to prevent damage or injury.
2. Safety Information
Observe the following safety precautions to ensure safe operation:
- Always operate the device on a stable, level surface.
- Ensure the power supply matches the device's requirements (100-120V AC or 200-240V AC, 50/60Hz).
- The drying chamber operates at high temperatures (up to 160 °C). Avoid direct contact with the heating element and hot parts during and immediately after operation.
- Do not place flammable or explosive materials in the drying chamber.
- Disconnect the power cord before cleaning or performing any maintenance.
- Only use original Radwag accessories and spare parts.
3. Product Overview
The Radwag MA 50.R is a precision moisture analyzer designed for accurate determination of moisture content in various samples. It features an infrared drying lamp, a backlit LCD display, and multiple connectivity options.
3.1 Key Features
- Maximum Capacity: 50g
- Moisture Content Readability: 0.001%
- Drying Lamp: 160 °C Infrared
- Display: Backlit LCD
- Functions: Dry Mass Determination, 4 Weighing Modes
- Connectivity: RS232, USB-A, USB-B, Wi-Fi
3.2 Components



4. Setup
Follow these steps to set up your MA 50.R Moisture Analyzer:
- Unpacking: Carefully remove the analyzer from its packaging. Retain all packaging materials for future transport or storage.
- Placement: Place the analyzer on a stable, level, and vibration-free surface. Ensure the operating temperature is between +10 °C and +40 °C. Avoid direct sunlight, drafts, and proximity to heat sources.
- Leveling: The analyzer features a manual leveling system. Adjust the leveling feet until the bubble in the spirit level (usually located near the display) is centered.
- Sample Pan Installation: Open the drying chamber lid. Carefully place the sample pan support and then the sample pan onto the weighing platform.
- Power Connection: Connect the power cord to the analyzer's power input and then to a suitable electrical outlet. The device supports 100-120V AC or 200-240V AC, 50/60Hz.
- Initial Power On: Press the On/Off button to power on the device. Allow a brief warm-up period for stable operation.

5. Operating Instructions
The MA 50.R Moisture Analyzer offers various modes and functions for precise moisture analysis.
5.1 Basic Operation
- Preparation: Ensure the analyzer is level and the sample pan is clean and correctly placed.
- Tare: Press the 'Tare' button to zero the display with the empty sample pan.
- Sample Loading: Carefully place your sample onto the sample pan. Ensure even distribution for optimal drying.
- Close Lid: Gently close the drying chamber lid.
- Select Drying Program: Use the control panel buttons to select the desired drying program. The MA 50.R supports 4 drying modes: Standard, Fast, Step, and Mild.
- Start Analysis: Press the 'Start/Stop' button to begin the drying process. The display will show real-time measurements such as temperature, time, and moisture content.
- End of Analysis: The analysis will conclude based on the selected finish mode (Automatic, Manual, Time-defined, User-defined). The final results will be displayed.

5.2 Data Management and Connectivity
The MA 50.R allows for comprehensive data management and connectivity:
- Databases: The system supports 6 databases for users (up to 10), products (up to 1000), weighings (up to 1000), tares (up to 100), drying programs (up to 100), and drying process reports (up to 1000).
- Direct Access: Use the keyboard soft keys for direct access to databases, functions, reports, profiles, and sample information.
- Interfaces: Utilize the RS232, USB-A, USB-B, and Wi-Fi interfaces for data transfer, external device connection, and software integration.
6. Maintenance
Regular maintenance ensures the longevity and accuracy of your moisture analyzer.
- Cleaning: Disconnect the power before cleaning. Use a soft, damp cloth with a mild detergent to clean the exterior. Do not use abrasive cleaners or solvents.
- Sample Pan: Clean the sample pan after each use to prevent contamination and ensure accurate readings.
- Drying Chamber: Periodically inspect the drying chamber and heating element for any residue. Clean carefully when the device is cool.
- Calibration: Regular calibration with certified weights is recommended to maintain measurement accuracy. Refer to the device's menu for calibration procedures.
7. Troubleshooting
This section addresses common issues you might encounter with the MA 50.R Moisture Analyzer.
| Problem | Possible Cause | Solution |
|---|---|---|
| Device does not power on | No power supply; faulty power cord; device malfunction | Check power cord connection and outlet. Try a different power cord. If issue persists, contact support. |
| Inaccurate readings | Unstable surface; uncalibrated device; dirty sample pan; drafts | Ensure stable placement. Perform calibration. Clean sample pan. Shield from drafts. |
| Drying process takes too long | Incorrect drying program; sample too large/thick | Select a more appropriate drying mode (e.g., Fast). Reduce sample size or thickness. |
| Connectivity issues (USB/Wi-Fi) | Incorrect settings; driver issues; network problems | Verify device settings. Install necessary drivers on PC. Check Wi-Fi network connection. |
8. Specifications
Detailed technical specifications for the Radwag MA 50.R Moisture Analyzer:
- Model: MA 50.R
- Manufacturer: Radwag
- Maximum Capacity: 50 g
- Readability (d): 1 mg
- Moisture Content Readability: 0.001%
- Moisture Content Repeatability: +/- 0.05% (sample ~2g), +/- 0.01% (sample ~10g)
- Heating Module: IR emitter
- Drying Temperature Range: Max 160 °C
- Drying Modes: Standard, Fast, Step, Mild
- Finish Modes: Automatic, Manual, Time-defined, User-defined
- Display Type: LCD (backlit)
- Protection Class: IP 43
- Weighing Pan Dimensions: Ø90 mm, h=8 mm
- Communication Interface: RS232, USB-A, USB-B, Wi-Fi
- Power Supply: 100-120V AC 50/60Hz or 200-240V AC 50/60Hz
- Power Consumption: 4 W (operating), 450 W (heating module)
- Operating Temperature: +10 °C to +40 °C
- Product Dimensions: 13 x 8 x 7 inches (330 x 203 x 178 mm)
- Net Weight: 10 Pounds (4.5 kg)
- Gross Weight: 13.2 Pounds (6 kg)



9. Warranty
For warranty information, please refer to the documentation included with your purchase or visit the official Radwag website. Warranty terms and conditions may vary by region and distributor.
10. Support
For technical assistance, service, or further inquiries, please contact your local Radwag distributor or visit the official Radwag website: